Table 3 Spatial resolution (FWHM), image noise (CVBG), and lung count error (ΔNLung) determined using the NEMA image quality phantom filled with 99mTc and 166Ho at 8:1 contrast with different scan times

From: Gamma camera imaging characteristics of 166Ho and 99mTc used in Selective Internal Radiation Therapy

Phantom setup

scan time per projection

FWHM (mm)

CVBG (%)

ΔNLung (%)

99mTc, 8:1 contrast

5 s

12.8

12.0

22.9

10 s

12.5

8.9

22.5

15 s

12.3

7.8

22.5

20 s

12.3

7.7

22.3

166Ho, 8:1 contrast

10 s

15.4

24.0

33.3

20 s

15.1

18.9

31.9

30 s

15.2

15.2

29.6