Quantitative SPECT/CT imaging of lead-212: a phantom study

A Correction to this article was published on 10 October 2022

This article has been updated

Abstract

Background

Lead-212 (212Pb) is a promising radionuclide for targeted therapy, as it decays to α-particle emitter bismuth-212 (212Bi) via β-particle emission. This extends the problematic short half-life of 212Bi. In preparation for upcoming clinical trials with 212Pb, the feasibility of quantitative single photon-emission computed tomography/computed tomography (SPECT/CT) imaging of 212Pb was studied, with the purpose to explore the possibility of individualised patient dosimetric estimation.

Results

Both acquisition parameters (combining two different energy windows and two different collimators) and iterative reconstruction parameters (varying the iterations x subsets between 10 × 1, 15 × 1, 30 × 1, 30 × 2, 30 × 3, 30 × 4, and 30 × 30) were investigated to evaluate visual quality and quantitative uncertainties based on phantom images. Calibration factors were determined using a homogeneous phantom and were stable when the total activity imaged exceeded 1 MBq for all the imaging protocols studied, but they increased sharply as the activity decayed below 1 MBq. Both a 20% window centred on 239 keV and a 40% window on 79 keV, with dual scatter windows of 5% and 20%, respectively, could be used. Visual quality at the lowest activity concentrations was improved with the High Energy collimator and the 79 keV energy window. Fractional uncertainty in the activity quantitation, including uncertainties from calibration factors and small volume effects, in spheres of 2.6 ml in the NEMA phantom was 16–21% for all protocols with the 30 × 4 filtered reconstruction except the High Energy collimator with the 239 keV energy window. Quantitative analysis was possible both with and without filters, but the visual quality of the images improved with a filter.

Conclusions

Only minor differences were observed between the imaging protocols which were all determined suitable for quantitative imaging of 212Pb. As uncertainties generally decreased with increasing iterative updates in the reconstruction and recovery curves did not converge with few iterations, a high number of reconstruction updates are recommended for quantitative imaging.

Background

Targeted radioligand therapy consists of selectively delivering radiation to cancer cells using targeting agents and has been proven a successful treatment for several cancers. Currently, beta (β)-particle emitters are the main radionuclides used; however, using alpha (α)-particle emitters in targeted alpha therapy (TAT) presents an appealing alternative. Compared to β-particles, α-particles have higher energies, shorter path lengths, and much higher probabilities of DNA double-strand breaks when interacting with cell nuclei [1, 2]. These are all properties desirable for targeting small tumours in metastasised cancer as they allow for high absorbed doses to cancerous tissue while avoiding toxicity in surrounding tissue. The cytotoxicity of α-particles is approximately 500 times higher than that of β-particles and a high level of accuracy in delivering the α-radiation is essential to avoid damage to non-targeted tissues [3]. This accuracy may be assessed with imaging, but imaging of α-particle emitters is difficult due to few photons of appropriate energies being emitted at the relevant activity levels.

Several α-particle emitters have already been studied in patients and many of them have been imaged. Radium-223 (223Ra) is an α-particle emitter currently used in the form of 223Ra-dichloride to treat bone metastases, and it is the most studied α-particle emitter in radiotherapy [4,5,6,7]. Quantitative planar scintigraphy and single photon-emission computed tomography (SPECT) imaging have been investigated for 223Ra and were found to be potentially useful tools for dosimetry [4]. α-particle therapies have been proposed for a variety of cancers with different targeting agents, and an overview of TATs on the market or in development was provided recently by Sgouros et al. [8]. Imaging of several of the α-particle emitting radionuclides has been shown to be feasible; for instance, thorium-227 (227Th) and actinium-225 (225Ac) can be imaged with planar scintigraphy [5, 6, 9,10,11,12,13] and SPECT [11,12,13,14]. Planar scintigraphy and SPECT scans of the liver and brain have been performed with bismuth-213 (213Bi) [14,15,16], while terbium-149 (149Tb) can be imaged with positron emission tomography (PET) [17]. 212Pb functions as an in vivo α-particle emitter generator as it decays to bismuth-212 (212Bi) with a half-life of 10.64 h. As 212Pb decays to 212Bi, imageable gammas of 238.6 keV, with an intensity of 43.3%, and X-rays (75–91 keV), with a total intensity of 36%, are emitted [18]. Planar scintigraphy images of 212Pb were obtained by Meredith et al. in a clinical study [19] and by Kasten et al. in a preclinical study [20]. Surrogate radiopharmaceuticals with more appropriate photon emissions are sometimes used [21], and lead-203 (203Pb) has been investigated as a surrogate isotope for 212Pb [22, 23]. As there are photon emissions in the 212Pb decay chain that are imageable, we here investigate the possibilities for direct quantitative SPECT imaging of 212Pb.

There are several good targets for targeted radioligand therapy using 212Pb. Prostate-specific membrane antigen (PSMA) TATs could be an improvement on current treatments available for patients with castration-resistant metastatic prostate cancer, and 212Pb is an advantageous choice of radionuclide due to the industrial-scale production methods available [24]. Encouraging preclinical results with PSMA-targeting 212Pb-NG001 show lower kidney uptake than 212Pb-labelled PSMA-617 in addition to good tumour uptake and inhibited tumour growth [25, 26], and a clinical trial is being planned. The current imaging study has been conducted using a Siemens Symbia Intevo Bold to prepare for imaging and dosimetry of patients injected with 212Pb-targeted radiotherapy. Collimators, energy windows, and reconstruction parameters have been compared using various phantoms. The aim of this study was to determine the feasibility of and to find the best protocol for quantitative SPECT/computed tomography (CT) imaging of 212Pb.

Materials and methods

Decay characteristics

212Pb decays via β-particle emission to α-particle emitter bismuth-212 (212Bi) with a half-life of 10.64 h. In the same transition, most of the photons in the decay chain are emitted, most notably one of 238.6 keV with an intensity of 43.3%. X-rays (75–91 keV) with a total intensity of 36% are also emitted during this transition. A high-energy photon emission of 2614.5 keV is emitted by daughter thallium-208 (208Tl) with 36% intensity per 212Pb decay and creates scatter which may also contribute to the SPECT energy spectrum [18]. The decay scheme and emitted photon, alpha, and beta energies are presented in Fig. 1.

Fig. 1
figure 1

Decay scheme of 212Pb. The mean alpha and beta energies are included. The relevant photons emitted and their energies are added to the transition. Photons with emission probabilities smaller than 1% per 212Pb decay or with energies below 70 keV are not included. The data are taken from ICRP 107 [18]

Imaging protocol

Four 30-min imaging protocols were examined on a Siemens Symbia Intevo Bold SPECT/CT with a 3/8″ crystal. Two collimators, High Energy (HE) and Medium Energy (ME), and two energy windows, 40% at 79 keV and 20% at 239 keV, were combined in the protocols. Dual scatter windows of 20% for the 79 keV peak centred on 55 keV and 103 keV and 5% for the 239 keV peak centred on 209 keV and 268 keV were used for scatter correction. The energy and scatter windows were chosen based on the energy spectra shown in Fig. 2. The SPECT images were acquired with body contouring orbits, a 256 × 256 matrix, and 60 views with acquisition during steps. The images were reconstructed with Flash-3D, a commercial OSEM reconstruction algorithm available on the Siemens SPECT, which includes depth-dependent 3D resolution recovery with Gaussian point spread function correction [27]. Reconstruction updates are here used for number of iterations multiplied with number of subsets, and images were reconstructed with seven different numbers of updates (iterations x subsets): 10 × 1, 15 × 1, 30 × 1, 30 × 2, 30 × 3, 30 × 4, and 30 × 30. Images were acquired separately for the two energy windows, and attenuation correction for the central energy based on the CT images was performed. Both images with a 12-mm Gaussian filter and without filters were analysed. These imaging protocols were used for all the experiments.

Fig. 2
figure 2

Energy spectra acquired on the SPECT scanner with a 212Pb source. The two investigated energy windows and their adjacent scatter windows are also illustrated. The spectrum with the ME collimator is shown in purple, the spectrum with the HE collimator is shown in light blue, and the spectrum acquired without a collimator is shown in dark blue

Calibration factors

Calibration factors (CFs) were determined by imaging a 6283 ml uniform cylindrical phantom filled with 0.5–8.5 MBq 212Pb eleven times. In Slicer (v4, http://www.slicer.org) [28], a cylindrical volume of interest (VOI) of approximately 4000 ml was placed well within the boundaries of the phantom to avoid contributions from edge effects. The mean counts per voxel was extracted and divided by the volume of a voxel to give mean counts per ml, \(\overline{C}_{{{\text{mean}}}}\), and the CF was calculated with

$$CF = \frac{{\overline{C}_{{{\text{mean}}}} }}{{t \times \overline{A}}} ,$$
(1)

where t is the scan duration and \(\overline{A}\) is the activity per ml.

Recovery coefficients and contrast

The National Electrical Manufacturers Association (NEMA) International Electrotechnical Commission (IEC) PET Body Phantom was used to find mean and maximum recovery coefficients, RCmean and RCmax, and to determine the contrast achieved with the different imaging protocols. The sphere volumes were 26.52, 11.49, 5.57, 2.57, 1.15, and 0.52 ml. The phantom was filled twice, with 212Pb in the spheres and non-radioactive water in the background. For each imaging protocol, the phantom was imaged with eight different activity concentrations in the spheres, between 17.2 kBq/ml and 104.2 kBq/ml, that is a total of 0.8–5.0 MBq. The maximum and mean counts per ml, \(\overline{C}_{{{\text{max}}}}\) and \(\overline{C}_{{{\text{mean}}}}\), in each sphere were obtained by placing spherical VOIs of the physical sphere volume using the CT images in Slicer, extracting maximum and mean counts per voxel, and dividing by voxel size. The mean counts per ml in the background, \(\overline{C}_{{{\text{background}}}}\), resulting from scatter and spill-out, were calculated by placing a cylinder VOI of radius 20 mm and height 18 mm in the volume between the spheres.

The data points for the recovery coefficient fit were calculated using

$$RC = \frac{{\left( {\overline{C} - \overline{C}_{{{\text{background}}}} } \right)}}{{CF \times t \times \overline{A}_{{\text{true, sph}}} }},$$
(2)

where t is the scan duration and \(\overline{A}_{{\text{true, sph}}}\) is the known, physical activity concentration in the sphere. RC is the recovery coefficient which is RCmax if \(\overline{C}_{{{\text{max}}}}\) is used for \(\overline{C}\) and RCmean if \(\overline{C}_{{{\text{mean}}}}\) is used for \(\overline{C}\). The uncertainties on the activity quantitation of the volumes were propagated from the standard deviations (SDs) of \(\overline{C}_{{\text{mean, max}}}\) and CFs. Fractional uncertainties were taken as this uncertainty divided by the average of the eight RCmean, max for each volume. The recovery coefficient data points were fitted to the function RCfit with

$$RC_{{{\text{fit}}}} = 1 - \frac{1}{{1 + \frac{v}{{b_{1} }}^{{b_{2} }} }},$$
(3)

where v is volume and b1 and b2 are fitting parameters [29]. The fit parameters and associated standard errors (SE) were calculated using MATLAB (R2017a, MathWorks). The eight RCmean data points for each volume were included in the fit, not the average used for fractional uncertainties.

The contrast was calculated as

$${\text{Contrast}} = \frac{{\overline{C}_{{{\text{mean}}}} - \overline{C}_{{{\text{background}}}} }}{{\overline{C}_{{{\text{mean}}}} }},$$
(4)

where the symbols are the same as defined above. The errors on contrast were taken as the SDs of measurements at different activity levels.

Results

The calibration factors calculated for the four imaging protocols with a filter are presented in Fig. 3. Corresponding results for the unfiltered images can be found in Additional file 1. The CFs stayed approximately constant for all protocols when more than 1 MBq (0.16 kBq/ml) 212Pb was imaged, but increased sharply below 1 MBq. Due to this, only data points obtained with more activity than 1 MBq were used when calculating the average CFs. The count rate was highest with the ME collimator and the 79 keV window, giving a CFME,79 keV = 0.51 ± 0.07 cps/kBq when taking the mean and SD of all the values obtained with activities higher than 1 MBq. The count rate decreased when the HE collimator was used with the same energy window, giving CFHE,79 keV = 0.34 ± 0.04 cps/kBq. As can be seen from Fig. 3a, the 239 keV window gave the lowest count rate and little difference between the collimators, giving CFME,239 keV = 0.08 ± 0.01 cps/kBq and CFHE,239 keV = 0.08 ± 0.02 cps/kBq. The CFs for the reconstructions without a filter and the other numbers of updates are given in Additional file 1 and show very similar values. As seen in Fig. 3a, there was little variation in CFs between reconstructions except a dip for the 10 × 1 and 15 × 1 reconstructions. The dip was most significant for image protocols using the 79 keV peak and largest for the ME collimator. The ME 79 keV was the image protocol which most consistently had the lowest coefficients of variation (CVs) on the CFs (Fig. 3b). This imaging protocol also had approximately constant CVs with reconstruction updates, as they stayed between 11 and 15% for all reconstructions when a filter was applied. The CVs for HE 79 keV only deviated from the ME 79 keV CVs level for 10 × 1 and 15 × 1, with 27% and 18%, respectively. The 239 keV window caused larger CVs, but while for the HE collimator the CVs varied between 20 and 39%, the ME collimator gave a 66% CV for 10 × 1 which steadily decreased with increasing iteration updates to 10% for 30 × 30. These trends were similar for the reconstructions without a filter.

Fig. 3
figure 3

Results for the CFs for the reconstructions with a filter applied with 212Pb. In panel a, the CFs are plotted for different reconstructions with error bars showing the SDs of the CFs calculated at activities above 1 MBq. In panel b, the plot shows CVs of the CFs plotted for different reconstructions. The corresponding results for images without a filter can be found in the supplementary material. In panel c, the data points from which the CFs were calculated for the 30 × 4 reconstruction are plotted against activity for all four image protocols with a filter applied. The sharp increase in CFs as activity decays below 1 MBq can be seen in the plot. In panel d, high activity maximum intensity projection images of the homogeneous phantom are presented for each of the imaging protocols with a 12-mm filter and the 30 × 4 reconstruction. The scale is set the same for all the images, to illustrate the different count rates, with the white set to 25 counts and the black being 0 counts

Using the calculated CFs, the measured mean counts in the six spheres of the NEMA phantom were converted to activity concentrations. The measured activity concentrations were divided by the known physical activity concentrations for the six spheres at different time points and plotted against activity concentration. The results for the 30 × 4 reconstruction with a filter applied are presented in Fig. 4, and the equivalent plots for the other reconstructions are included in Additional file 1.

Fig. 4
figure 4

Activity in each sphere measured from the mean counts in the images divided by the known, physical activity in each sphere plotted against activity concentration. The figure shows the results obtained with the filtered 30 × 4 reconstruction, with ME 239 keV in panel a, ME 79 keV in panel b, HE 239 keV in panel c, and HE 79 keV in panel d. Equivalent figures for the other reconstructions with and without filters applied can be found in the supplementary material. The mean activity ratio for the sphere volume is shown as a line. Of note, the x-axes are the same, but the values on the y-axes differ

Recovery coefficient (RC) curves are shown for the 30 × 4 reconstruction with a filter applied in Fig. 5. Both RCmean and RCmax values are plotted, and the curves fit to RCmean are also shown. The intensity spread, used here for all effects that in an image contribute to spread of intensity around a source such as spill-out and partial volume effects, appeared to be more significant for the 79 keV window. The curve fit parameters and the associated SE for the 30 × 4 reconstruction with a filter applied are presented in Table 1. The fractional SE on the curve fit parameters were smaller than or equal to 16% for all protocols with the 30 × 4 reconstruction. The fractional SE were smaller for the 239 keV window than the 79 keV window for both collimators. Corresponding figures and tables can be found for the other reconstructions, with and without a filter applied, in the supplementary material. The fractional SE were similar for the other reconstructions, except for the 30 × 30 reconstruction without a filter applied which had fractional SE exceeding 50% on b1 with the 79 keV window. The fitted functions were different with fewer reconstruction updates, as the steepness of the curve at small volumes increased with the number of updates. With few reconstruction updates, the recovery curves did not converge. Without a filter applied, there was more variation in RCmax values, especially with high numbers of updates.

Fig. 5
figure 5

RCs and curve fits for the four image protocols with the 30 × 4 reconstruction and a 12-mm Gaussian filter. The equivalent figures for the other reconstructions can be found in the supplementary material. ME 239 keV is in panel a, ME 79 keV is in panel b, HE 239 keV is in panel c, and HE 79 keV is in panel d. Both RCmax and RCmean data points are plotted, in addition to a curve fit to RCmean

Table 1 Fitting parameters for the RC curve for the 30 × 4 reconstruction with a 12-mm Gaussian filter

The fractional uncertainties on the activity quantitation in the spheres were taken as the combined uncertainties from the CFs and measured counts in the spheres, divided by the average RC for the volume, and they are plotted against reconstruction updates for different volumes in Fig. 6. The fractional uncertainties were similar for ME 239 keV, ME 79 keV, and HE 79 keV with the three reconstructions with most iteration updates. The HE 239 keV protocol consistently gave larger fractional uncertainties. With fewer reconstruction updates than 30 × 3, the fractional uncertainties of both HE 239 keV and ME 239 keV were markedly larger than those of the 79 keV window. With the 30 × 4 reconstruction, the fractional uncertainty on activity quantitation when using maximum counts for the 2.57 ml sphere was 15% for the ME 79 keV protocol and 18% for the HE 79 keV protocol with the 30 × 4 reconstruction. For activity quantitation with mean counts, with the same sphere and reconstruction, the fractional uncertainty was 16% for ME 79 keV and 18% for HE 79 keV. 2.57 ml was the smallest sphere for which activity quantitation with fractional uncertainties below 20% was possible with 212Pb. The corresponding plots for reconstructions without a filter applied are included in Additional file 1. The trends and values were similar with and without a filter, but without a filter the fractional uncertainties for activity quantitation using maximum counts with the 30 × 30 reconstruction increased for all volumes.

Fig. 6
figure 6

Fractional uncertainties for activity quantitation in spheres plotted against reconstruction updates for different sphere volumes. Results from using the mean counts are on the left (ae), and results from using maximum counts are on the right (fj). The smallest sphere (0.52 ml) is not included as all the uncertainties exceeded 60%

Some of the filtered reconstructed images of the NEMA phantom filled with the highest imaged concentration of 212Pb, 88–100 kBq/ml, are shown in Fig. 7. The smallest clearly visible sphere was the 13-mm-diameter sphere for all protocols, except the 30 × 30 reconstruction with ME 239 keV which also clearly showed the 10-mm sphere. One could make out the 10-mm sphere with the 79 keV window with the 30 × 4 reconstruction as well, but it was not clearly defined above the noise. The images appear to converge with fewer reconstruction updates for the 79 keV energy window.

Fig. 7
figure 7

Some of the filtered reconstructed images of the NEMA phantom filled with 88–100 kBq/ml 212Pb for all the imaging protocols, but the maximum counts illustrated as white differs and is given in the top right corner of each image. Reconstructions 15 × 1, 30 × 4, and 30 × 30 were chosen as examples to illustrate the change with reconstruction updates

Selected filtered reconstructed images of the NEMA phantom with a low concentration of 212Pb, 17–19 kBq/ml, are shown in Fig. 8. The noise level increased compared to the high activity images for all protocols. The 30 × 30 reconstruction contained the most noise for all protocols, and it was difficult to distinguish true activity regions from noise. For the 30 × 4 reconstruction, there was also significant noise, but for the HE 79 keV images five of the spheres could be distinguished. The ME 239 keV images had the least noise and four spheres could be visualised. With the 15 × 1 reconstruction, only two spheres were visible on the ME 239 keV images and three spheres were visible on the HE 239 keV images. Four and five spheres were visualised on the ME and HE 79 keV images, respectively, again indicating that the images converged with fewer reconstruction updates for the 79 keV window. The equivalent images without a filter can be found in Additional file 1.

Fig. 8
figure 8

Selected filtered reconstructed images of the NEMA phantom filled with 17–19 kBq/ml 212Pb shown as maximum intensity projections. The scale bar is the same for all images, but the maximum counts illustrated in white differs and is given in the top right corner of each image

In Fig. 9, contrast is shown for the four imaging protocols for all filtered reconstructions. The error bars show SDs of measurements repeated at different activity concentrations. As sphere volumes decreased, reconstructions with few reconstruction updates gave poorer contrasts. For all four protocols, the contrast was close to one for the four largest spheres. For the two smallest spheres, the contrast became more variable and error bars became notably larger.

Fig. 9
figure 9

Contrast plotted against sphere diameter for all the image protocols for the different reconstructions with a 12-mm Gaussian filter. Contrast was calculated with Eq. 4, except for when the source VOI included no counts. In cases with no counts in the source VOI, contrast was set to 0 to find averages and SDs. The equivalent plots based on unfiltered images can be found in the supplementary material. The ME 239 keV results are shown in panel a, the ME 79 keV results are shown in panel b, the HE 239 keV results are shown in panel c, and the HE 79 keV results are shown in panel d

Discussion

The CFs for 212Pb were stable when the total imaged activity exceeded 1 MBq (0.16 kBq/ml) for all four imaging protocols studied with an acquisition time of 30 min. Additionally, it was found that the activity in volumes of 2.6 ml could be quantified with a fractional uncertainty around 20% with 212Pb with all protocols except HE 239 keV. This uncertainty includes contributions from the calibration factors and from small volume effects, but other sources of uncertainty, such as from a dose calibrator, were not included. All the imaging protocols had a large increase in fractional uncertainty for the two smallest volumes, and volumes of 1.15 ml were considered too small for reliable activity quantitation with 212Pb. When evaluating the visual quality of the images, the 79 keV window gave images which converged with fewer reconstruction updates, but the images were noisier than the 239 keV window images. The HE 79 keV stood out as the visually best protocol due to more visible spheres in the images with a lower activity concentration. There were only small differences in the quantitative uncertainties and visual quality of the protocols, which all showed promise for clinical imaging.

The limit of quantitative ability was determined to be 1 MBq (or 0.16 kBq/ml), based on the CFs no longer being stable. This limit might be lower if the acquisition time is increased. While the limit is based on the total activity in the field of view, it is not a hard limit and smaller total activities with much higher activity concentrations can still be quantified. This can be seen from Fig. 4, where the lowest activity concentration is 17.2 kBq/ml, corresponding to a total activity of 0.8 MBq, but the data points still lie close to the mean. Depending on the treatment (amount of activity administered, biodistribution, biological half-life, etc.), this lower limit can be used to estimate the latest feasible imaging time points in upcoming trial protocols. For example, for the PSMA-targeting 212Pb-NG001, if it has a similar biological half-life as PSMA-617, the effective half-life of 212Pb-NG001 should be around 8.7 h. The effective half-life with [lutetium-177 (177Lu)]Lu-PSMA-617 was calculated to be between 35 and 40 h for most patients [30], giving a biological half-life of around 49 h. As clinically relevant administered activity is likely to be 60–80 MBq [25], there will probably be around 1 MBq 212Pb-NG001 left in the patient after 48 h. However, since close to 1 MBq is probably needed in the camera field of view to obtain reliable quantitative results and the activity will be distributed throughout the body, the final imaging time point should be less than 48 h post-injection.

As stable quantitative results can be obtained with around 1 MBq in the field of view, spatial resolution is likely a larger issue than sensitivity, as the primary target for most 212Pb-based therapies will be micrometastases. If the tumours are smaller than 2.6 ml, one cannot expect quantitative tumour dosimetry based on these results. However, as long as the volume is larger than 2.6 ml, it should be possible to calculate the activity in organs at risk and larger tumours with a reasonable uncertainty of around 20% from the CFs and small volume effects. The uncertainty did not decrease with increasing volume, and an uncertainty around 20% should be expected for all volumes larger than 2.6 ml. From Fig. 4, it could, however, be seen that the variation in measured to known activity in the spheres increased at lower activity concentrations. Likely, greater accuracy than 20% can be achieved for VOIs with high activity. Performing dosimetry on these volumes is therefore feasible, but factors such as daughter nuclides released from the carrier molecule, the complex dose dependence of relative biological effectiveness, and non-uniform radiation at the micro- and multicellular levels will also add uncertainty to the dosimetry calculations [31,32,33,34]. The fractional uncertainties on activity quantitation in spheres were very similar between using mean counts (on the left of Fig. 6) and maximum counts (on the right of Fig. 6), but these results were obtained with the same activity concentration in the entire spherical volume. This is rarely the case in patient tumours and the uncertainties associated with a maximum rather than a mean value are expected to increase.

In Fig. 8, 17–19 kBq/ml is referred to as a low activity concentration. In terms of quantitative imaging, this is a low activity concentration, as it corresponds to a total activity of 0.8 MBq in the phantom, which is just below the limit for stable CFs. 17–19 kBq/ml is likely also a low activity concentration in terms of tumour concentration. For example, for [177Lu]Lu-PSMA-617 it has been shown that a typical tumour reaches its maximum activity concentration around 10 h post-injection [35]. Correcting for physical half-lives and assuming similar pharmacokinetic properties of the radiotherapeutics when replacing 177Lu with 212Pb, this would give an activity concentration around 32 kBq/ml with 212Pb after a 100 MBq injection. While there are possibly large variations, low activity images in Fig. 8 might still give an indication of the visibility of a typical tumour at the later imaging time points.

Whether monitoring dose to organs at risk is more important than tumour dosimetry is debatable. Quantitation of tumour absorbed doses are very useful for dose–effect estimates and to avoid undertreating, but doses to organs at risk are important to ensure the safety of the radiopharmaceutical. Kidneys are considered one of the main organs at risk for many radionuclide therapies including 212Pb-NG001 [25] and clinical studies with other PSMA-targeting radiopharmaceuticals have also shown salivary glands, liver, and bone marrow as normal tissues at risk [35,36,37]. Kidney dosimetry should be achievable with 212Pb, but potential non-uniform uptake in the kidneys will be difficult to visualise and quantify. Localisation of the radiopharmaceutical in radiosensitive subregions of the kidneys might give whole organ toxicity that is poorly related to mean absorbed kidney dose [38]. This will be difficult to predict solely on the 212Pb images, and therefore preclinical data or imaging of surrogate diagnostic compounds in combination with pharmacokinetic models should be used for dosimetry calculations on that level [34, 39].

A minimised uncertainty on the activity quantitation was the primary goal when comparing imaging protocols. For instance, the ME 239 keV protocol had the highest activity recovery and was hence the most resilient against intensity spread, but since this can be accounted for with RCs this characteristic was deemed less substantial. The uncertainties of the curve fits are hence a more relevant property for comparison. Importantly, the uncertainties given in Table 1 reflect the appropriateness of the fitted function, not the quantitative ability of the protocol. The fractional errors on the fitting parameters were smaller for the 239 keV window, but the fractional uncertainties on activity quantitation in spheres were smaller for the 79 keV window. After inspecting the RC curves, it is not surprising that the fits for the 79 keV window are poorer, as the fitted curve seems to reach towards 1 while the data points to a larger extent flatten out. Attenuation plays a larger role for 79 keV photons than for 239 keV photons and is likely what causes the lower activity recovery seen for 79 keV. The photons lost to attenuation in the NEMA phantom have not been accounted for with the CFs, since they were calculated from homogeneously distributed 212Pb. When adding b3 as a third fitting parameter to the RC function,

$$RC_{{{\text{fit}}}} = b_{3} - \frac{1}{{1 + \frac{v}{{b_{1} }}^{{b_{2} }} }},$$
(5)

b3 was less than 1 for all the filtered reconstructions, expect 10 × 1, for all four imaging protocols. For the reconstructions 30 × 2, 30 × 3, and 30 × 4, 1 was more than two SE from b3. Hence, it might be inappropriate to force the RC function to 1. An additional challenge posed by increased attenuation of the lower energy photons is that although the NEMA phantom to some degree simulates a body, it would correspond to a small patient. Thus, it is possible, with the small differences observed, that the 239 keV window would give better quantitative results if the circumference of the NEMA phantom was larger.

Since the quantitative ability was similar between the protocols, visual quality more strongly impacts the choice for clinical imaging. No activity was added to the background when imaging the NEMA phantom, based on the assumption that a high tumour to background contrast will be seen in patients treated with 212Pb-NG001, as has been observed in imaging studies with gallium-68-labelled PSMA [40]. For both energy windows, false positives appeared with increased numbers of updates, which can be regarded more concerning than the noise seen in the images. This may suggest making reconstructions tailored to the patient-specific activity concentration and distribution. A large amount of statistical noise was expected in 212Pb images due to the limited number of photons and the large amount of scatter [41] from the 208Tl emission of 2.6 MeV. Filters compensate for the noise from scatter and low count statistics; the filtered 212Pb images being of higher visual quality were therefore expected. Since the quantitative uncertainties were similar in the filtered and unfiltered images, filtering is recommended. Also, since no large quantitative differences were found, only a 12-mm filter was investigated in this study and the filter strength was not optimised. Likely different reconstructions benefit from different filters. In general, the uncertainties found decreased with increasing number of updates. This, in addition to recovery curves not converging with few iterative updates encourages a higher number of reconstruction updates than what is typical in the clinic. If the same images are used for quantitation and visual interpretation, this might also motivate stronger filters.

The question of whether 212Pb can be quantitatively imaged was investigated in this work. As far as we know, no previously published article has undertaken this problem, but an abstract was published in 2019 which compared imaging of one sphere of 212Pb with an ME collimator to simulations [42]. The authors concluded that imaging of 212Pb could be feasible, but features such as intensity spread in small volumes were not discussed. However, quantitative imaging of α-particle emitter 223Ra has been extensively studied. With planar images a 200 ml volume was quantified with 10% uncertainty and a 0.5 ml volume with 40% uncertainty [4]. A SPECT study with 223Ra used an ME collimator, a 5/8″ crystal, 2 iterations and 10 subsets with a Butterworth filter and added three energy windows (40% centred on 85 keV, 20% on 154 keV, and 20% on 270 keV). Similarly to our findings with 212Pb, they achieved quantitation of 5.6 ml volumes with an error smaller than 19% [43]. The count rate achieved with 212Pb with the 79 keV energy window was much higher than when imaging 223Ra [4, 6, 13, 43]. However, a lot of photons contributing to the 79 keV energy peak are characteristic X-rays from the lead collimator and scattered photons from the 2.6 MeV emission of 208Tl, and hence it would be more relevant to report the number of primary photons detected to compare sensitivity. Still, almost 25 times more activity is expected to be administered with 212Pb compared to 223Ra therapy, and it is therefore expected that images of reasonable quality can be obtained. Count rates with 227Th and 225Ac were lower than with 223Ra and hence also lower than with 212Pb [13]. Other studies have attempted to quantify 227Th and 223Ra in combination with planar imaging and found that 227Th can be separated from 223Ra in the images. Spatial filtering improved the images visually, similarly to 212Pb [5]. Using a spectral analysis technique modelling energy spectra and scattered photons, Murray et al. calculated that the average difference between known and estimated activity was 5.1% for 227Th and 3.4% for 223Ra, but differences of 50% were observed [6]. However, even with measurements differing from the known activity with 50%, they obtained differences smaller than 10% when calculating the time integrated activity. Hence, they found that for dosimetry, “uncertainty in individual data measurements may be mitigated by carrying out multiple measurements over several time points” [6]. This is likely also applicable to 212Pb, but will require more imaging time points than usually acquired for clinical dosimetry.

Conclusion

In conclusion, calibration factors for quantitative imaging of 212Pb were stable when the total imaged activity exceeded 1 MBq. All imaging protocols studied showed promise for quantitative imaging. For example, for three of the four protocols the fractional uncertainties for the activity quantitation in the 2.6-ml sphere ranged from 16 to 21% for the 30 × 4 filtered reconstruction, accounting for uncertainties introduced by the calibration factors and intensity spread in small volumes. Regardless of the protocol used, the lack of convergence of recovery curves for few reconstruction iterations and decreasing CVs with increasing reconstruction updates encourage reconstructions with a high number of updates. Visual quality at low activity concentrations was better with the 79 keV peak and HE collimator for the geometry investigated, a phantom with spheres. Overall, the results indicate that it could be possible to perform patient-specific dosimetry with SPECT/CT imaging.

Availability of data and materials

Freely available upon request.

Change history

Abbreviations

α:

Alpha

β:

Beta

212Pb:

Lead-212

223Ra:

Radium-223

212Bi:

Bismuth-212

208Tl:

Thallium-208

227Th:

Thorium-227

213Bi:

Bismuth-213

225Ac:

Actinium-225

149Tb:

Terbium-149

203Pb:

Lead-203

177Lu:

Lutetium-177

SPECT:

Single photon-emission computed tomography

CT:

Computed tomography

PET:

Positron emission tomography

CF:

Calibration factor

RC:

Recovery coefficient

CV:

Coefficient of variation

PSMA:

Prostate-specific membrane antigen

TAT:

Targeted alpha therapy

HE:

High energy

ME:

Medium energy

NEMA:

National Electrical Manufacturers Association

IEC:

International Electrotechnical Commision

VOI:

Volume of interest

SD:

Standard deviation

SE:

Standard error

References

  1. Haberkorn U, Giesel F, Morgenstern A, Kratochwil C. The future of radioligand therapy: α, β, or both? J Nucl Med. 2017;58(7):1017–8.

    Article  PubMed  CAS  Google Scholar 

  2. King AP, Lin FI, Escorcia FE. Why bother with alpha particles? Eur J Nucl Med Mol Imaging. 2021;49(1):7–17.

    Article  PubMed  Google Scholar 

  3. Riihimäki M, Hemminki A, Sundquist K, Sundquist J, Hemminki K. The epidemiology of metastases in neuroendocrine tumors. Int J Cancer. 2016;139(12):2679–86.

    Article  PubMed  CAS  Google Scholar 

  4. Hindorf C, Chittenden S, Aksnes AK, Parker C, Flux GD. Quantitative imaging of 223Ra-chloride (Alpharadin) for targeted alpha-emitting radionuclide therapy of bone metastases. Nucl Med Commun. 2012;33(7):726–32.

    Article  PubMed  CAS  Google Scholar 

  5. Larsson E, Brolin G, Cleton A, Ohlsson T, Lindén O, Hindorf C. Feasibility of thorium-227/radium-223 gamma-camera imaging during radionuclide therapy. Cancer Biother Radiopharm. 2020;35(7):540–8.

    PubMed  CAS  Google Scholar 

  6. Murray I, Rojas B, Gear J, Callister R, Cleton A, Flux GD. Quantitative dual-isotope planar imaging of thorium-227 and radium-223 using defined energy windows. Cancer Biother Radiopharm. 2020;35(7):530–9.

    PubMed  PubMed Central  CAS  Google Scholar 

  7. Pacilio M, Ventroni G, De Vincentis G, Cassano B, Pellegrini R, Di Castro E, et al. Dosimetry of bone metastases in targeted radionuclide therapy with alpha-emitting (223)Ra-dichloride. Eur J Nucl Med Mol Imaging. 2016;43(1):21–33.

    Article  PubMed  CAS  Google Scholar 

  8. Sgouros G, Bodei L, McDevitt MR, Nedrow JR. Radiopharmaceutical therapy in cancer: clinical advances and challenges. Nat Rev Drug Discov. 2020;19(9):589–608.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  9. Usmani S, Rasheed R, Alkandari F, Marafi F, Naqvi SAR. 225Ac prostate-specific membrane antigen posttherapy α imaging: comparing 2 and 3 photopeaks. Clin Nuclear Med. 2019;44(5):558.

    Article  Google Scholar 

  10. Ocak M, Toklu T, Demirci E, Selçuk N, Kabasakal L. Post-therapy imaging of (225)Ac-DOTATATE treatment in a patient with recurrent neuroendocrine tumor. Eur J Nucl Med Mol Imaging. 2020;47(11):2711–2.

    Article  PubMed  Google Scholar 

  11. Gosewisch A, Schleske M, Gildehaus FJ, Berg I, Kaiser L, Brosch J, et al. Image-based dosimetry for 225Ac-PSMA-I&T therapy using quantitative SPECT. Eur J Nucl Med Mol Imaging. 2021;48(4):1260–1.

    Article  PubMed  CAS  Google Scholar 

  12. Kamaleshwaran KK, Suneelkumar M, Madhusairam R, Radhakrishnan EK, Arunpandiyan S, Arnold VJ. Whole-body and single-photon emission computed tomography/computed tomography postpeptide receptor alpha radionuclide therapy images of actinium 225-tetraazacyclododecanetetraacetic acid-octreotide as a primary modality of treatment in a patient with advanced rectal neuroendocrine tumor with metastases. Indian J Nucl Med. 2020;35(3):226–8.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Benabdallah N, Scheve W, Dunn N, Silvestros D, Schelker P, Abou D, et al. Practical considerations for quantitative clinical SPECT/CT imaging of alpha particle emitting radioisotopes. Theranostics. 2021;11(20):9721–37.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  14. Sgouros G, Ballangrud AM, Jurcic JG, McDevitt MR, Humm JL, Erdi YE, et al. Pharmacokinetics and dosimetry of an alpha-particle emitter labeled antibody: 213Bi-HuM195 (anti-CD33) in patients with leukemia. J Nucl Med. 1999;40(11):1935–46.

    PubMed  CAS  Google Scholar 

  15. Kratochwil C, Giesel FL, Bruchertseifer F, Mier W, Apostolidis C, Boll R, et al. 213Bi-DOTATOC receptor-targeted alpha-radionuclide therapy induces remission in neuroendocrine tumours refractory to beta radiation: a first-in-human experience. Eur J Nucl Med Mol Imaging. 2014;41(11):2106–19.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  16. Cordier D, Forrer F, Bruchertseifer F, Morgenstern A, Apostolidis C, Good S, et al. Targeted alpha-radionuclide therapy of functionally critically located gliomas with 213Bi-DOTA-[Thi8, Met(O2)11]-substance P: a pilot trial. Eur J Nucl Med Mol Imaging. 2010;37(7):1335–44.

    Article  PubMed  CAS  Google Scholar 

  17. Müller C, Vermeulen C, Köster U, Johnston K, Türler A, Schibli R, et al. Alpha-PET with terbium-149: evidence and perspectives for radiotheragnostics. EJNMMI Radiopharm Chem. 2016;1(1):5.

    Article  PubMed  PubMed Central  Google Scholar 

  18. ICRP. Nuclear Decay Data for Dosimetric Calculations. ICRP Publication 107 Annals of the ICRP. 2008;38(3).

  19. Meredith RF, Torgue J, Azure MT, Shen S, Saddekni S, Banaga E, et al. Pharmacokinetics and imaging of 212Pb-TCMC-trastuzumab after intraperitoneal administration in ovarian cancer patients. Cancer Biother Radiopharm. 2014;29(1):12–7.

    PubMed  PubMed Central  CAS  Google Scholar 

  20. Kasten BB, Azure MT, Schoeb TR, Fisher DR, Zinn KR. Imaging, biodistribution, and toxicology evaluation of 212Pb-TCMC-trastuzumab in nonhuman primates. Nucl Med Biol. 2016;43(7):391–6.

    Article  PubMed  CAS  Google Scholar 

  21. Seo Y. Quantitative imaging of alpha-emitting therapeutic radiopharmaceuticals. Nucl Med Mol Imaging. 2019;53(3):182–8.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  22. Máthé D, Szigeti K, Hegedűs N, Horváth I, Veres DS, Kovács B, et al. Production and in vivo imaging of (203)Pb as a surrogate isotope for in vivo (212)Pb internal absorbed dose studies. Appl Radiat Isot. 2016;114:1–6.

    Article  PubMed  CAS  Google Scholar 

  23. Dos Santos JC, Schäfer M, Bauder-Wüst U, Lehnert W, Leotta K, Morgenstern A, et al. Development and dosimetry of (203)Pb/(212)Pb-labelled PSMA ligands: bringing “the lead” into PSMA-targeted alpha therapy? Eur J Nucl Med Mol Imaging. 2019;46(5):1081–91.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  24. Juzeniene A, Stenberg VY, Bruland ØS, Larsen RH. Preclinical and clinical status of PSMA-targeted alpha therapy for metastatic castration-resistant prostate cancer. Cancers (Basel). 2021;13(4):2238.

    Article  CAS  Google Scholar 

  25. Stenberg VY, Larsen RH, Ma L-W, Peng Q, Juzenas P, Bruland ØS, et al. Evaluation of the PSMA-binding ligand 212Pb-NG001 in multicellular tumour spheroid and mouse models of prostate cancer. Int J Mol Sci. 2021;22(9):4815.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  26. Stenberg VY, Juzeniene A, Chen Q, Yang X, Bruland ØS, Larsen RH. Preparation of the alpha-emitting prostate-specific membrane antigen targeted radioligand [212Pb]Pb-NG001 for prostate cancer. J Labelled Compd Radiopharm. 2020;63(3):129–43.

    Article  CAS  Google Scholar 

  27. Tran-Gia J, Lassmann M. Characterization of noise and resolution for quantitative 177Lu SPECT/CT with xSPECT quant. J Nucl Med. 2019;60(1):50–9.

    Article  PubMed  CAS  Google Scholar 

  28. Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, Pujol S, et al. 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging. 2012;30(9):1323–41.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Gear JI, Cox MG, Gustafsson J, Gleisner KS, Murray I, Glatting G, et al. EANM practical guidance on uncertainty analysis for molecular radiotherapy absorbed dose calculations. Eur J Nucl Med Mol Imaging. 2018;45(13):2456–74.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  30. Kurth J, Krause BJ, Schwarzenböck SM, Stegger L, Schäfers M, Rahbar K. External radiation exposure, excretion, and effective half-life in (177)Lu-PSMA-targeted therapies. EJNMMI Res. 2018;8(1):32.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  31. de Kruijff RM, Wolterbeek HT, Denkova AG. A critical review of alpha radionuclide therapy-how to deal with recoiling daughters? Pharmaceuticals (Basel). 2015;8(2):321–36.

    Article  CAS  Google Scholar 

  32. Sato T, Furuta T, Liu Y, Naka S, Nagamori S, Kanai Y, et al. Individual dosimetry system for targeted alpha therapy based on PHITS coupled with microdosimetric kinetic model. EJNMMI Phys. 2021;8(1):4.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Hobbs RF, Howell RW, Song H, Baechler S, Sgouros G. Redefining relative biological effectiveness in the context of the EQDX formalism: implications for alpha-particle emitter therapy. Radiat Res. 2014;181(1):90–8.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  34. Sgouros G, Hobbs RF, Song H. Modelling and dosimetry for alpha-particle therapy. Curr Radiopharm. 2011;4(3):261–5.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  35. Peters SMB, Privé BM, de Bakker M, de Lange F, Jentzen W, Eek A, et al. Intra-therapeutic dosimetry of [(177)Lu]Lu-PSMA-617 in low-volume hormone-sensitive metastatic prostate cancer patients and correlation with treatment outcome. Eur J Nucl Med Mol Imaging. 2021;5:889.

    Google Scholar 

  36. Hohberg M, Eschner W, Schmidt M, Dietlein M, Kobe C, Fischer T, et al. Lacrimal glands may represent organs at risk for radionuclide therapy of prostate cancer with [177Lu]DKFZ-PSMA-617. Mol Imag Biol. 2016;18(3):437–45.

    Article  CAS  Google Scholar 

  37. Peters SMB, Hofferber R, Privé BM, de Bakker M, Gotthardt M, Janssen M, et al. [(68)Ga]Ga-PSMA-11 PET imaging as a predictor for absorbed doses in organs at risk and small lesions in [(177)Lu]Lu-PSMA-617 treatment. Eur J Nucl Med Mol Imaging. 2021;5:58.

    Google Scholar 

  38. Wessels BW, Konijnenberg MW, Dale RG, Breitz HB, Cremonesi M, Meredith RF, et al. MIRD pamphlet No. 20: the effect of model assumptions on kidney dosimetry and response—implications for radionuclide therapy. J Nuclear Med. 2008;49(11):1884–99.

    Article  Google Scholar 

  39. Zaid NRR, Kletting P, Winter G, Prasad V, Beer AJ, Glatting G. A physiologically based pharmacokinetic model for in vivo alpha particle generators targeting neuroendocrine tumors in mice. Pharmaceutics. 2021;13(12):2132.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  40. Afshar-Oromieh A, Malcher A, Eder M, Eisenhut M, Linhart HG, Hadaschik BA, et al. PET imaging with a [68Ga]gallium-labelled PSMA ligand for the diagnosis of prostate cancer: biodistribution in humans and first evaluation of tumour lesions. Eur J Nucl Med Mol Imaging. 2013;40(4):486–95.

    Article  PubMed  CAS  Google Scholar 

  41. Lyra M, Ploussi A. Filtering in SPECT image reconstruction. Int J Biomed Imaging. 2011;2011:693795.

    Article  PubMed  PubMed Central  Google Scholar 

  42. He B, Frey E, Sgouros G, Ghaly M, Tworowska I, Delpassand E. Development and validation of methods for quantitative in vivo SPECT of Pb-212. J Med Imaging Radiat Sci. 2019;50(1, Supplement):S33.

    Article  Google Scholar 

  43. Benabdallah N, Bernardini M, Bianciardi M, de Labriolle-Vaylet C, Franck D, Desbrée A. 223Ra-dichloride therapy of bone metastasis: optimization of SPECT images for quantification. EJNMMI Res. 2019;9(1):20.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

Download references

Acknowledgements

We would like to thank Roy Larsen for providing the lead-212 and Asta Juzeniene for initial discussions. We would also like to thank the anonymous reviewers for excellent suggestions.

Funding

The present work was financially supported by the South-Eastern Norway Regional Health Authority.

Author information

Authors and Affiliations

Authors

Contributions

CS planned the study, supervised the project, and interpreted the findings. MK carried out the experiments, analysed the data, interpreted the findings, and took lead in writing the manuscript. MER was involved in planning the study, supervised the project, and interpreted the findings. All authors discussed the results and contributed to the final version of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Monika Kvassheim.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

None of the authors have any competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

The original version of this article was revised: there was an error in Fig. 1.

Supplementary Information

Additional file 1:

 All supplementary materials with results from reconstructions not shown in the main paper are found in Additional file 1. Pages 1–12 include all materials from the reconstructions with a 12-mm Gaussian filter applied which were not included in the paper. Pages 13–34 include all materials from the reconstructions without filters applied. Calibration factors for all reconstructions are given in the tables on pages 1 (filtered) and 13 (unfiltered). The equivalent plots as that shown in Fig. 2 c are shown for the other reconstructions on page 2 (filtered) and pages 14–15 (unfiltered). The unfiltered versions of Fig. 3 a and b are found on page 13. Versions of Fig. 4 for other reconstructions are shown on pages 3–5 (filtered) and 15–18 (unfiltered). Recovery plots as shown in Fig. 5 and tables equivalent to Table 1 for other reconstructions are given on pages 6–11 (filtered) and 19–29 (unfiltered). The equivalent plots as those given in Fig. 6 without filters are given on pages 30–31 and the filtered version for the smallest sphere is given on page 12. The unfiltered versions of the images in Fig. 7 and Fig. 8 are shown on page 32 and 33, respectively. The version of Fig. 9 without a filter is given on page 34.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kvassheim, M., Revheim, ME.R. & Stokke, C. Quantitative SPECT/CT imaging of lead-212: a phantom study. EJNMMI Phys 9, 52 (2022). https://doi.org/10.1186/s40658-022-00481-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-022-00481-z

Keywords