Table 4 Spatial resolution (FWHM), image noise (CVBG), and lung count error (ΔNLung) determined using the NEMA image quality phantom filled with 99mTc and 166Ho in different phantom setups (spheres without background, contrast 8:1 and 4:1, normal and obese phantom setup) with standard and adapted acquisition parameters (collimator type, reduced projection time). For measurements performed twice, results are presented as the mean and standard deviation

From: Gamma camera imaging characteristics of 166Ho and 99mTc used in Selective Internal Radiation Therapy

Phantom setup

FWHM (mm)

CVBG (%)

ΔNLung (%)

8:1 contrast

99mTc

12.2 ± 0.1

7.6 ± 0.1

23.5 ± 1.2

99mTc obese

12.6

8.4

26.9

99mTc MELP

14.3 ± 0.2

6.8 ± 0.3

26.8 ± 1.4

99mTc MELP 1.5 s

14.3

15.8

23.8

166Ho

15.2 ± 0.1

18.0 ± 0.9

30.9 ± 1.1

166Ho obese

15.6

22.6

33.3

4:1 contrast

99mTc

12.7

7.2

23.1

99mTc MELP

15.4

7.2

24.6

99mTc MELP 1.5 s

15.3

14.5

23.6

166Ho

16.0

17.7

31.8

Spheres without background

99mTc

10.1

-

-

99mTc MELP

12.0

-

-

99mTc MELP 1.5 s

12.2

-

-

166Ho

12.5

-

-