Gamma camera imaging characteristics of 166Ho and 99mTc used in Selective Internal Radiation Therapy

Abstract

Background

The administration of a 166Ho scout dose is available as an alternative to 99mTc particles for pre-treatment imaging in Selective Internal Radiation Therapy (SIRT). It has been reported that the 166Ho scout dose may be more accurate for the prediction of microsphere distribution and the associated therapy planning. The aim of the current study is to compare the scintigraphic imaging characteristics of both isotopes, considering the objectives of the pre-treatment imaging using clinically geared phantoms.

Methods

Planar and SPECT/CT images were obtained using a NEMA image quality phantom in different phantom setups and another body-shaped phantom with several inserts. The influence of collimator type, count statistics, dead time effects, isotope properties and patient obesity on spatial resolution, contrast recovery and the detectability of small activity accumulations was investigated. Furthermore, the effects of the imaging characteristics on personalized dosimetry are discussed.

Results

The images with 99mTc showed up to 3 mm better spatial resolution, up to two times higher contrast recovery and significantly lower image noise than those with 166Ho. The contrast-to-noise ratio was up to five times higher for 99mTc than for 166Ho. Only when using 99mTc all activity-filled spheres could be distinguished from the activity-filled background. The measurements mimicking an obese patient resulted in a degraded image quality for both isotopes.

Conclusions

Our measurements demonstrate better scintigraphic imaging properties for 99mTc compared to 166Ho in terms of spatial resolution, contrast recovery, image noise, and lesion detectability. While the 166Ho scout dose promises better prediction of the microsphere distribution, it is important to consider the inferior imaging characteristics of 166Ho, which may affect individualized treatment planning in SIRT.

Background

Selective Internal Radiation Therapy (SIRT) is a well-established nuclear medicine treatment for primary and secondary liver malignancies. Radioactively labeled microspheres are injected into the hepatic artery and then accumulate predominantly in the tumor tissue. Currently, three different types of microspheres with different physical properties are commercially available, using the high-energy beta-emitters Yttrium-90 (90Y) and Holmium-166 (166Ho).

Comprehensive treatment planning, including therapy simulation and scintillation camera imaging, is usually performed using Technetium-99m-labeled macroaggregated albumin (99mTc-MAA) or human serum albumin microspheres (99mTc-HSA) as a surrogate for the therapeutic microspheres. The objectives of the pre-treatment workup and scintigraphy are the determination of the liver-lung shunt (LLS) and the detection of abdominal extrahepatic microsphere deposition as well as the prediction of intrahepatic activity distribution as a measure of regional dose. Several studies have shown a limited predictive value of 99mTc-MAA, which may be due to differences in size, shape, and density of the injected particles [1, 2]. The administration of a scout dose of 166Ho poly-L-lactic-acid (PLLA) microspheres has recently become available as an alternative to 99mTc particles for SIRT with 166Ho. This displays the theoretical advantage since the same microspheres can be used for diagnostics and therapy. It is reported that the 166Ho scout dose may be more accurate for the prediction of microsphere distribution and the associated therapy planning [3,4,5]. However, further investigation regarding the imaging properties of 166Ho compared to 99mTc is required with respect to the different physical properties, e.g. gamma energy and emission probability (Table 1). Due to the low gamma emission probability of 166Ho (6.7%) and the need for a medium-energy collimator due to the higher energy bremsstrahlung photons, scintigraphic imaging is expected to be inferior to 99mTc in terms of sensitivity, spatial resolution and lesion detectability.

The aim of the current study was to compare the scintigraphic imaging characteristics of both isotopes, considering the objectives of the pre-treatment imaging (liver-lung shunt, extrahepatic deposition, intrahepatic distribution). The influence of collimator type, count statistics, dead time effects, physical properties and patient obesity on spatial resolution, contrast recovery and the detectability of small activity accumulations was investigated. For this purpose, planar and tomographic images of clinically geared phantoms were acquired. In addition, the influence of the imaging characteristics on treatment planning and personalized dosimetry is discussed.

Table 1 Physical properties of the isotopes 166Ho and 99mTc

Methods

All measurements were performed on a dual-head Symbia Intevo 6 SPECT/CT system (Siemens Healthineers) with 3/8 inch NaI(Tl) crystals. The rectangular field of view (FOV) was 53.3 cm x 38.7 cm. Data was acquired using a low-energy high-resolution (LEHR) collimator (hole length: 24.05 mm, septal thickness: 0.16 mm, hole diameter: 1.11 mm) for 99mTc and a medium-energy low-penetration (MELP) collimator (hole length: 40.64 mm, septal thickness: 1.14 mm, hole diameter: 2.94 mm) for 166Ho and 99mTc to investigate the influence of the collimator. Dual-energy windows were used for both isotopes. For 99mTc, the photopeak window was centered at 140 keV (15% width) with an adjacent lower scatter window at 119 keV (17.6% width). The 166Ho photopeak window was set to 81 keV (15% width). An additional scatter window at 118 keV (12% width) was used to correct for down-scattered high-energy photons. Activity for the phantom experiments was measured using a dose calibrator (ISOMED 2010, NuviaTech Healthcare).

Sensitivity and count rate performance

The sensitivity was measured using a point source placed in the center of the FOV without attenuation and scatter with activities of 146 MBq and 222 MBq for 99mTc and 166Ho, respectively. Planar images were acquired for 5 min on a 256 × 256 matrix. The source-detector distance was set to 10 cm.

The effects of dead time and the resulting deviations of count rate linearity were analyzed depending on the activity at scan time. A custom-made phantom with a body-shaped cross section (diameter 30 cm, width 8 cm), including parts of the ribs and spine with bone-equivalent tissue (Fig. 1a), was used to simulate patient attenuation and scattering conditions. An activity-filled vial was placed in a hole at the center of the phantom. Activities ranging from 190 MBq to 2835 MBq and 154 MBq to 6780 MBq were measured for 99mTc and 166Ho, respectively. Planar images were acquired as mentioned above. Anterior and posterior detector-phantom center distances were set to 22 cm and 17 cm, respectively, similar to patient settings. For data analysis, the counts in the photopeak window were plotted against the activity in the phantom. For each isotope and collimator, a data fit was performed according to the paralyzable detector model as follows:

Fig. 1
figure 1

(a) Custom-made phantom with a body-shaped cross section, including parts of the ribs and spine with bone-equivalent tissue used for count rate performance measurements. (b) NEMA image quality phantom wrapped in cooling gel packs to mimic the attenuation and scattering conditions of an obese patient. The NEMA image quality phantom was used for quantitative evaluation of image quality and investigation of the influence of patient obesity. (c) CT image of the NEMA image quality phantom. Transverse slice at the level of the spheres. Background ROIs (blue) and lung insert ROI (red)

$$C\left( A \right) = \alpha \cdot A \cdot {e^{ - \tau \cdot \alpha \cdot A}}$$
(1)

where C is the measured count rate, A is the activity at scan time, α describes the linear detector performance and τ is the detector dead time. It was specified that the curve passes through the axis origin. In addition, energy spectra were measured and analyzed.

Quantitative evaluation of image quality

Phantom setups

The torso-shaped NEMA image quality (IQ) phantom, containing a fillable background compartment, six fillable coplanar spheres (inner diameter = 10, 13, 17, 22, 28, and 37 mm), and a cylindrical lung insert, was used to evaluate the image quality and spatial resolution. It was filled with three different sphere-to-background activity concentration ratios to resemble clinical setups. To mimic high extrahepatic abdominal deposition, which is characterized by small activity spots with no background activity, the phantom was measured with activity in the six spheres only. Spheres and the background compartment were filled to mimic the activity distribution in the liver with different activity accumulation in the tumor and parenchymal tissue. A sphere-to-background activity concentration ratio of 4:1 was used to mimic liver metastases with moderate uptake, and a sphere-to-background ratio of 8:1 to mimic lesions of hepatocellular carcinoma (HCC) with higher activity uptake relative to the parenchyma. The activity concentrations in the spheres and background were chosen to be similar to those in clinical practice, assuming a patient with a liver volume of 2000 ml containing 150 MBq of 99mTc or 250 MBq of 166Ho, which are standard activity values for pre-treatment SIRT diagnostics. A tumor fraction of 10% (200 ml) and a homogeneous activity distribution in tumors and liver parenchyma were assumed. The actual activity concentrations at the time of imaging are shown in Table 2. The phantom was also measured after being wrapped in cooling gel packs (Fig. 1b) to mimic the attenuation and scattering conditions of an obese patient. The cooling packs form an additional thickness of approximately 3 cm around the phantom. The attenuation coefficient of the cooling packs was similar to that of water.

Table 2 Information on the measurements with the NEMA image quality phantom filled with 99mTc and 166Ho in different phantom setups (spheres without background; sphere-to-background activity concentration ratio 8:1 and 4:1)

Planar imaging of the NEMA image quality phantom

For the evaluation of the planar scintigraphic imaging characteristics of 99mTc and 166Ho, planar anterior and posterior images were acquired for each phantom setup with an acquisition time of 5 min and an image matrix of 256 × 256. Additional measurements of 99mTc with the MELP collimator and a reduced acquisition time of 22 s were performed to investigate the influence of the used collimator and the low gamma emission probability of 166Ho compared to 99mTc. The reduced acquisition time was chosen based on the ratio of the gamma emission probabilities of 166Ho to 99mTc. Planar scatter correction was applied for 166Ho. The images were evaluated with regard to the pre-treatment determination of the liver-lung shunt in SIRT, which is usually based on planar scans. A region of interest (ROI) was placed in the phantom simulating the liver and another ROI was placed above the phantom in an area of no activity. The ratio of the counts in the ROI outside the phantom to the sum of the counts in both ROIs was calculated as a measure of apparent lung shunt.

SPECT/CT acquisition and reconstruction parameters

All SPECT data was acquired with 120 projections (60 per head, 20 s per projection) over a non-circular 360° orbit using step-and-shoot mode, followed by a low-dose CT scan (130 kV, 20 mAs, 2.5 mm slice thickness) for attenuation correction. The images were reconstructed to a voxel size of 2.4 × 2.4 × 2.4 mm3 using the 3D OSEM algorithm (Flash 3D; Siemens Healthineers) with 8 iterations and 8 subsets for 99mTc and 16 iterations and 8 subsets for 166Ho according to our clinical routine. Gaussian post-reconstruction filtering of 9 mm was applied. Scatter correction for 99mTc was performed using the dual-energy window method with an adjacent lower scatter window and a scatter fraction k-factor of 0.5. For 166Ho the photopeak was corrected for scatter with the upper scatter window at 118 keV and a k-factor of 1.4, as previously investigated [6]. Additional measurements of 99mTc with the MELP collimator and a reduced scan time of 1.5 s per projection were performed to investigate the influence of the used collimator and the low gamma emission probability of 166Ho compared to 99mTc. The acquisition times were adjusted over time according to the respective half-life to obtain comparable count statistics. To obtain information on statistical variation and reproducibility of the results, measurements of the NEMA IQ phantom at 8:1 contrast and standard phantom setup were performed exemplarily twice. The average and standard deviation of both measurements were used for further analysis.

Influence of count statistics

The influence of count statistics on SPECT image quality was evaluated by examining the reconstructed images of the NEMA IQ phantom with a sphere-to-background ratio of 8:1 for different acquisition times. SPECT images with projection times of 5 s, 10 s, 15 s, and 20 s for 99mTc and 10 s, 20 s, and 30 s for 166Ho were analyzed.

Image analysis

The SPECT images were evaluated in terms of contrast recovery, image noise, detectability, relative count error in the lung insert, and spatial resolution. The six hot spheres were segmented using 3D isocontours with a 50% background-adapted threshold for each sphere according to [7]. Three cylindrical volumes of interest (VOI) (45 mm diameter, 150 mm length) were defined in the phantom background (Fig. 1c). The lung insert was delineated with a cylindrical centered VOI (30 mm diameter, 130 mm length) similar to the NEMA NU 2-2018 protocol [8].

The contrast recovery coefficient (CRC) was calculated for each of the six hot spheres as follows:

$$CRC = \frac{{\frac{{{{\mathop N\limits^ - }_{\text{S}}}}}{{{{\mathop N\limits^ - }_{{\text{BG}}}}}} - 1}}{{R - 1}} \cdot 100\%$$
(2)

where \({\stackrel{-}{N}}_{\text{S}}\) is the mean number of counts in the sphere VOI, \({\stackrel{-}{N}}_{\text{BG}}\) is the mean number of counts in the background VOIs, and \(R\) is the true sphere-to-background activity concentration ratio. The relative count error in the lung insert (\(\varDelta {N}_{\text{lung}})\) was determined as:

$$\Delta {N_{{\text{lung}}}} = \frac{{{{\mathop N\limits^ - }_{{\text{lung}}}}}}{{{{\mathop N\limits^ - }_{{\text{BG}}}}}} \cdot 100\%$$
(3)

where \({\stackrel{-}{N}}_{\text{lung}}\) is the mean number of counts in the lung insert VOI. To evaluate the image noise, the noise coefficient of variation (\({CV}_{\text{B}\text{G}}\)) was calculated using:

$$C{V_{{\text{BG}}}} = \frac{{{\sigma _{{\text{BG}}}}}}{{{{\mathop N\limits^ - }_{{\text{BG}}}}}} \cdot 100\%$$
(4)

where \({\sigma }_{\text{BG}}\) is the standard deviation of all voxels within the three background VOIs. The contrast-to-noise ratio (CNR) was used to asses object detectability. It was calculated for each sphere as follows:

$$CNR=\frac{{\stackrel{-}{N}}_{\text{S}}-{\stackrel{-}{N}}_{\text{BG}}}{{\sigma }_{\text{BG}}}$$
(5)

The tomographic spatial resolution was determined based on the analysis of radial profiles through the homogeneously filled phantom spheres in the reconstructed images according to [9]. The full width at half maximum (FWHM) of the point spread function was assessed using the software Rover (version 3.0.60h, ABX, Germany).

Results

Sensitivity and count rate performance

The sensitivities measured for the photopeak window for 99mTc were 93.1 cps/MBq and 125.3 cps/MBq for the LEHR and MELP collimator, respectively. For 166Ho the sensitivity was much lower at 16.4 cps/MBq for the MELP collimator.

Figure 2a shows the count rate response for 99mTc measured with the LEHR and MELP collimator and for 166Ho measured with the MELP collimator. With increasing activity, there was an increasing deviation from the ideal linear count rate response. The count loss was higher for 166Ho than for 99mTc. The activity leading to a 5% count loss was 169 MBq for 166Ho with MELP collimator and 3141 MBq and 1611 MBq for 99mTc with LEHR and MELP collimator, respectively. In Fig. 2b, the measured energy spectra of 99mTc and 166Ho normalized to the sum of the counts in each photopeak window are shown. It illustrates the photopeak of 99mTc at 141 keV and the characteristic Kα and Kβ radiation of lead at approximately 74 keV and 85 keV. In the energy spectrum of 166Ho, the photopeak at 81 keV, the characteristic X-ray radiation of 166Ho at 49 keV and a pronounced bremsstrahlung continuum above the photopeak window can be seen. For the equal number of counts in the photopeak window, approximately six times more counts were detected for 166Ho than for 99mTc over the entire detector energy range.

Fig. 2
figure 2

(a) Count rate performance measured for 99mTc with LEHR and MELP collimator and for 166Ho measured with MELP collimator. (b) 99mTc and 166Ho spectrum obtained in phantom setup used for count rate performance. The curves are normalized to the sum of the counts in each photopeak window. The dashed lines represent the 99mTc photopeak window centered at 140 keV and the 166Ho photopeak window at 81 keV

Quantitative evaluation of image quality

Planar imaging of the NEMA image quality phantom

Figure 3 shows the NEMA IQ phantom planar images of 99mTc and 166Ho with a sphere-to-background ratio of 8:1. The 99mTc images featured a better detectability of the spheres and a lower image noise compared to 166Ho. The uncorrected image with 166Ho showed many artificial counts in the activity-free background, whereas the scatter-corrected image showed reduced number of counts in the activity-free background.

Fig. 3
figure 3

Planar posterior gamma camera images of the NEMA image quality phantom at a sphere-to-background activity concentration ratio of 8:1. (a) 99mTc with LEHR collimator, (b) 99mTc with MELP collimator, (c) 99mTc with MELP collimator and reduced scan time, (d) 166Ho with MELP collimator, (e) 166Ho with MELP collimator and planar scatter correction

The calculated apparent lung shunt was 2.0% for the 99mTc scans with the LEHR collimator and 1.9% and 1.7% for the 99mTc scans with the MELP collimator with standard and reduced scan time, respectively. For 166Ho, the apparent lung shunt was 13.0% for the scans without scatter correction and was reduced to 2.3% with planar scatter correction.

Influence of count statistics

In Fig. 4, contrast recovery coefficients and contrast-to-noise ratios are shown as a function of sphere diameter for measurements with different acquisition times. For both isotopes, the CRC was almost identical for measurements with different projection times and decreased continuously with decreasing sphere diameter. Spatial resolution, image noise, and lung count error for the different acquisition times are listed in Table 3. The background noise level decreased with increasing acquisition time for 99mTc and 166Ho, but was more than twice as high for 166Ho compared to 99mTc. For 99mTc, there was no visible improvement in the noise level for the measurement with 20 s projection time compared to the measurement with 15 s projection time. The spatial resolution was comparable between the images of different acquisition times and was approximately 12.5 mm for 99mTc and 15.2 mm for 166Ho. For the measurement with 166Ho and an acquisition time of 10 s per projection, the smallest sphere (d = 10 mm) could not be evaluated due to the high image noise in the background.

Fig. 4
figure 4

Contrast recovery coefficients (%) and contrast-to-noise ratios as a function of sphere diameter for 99mTc (upper row) and 166Ho (lower row) for different projection times. Data was determined using the NEMA image quality phantom at a sphere-to-background activity concentration ratio of 8:1

Table 3 Spatial resolution (FWHM), image noise (CVBG), and lung count error (ΔNLung) determined using the NEMA image quality phantom filled with 99mTc and 166Ho at 8:1 contrast with different scan times

Influence of collimator type

Table 4 shows the spatial resolution, image noise, and lung count error for 99mTc measurements with LEHR and MELP collimator for the different phantom setups. The measurements with the MELP collimator showed a slightly higher count error in the lung insert than the measurements with the LEHR collimator. Image noise was slightly lower with the MELP collimator than with the LEHR collimator at 8:1 contrast, and identical for both collimators at 4:1 contrast, with a noise level of 7.2%. The spatial resolution was approximately 2 mm worse when using the MELP collimator.

Table 4 Spatial resolution (FWHM), image noise (CVBG), and lung count error (ΔNLung) determined using the NEMA image quality phantom filled with 99mTc and 166Ho in different phantom setups (spheres without background, contrast 8:1 and 4:1, normal and obese phantom setup) with standard and adapted acquisition parameters (collimator type, reduced projection time). For measurements performed twice, results are presented as the mean and standard deviation

Comparison of the imaging of 99mTc and 166Ho with different phantom setups

Figure 5 shows the CRC and the CNR as a function of sphere diameter for the measurements using the NEMA IQ phantom with 8:1 contrast of 99mTc and 166Ho with standard acquisition parameters and for 99mTc with MELP collimator and reduced scan time. For the measurements with 99mTc, the CNR was higher for all spheres compared to 166Ho. The CNRs of the largest sphere was 79.4 and 42.1 for 99mTc with standard and adjusted parameters, respectively, and 28.8 for 166Ho. The CRC decreased continuously with decreasing sphere diameter for 99mTc. This was similar for 166Ho, except that the CRC of the 17 mm and the 22 mm sphere was similar. The smallest sphere (d = 10 mm) could not be evaluated for the measurement with 99mTc and adapted parameters due to the high noise level. Spatial resolution, image noise, and lung count error for the different phantom setups are listed in Table 4. The spatial resolution for the 99mTc images with standard parameters was up to 3 mm better than for 166Ho. The spatial resolution was better in the images of the phantom with no background activity and a sphere-to-background ratio of 8:1 compared to 4:1. The image noise for the 99mTc measurement with adjusted parameters was worse than the 99mTc standard (noise level 15.8% vs. 7.6%), but better than 166Ho with a noise level of 18.0%. The results of the measurements performed twice showed only slight deviations, which are mostly not visible in the CRC and CNR figures (Fig. 5).

Fig. 5
figure 5

Contrast recovery coefficients (%) and contrast-to-noise ratios as a function of sphere diameter determined using the NEMA image quality phantom at a sphere-to-background activity concentration ratio of 8:1. 99mTc acquired with LEHR collimator and with MELP collimator and reduced scan time (1.5 s projection time) and 166Ho acquired with MELP collimator. 99mTc and 166Ho acquired with the modified phantom setup mimicking patient obesity. For measurements performed twice, results are presented as the mean and standard deviation (error bars)

Figure 6a shows the reconstructed transverse images of all NEMA IQ phantom setups of 99mTc and 166Ho with standard acquisition parameters and for 99mTc with MELP collimator and reduced scan time. For the measurement with activity in the spheres only, all spheres were visible for both isotopes. When comparing the 8:1 contrast images, for 166Ho and 99mTc measured with the MELP collimator, the smallest sphere (d = 10 mm) was not detectable and the high background noise level is evident. In the 4:1 contrast images, the smallest sphere was not detectable for 99mTc and only the three largest spheres (d ≥ 22 mm) were detectable for 166Ho. The images with activity in the background showed a distortion of the spheres for the measurements with 99mTc with reduced scan time and for 166Ho.

Fig. 6
figure 6

(a) SPECT images of the NEMA image quality phantom filled with 99mTc and 166Ho in different phantom setups (spheres without background activity, sphere-to-background activity concentration ratio 8:1 & 4:1). For 99mTc, images were acquired with LEHR collimator, MELP collimator and reduced scan time. 166Ho images were acquired using the MELP collimator. Transverse slices at the level of the spheres are shown. (b) SPECT images of the NEMA image quality phantom filled with 99mTc and 166Ho at a sphere-to-background activity concentration ratio of 8:1, acquired with the standard phantom setup and the modified phantom mimicking patient obesity. Transverse slices at the level of the spheres are shown

Influence of an obese body

The use of the obese phantom setup resulted in an increase in the lung count error and the image noise and a slight decrease in the CRC for both isotopes (Table 4). This has resulted in a decrease in the CNR (Fig. 5) and therefore a decrease in the detectability of the hot spheres. For the measurement with 166Ho in the obese phantom setup, the smallest sphere (d = 10 mm) could not be evaluated due to the high noise level. The spatial resolution was slightly worse when using the obese phantom setup (99mTc: 12.6 mm; 166Ho: 15.6 mm) compared to the standard phantom setup (99mTc: 12.2 mm; 166Ho: 15.2 mm). The reconstructed transverse images of the NEMA IQ phantom measurements of 99mTc and 166Ho with standard and obese phantom setup are shown in Fig. 6b. The measurement of 166Ho with the obese phantom setup showed a significantly worse detectability of the spheres and a blurring at the edge of the phantom compared to the standard phantom setup.

Discussion

The scintigraphic imaging properties of 99mTc and 166Ho have been well studied and characterized [10,11,12,13], and there also is a comparison of the three isotopes (99mTc, 90Y, and 166Ho) used in radioembolization [14]. However, these studies are primarily oriented towards the NEMA performance measurement guidelines and do not consider the clinical imaging aspects of SIRT. Our study focused on the scintigraphic imaging characterization of 99mTc and 166Ho using phantom geometries, considering the aspects of pre-treatment imaging in radioembolization. The influence of the collimator type, count statistics, dead time effects isotope properties and patient obesity on image quality (spatial resolution, contrast recovery, image noise, lesion detectability) was evaluated in detail.

As expected, the sensitivity for 166Ho was significantly lower than for 99mTc, but higher than the expected sensitivity according to the low gamma emission probability of 166Ho compared to 99mTc. This is due to the presence of lead X-rays and bremsstrahlung photons in the 81 keV photopeak window. The MELP collimator showed a higher planar sensitivity than the LEHR collimator for 99mTc due to the larger hole diameter and the resulting larger acceptance angle. The count rate performance of 99mTc showed an almost linear detector response and there is no significant count loss for the clinically used activity range. The activity resulting in a 2% count loss was 1237 MBq for 99mTc with LEHR collimator. This is in a similar order of magnitude as the activity values of about 1 GBq reported by Ryu et al. [12] and Elschot et al. [14]. The progressive count loss for 166Ho with increasing activity is caused by dead time effects due to the high sum of counts from detected bremsstrahlung photons over the entire energy range of the detector. In this study, a value of 736 MBq was found, leading to a count rate loss of 20% for 166Ho. The work of Stella et al. [15] reported similar count loss values. For a 166Ho scout dose of 250 MBq, count loss can be neglected, but at high activity levels measured after radioembolization, dead time effects must be considered, especially for quantitative SPECT imaging. Sensitivity and count rate performance are directly related to count statistics and image noise. Analyses of the effect of count statistics showed a continuous decrease in image noise with increasing count statistics. Due to the low gamma emission probability of 166Ho, the noise level is significantly higher than for 99mTc. It was found that the count statistics did not have a significant effect on spatial resolution and CRC. However, image noise, CNR, and therefore lesion detectability decrease with lower count statistics. Except for the measurements of 99mTc with 15 s and 20 s projection time, where similar values were found for image noise and CNR. Therefore, these results may be useful for clinical routine, as the 15 s projection time images provide an equivalent image quality. The use of the MELP collimator instead of the LEHR collimator degraded the resolution by approximately 2 mm FHWM for 99mTc. This is a consequence of the wider acceptance angle of the MELP collimator and the larger source to detector (crystal surface) distance due to the thicker MELP collimator. The use of the MELP collimator for 99mTc was not considered for clinical routine, but only for investigation of the collimator influence on image quality.

The qualitative and quantitative evaluation of image quality revealed that planar and SPECT imaging was worse for 166Ho compared to 99mTc regarding spatial resolution, contrast recovery, image noise, lesion detectability, and lung count error (Figs. 3, 5 and 6; Table 4). To investigate the influence of the properties of the two isotopes on image quality, measurements with 99mTc were performed using the MELP collimator and reduced acquisition time to achieve similar count statistics. These measurements also showed better image characteristics compared to 166Ho, demonstrating that additional aspects affect the inferior image quality of 166Ho. The lower gamma energy of 166Ho (81 keV) compared to 99mTc (141 keV) leads to a decrease of the spatial resolution and a noticeably greater blurring. The effect of gamma energy on gamma camera spatial resolution was previously investigated by Holstensson et al. [16]. The high energy bremsstrahlung photons of 166Ho lead to septal penetration and crystal scatter and they produce lead X-rays in the collimator which are detected in the 81 keV energy window. These photons have a loss of spatial information and result in a worse image quality regarding image noise and contrast recovery. There is a high percentage of scatter when measuring 166Ho, leading to a high lung count error and the detection of scattered photons in compartments with no activity. The high noise level in the images with 99mTc with reduced scan time and with 166Ho results in a distortion of the spheres as they merge with the noisy background [17]. For both isotopes, the CRC decreased continuously with decreasing sphere diameter due to the partial volume effect (Fig. 5). For 166Ho, the 22 mm sphere unexpectedly had a similar CRC value as the 17 mm sphere. This could be due to the sphere position in the phantom and the larger detector radius in the lateral position when using auto-contouring [18]. For the 166Ho measurement with the obese body phantom setup the CRC of the 22 mm sphere was similar to the CRC of the standard phantom setup measurement, but the CRC of the 17 mm sphere was lower. This is most likely caused by the larger detector radius in the anterior position while the lateral radius is the same due to the patient couch. The exemplary twice-performed measurements show a low statistical variation. This indicates that the results are reproducible and are not coincidental findings.

The measurements with the NEMA IQ phantom mimicking an obese patient resulted in a degraded image quality in terms of CRC, CNR, image noise, lung count error, and spatial resolution for both isotopes. This is a result of the additional material around the phantom leading to increased photon attenuation and scattering. In addition, the source-detector-distance is larger due to the additional volume, which may explain the worse spatial resolution [19]. The reduced CNR indicates that the detectability of small lesions is more challenging in obese patients. Reduced lesion detectability in obese patients is already reported by Lin et al. [20] for 99mTc planar imaging. Peters et al. [21] previously investigated the influence of an obese body on SPECT imaging with 99mTc using different phantoms and reported slightly reduced recovery coefficients similar to our results. Degraded image quality for an obese phantom setup was also reported for PET imaging by Braune et al. [22]. The reduction in image quality for an obese body was greater for 166Ho than for 99mTc due to the lower gamma energy and therefore greater attenuation in the material. The half-value layers for 99mTc and 166Ho in water are 4.59 mm and 3.96 mm, respectively [23].

The different imaging characteristics can have an impact on pre-treatment diagnostic and treatment planning in SIRT. The planar phantom images showed a high apparent lung shunt for 166Ho due to the high number of scattered photons. This can lead to an overestimation of the LLS and therefore an overestimation of the predicted dose to the lungs. Planar scatter correction was able to reduce the apparent lung shunt to values comparable to those measured with 99mTc. Other studies have also found an overestimation of the LLS using 166Ho planar images [3, 15]. In contrast to the study by Stella et al., we found smaller apparent lung shunt values, which could be due to the use of different phantoms. Measurements of the NEMA IQ phantom with activity only in the spheres suggest that small extrahepatic depositions can be detected with 99mTc as well as 166Ho. The NEMA IQ phantom images with 8:1 and 4:1 sphere-to-background activity concentration ratios showed worse detectability of small spheres with 166Ho compared to 99mTc. This means that the activity accumulation in small liver metastases may not be detectable in SPECT/CT images with 166Ho and could not be considered for treatment planning and dosimetry. For both isotopes, the resolution and the large partial volume effect can have an impact on predictive voxel-based dosimetry. Predicted tumor absorbed doses may be underestimated in small lesions due to underestimation of activity in small volumes. This underestimation of activity is less pronounced for 99mTc than for 166Ho due to the better resolution and contrast recovery. All results in this study are based on phantom measurements which represent ideal measuring conditions, meaning homogenous activity distribution and spherical activity spots. In vivo imaging conditions such as irregularly shaped lesions and inhomogeneous distribution may have additional influences on scintigraphic imaging that could not be observed in this study. Furthermore, the influence of the reconstruction algorithm and the reconstruction parameters were not investigated. Monte Carlo-based reconstruction algorithms, including simulation of photon scattering and collimator-detector response, can improve image quality, but are usually not available for clinical use [11].

Personalized dosimetry, taking into account the activity distribution in the tumor and parenchyma, is becoming increasingly important, as a significant dose-response relationship has been confirmed for radioembolization [24,25,26,27]. Prerequisites for individualized dosimetry are a good agreement between the microsphere distribution in diagnostics and therapy, and additionally an accurate imaging of the microsphere distribution. In addition to the type of particles used, intrahepatic activity distribution is significantly influenced by catheter position, regional vasculature, and hemodynamics. The 166Ho scout dose provides a good prediction of particle distribution because the same microspheres are used for diagnostics and therapy [5]. However, the poor imaging characteristics of 166Ho can have a negative impact on individualized dosimetry. Therefore, it has to be analyzed whether the use of 99mTc-labeled PLLA microspheres allows a superior imaging of the particle distribution and therefore a more accurate predictive dosimetry. In future studies the labeling of PLLA microspheres with 99mTc and a comparison of the in vivo distribution of 99mTc-PLLA and 166Ho-PLLA should be investigated.

Conclusion

In summary, our measurements revealed that planar and SPECT imaging is better with 99mTc compared to 166Ho regarding spatial resolution, contrast recovery, image noise, and lesion detectability. Therefore, 99mTc is the more appropriate isotope for pre-treatment imaging in SIRT. The worse imaging characteristics of 166Ho can have a negative influence on individualized treatment planning. However, 166Ho scout dose provides a good prediction of particle distribution since the same microspheres are used for therapy. Further investigations regarding 99mTc-labeled PLLA microspheres, combining the better scintigraphic imaging properties of 99mTc with the advantages of the PLLA microspheres, have to be performed.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

SIRT:

Selective Internal Radiation Therapy

LLS:

Liver lung shunt

PLLA:

Poly–L–lactic–acid

FOV:

Field of view

LEHR:

Low–energy high–resolution

MELP:

Medium–energy low–penetration

IQ:

Image quality

ROI:

Region of interest

OSEM:

Ordered Subset Expectation Maximization

VOI:

Volume of Interest

CRC:

Contrast Recovery Coefficient

CNR:

Contrast–to–noise ratio

FWHM:

Full width at half maximum

cps:

Counts per second

References

  1. Haste P, Tann M, Persohn S, LaRoche T, Aaron V, Mauxion T, et al. Correlation of Technetium-99m Macroaggregated Albumin and Yttrium-90 glass Microsphere Biodistribution in Hepatocellular Carcinoma: a retrospective review of pretreatment single Photon Emission CT and Posttreatment Positron Emission Tomography/CT. J Vasc Interv Radiol. 2017;28:722–e7301. https://doi.org/10.1016/j.jvir.2016.12.1221.

    Article  PubMed  Google Scholar 

  2. Ilhan H, Goritschan A, Paprottka P, Jakobs TF, Fendler WP, Todica A, et al. Predictive value of 99mTc-MAA SPECT for 90Y-Labeled Resin Microsphere distribution in radioembolization of primary and secondary hepatic tumors. J Nucl Med. 2015;56:1654–60. https://doi.org/10.2967/jnumed.115.162685.

    Article  CAS  PubMed  Google Scholar 

  3. Elschot M, Nijsen JFW, Lam MGEH, Smits MLJ, Prince JF, Viergever MA, et al. (99m)Tc-MAA overestimates the absorbed dose to the lungs in radioembolization: a quantitative evaluation in patients treated with 166Ho-microspheres. Eur J Nucl Med Mol Imaging. 2014;41:1965–75. https://doi.org/10.1007/s00259-014-2784-9.

  4. Braat AJAT, Prince JF, van Rooij R, Bruijnen RCG, van den Bosch MAAJ, Lam MGEH. Safety analysis of holmium-166 microsphere scout dose imaging during radioembolisation work-up: a cohort study. Eur Radiol. 2018;28:920–8. https://doi.org/10.1007/s00330-017-4998-2.

    Article  PubMed  Google Scholar 

  5. Smits MLJ, Dassen MG, Prince JF, Braat AJAT, Beijst C, Bruijnen RCG, et al. The superior predictive value of 166Ho-scout compared with 99mTc-macroaggregated albumin prior to 166Ho-microspheres radioembolization in patients with liver metastases. Eur J Nucl Med Mol Imaging. 2020;47:798–806. https://doi.org/10.1007/s00259-019-04460-y.

    Article  CAS  PubMed  Google Scholar 

  6. Kästner D, Oehme L, Freudenberg R, Kotzerke J. Optimierung der szintigraphischen Bildgebung bei der Selektiven Internen Radiotherapie (SIRT) an einem SPECT/CT. NuklearMedizin. Leipzig. © Georg Thieme Verlag KG; 2020. https://doi.org/10.1055/s-0040-1708254.

  7. Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging. 2010;37:181–200. https://doi.org/10.1007/s00259-009-1297-4.

    Article  PubMed  Google Scholar 

  8. National Electrical Manufacturers Association. NEMA NU 2-2018 Performance Measurements of Positron Emission Tomographs (PET) 2018.

  9. Hofheinz F, Dittrich S, Pötzsch C, van Hoff J. den. Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. Phys Med Biol. 2010;55:1099–113. https://doi.org/10.1088/0031-9155/55/4/013.

  10. de Wit TC, Xiao J, Nijsen JFW, van het Schip FD, Staelens SG, van Rijk PP, Beekman FJ. Hybrid scatter correction applied to quantitative holmium-166 SPECT. Phys Med Biol. 2006;51:4773–87. https://doi.org/10.1088/0031-9155/51/19/004.

    Article  PubMed  Google Scholar 

  11. Elschot M, Smits MLJ, Nijsen JFW, Lam MGEH, Zonnenberg BA, van den Bosch MAAJ, et al. Quantitative Monte Carlo-based holmium-166 SPECT reconstruction. Med Phys. 2013;40:112502. https://doi.org/10.1118/1.4823788.

    Article  CAS  PubMed  Google Scholar 

  12. Ryu H, Meikle SR, Willowson KP, Eslick EM, Bailey DL. Performance evaluation of quantitative SPECT/CT using NEMA NU 2 PET methodology. Phys Med Biol. 2019;64:145017. https://doi.org/10.1088/1361-6560/ab2a22.

    Article  CAS  PubMed  Google Scholar 

  13. Kupitz D, Wissel H, Wuestemann J, Bluemel S, Pech M, Amthauer H, et al. Optimization of SPECT/CT imaging protocols for quantitative and qualitative 99mTc SPECT. EJNMMI Phys. 2021;8:57. https://doi.org/10.1186/s40658-021-00405-3.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Elschot M, Nijsen JFW, Dam AJ, de Jong HWAM. Quantitative evaluation of scintillation camera imaging characteristics of isotopes used in liver radioembolization. PLoS ONE. 2011;6:e26174. https://doi.org/10.1371/journal.pone.0026174.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Stella M, Braat AJAT, Lam MGEH, de Jong HWAM, van Rooij R. Gamma camera characterization at high holmium-166 activity in liver radioembolization. EJNMMI Phys. 2021;8:22. https://doi.org/10.1186/s40658-021-00372-9.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Holstensson M, Partridge M, Buckley SE, Flux GD. The effect of energy and source location on gamma camera intrinsic and extrinsic spatial resolution: an experimental and Monte Carlo study. Phys Med Biol. 2010;55:1735–51. https://doi.org/10.1088/0031-9155/55/6/013.

    Article  PubMed  Google Scholar 

  17. Alqahtani MM, Willowson KP, Constable C, Fulton R, Kench PL. Optimization of 99m tc whole-body SPECT/CT image quality: a phantom study. J Appl Clin Med Phys. 2022;23:e13528. https://doi.org/10.1002/acm2.13528.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Dickson JC, Armstrong IS, Gabiña PM, Denis-Bacelar AM, Krizsan AK, Gear JM, et al. EANM practice guideline for quantitative SPECT-CT. Eur J Nucl Med Mol Imaging. 2023;50:980–95. https://doi.org/10.1007/s00259-022-06028-9.

    Article  PubMed  Google Scholar 

  19. Kappadath SC. Effects of Voxel size and iterative reconstruction parameters on the spatial resolution of 99mTc SPECT/CT. J Appl Clin Med Phys. 2011;12:3459. https://doi.org/10.1120/jacmp.v12i4.3459.

    Article  PubMed  Google Scholar 

  20. Lin P-Y, Jhan K-J, Ko K-Y, Yang C-C. Investigating the lesion detectability of Tc-99m planar scintigraphy acquired with LEHRS collimator for patients with different body sizes: a phantom study. J Appl Clin Med Phys. 2022;23:e13744. https://doi.org/10.1002/acm2.13744.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Peters SMB, van der Werf NR, Segbers M, van Velden FHP, Wierts R, Blokland KJAK, et al. Towards standardization of absolute SPECT/CT quantification: a multi-center and multi-vendor phantom study. EJNMMI Phys. 2019;6:29. https://doi.org/10.1186/s40658-019-0268-5.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Braune A, Oehme L, Freudenberg R, Hofheinz F, van den Hoff J, Kotzerke J, Hoberück S. Comparison of image quality and spatial resolution between 18F, 68Ga, and 64Cu phantom measurements using a digital Biograph Vision PET/CT. EJNMMI Phys. 2022;9:58. https://doi.org/10.1186/s40658-022-00487-7.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Seltzer S. Tables of X-Ray Mass Attenuation coefficients and Mass Energy-absorption coefficients, NIST Standard Reference Database. Volume 126. National Institute of Standards and Technology; 1995.

  24. Bastiaannet R, van Roekel C, Smits MLJ, Elias SG, van Amsterdam WAC, Doan D, et al. First evidence for a dose-response relationship in patients treated with 166Ho radioembolization: a prospective study. J Nucl Med. 2020;61:608–12. https://doi.org/10.2967/jnumed.119.232751.

    Article  CAS  PubMed  Google Scholar 

  25. van Roekel C, Bastiaannet R, Smits MLJ, Bruijnen RC, Braat AJAT, de Jong HWAM, et al. Dose-Effect relationships of 166Ho Radioembolization in Colorectal Cancer. J Nucl Med. 2021;62:272–9. https://doi.org/10.2967/jnumed.120.243832.

    Article  CAS  PubMed  Google Scholar 

  26. Hermann A-L, Dieudonné A, Ronot M, Sanchez M, Pereira H, Chatellier G, et al. Relationship of Tumor Radiation-absorbed dose to survival and response in Hepatocellular Carcinoma Treated with Transarterial Radioembolization with 90Y in the SARAH Study. Radiology. 2020;296:673–84. https://doi.org/10.1148/radiol.2020191606.

    Article  PubMed  Google Scholar 

  27. Roosen J, Klaassen NJM, Westlund Gotby LEL, Overduin CG, Verheij M, Konijnenberg MW, Nijsen JFW. To 1000 gy and back again: a systematic review on dose-response evaluation in selective internal radiation therapy for primary and secondary liver cancer. Eur J Nucl Med Mol Imaging. 2021;48:3776–90. https://doi.org/10.1007/s00259-021-05340-0.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

D.K., A.B., C.B., R.F., J.K. and E.M. contributed to the design of the study and discussed the data. D.K. planned, performed, and analyzed the measurements. D.K. was a major contributor in writing the manuscript. E.M. is the senior author. All authors discussed the results and implications and commented on the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to David Kästner.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kästner, D., Braune, A., Brogsitter, C. et al. Gamma camera imaging characteristics of 166Ho and 99mTc used in Selective Internal Radiation Therapy. EJNMMI Phys 11, 35 (2024). https://doi.org/10.1186/s40658-024-00633-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-024-00633-3

Keywords