Skip to main content

Table 2 Information on the measurements with the NEMA image quality phantom filled with 99mTc and 166Ho in different phantom setups (spheres without background; sphere-to-background activity concentration ratio 8:1 and 4:1)

From: Gamma camera imaging characteristics of 166Ho and 99mTc used in Selective Internal Radiation Therapy

Nuclide

Phantom setup

Activity concentration in spheres (kBq/ml)

Activity concentration in background (kBq/ml)

Sphere-to-background activity concentration ratio

99mTc

spheres without background

698.0

-

-

8:1 contrast

(performed twice)

338.8

419.6

39.6

44.6

8.56:1

7.86:1

4:1 contrast

213.7

52.4

4.08:1

166Ho

spheres without background

1376.0

-

-

8:1 contrast

(performed twice)

719.2

589.5

85.7

74.4

8.39:1

7.92:1

4:1 contrast

386.6

91.5

4.23:1