Dual energy window imaging for optimisation of P/V ratios in VP SPECT

Abstract

Purpose

Ventilation–perfusion single-photon emission computed tomography (VP SPECT) plays an important role in pulmonary embolism diagnosis. Rapid results may be obtained using same-day ventilation followed by perfusion imaging, but generally requires careful attention to achieving an optimal count rate ratio (P/V ratio) of ≥ 3:1. This study investigated whether the ratio of counts simultaneously acquired in adjacent primary and Compton scatter energy windows (Eratio) on V SPECT was predictive of final normalised perfusion count rate (PCRnorm) on P SPECT using [99mTc]Tc-macroaggregated albumin (MAA), thus allowing for optimisation of P/V ratios.

Methods

Same-day VP SPECT studies acquired using standard protocols in adult patients during a 2-year period (training dataset) were assessed. Studies were included provided they were acquired with correct imaging parameters, and injection site imaging and laboratory records were available for quality control and normalised count rate corrections. Extraction of DICOM information, and linear regression were performed using custom Python and R scripts. A predictive tool was developed in Microsoft Excel. This tool was then validated using a second (validation) dataset of same-day studies acquired over a subsequent 7-month period. Accuracy of the prediction tool was assessed by calculating the mean absolute percentage error (MAPE).

Results

Of 643 studies performed, the scans of 342 participants (median age 30.4 years, 318 female) were included in the training dataset, the analysis of which yielded a significant regression equation (F(1,340) = 1057.3, p < 0.0001), with an adjusted R2 of 0.756 and MSE of 0.001089. A prediction tool designed for routine clinical use was developed for predicting final P/V ratio. Of an additional 285 studies, 198 were included in the second (validation) dataset (median age 29.7 years, 188 female). The Excel-based tool was shown to be 91% accurate (MAPE: 9%) in predicting P/V ratio.

Conclusion

The relationship between the ratio of simultaneously acquired counts in adjacent energy windows on V SPECT and perfusion count rate after administration of a known activity of [99mTc]Tc-MAA can be linearly approximated. A predictive tool based on this work may assist in optimising the dose and timing of [99mTc]Tc-MAA administration in same-day studies to the benefit of patients and workflows.

Introduction

The perfusion component of a ventilation–perfusion single-photon emission computed tomography (VP SPECT) study is typically undertaken on the same day as, and frequently immediately following the ventilation component. When using technetium-99m (99mTc)-based radiopharmaceuticals in a same-day imaging strategy (generally recommended, especially to obtain rapid diagnosis of pulmonary embolism) [1], residual ventilation counts contribute to the acquired perfusion images and may obscure perfusion defects leading to misdiagnosis. While software can perform correction for residual ventilation counts in the final perfusion image, slight differences in radiopharmaceutical distributions, and imperfect co-registration can lead to significant artefacts, and it is therefore recommended that the perfusion count rate (PCR—corrected for residual ventilation counts) exceeds the ventilation count rate (VCR) by a factor of at least 3 [2,3,4]. This is the equivalent, when a perfusion SPECT is performed immediately after the ventilation SPECT, of the uncorrected count rate exceeding the VCR by a minimum factor of 4. At the same time, higher ratios may also be considered undesirable since these would require high doses of the perfusion radiopharmaceutical [99mTc]Tc—macroaggregated albumin ([99mTc]Tc-MAA), which would result in higher radiation dose, and increased risks posed by the administration of a higher number of [99mTc]Tc-MAA particles. The former is especially relevant in VP SPECT studies performed to diagnose pulmonary embolism (PE) in pregnant or lactating women in whom radiation exposure poses an additional risk; while the latter is relevant especially when administering [99mTc]Tc-MAA in patients with known or occult disorders of the pulmonary vasculature.

The targeting of an optimal perfusion/ventilation count rate ratio (P/V ratio) is complicated principally by the fact that it is often difficult to quantify the amount of ventilation activity that has been administered in any individual patient. In patients with a large amount of soft tissue, higher amounts of administered activity may be required to obtain a sufficiently high VCR to guarantee a ventilation SPECT of sufficient quality. This is mainly because as the amount of soft tissue increases, there is increased attenuation due to Compton scatter [5]. Post hoc calculation of administered ventilation activity using knowledge of the imaging system sensitivity alone is not possible due to the frequently significant amount of attenuation by thoracic tissue. While absolute quantification of administered ventilation activity is possible using quantitative SPECT, this remains experimental and requires specialised software and a SPECT/CT system [6]. As a result, the dose and timing of [99mTc]Tc-MAA administration after the ventilation SPECT is frequently suboptimal, resulting in PCRs that are either too low to achieve an adequate P/V ratio, or the administration of activities exceeding those recommended or required, and thus not meeting the principle of “as low as reasonably achievable” (ALARA).

It is possible to account for photon attenuation using information from a second, simultaneously acquired Compton scatter window [7]. Using various techniques in phantom-based studies, researchers have demonstrated linear relationships between the number of photons detected in scatter windows and voxel attenuation coefficients [8] or attenuation depth [9].

This study aimed to investigate whether a simple ratio of counts acquired in two adjacent energy windows on the ventilation SPECT (Eratio) is predictive of the final perfusion count rate normalised for [99mTc]Tc-MAA activity (PCRnorm) on the perfusion SPECT; and if so, whether such a relationship can be used to develop a simple tool to adjust either the activity (dose) of [99mTc]Tc-MAA administered or the timing (or both) of the perfusion SPECT, such that an optimal P/V ratio may be achieved.

Material and methods

Participant selection

All patients referred to the Tygerberg Hospital Division of Nuclear Medicine for VP SPECT studies between July 2017 and July 2019 were eligible for inclusion. Only patients with same-day studies (ventilation followed by perfusion) and who were older than 18 years were included in this retrospective analysis. Patients in whom no [99mTc]Tc-MAA injection site imaging was performed; or for whom the [99mTc]Tc-MAA dispensing time was not recorded (to correct [99mTc]Tc-MAA dose for residual counts at administration site, and for decay) were excluded, as were the scans of patients in whom the imaging acquisition differed from the divisional protocol; or in cases in which a repeat SPECT acquisition was performed. A second sample (validation dataset), subsequently drawn from scans performed between July 2019 and March 2020, with the same inclusion and exclusion criteria, was prepared to validate the tool developed with the original sample (training dataset).

Imaging procedure

Ventilation was performed with the patient lying supine, using a [99mTc]Tc-Technegas generator (Vita Medical Ltd, Lucas Heights, New South Wales, Australia). After each breath, tubing of the ventilation system was set aside and a Geiger-Müller (GM) detector (Radiation Alert® – Inspector, SE International Inc, Summertown, TN, USA) positioned on the sternum. Technicians were instructed to target an approximate stabilised probe count rate of ~ 2500 counts per minute (cpm). Immediately after ventilation, SPECT was performed on a dual-head Siemens Symbia SPECT camera (Siemens Healthineers, Hoffman Estates, IL, USA) with measured detector energy resolution of 8.5%, or a dual-head GE Infinia Hawkeye SPECT-CT camera (General Electric Medical Systems, Milwaukee, WI, USA) with measured detector energy resolution of 11.1%. Acquisition protocols (Table 1) were chosen to match most closely those recommended by Palmer et al. [4]. All SPECT scans were acquired with a primary energy window centred on a peak of 140 keV (width ± 7.5%); and a symmetrical down-scatter window, centred on a peak of 119 keV, of exactly the same width immediately beneath this. The latter window has been routinely included in VP SPECT acquisitions in the Tygerberg Division of Nuclear Medicine since 2015, at which time image reconstruction software (Hermes Gold 3; Nuclear Diagnostics, Stockholm Sweden) used this data to construct a synthetic attenuation map.

Table 1 SPECT acquisition parameters

The perfusion study was acquired either immediately after the ventilation study or after a delay to allow for some decay of ventilation counts using ~ 120 MBq of [99mTc]Tc-MAA (Pulmocis®, Cisbio). All dispensed [99mTc]Tc-MAA doses had a radiochemical purity of ≥ 95%. Static acquisitions (64 × 64 matrix, 60 s, 129.5–150.5 keV), using the same LEAP collimator as used for SPECT, were made of the [99mTc]Tc-MAA injection site to exclude major misinjection and to quantify residual activity in the injection portal. Residual activity in the syringe was not measured.

Data extracted from DICOM files and paper records

From the radiopharmacy and patient scan records, the time [99mTc]Tc-MAA was dispensed and the activity at dispense time were recorded.

All raw tomographic VP SPECT image data, as well as corresponding injection site images for the periods in question were downloaded from the divisional PACS in DICOM format.

A custom script written in Python version 3.8.2 [10] using the pydicom package version 1.4.2 [11] was used to extract information from the DICOM image and header tags, both to assure compliance with divisional acquisition protocols, and to obtain the necessary image measures for analysis purposes.

Information to ensure consistency of the acquisition protocols included width and centring of the energy windows; scan arc; number of frames; frame duration; and camera and collimator identifiers. Scan order (ventilation followed by perfusion) was also validated by checking scan dates and times.

Measures extracted for analysis purposes included patient name, surname, and folder number (subsequently deidentified); patient gender and date of birth (for demographic purposes); scan dates and times (for decay correction); and total image counts for both energy windows (for tomographic data) or in the primary energy window for the injection site images.

Derived image measures and how they were calculated are summarised in Table 2.

Table 2 Derived image measures

Modelling

Based on visual analysis of the initial scatter plot of the training dataset, a linear regression model of Eratio and PCRnorm was built in RStudio version 1.3.1093 [12] using R version 4.0.2 [13]. Diagnostic plots were generated using native, ggfortify [14], and nortest packages. A threshold of p = 0.05 was considered statistically significant. Calculation of 95% prediction intervals for graphing purposes was performed using the native predict.lm function. Normality of the residuals was assessed using graphical methods.

Prediction tool

Using the results of the linear regression performed on the training dataset, and the mean squared error (MSE), a tool to predict final P/V ratio was developed in Microsoft® Excel® for Microsoft 365 MSO (16.0.13127.20620) 64-bit.

Validation

Validation of the Excel-based tool was performed by comparing the model-predicted P/V ratio values with the actual P/V ratios for studies in the validation dataset. Accuracy of the model in predicting P/V ratio was assessed by calculating the mean absolute percentage error (MAPE) using the MLmetrics package in R.

Clinical impact

The practical implications of using the tool on studies included in the validation dataset was also simulated for a sample strategy that favoured dose sparing. First, VCR values were normalised such that the summed ventilation counts equalled the recommended value of 2 million [4]. Because the actual PCR per MBq of administered [99mTc]Tc-MAA could be retrospectively calculated, it was possible to (1) calculate the [99mTc]Tc-MAA dose that would have been required to achieve a minimum P/V ratio of 3 had the perfusion component immediately followed the ventilation component (i.e. no delay), and (2) to calculate the true delay that would have been required after an optimal ventilation study before it would have been possible to proceed with a target dose (120 MBq) of [99mTc]Tc-MAA and still achieve an adequate P/V ratio. These results were then compared to the delay recommended by the prediction tool, assuming a [99mTc]Tc-MAA dose of 120 MBq and the same minimum P/V ratio. The 120 MBq target dose is the same as that recommended in a same-day protocol in pregnant patients [15] and provides a margin of safety should the actual dose required while injecting under the detector be greater (up to the limit of 185 MBq according to the package insert for the Pulmocis® product). By comparing the true delay required and the predicted delay required, unnecessary delay (delay suggested by model exceeding actual delay required) and insufficient delay (delay suggested by model inadequate) could also be quantified. Benefits of using the prediction tool were quantified in terms of dose of [99mTc]Tc-MAA spared. Costs of using the prediction tool were quantified in terms of unnecessary delay. Since the demonstration of clinical impact was performed purely for illustrative purposes using one of many possible strategies, confidence intervals for the predicted delays required were ignored for the sake of simplicity.

Results

Selection of participant scans

Training dataset

A total of 643 studies (ventilation SPECT followed by same-day perfusion SPECT) were identified in the study period. Of these, 22 were of patients under the age of 18 years and were excluded. An additional 199 studies were excluded due to missing injection site image; one significant misinjection (83 MBq); 69 due to missing radiopharmacy records ([99mTc]Tc-MAA dispense time not recorded); four due to incorrect dispense time records (recorded time later than scan time); and one due to deviation from the divisional protocol (incorrect collimator used). Four participants had repeat studies (10.1–58.6 weeks later), and for these participants, only the first study was included in the analysis. An initial analysis was performed on 343 imaging studies. One imaging study was retrospectively excluded after it became apparent that it was an outlier with high influence and an obvious cause was identified (details below). The sample size for the final analysis of the training dataset was thus of 342 imaging studies.

Validation dataset

A total of 285 additional same-day studies were identified in the second period for use in the validation dataset. Of these, six were of patients under the age of 18 years and were excluded. An additional 71 studies were excluded due to missing radiopharmacy records ([99mTc]Tc-MAA dispense time not recorded); eight due to incorrect dispense time records (recorded time later than scan time); one significant misinjection (36 MBq); and one due to a missing injection site image. No participants in this sample had repeat studies. The sample size for the validation dataset was thus 198 studies.

Demographics and summary statistics

Of the 342 imaging studies included in the final analysis of the training dataset, 317 were performed in female patients. The median patient age was 30.1 years (IQR: 25–37.2 years). Mean Asys was 114.8 MBq (SD: 8 MBq). Mean Eratio was 1.83 (SD: 0.21), and mean PCRnorm was 0.313 (SD 0.067). Mean P/V ratio was 4.01 (SD 1.09).

Of the 198 imaging studies included in the final analysis of the validation dataset, 188 were performed in female patients. The median patient age was 29.7 years (IQR: 24.6–37.5 years). Mean Asys was 118.4 MBq (SD: 9.9 MBq). Mean Eratio was 1.8 (SD: 0.23), and mean PCRnorm was 0.311 (SD 0.072). Mean P/V ratio was 4.98 (SD 1.44).

A breakdown of the final P/V ratio values in the training and validation datasets appears in Table 3.

Table 3 P/V ratio values of the studies included in the training and validation datasets

Linear regression

A simple linear regression of the training data was calculated to predict PCRnorm, given Eratio. The initial scatter plot and diagnostic charts (Additional file 1: Fig. S1) identified an obvious outlier (datapoint 223: x = 1.89, y = 0.62). Further investigation revealed that the ventilation SPECT in this case had a very high amount of activity outside of the patient which very likely influenced the Eratio value (Additional file 2: Fig. S2). Given that this observation was both an outlier and highly influential, and because an obvious cause was identified on review of the raw data, this observation was retrospectively excluded.

After exclusion of the outlier, a significant regression equation was found (F(1,340) = 1057.3, p < 0.0001), with an adjusted R2 of 0.756 and MSE of 0.001089. After adjusting for any specific camera-collimator sensitivity factor, the relationship to yield a re-normalised (adjusted) perfusion count rate was represented by the following equation:

$${\text{PCR}}\, = \,A \times S \times \left( {0.28\left( {E_{{{\text{ratio}}}} } \right) - 0.198} \right)$$

where PCR represents the expected perfusion count rate (in cps), for any administered activity (A) of [99mTc]Tc-MAA (in MBq), re-normalised (adjusted) to account for the user-specific sensitivity of the camera-collimator combination \(S\), and predicted by Eratio.

The final model, including confidence and prediction intervals is plotted in Fig. 1. The relevant diagnostic plots for the final regression appear in Additional file 3: Fig. S3 and Additional file 4: Fig. S4.

Fig. 1
figure 1

Final linear regression of perfusion count rate normalised to 1 MBq of administered [99mTc]Tc-MAA, and a camera-collimator sensitivity of 1 cps/MBq (PCRnorm) as a function of the dual energy window count ratio of the ventilation study (Eratio). Regression line (blue solid line), 95% confidence band (dark grey), and 95% prediction bands (red dashed lines). Image generated in RStudio

Prediction tool

The output of the linear regression enables calculation of the prediction interval around y, for any future value of x using the equation:

$$\hat{y} \pm t_{{(\frac{\alpha }{2},n - 2)}} \times \sqrt {{\text{MSE}} \times \left( {1 + \frac{1}{n} + { }\frac{{\left( {x_{h} - \overline{x}} \right)^{2} }}{{\sum \left( {x_{i} - \overline{x}} \right)^{2} }}} \right){ }}$$

where \(\hat{y}\) is the predicted fitted value of \(y\) for a particular value of \(x\); \({t}_{(\frac{\alpha }{2},n-2)}\) represents the two-tailed inverse of the Student's t distribution with significance level \(\alpha\) for \(n\) number of observations in the regression dataset; MSE is the mean squared error; \({x}_{h}\) is the value of a hypothetical future observation of \(x\); \({\bar{x}}\) is the mean value of \(x\) in the regression dataset; and \({x}_{i}\) represents the respective values of \(x\) in the regression dataset [16].

Using the above formula, a practical tool to predict the final P/V ratio (with 90% and 95% prediction intervals) was developed in Excel (Additional file 6). A video demonstrating practical implementation of the tool is included in Additional file 7.

Validation

A scatter plot of predicted P/V ratios for the validation dataset overlayed on the linear model developed using the training dataset is shown in Fig. 2. A plot comparing actual P/V ratios to predicted values appears in Fig. 3. The prediction tool was 91% accurate in predicting P/V ratios in the validation dataset (MAPE of 9%).

Fig. 2
figure 2

Final linear regression as shown in Fig. 1, with training dataset values in grey and superimposed validation dataset values in blue. Image generated in RStudio

Fig. 3
figure 3

Scatter plot of P/V ratio values predicted with the Excel-based tool versus actual values (validation dataset). The straight line indicates points for which P/V ratio predicted = P/V ratio actual. Image generated in RStudio

Clinical impact

In the simulation using data from the validation dataset, had the [99mTc]Tc-MAA dose been injected blindly (without using the prediction tool) immediately after an optimal ventilation study, the required activity would have exceeded the target dose of 120 MBq in 56/198 cases, of which seven would have exceeded the maximum recommended dose of 185 MBq.

A breakdown of the impact of using the prediction tool is shown in Fig. 4. Overall, use of the prediction tool would have delayed the perfusion study in 62 patients resulting in a dose sparing of 0.7–73.5 (median 19.7) MBq at a cost of 3.2–158.6 (median 20.2) minutes. Following the recommendation of the prediction tool would have resulted in 14 additional patients requiring a [99mTc]Tc-MAA dose less than 120 MBq and six patients being spared excessive doses above 185 MBq.

Fig. 4
figure 4

Results of applying the prediction tool to ventilation studies in the validation dataset using a dose-sparing strategy (targeting a [99mTc]Tc-MAA dose of 120 MBq, and a P/V ratio of 3). In the stimulation, values predicted by the tool were compared to equivalent values calculated directly from actual perfusion data

Discussion

This study found that the relationship between the ratio of ventilation counts in the primary energy window and an adjacent Compton scatter window and the final perfusion count rate after the administration of a given dose of [99mTc]Tc-MAA could be linearly approximated. Results of the analysis were used to develop a practical, Excel-based tool to predict the adequacy of the P/V ratio in same-day VP SPECT studies, which nuclear medicine practitioners can use to optimise the trade-offs that must frequently be made in these studies. This tool was validated using a second, training dataset and found to be accurate in predicting final P/V ratio. The clinical impact of using the tool was illustrated using a dose-sparing strategy.

In a same-day VP SPECT study, the ventilation component is followed (ideally immediately after) by the perfusion component. Different strategies are used to increase the likelihood of obtaining ventilation images that are of sufficiently high diagnostic quality, while still meeting a final target P/V ratio of ≥ 3, and adhering to recommended [99mTc]Tc-MAA doses. Most divisions target a specific count rate (e.g. 2000–2500 cpm) [17, 18] measured using a GM detector after ventilation, and having achieved this, proceed to ventilation imaging. Depending on the count rate on the final ventilation SPECT, centres either proceed immediately with injection of [99mTc]Tc-MAA under the camera detector, usually checking for or titrating to a fourfold increase in count rate, or delay the perfusion component of the study until such a time that ventilation count rate is judged to have sufficiently decayed.

There are several problems with these strategies. Foremost, that the VCR is attenuated by variable amounts of soft tissue. This has the effect that a given activity (e.g. 40 MBq) of administered ventilation agent might result in a very high VCR (in a thin patient with minimal attenuating soft tissue), or a very low VCR (in a patient with more thoracic soft tissue), typically requiring the additional administration of ventilation agent. Very high count rates on the final ventilation study may lead to unnecessary deferment of the perfusion study in the first patient (in whom immediately proceeding with, e.g. a 120 MBq dose of [99mTc]Tc-MAA might in actual fact have been sufficient), or lead to an inadequate P/V ratio in the second patient (who might require, e.g. 80 MBq of administered ventilation agent to achieve adequate ventilation counts) when their perfusion study is undertaken immediately, even with the maximum [99mTc]Tc-MAA dose of 185 MBq recommended by the Pulmocis® package insert. Titrating the dose of [99mTc]Tc-MAA administered to count rate under the camera is a useful compromise [19] but may require excessive amounts of activity (and particles) with attendant risks. It may be speculated that the difficulties in optimising the trade-offs between dose, count adequacy, sufficient P/V ratio, and reasonable patient turnaround times, are greatest in units caring for patients with wide variability in body habitus. Certainly, these trade-offs are frequently imperfect: a multicentre survey including the results of 286 VP SPECT studies in Germany reported inadequate P/V ratios in 25% of studies [19] which is similar to the findings of the current study, in which 18% of studies in the training dataset did not meet the recommended P/V ratio of > 3.

A P/V ratio that is too high is also problematic. While the optimal P/V ratio is uncertain, it seems unlikely that much additional diagnostic benefit is achieved with ratios above 5 (44% of cases in the validation dataset). While [99mTc]Tc-MAA is generally accepted as a safe compound, it is not entirely innocuous, with reports of respiratory complications and death when high particle numbers have been administered to patients with underlying pathology of the pulmonary vasculature [20,21,22,23]. The risks posed by exposure to low doses of ionising radiation are more controversial [24, 25] but certainly it is reasonable to try and minimise the risks conferred by both particle number and radiation when reasonably achievable.

Most SPECT cameras can be easily configured to simultaneously image two adjacent energy windows. Once the ventilation SPECT is complete, it is a simple matter to perform a quick quality check of the raw data, and to use the perfusion adequacy prediction tool to predict the final P/V ratio. Moreover, this tool allows users to prioritise lower radiation dose or higher scan quality by modelling the impacts of different [99mTc]Tc-MAA doses and/or delays before performing the perfusion SPECT on the P/V ratio.

Compton scatter is the predominant mode of gamma photon interaction with matter when imaging with [99mTc]Tc-based tracers and increases as the depth of attenuating soft tissue increases [9]. This is evidenced by plotting Eratio values against body mass index (a crude measure of body habitus) for a subset (n = 291) of participants from the training dataset in whom this information was available (Additional file 5: Fig. S5a). This also explains why Eratio for any individual patient is constant, independent on the amount of administered activity (dose) since the attenuation profile remains constant—illustrated by plotting Eratio values derived from the V SPECT study with Eratio values derived from the corresponding VP SPECT for the combined dataset (Additional file 5: Fig. S5b).

A linear model to describe the relationship between Eratio and PCRnorm was selected based on the scatter plot of data in the training dataset but the basis for such a relationship is uncertain. Researchers have previously demonstrated an approximately linear relationship between scatter fraction and attenuation: Sitek et al. demonstrated a linear relationship between the proportion of singly scattered photons and voxel attenuation coefficients, which they used to calculate mu-maps in a [99mTc]Tc torso phantom study [8]. Similarly, also using phantoms, another group showed that ratios of scatter photons to photons detected in the primary energy window can be linearly correlated with tissue depth [9]. These linear relationships, assuming they translate to real world conditions, would however still not explain a linear relationship between Eratio (calculated on the ventilation study raw data) and PCR in the current study, since the latter would still be subject to the phenomenon of exponential attenuation of photon transmission in broad beam conditions. Rather, it is speculated that the relationship is in fact exponential, and that the success of the linear model in this study is because a linear approximation is sufficiently accurate over the range of tissue depths encountered in the sample (not quantified in this experiment).

In addition to enabling the validation of the model, the second dataset allowed a simulation of clinical impact when using the prediction tool. For the purpose of illustration, a dose-sparing strategy was chosen, mimicking the preferred practice of the Tygerberg Hospital Division of Nuclear Medicine. It is however important to emphasise that the trade-off between dose and delay can be set to any user’s institutional or personal preference when using the tool. Using a dose-sparing strategy in this simulation, compared to values derived from the actual measures, the Excel-based tool would have appropriately deferred the majority of perfusion studies that would otherwise have required activities of [99mTc]Tc-MAA exceeding the target dose of 120 MBq, including those exceeding the 185 MBq maximum limit recommended by the manufacturer, and this would have been achieved with only the minimal cost of excessive delay in a few cases.

This study suffered from several limitations. Many scans had to be excluded from the analysis due to missing injection site imaging or radiopharmacy records, which affected the final sample size. The sample was mostly composed of young (adult), pregnant women which means that caution should be exercised when applying the results to other patient demographics, although no reason for deviation from the model is expected on in other patient groups. More importantly, the linear relationship between Eratio and PCRnorm at extreme values of Eratio was based on only a few observations, and the validity of the linear model at these extremes is less certain. Confidence and prediction intervals may have been narrower had this study included additional corrections potentially affecting the accurate calculation of the amount of systemically administered [99mTc]Tc-MAA(Asys): (a) for residual activity in the syringe after administration, and (b) for the interval between injection and the perfusion SPECT. The latter limitation is however not expected to be a major contributor to error, given the divisional policy of commencing the perfusion SPECT immediately post-injection. Additional sources of error not accounted for include non-correction for out-of-field-of-view scatter, and variation in the radiochemical purity of [99mTc]Tc-MAA between 95 and 100% (which may have contributed slightly to error in PCRnorm). Another, inherent limitation in predicting PCR is presented by variability in right–left shunting (including physiological shunting) which could not be accounted or corrected for [26]. Finally, it is also important to acknowledge that these results are specific for the energy window settings used on systems with an ~ 10% energy resolution, and using [99mTc]Tc-Technegas, which has minimal early clearance after ventilation [27]. Is not clear how changes in these acquisition setting or of the ventilation agent used, might affect the reliability of the model.

Conclusion

The expected perfusion count rate for a given dose of [99mTc]Tc-MAA can be linearly predicted by the ratio of ventilation counts in two adjacent energy windows. This relationship allows for the application of a simple tool to predict and/or optimise the final P/V ratio. In addition to the practice of directly monitoring count rate increases during [99mTc]Tc-MAA injection, this tool may assist in optimising the trade-offs between study quality, [99mTc]Tc-MAA dose, and patient turnaround time while still achieving diagnostic quality VP SPECT studies.

Availability of data and materials

Regression data are included in additional files. Other non-identifying data are available from the corresponding author on reasonable request.

Code availability

Python and R scripts are available from the corresponding author on reasonable request.

References

  1. Bajc M, Schümichen C, Grüning T, Lindqvist A, Le Roux P-Y, Alatri A, et al. EANM guideline for ventilation/perfusion single-photon emission computed tomography (SPECT) for diagnosis of pulmonary embolism and beyond. Eur J Nucl Med Mol Imaging. 2019;46:2429–51.

    Article  CAS  Google Scholar 

  2. Bajc M, Neilly JB, Miniati M, Schuemichen C, Meignan M, Jonson B, et al. EANM guidelines for ventilation/perfusion scintigraphy: part 1. Pulmonary imaging with ventilation/perfusion single photon emission tomography. Eur J Nucl Med Mol Imaging. 2009;36:1356–70.

    Article  CAS  Google Scholar 

  3. Parker JA, Coleman RE, Grady E, Royal HD, Siegel BA, Stabin MG, et al. SNM practice guideline for lung scintigraphy 4.0. J Nucl Med Technol. 2012;40:57–65.

    Article  Google Scholar 

  4. Palmer J, Bitzén U, Jonson B, Bajc M. Comprehensive ventilation/perfusion SPECT. J Nucl Med. 2001;42:1288–94.

    PubMed  CAS  Google Scholar 

  5. Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear medicine. Elsevier Health Sciences; 2012.

  6. Dickson J, Ross J, Vöö S. Quantitative SPECT: the time is now. EJNMMI Phys. 2019;6:1–7.

    Article  Google Scholar 

  7. Tadesse GF, Geramifar P, Tegaw EM, Ay MR. Techniques for generating attenuation map using cardiac SPECT emission data only: a systematic review. Ann Nucl Med. 2019;33:1–13.

    Article  Google Scholar 

  8. Sitek A, Moore SC, Kijewski MF. Correction for photon attenuation without transmission measurements using compton scatter information in SPECT. In: 2007 IEEE nuclear science symposium conference record. Honolulu, HI, USA: IEEE; 2007. p. 4210–2

  9. Ashoor M, Asgari A, Khorshidi A, Rezaei A. Evaluation of Compton attenuation and photoelectric absorption coefficients by convolution of scattering and primary functions and counts ratio on energy spectra. Indian J Nucl Med. 2015;30:239–47.

    Article  Google Scholar 

  10. van Rossum G. Python tutorial. Technical Report CS-R9526, Centrum voor Wiskunde en Informatica (CWI), Amsterdam. 1995

  11. Mason D, scaramallion, rhaxton, mrbean-bremen, Suever, Vanessasaurus, et al. pydicom/pydicom: v1.4.2 [Internet]. Zenodo; 2020. https://zenodo.org/record/3614067

  12. RStudio Team. RStudio: Integrated Development Environment for R [Internet]. Boston: RStudio, PBC; 2020. http://www.rstudio.com/

  13. R Core Team. R: A language and environment for statistical computing [Internet]. Vienna, Austria: R Foundation for Statistical Computing; 2020. https://www.R-project.org/

  14. Tang Y, Horikoshi M, Li W. ggfortify: unified interface to visualize statistical results of popular R packages. R J. 2016;8:474.

    Article  Google Scholar 

  15. Bajc M, Olsson B, Gottsäter A, Hindorf C, Jögi J. V/P SPECT as a diagnostic tool for pregnant women with suspected pulmonary embolism. Eur J Nucl Med Mol Imaging. 2015;42:1325–30.

    Article  Google Scholar 

  16. Introduction. Modern regression methods. 2nd ed. Wiley. Hoboken; 2009. p. 1–59.

  17. Recommended Technegas Delivery Protocol. The Crucible [Internet]. 2013 Aug;6. http://www.cyclomedica.com.au/wNewsletters/Newsletter1308Vol6No3.pdf

  18. TechnegasPlus: Technegas Generator User Manual [Internet]. Australia: Cyclomedica; 2007. http://www.cyclomedica.com/wp-content/uploads/sites/20/2019/05/UM.EU-1EN.pdf

  19. Schaefer WM, Meyer A, Knollmann D. Lung ventilation/perfusion SPECT for diagnosing pulmonary embolism. Is the required minimal P/V count rate ratio fulfilled in routine situation? Results from a multicentre survey. Nuklearmedizin. 2015;54:217–22.

    Article  CAS  Google Scholar 

  20. Roberts HJ. Fatal hemoptysis in pulmonary embolism probably precipitated by pulmonary scanning. Report of a case and suggested precautions. Angiology. 1970;21:270–4.

    Article  CAS  Google Scholar 

  21. Dworkin HJ, Smith JR, Bull FE. Reaction after administration of macroaggregated albumin for a lung scan. N Engl J Med. 1966;275:376.

    Article  CAS  Google Scholar 

  22. Vincent WR, Goldberg SJ, Donald D. Fatality immediately following rapid infusion of macroaggregates of 99mTc albumin (MAA) for lung scan 1. Radiology. 1968;91:1181–4.

    Article  Google Scholar 

  23. Williams JO. Death following injection of lung scanning agent in a case of pulmonary hypertension. BJR. 1974;47:61–3.

    Article  CAS  Google Scholar 

  24. Vaiserman A, Koliada A, Zabuga O, Socol Y. Health impacts of low-dose ionizing radiation: current scientific debates and regulatory issues. Dose Response. 2018;16.

  25. Prasad KN, Cole WC, Hasse GM. Health risks of low dose ionizing radiation in humans: a review. Exp Biol Med (Maywood). 2004;229:378–82.

    Article  CAS  Google Scholar 

  26. Ito K, Kurihara K, Ishibashi A, Morooka M, Mitsumoto T, Minamimoto R, et al. Cut-off value for normal versus abnormal right-to-left shunt percentages using 99mTc-macroaggregated albumin. Nucl Med Commun. 2011;32:936–40.

    Article  Google Scholar 

  27. Burch WM, Sullivan PJ, McLaren CJ. Technegas—a new ventilation agent for lung scanning. Nucl Med Commun. 1986;7:865–71.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors wish to thank the National Research Foundation of South Africa, which wholly funded this work.

Funding

This work is based on the research supported wholly by the National Research Foundation of South Africa (Grant Number 127484).

Author information

Authors and Affiliations

Authors

Contributions

AD contributed to the study conception and design. Material preparation, data collection and analysis were performed by AD and JH. The first draft of the manuscript was written by AD and both authors commented on previous versions of the manuscript. Both authors read and approved the final manuscript.

Corresponding author

Correspondence to A. G. G. Doruyter.

Ethics declarations

Ethics approval

This study was approved by the Human Research Ethics Committee of Stellenbosch University (Ref # N20/09/096).

Consent to participate

Due to the low risk of this retrospective analysis, a waiver of informed consent was granted by the Stellenbosch University Human Research Ethics Committee.

Consent for publication

Not applicable.

Competing interests

The authors have no relevant financial or non-financial interests to disclose.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Fig 1

Diagnostic plots of initial training dataset. a Scatter plot of observations in which Eratio represents the calculated ratio between acquired counts in the primary energy window divided by the down-scatter window on the ventilation study and PCRnorm represents the perfusion count rate per megabecquerel of administered [99mTc]Tc-MAA (cps/MBq), normalised to a camera-collimator sensitivity of 1 cps/MBq. b Chart of standardised residuals identified an obvious outlier (observation 223), which was also shown to be highly influential on c Cooks’ distance chart. Investigation of this observation identified an obvious cause (Figure 2), and on this basis the case was excluded, and the regression repeated. Images generated in RStudio.

Additional file 2: Fig 2

Reconstructed ventilation SPECT of participant 223 (anterior and lateral projections), exhibiting high amount of extracorporeal radiopharmaceutical (red crosshair), very likely on the basis of collimator/detector contamination, resulting in an unreliable Eratio value. This participant was excluded from the final analysis. Image generated in Hermes Hybrid Viewer PDR 4.0.1 with linear BW inverse colour table, relative scaling.

Additional file 3: Fig 3

Diagnostic plots of final training dataset, exhibiting a linearity of the data, b normality of the residuals, c homogeneity of variance, and d outlier observations and points with high leverage. Abbreviations: lm: linear model; PCRnorm: normalised perfusion count rate; Eratio: energy window count ratio of the ventilation study. Images generated in RStudio.

Additional file 4: Fig 4

Q-Q (top row) and density (bottom row) plots of the residuals (linear regression of PCRnorm on Eratio), for the training (left), validation (middle) and combined (training+validation, right) datasets. Results illustrate the close adherence to normal distribution of the residuals and support the use of parametric methods to build a prediction tool using the training dataset. Images generated in RStudio.

Additional file 5: Fig 5

Scatter plots illustrating the relationship between body habitus (body mass index—BMI) and Eratio (a), as well as that Eratio values for any individual subject on V SPECT and VP SPECT are equal (R2 = 0.98) (b). The straight line indicates points for which Eratio values calculated on the two studies are equal. Images generated in RStudio.

Additional file 6

. Perfusion adequacy prediction tool.

Additional file 7. Explanatory video.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Doruyter, A.G.G., Holness, J.L. Dual energy window imaging for optimisation of P/V ratios in VP SPECT. EJNMMI Phys 8, 69 (2021). https://doi.org/10.1186/s40658-021-00417-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-021-00417-z

Keywords