Table 1 SPECT acquisition parameters

From: Dual energy window imaging for optimisation of P/V ratios in VP SPECT

Camera model

Siemens Symbia—dual detector

Infinia Hawkeye—dual detector

Collimators

LEAP

LEAP

Collimator sensitivity (measured) (cps/MBq)

133

119

Matrix

64 × 64

64 × 64

Zoom

1.45

1.3

Pixel size (mm)

6.8

6.8

Primary energy window centre (keV)

140

140

Primary energy window lower threshold (keV)

129.5

129.5

Primary energy window upper threshold (keV)

150.5

151

Secondary energy window centre (keV)

119

119

Secondary energy window lower threshold (keV)

108.5

108.5

Secondary energy window upper threshold (keV)

129.5

129.5

Frames

120

120

Angle step (degrees)

3

3

Rotation (degrees)

360

360

Frame duration ventilation SPECT (sec)

10

12

Frame duration perfusion SPECT (sec)

5

5

  1. LEAP, Low-energy, all purpose; keV, kiloelectron volts; sec, seconds