Abstract

Background

Aiming to measure the difference in arrival times of two coincident γ-photons with an accuracy in the order of 200ps, time-of-flight positron emission tomography systems commonly employ silicon photomultipliers (SiPMs) and high-resolution digitization electronics, application specific integrated circuits (ASICs). This work evaluates the performance of the TOFPET2 ASIC, released by PETsys Electronics S.A. in 2017, dependent on its configuration parameters in multi-channel coincidence measurements.

Methods

SiPM arrays fabricated by different vendors (KETEK, SensL, Hamamatsu, Broadcom) were tested in combination with the ASIC. Scintillator arrays featuring different reflector designs and different configurations of the TOFPET2 ASIC software parameters were evaluated. The benchtop setup used is provided with the TOFPET2 ASIC evaluation kit by PETsys Electronics S.A.

Results

Compared to existing studies featuring the TOFPET2 ASIC, multi-channel performance results dependent on a larger set of ASIC configuration parameters were obtained that have not been reported to this extend so far. The ASIC shows promising CRTs down to 219.9 ps in combination with two Hamamatsu S14161-3050-HS-08 SiPM arrays (128 channels read out, energy resolution 13.08%) and 216.1 ps in combination with two Broadcom AFBR-S4N44P643S SiPM arrays (32 channels read out, energy resolution 9.46%). The length of the trigger delay of the dark count suppression scheme has an impact on the ASIC performance and can be configured to further improve the coincidence resolution time. The integrator gain configuration has been investigated and allows an absolute improvement of the energy resolution by up to 1% at the cost of the linearity of the energy spectrum.

Conclusion

Measuring up to the time-of-flight performance of state-of-the-art positron emission tomography (ToF-PET) systems while providing a uniform and stable readout for multiple channels at the same time, the TOFPET2 ASIC is treated as promising candidate for the integration in future ToF-PET systems.

Background

A functional imaging modality widely used in the diagnosis and staging of cancer as well as in cardiology or neurology is positron emission tomography (PET) [13]. After an injected tracer undergoing a β+-decay, the emitted positron annihilates with an electron in the surrounding tissue, resulting in the back-to-back release of two γ-photons. The γ-photons are detected by two opposing elements of a ring-shaped detector, each consisting of a scintillator array coupled to a photo-detector array. Via a scintillation process, the γ-photons are stopped and converted in to optical photons, which then reach the employed photo-detector.

In time-of-flight positron emission tomography (ToF-PET), the difference in arrival times of the two detected γ-photons can be resolved, which results in a more precise localization of the annihilation event along the line of response (LOR) connecting the two points of detection [4, 5]. The coincidence resolution time (CRT) assesses the capability of a PET system to resolve this ToF information. Studies have shown that incorporating ToF information into the image reconstruction process increases the signal-to-noise ratio (SNR) of a PET image and therefore improves the image quality [68]. For a precise measurement of the ToF information, CRTs in the order of few hundred picoseconds are required. A low energy resolution in the order of 10% around 511 keV is of advantage to filter true coincidences from scattered events.

State-of-the-art clinical PET systems reach CRTs ranging from 214 ps to 500 ps. The achieved energy resolutions lay between 9% and 12% [5, 9, 10]. For prototype PET systems, CRTs between 200 ps to 450 ps and energy resolutions between 11 and 12% are reported [5, 11]. For both cases, future systems aim to reach CRTs below 200 ps [1214]. On benchtop level, this limit has already been excelled by various setups reaching CRTs below 100 ps [1517].

Scintillators used in PET systems require a high stopping power, a short scintillation decay time, and high photon statistics [18]. Reflective foils or mixtures of glues and powders can be used to optically segment an array of scintillators. Here, various scintillator topologies can be considered, namely one-to-one coupling between single scintillator needles and photo-sensor channels, multiple scintillator needles on one photo-sensor channel as well as monolithic scintillator blocks or slabs on an array of photo-sensor channels. The latter ones allow for depth-of-interaction (DOI) positioning [1924].

In common state-of-the-art systems, photo-detectors such as photo-multiplier tubes (PMTs) and analog or digital silicon photomultipliers (SiPMs) and custom-designed readout electronics are employed. SiPMs became popular due to the high photo-detection efficiency (PDE) (up to 50%, some up to 65%), their high internal gain, their compactness, their fast response time, and their compatibility with magnetic fields, allowing simultaneous PET/MR imaging [17, 2528]. SiPMs consist of several thousands single-photon avalanche diodes (SPADs) which are connected in parallel and operated in Geiger-mode. An incident optical photon hitting a SPAD causes a self-sustaining charge carrier avalanche. The avalanche effect is used for signal amplification enabling the detection of single optical photons, whereby individual SPAD signals are overlaid to a sum signal forming an SiPM pulse with a steep rising edge [25]. The first prototype system making use of digital SiPMs (dSiPMs) featuring direct binary digitization has been developed in recent years [2933].

Typically, application-specific integrated circuits (ASICs) are employed to precisely digitize analog SiPM signals. This study specifically focuses on the application of ASICs in ToF-PET systems. Apart from exclusive ToF-PET systems, also hybrid imaging systems combining ToF-PET with other modalities, such as magnetic resonance imaging (MRI), electroencephalography (EEG) or ultrasound imaging (US), feature ASICs as readout and digitization electronics (MADPET4 [34, 35], TRIMAGE [3638], EndoToFPET-US [13, 39]). ASICs typically feature a comparator with a low threshold to trigger on the rising edge of the SiPM pulse, i.e., on the first optical photons arriving, and a time-to-digital converter (TDC) to assign a timestamp to an event. The energy of the respective event can then be measured either by signal integration featuring capacitors (qdc-method) or by measuring the time over a specified threshold (tot-method) [13, 40, 41]. The Weeroc series [36, 37, 42], the PETA series [27, 4348] and the TOFPET2 series by PETsys Electronics S.A. [49, 50] use such a qdc-method to measure the signal energy.

In this work, we evaluate the multi-channel performance of the TOFPET2 ASIC (version 2b), released by PETsys Electronics S.A. in 2017 [5153]. Existing studies on the TOFPET2 ASIC performance deal with simulations or experimental results for single SiPMs in combination with the ASIC, which greatly differs to the experimental multi-channel results we are providing [12, 52, 54]. Further studies show multi-channel results that are restricted to a very small parameter space or focus on the DOI resolution of a certain scintillator topology obtained with TOFPET2 readout [53, 5557]. They do not report experimental performance results dependent on the TOFPET2 configuration parameters, e.g., on the delay line configuration or the integrator gain, to the extend we do.

The goal of this study is to assess whether the TOFPET2 ASIC should be considered as solution for digitizing and processing analog SiPM signals in future (whole-body) ToF-PET systems. The presented multi-channel studies are necessary to investigate detector-related effects such as crosstalk and light-sharing as well as the channel-spread of the ASIC performance handling data from multiple channels at once, which cannot be assess in single-channel measurements. The compatibility of the ASIC with different SiPM types in combination with different materials for scintillator segmentation was tested. In addition, the parameters of the ASIC configuration related to the trigger circuit were varied. The impact of the hard- and software configuration on the ASIC performance was quantified.

Materials and methods

TOFPET2 ASIC

The TOFPET2 ASIC is characterized by its compactness (14 mm × 14 mm chip size including bonding area [58]), 128 readout channels, low power consumption [49, 59] and a high hit rate of up to 480 kcps per channel, which corresponds to a data rate of 640 Mbit/s [49, 54]. The maximum output data rate of the ASIC, i.e., 64 channels, is specified as 2.8 Gbit/s [49].

Each of the 64 individual channels is multi-buffered by four analog buffers and employs a three-threshold trigger logic with two discriminators D_T1 and D_T2 in the timing and one discriminator D_E in the energy branch [49, 60, 61]. A global- and a channel-specific configuration register allow to change the ASIC configuration. The ASIC can be operated in a time-over-threshold (tot) or an energy integration (qdc) mode for energy measurement. The charge-to-digital converter (QDC) used in qdc mode behaves linear for integration charges up to 1500 pC [54]. The TDC has a resolution of 30 ps. The chip runs with a clock cycle of 200 MHz.

Incoming SiPM signals are amplified by a trans-impedance ampliflier in the timing branch (nominal gain 3000 Ω) and a transimpedance ampliflier energy branch (nominal gain 300 Ω), respectively [49]. As determined in an oscilloscope measurement, the pre-amplified SiPM pulses in the timing branch have a signal height of about 300 mV. The thresholds of the three discriminators can be adjusted via the three dimensionless parameters vth_t1, vth_t2, and vth_e in the ASIC configuration. Increasing these parameters by one digital-to-analog converter (DAC) step is equal to increasing the trigger level by approximately 2.5 mV, 15 mV, and 20 mV, respectively, over a baseline set during calibration [62]. For a proper operation of the trigger logic [52], it has to be ensured that the voltage thresholds at the discriminators Vth_T1 and Vth_T2 fulfill Vth_T1<Vth_T2. The trigger circuit of each channel enables dark count rejection and high timing resolution by triggering on a low voltage threshold with the first discriminator D_T1 at a very early point in time, i.e., on the first optical photons hitting the SiPM. This trigger is delayed by a specified delay period and passes an AND gate opened by a second trigger activated on a higher voltage threshold with a second discriminator D_T2 (see Fig. 1). The delay period is in the order of few nanoseconds and can be configured by adjusting the parameter fe_delay of the ASIC configuration [49]. The design of the trigger logic allows the rejection of small noise pulses, but is associated with the generation of satellite peaks in the coincidence time difference spectra which are caused by a shift in timestamp generation from the delayed D_T1 to the non-delayed D_T2. A detailed description on the operation of the trigger circuit and the generation of satellite peaks is given in [52].

Fig. 1
figure 1

Schematic drawing of the TOFPET2 channel trigger circuit. In the timing branch of the trigger circuit, the output of the first discriminator D_T1 is delay by a configurable delay element with respect to the output of the second discriminator D_T2. The design allows the rejection of small noise pulses

Benchtop setup

To evaluate the TOFPET2 ASIC in combination with different SiPM types, we used the TOFPET2 ASIC evaluation kit designed by PETsys Electronics S.A. as it enables the user to test various ASIC-SiPM combinations under benchtop conditions. The kit comes along with two ASIC test boards each allowing to connect an SiPM array to a TOFPET2 ASIC via two SAMTEC connectors. A high-voltage digital-to-analog converter (HV-DAC) mezzanine board (version 08/2016) providing the bias voltage for the employed SiPMs, and a front end board (FEB/D) generating the global clock signal and holding the main power supply for the ASIC test boards are included as well. Data transmission to a readout computer is conducted via a 1-Gigabit Ethernet (GbE) mezzanine board. The ASIC test boards provide direct coupling between the employed SiPMs and the ASIC.

Additionally, a breadboard for mounting the setup is provided with the kit (see Fig. 2). To conduct coincidence experiments, the breadboard allows to position both ASIC test boards face-to-face in different distances to the source holder mounting position. Featuring two cable inlets, which were used to connect the ASIC test boards to the FEB/D board via two flexible cables, the breadboard can be encapsulated light-proof by a top cover. A temperature sensor and a proportional–integral–derivative (PID) controller regulating a thermo-element were used to adjust the temperature insight the box. In addition, the whole benchtop setup was placed in a climate chamber to ensure a stable temperature control. A custom-designed source holder was employed to assemble the multi-source geometry. Molds were used to assemble the crystal-SiPM array configurations (see Fig. 3).

Fig. 2
figure 2

Benchtop setup. Optical breadboard holding two ASIC test boards equipped with KETEK PA3325-WB-0808 SiPMs for coincidence experiments. The whole setup can be enclosed with a top cover featuring thermo-control and operated inside a climate chamber. Graphic reprinted from [59]

Fig. 3
figure 3

Assembly tools. Molds used to assemble the detector stacks composed of a segmented scintillator array, a coupling material, and an SiPM array. Each stack was wrapped in teflon tape to prevent light-loss

Detectors

To test the compatibility and performance of the TOFPET2 ASIC with different analog SiPM types, we used 8 ×8 SiPM arrays fabricated by different vendors. Two samples each of KETEK PA3325-WB-0808, SensL ArrayJ-30020-64P-PCB, Hamamatsu S14161-3050-HS-08, and Broadcom AFBR-S4N44P643S (see Table 1) were used to perform coincidence experiments. A multi-source geometry of five 22Na NEMA cube point sources with a total activity of approximately 3 MBq was placed in the center of the setup during these experiments. For the SensL ArrayJ-30020-64P-PCB SiPM arrays, additional adapter boards needed to be employed to establish the connection to the ASIC test board.

Table 1 Characteristics of the investigated analog SiPM types. Parameters are taken from the respective data sheet

Using Sylgard® 527, a two-component dielectric gel fabricated by Dow Corning for optical coupling, each SiPM array was successively one-to-one-coupled to 8 × 8 or 4 × 4 LYSO scintillator arrays. These arrays feature different segmentation layers (360 µm or 110 µm BaSO4 powder mixed with epoxy and 67 µm glued enhanced specular reflector (ESR) foil. The size of an individual scintillator needle in the 8 × 8 ESR array is 3.28 mm × 3.28 mm × 12.00 mm. The size of an individual scintillator needle in the 4 × 4 ESR array is 3.85 mm × 3.85 mm × 12.00 mm. For thicker reflector layers, the scintillator width is reduced by the additional width of the reflector. The selected scintillator length was chosen to reduce the influence of time jitter related light-transport effects accompanying long scintillator needles [63], and thus, be more sensitive to ASIC-related effects on the timing performance. The chosen scintillator length can be found in some pre-clinical and clinical PET systems [5, 33]. A performance degradation is expected if operating the TOFPET2 ASIC in combination with longer scintillator needles. To give a benchmark of this degradation, additional measurements were conducted using a 4 × 4 ESR array consisting of 2.62 mm × 2.62 mm × 19.00 mm scintillator needles.

Setup calibration

The setup was calibrated using the PETsys calibration routine implemented in the software coming along with the evaluation kit [61]. For each SiPM type, a calibration was run once at default ASIC configuration applying an overvoltage of 4 V at an environment temperature of 16C. The required bias voltages were specified depending on the SiPM type employed. It is not necessary to calibrate the setup for every threshold or overvoltage change. A radioactive source can be left in the setup during calibration since this does not affect the determined parameters [62].

Parameter studies

For each SiPM-crystal configuration, the discriminator threshold vth_t1 was varied between 10 and 50 in steps of 10, while vth_t2 and vth_e were kept at constant values (vth_t2=20, vth_e=15). This roughly corresponds to triggering between the 1 p.e. to 3 p.e. (photoelectron) as it is reported for the example of a Hamamatsu S14161-3050-HS-08 SiPM array in [59]. For each setting, raw data were acquired for 120 s at an overvoltage of 2.75 V to 7.75 V. Additionally, the crystal top-to-source distance was varied between 18 mm to 58 mm in steps of 20 mm.

For the KETEK PA3325-WB-0808 arrays coupled to the scintillator arrays featuring 360 µm BaSO4 as segmentation layer, the trigger delay period was configured as different lengths in the range of few nanoseconds (0.0 ns to 12.9 ns) for vth_t1=10 and vth_t1=50 and overvoltages between 2.75 V to 7.75 V. For the Hamamatsu S14161-3050-HS-08 arrays coupled to the scintillator arrays featuring 360 µm BaSO4 as segmentation layer, the integrator gain settings GQ1 and GQ2 were varied between 0.32-2.25 and 1.00-1.68, respectively, to evaluate the linearity of the acquired energy value spectra and the influence on the resulting energy resolution. For each setting, raw data were acquired for 120 s at an overvoltage of 4.75 V and with vth_t1=20, vth_t2=20, and vth_e=15. Effects due to the assembly of SiPM and scintillator array can be excluded since groups of measurements varying the trigger delay and gain configuration as well as the discriminator thresholds are performed with the same assembly directly after each other.

Data collection and processing

For data collection, we used the data acquisition routine implemented by PETsys. We prepared the acquired raw data with the convert_raw_to_singles method implemented by PETsys and provided with the evaluation kit [61]. Using this routine, raw data were converted into single-raw-hit information by applying the acquired calibration data. A table containing a timestamp, an energy value, and a channel id for each single event acquired was returned. The returned energy value does not correspond to ADC bins on hardware level, but is given in an arbitrary unit.

These single events were further processed using an analysis software developed at our institute [52]. The energy values returned by the PETsys routine were sorted into channel-individual energy value histograms. Afterwards, the noise background of this spectrum was estimated and subtracted [6466]. The positions of the 511 keV and 1274.5 keV peaks were determined applying a Gaussian peak finder routine and an iterative Gaussian fitting method after subtracting the background of the spectrum [67]. A simple saturation model neglecting offsets and optical crosstalk was applied to the found positions, parametrizing the saturation corrected energy as

$$ E = c \cdot s \cdot \log \left(\frac{1}{1 - \frac{e}{s}} \right) $$
(1)

where E is the energy in keV, c is a correction factor in calibrated energy units −1, s is a saturation factor in calibrated energy units, and e is the acquired energy value in calibrated energy units [52], similar to [12]. Now, all acquired energy values are converted and sorted into channel-individual energy histograms. An energy filter ranging from 400 keV to 700 keV was applied to filter true coincidences, which should have an energy around 511 keV, from scattered events. A Gaussian was fitted to the remaining histogram data corresponding to the 511 keV peak. The energy resolution was determined as the full width at half maximum (FWHM) of the fitted Gaussian summing up all channel-individual energy histograms to a global energy histogram. The fit range was iteratively adjusted by 10% of the FWHM. Filtered events were checked for coincidences applying a coincidence window of 7.5 ns. To evaluate the effects of different trigger delay configurations, this window was extended to 35 ns. The time difference between two events matched as a coincidence was calculated. All computed time differences were filled into a time difference histogram. A Gaussian was fitted to the histogram peak while iteratively adjusting the fit range by 10% of the FWHM. The coincidence resolution time (CRT) was determined as the FWHM of the fitted Gaussian. Additionally, the full width at tenth maximum (FWTM) is computed to get information about the tail behavior of the peak. A Gaussian excess factor gaussexc is used to specify the outgrow of the tails with reference to their expected behavior. This factor is defined as [24]

$$ \mathsf{gaussexc} = \frac{\mathsf{FWTM} - \frac{\mathsf{FWHM}}{2.355} \cdot 4.294}{\mathsf{FWTM}} $$
(2)

where FWHM/2.355·4.294 computes the theoretical value of the FWTM for a truly Gaussian distribution. The obtained performance parameters were always plotted as a function of the offset-corrected overvoltage \(U_{\mathsf {ov\_cor}}\) computed via

$$ U_{\mathsf{ov\_cor}} = U_{\mathsf{ov,set}} - U_{\mathsf{off}} $$
(3)

The overvoltage voltage set Uov,set via the acquisition routine was corrected by an offset Uoff (approximately 750 mV [68]).

Results

The energy value spectra of different SiPM types allow to determine 511-keV and 1275-keV peak positions for all operated channels. The peaks are shifted to higher energy values for higher overvoltages (see Fig. 4a, b). The spread of the positions observed for individual channels increases for higher overvoltages and the peak positions are sorted according to the SiPM gains (see Table 1). In addition, the peaks are shifted to higher energy values if a thicker segmentation layer is used in the scintillator array and if BaSO4 is used instead of ESR (see Fig. 4c, d). The spatial structure of the channel spread of the peak positions as well as the determined energy resolution and CRT is shown exemplary for a Hamamatsu S14161-3050-HS-08 array in Fig. 5. The channel spread in peak position and energy resolution shows a Gaussian behavior with few outliers. For the CRTs, which were calculated as mean of the CRTs of the respective channel with its coincident channels, a longer tail of the distribution is visible towards lower values.

Fig. 4
figure 4

Spread in gain for 128 ASIC channels coupled to SiPM arrays. A geometry of five NEMA cubes (22Na sources) with a total activity of approximately 3 MBq is employed. a Position of the 511-keV peak in the energy value spectrum for different SiPM types. b Position of the 1275-keV peak in the energy value spectrum. Measurements were conducted using 110 µm BaSO4 as segmentation layer and SiPM arrays from Broadcom, KETEK, SensL, and Hamamatsu. c Position of the 511-keV peak in the energy value spectrum for different segmentation layers. d Position of the 1275-keV peak in the energy value spectrum for different segmentation layers. Measurements were conducted with two KETEK PA3325-WB-0808 SiPM arrays. All positions are plotted against the applied offset-corrected overvoltage \(\mathsf {U_{ov\_cor}}\)

Fig. 5
figure 5

Uniform channel performance. A geometry of five NEMA cubes (22Na sources) with a total activity of approximately 3 MBq is employed. a Position of the 511-keV peak in the energy value spectrum. b Position of the 1275-keV peak in the energy value. c Coincidence resolution time (CRT) computed as the mean of the CRTs of this channel with all its coincident channels. d Energy resolution (dE/E). Measurements were conducted with two Hamamatsu S14161-350-HS-08 SiPM arrays at 4 V overvoltage and vth_t1=50. BaSO4 powder (360 µm) mixed with an epoxy was used as segmentation layer. The scintillator array used consists of 8 × 8 needles with a size of 3 mm × 3 mm × 12 mm

Determining these performance parameters globally, operating two Hamamatsu S14161-3050-HS-08 SiPM arrays at a range of overvoltages shows an improvement in CRT (FWHM and FWTM) and energy resolution for higher overvoltages before deteriorating again (see Fig. 6). The reported Gaussian excess factor gaussexc show an increased contribution of non-Gaussian tails to the time difference spectra for higher overvoltages (see Fig. 6c). Triggering on higher thresholds vth_t1 further improves the CRT (FWHM and FWTM). The energy resolution is rarely affected by this parameter change. Generally, these effects are observed for all investigated SiPM types and segmentation layers. The point of transition from the negative to the positive influence of higher overvoltages and discriminator thresholds vth_t1 varies according to the SiPM type and scintillator array used.

Fig. 6
figure 6

Impact of the trigger threshold on CRT and energy resolution. Measurements were conducted with two Hamamatsu S14161-3050-HS-08 each coupled to an 8 × 8 12-mm-high LYSO scintillator array featuring 360 µm BaSO4 powder mixed with epoxy as segmentation layer. Single scintillator needles had a size of 3 mm × 3 mm × 12 mm. A geometry of five NEMA cubes (22Na sources) with a total activity of approximately 3 MBq is employed. a Coincidence resolution time (CRT - FWHM). b Coincidence resolution time (CRT - FWTM). c Gaussian excess factor (gaussex). d Energy resolution (dE/E)

To compare the influence of the investigated parameters, the lowest CRT (FWHM) achieved with each configuration along the investigated range of overvoltages is reported together with the corresponding energy resolution in Tables 2 and 4. Depending on the SiPM type, changing the threshold vth_t1 results in a CRT (FWHM) improvement of approximately 3 ps to 9 ps (approximately 1% to 3% relative improvement). Triggering on a higher threshold, also leads to slightly lower Gaussian excess factors, i.e., a more Gaussian behavior of the tails of the coincidence time difference histogram, especially for KETEK and SensL SiPMs (see Table 3). Among the different scintillator arrays, both scintillator types employing BaSO4 powder mixed with epoxy as segmentation layer outperform the type featuring ESR foil. A segmentation layer of 360 µm BaSO4 reaches approximately 100 ps to 200 ps lower CRTs (FWHM) than a segmentation layer of 67 µm ESR foil (see Table 2). This is equal to a relative performance gain of approximately 29% to 43% depending on the SiPM type. The CRT (FWTM) shows a similar behavior (see Table 3). Comparing the corresponding energy resolutions, an absolute improvement of up to approximately 2.5% can be reported (see Table 4). Here, the thicker BaSO4-layer of 360 µm shows better performance results (approximately 6% to 14% relative performance gain) than the thinner layer of 110 µm. For different segmentation layers, different mean channel raw hit rates and mean channel coincidence rates, respectively, are reported (see Fig. 7a, b).

Fig. 7
figure 7

Data rates for different setup hardware configurations. a Mean raw hit rate per channel for different reflector materials. b Mean coincidence rate per channel for different reflector materials. c Mean raw hit rate per channel for different distances of the detectors to the source. d Mean coincidence rate per channel for different distances of the detectors to the source

Table 2 Coincidence resolution time (FWHM)
Table 3 Coincidence resolution time (FWTM) and Gaussian excess factors gaussexc
Table 4 Energy resolution

Triggering at larger distances between sensor and the employed multi-source geometry leads to an improvement of the CRT (FWHM) and energy resolution by about 10 ps to 20 ps and 1.7% for a distance variation of 20 mm (see Tables 2 and 4). Moving the detector stacks from the farthest to the closest distance corresponds to a three- to fourfold increase of the mean channel raw hit rate and an about seven-fold increase of the mean channel coincidence rate, measured at 2 V (see Fig. 7c, d).

Due to limited availability of materials, the parameter study was only partly performed for the Broadcom AFBR-S4N44P643S SiPM arrays. Here, triggering on a higher threshold vth_t1 improved the lowest CRT (FWHM) achieved by up to approximately 25 ps, which corresponds to approximately 9% relative performance gain (see Table 5). The corresponding CRT (FWTM) is improved as well. The corresponding energy resolution only showed minor fluctuations less than 0.3% (absolute change). Again, the scintillator type employing BaSO4 powder mixed with epoxy as segmentation layer outperform the type featuring ESR foil regarding the CRT.

Table 5 Supplementary results contributing to the parameter study

Among the four SiPM types tested in combination with the TOFPET2 ASIC, the 8 × 8 Hamamatsu S14161-3050-HS-08 SiPM arrays outperform the other SiPM arrays achieving CRTs (FWHM) down to 219.9 ± 0.7 ps (dE/E = 13.08 ± 0.03%) with an 8 × 8 12-mm-high scintillator array featuring 360 µm BaSO4 mixed with epoxy as segmentation layer for 128 channels read out. Broadcom AFBR-S4N44P643S SiPM arrays reached comparable CRTs down to 216.1 ± 2.6 ps (dE/E = 9.46 ± 0.09%) with a 4 × 4 12-mm-high scintillator array featuring a thinner BaSO4 layer (110 µm) for 32 channels read out. With a 4 × 4 19-mm-high scintillator array featuring 155 µm air-coupled ESR-foil as the inter-crystal layer, the Hamamatsu arrays reached CRTs (FWHM) down to 247.5 ± 2.4 ps (dE/E = 10.46 ± 0.08%) for 32 channels read out.

Satellite peaks, which were observed for measurements with single SiPMs [52], also appear in the coincidence time difference spectra of multi-channel measurements (see Fig. 8). The peaks are generated by small noise pulses that trigger the first discriminator and are validated by a true coincidence event triggering on the second discriminator occurring shortly after the noise event and while the AND gate connecting first and second discriminator is still active. As Fig. 8 depicts, the shift of the peaks from the center peak in the time difference histogram is related to the configured trigger delay between the first and second discriminator of the ASIC channel circuit. The fraction of events causing the formation of satellite peaks is higher for higher overvoltages and lower discriminator thresholds vth_t1. A broader explanation is given in [52]. As Fig. 9 depicts, changing the delay period between the first and second discriminator, not only leads to the shift of satellite peaks over the range of the coincidence time difference spectrum but also causes a systematic deterioration in energy resolution for shorter delay periods (relative deterioration of about 10%). Additionally, the lowest CRTs (FWHM) achieved vary by up to approximately 30 ps (up to approximately 11% relative change). The lowest FWTM of the coincidence time difference histograms varies by up to approximately 50 ps. For higher overvoltages, the effect of a narrower peaks and less distinct tails in the time difference spectra depending on the delay configurations reverses itself (see Fig. 9b). The computed Gaussian excess factors show that especially for higher overvoltage the influence of non-Gaussian histogram tails become more and more evident (see Fig. 9c). Repeating the measurements and evaluation with different discriminator thresholds vth_t1 or narrower coincidence windows did not lead to a change of these effects.

Fig. 8
figure 8

Satellite peaks in the coincidence time difference spectra for different delay periods. Measurements were conducted with two KETEK PA3325-WB-0808 each coupled to an 8 × 8 12-mm-high LYSO scintillator array featuring 360 µm BaSO4 powder mixed with epoxy as segmentation layer. Data are collected at 4.75 V overvoltage and with vth_t1=10. A geometry of five NEMA cubes (22Na sources) with a total activity of approximately 3 MBq is employed. Red lines indicate the configured delay period [49, 62]. a Delay line bypassed. b 0.39 ns. c 2.95 ns. d 5.8 ns. e 8.4 ns. f 12.9 ns

Fig. 9
figure 9

Performance results dependent on the configured delay period. Data are acquired with two KETEK PA3325-WB-0808 in multi-channel coincidence experiments with vth_t1=50. Each SiPM array is coupled to an 8 × 8 LYSO of 12 mm height scintillator array featuring 360 µm BaSO4 powder mixed with epoxy as the inter-crystal layer. A geometry of five NEMA cubes (22Na sources) with a total activity of approximately 3 MBq is employed. a Coincidence resolution time (CRT - FWHM). b Coincidence resolution time (CRT - FWTM). c Gaussian excess factor (gaussex). d Energy resolution (dE/E)

Figure 10 depicts the linearity ratio computed as the ratio of the 511 keV and 1275 keV peak positions in the ADC value spectra as well as the energy resolution for different integrator gains GQ1 and GQ2. We find that the default integrator gain setting GQ1=1.0 and GQ2=1.0 results in the most linear energy spectra (linearity ratio of 0.71, dE/E ≈ 13%). The energy resolution can be improved by approximately 5% (relative improvement) changing the GQ1-setting to GQ1=2.5 at the cost of energy linearity. The absolute spread in energy resolution for both of these settings is on the order of approximately 0.6%. The spread is generally increased by higher gains, which also result in a deterioration of the energy resolution to up to 14%.

Fig. 10
figure 10

Energy linearity and energy resolution for different integrator gain configurations. The default configuration is GQ1·GQ2=1.0. a Energy linearity computed as the ratio of the 511 keV and 1274.5 keV peak positions in the energy value histogram. b Energy resolution (dE/E) computed using the FWHM of the 511 keV peak in the energy histogram

Discussion

The peak positions reported for the energy value spectra (see Figs. 4 and 5) suggest that all channels on one ASIC operate uniformly. Since especially in Fig. 4 128 channel, i.e., two ASICs, are considered when computing the channel spread, the operation of two different ASICs is uniform as well. Since single channels not matching the peak position pattern were only identified for one of the Hamamatsu S14161-3050-S-08 SiPM arrays (outliers in Figs. 4a, b), but not for other SiPM arrays, this behavior is probably due to the connected SiPMs and not the operation of ASIC channels. This channel was excluded from further evaluation steps. Since all channels operate uniformly, reporting the CRT and energy resolution as globally computed parameters is justified.

In single- and multi-channel studies, a similar behavior of the CRT and energy resolution dependent on the applied overvoltage was observed. With the selected thresholds, a convergence of the CRT (FWHM) can already be observed for SensL and Hamamatsu SiPMs when setting vth_t1=40 and vth_t1=50 (see Table 2). The improvement observed for KETEK SiPMs is less than 2 ps. For Broadcom SiPMs, a slight convergence is visible yet, as the improvement becomes less for higher thresholds (see Table 5). Additionally, one can observe a slight decrease in energy resolution for higher thresholds (see Tables 4 and 5). This can be explained by the fact that only higher pulses, i.e., events with a higher energy, are triggered on with higher thresholds, leading to more compressed energy spectra. Since for most cases, a convergence of the CRT and a deterioration of the energy resolution is observed, we refrained from triggering on higher thresholds. The non-Gaussian tails of the peak in the coincidence time difference histogram remain below extension 10% of the Gaussian peak at the optimal operation point of the respective SiPM and are suspected to be caused by the detection of scattered events. This matches the observation that the tails are more prone for low thresholds vth_t1 and higher overvoltages, increasing the gain of the SiPM, making it more sensitive to noise and low-energy events. Thus, we suspect that the tails could be reduced by applying a narrower energy window. This effect of the energy window on the CRT (FWTM) was investigated for digital SiPMs in [24]. The impact of the discriminator threshold vth_t1 on the achieved performance is stronger in multi-channel than in previous single-channel investigations [52]. While in single-channel measurements with two KETEK PM3325-WB-A0 the effect was only visible for overvoltages lower than 3 V and no reverse effect was observed for higher overvoltages, for multi-channel measurements with two KETEK PA3325-WB-0808, the impact of vth_t1 was visible over the whole investigated overvoltage range and reversed itself for overvoltages higher than 4 V. The observed differences in behavior could be due to effects on the SiPM array such as crosstalk and light-sharing.

Among the different parameters changed, the scintillator segmentation layer has the largest relative impact on the achieved performance (29% to 43%). The effect of improved performance for BaSO4 as reflector material is likely due to a lower probability of light-sharing between channels and thus, a higher fraction of γ-events depositing their whole energy on one SiPM channel. This leads to a higher number of small pulses being acquired resulting in a higher mean raw hit rate and especially mean coincidence rate per channel. The observed higher raw data rate with increasing overvoltage is expected due to the increase of the SiPM gain, crosstalk and photo-detection efficiency with the overvoltage (see Fig. 7a). The coincidence rate is expected and observed to be stable over the investigated overvoltage range (see Fig. 7b). Since no saturation is observed, the performance dependency on the reflector material is probably not a γ-rate effect. In addition, higher energy values are observed for individual channels if thicker segmentation layers are used as shown in Fig. 4c and d. The different reflective behaviors of BaSO4 (diffusive reflection) and ESR foil (specular reflexion) also suspected to contribute to the reduced and increased event rate and more or less light-sharing between single scintillator needles. Setting higher thresholds vth_t2 and vth_e could be used to filter out these events, while still triggering on the first optical photons with a low thresholds vth_t1, which preserve the ToF information of the detected event. An improved performance seen in experiments with larger distances to the employed sources, which also is associated with a lower mean raw hit rate per channel and lower mean coincidence rate per channel (see Fig. 7c, d), additionally indicates a raw-data-rate-dependent performance. Here, a saturation of the mean raw hit rate per channel is visible for the detector position closest to the sources at higher overvoltage (see Fig. 7c). The detected coincidence rate drops accordingly (see Fig. 7d). Since here, the detector configuration stayed the same and only the distance to the source was varied, this is a γ-rate effect. Since the origin of the loss of events is not flagged, it cannot be determined if this drop is due to the ASIC or other parts of the signal chain. PETsys Electronics S.A. states a maximum event rate of up to 480 kcps per channel [49], which is not reached by the detected mean raw hit rate per channel.

Communication with PETsys lead to the assumption that the systematic deterioration in energy resolution with changing the trigger delay period is probably related to the configuration of the energy integration window. Acquiring data in qdc-mode, the integrated charge is corrected by an estimate for a charge offset related to the time difference between trigger_Q, which starts the energy integration, and the delayed trigger_T1, which sets the event timestamp, i.e., the time of arrival of the optical photons hitting the SiPM channel. The latter one is prolonged or shortened with respect to the configuration of the trigger delay period, which results in an adaptation of the charge offset applied as a correction. The energy integration window is not adapted for different trigger delay periods and thus, needs to be adjusted manually. In all investigations so far, this window has been kept at a fixed default value of approximately 290 ns. Further experiments to investigate the influence of manually adjusting the window are on-going. The difference in behavior as a function of overvoltage cannot be understood without profound knowledge of the details of the ASIC implementation. Discussions with PETsys regarding this matter are on-going.

Investigations with two Hamamatsu S14161-3050-HS-08 arrays show that the integrator gain setting can be used to linearize the acquired energy spectra and optimize the energy resolution. The absolute spread in the energy resolution of all investigated channels is in the order of less than 1%. Thus, it can be concluded that the SiPM and ASIC channels operated uniformly in this experiment. In this study, the ASIC is already operated at the most linear setting possible. This setting is recommended to be used for further investigations featuring high-gain SiPMs. Switching to a more non-linear setting to improve the energy resolution at 511 keV is possible as long as the saturation correction method can be applied, i.e., as long as the two peaks of the Na22-spectrum remain separable.

Comparing the performance of different analog SiPM types, Hamamatsu S14161-3050-HS-08 and Broadcom AFBR-S4N44P643S show the lowest CRTs in combination with the TOFPET2 ASIC. Both SiPM types feature a higher gain and a larger SPAD size than the investigated KETEK and SensL SiPMs (see Table 1). However, it has to be kept in mind that this study only reports performance results for the combination of SiPM type and ASIC using the delivered benchtop setup “as-is.” Since the coupling between SiPM and ASIC was not individually optimize for each SiPM type, this study cannot be used to compare the performance of solely the SiPM types among each other, but only their performance in combination with the TOFPET2 ASIC for the specific coupling scheme given by the electronics provided. Individual optimization of the SiPM-ASIC coupling is expected to improve the reported performance [69, 70].

On benchtop level, the TOFPET2 ASIC measures up to the performance of state-of-the-art ASICs in single-channel coincidence experiments. An overview on single-channel experiments is provided in [52]. In comparison with other multi-channel performance experiments, the TOFPET2 ASIC measures up to state-of-the-art system performance regarding its ToF capability. An 18-channel ASIC using a tot-method to digitize energies achieved a CRT of 275 ps and an energy resolution of 11.8% in coincidence experiments with two 12 × 12 Hamamatsu C13500-4075LC-12 modules coupled to 20-mm-high scintillator arrays and read out by 8 ASICs in total [71]. Module tests with the PETA5 ASIC and SiPMs fabricated by Fondazione Bruno Kessler (FBK, RGB-HD technology) achieved and average CRT of 230 ps and an energy resolution of about 14% when one-to-one-coupled to 10-mm-high LYSO scintillator arrays [47]. PETsys reports a CRT of 260 ps between two Hamamatsu S13361-3050AE-04 arrays each one-to-one-coupled to an array of 15-mm-high LYSO needles [12]. On system level, the TRIMAGE scanner equipped with the TRIROC ASIC reaches CRTs of 515 ps and an energy resolution between 20% to 22% employing a two-layered scintillator geometry consisting of LYSO needles with a total height of 20 mm on FBK SiPMs (NUV-HD technology) [72]. A study on a prototype PET scanner employing digital SiPMs (DPC 3200-22 by Philips Digital Photon Counting) coupled to 10-mm-high LYSO scintillator arrays reports CRTs down to 215.2 ± 0.4 ps (energy resolution: 11.381 ± 0.007%) [11]. The Biograph Vision PET/CT (computed tomography) scanner recently released by Siemens also achieves 214 ps on system level, using 3.2 mm × 3.2 mm × 20 mm LSO needles [73, 74]. Further performance studies should include monolithic scintillator blocks as well as slabs to obtain DOI information with the TOFPET2 ASIC. First experiments featuring DOI-capable scintillator geometries have been conducted [22, 55, 56].

In addition, when designing a PET system or a MR-compatible PET insert, one has to keep in mind other system requirements such as power supply limitations or interference problems [75]. The power consumption of the TOFPET2 ASIC and its influence on the ASIC performance were already evaluated in an additional study [59]. Regarding MR-compatibility tests, similar test protocols as in [28, 76, 77] need to be defined and executed.

Outlook and conclusion

Generally, the TOFPET2 ASIC evaluation kit enables the user to easily test various SiPM types and scintillator topologies in combination with the TOFPET2 ASIC.

The ASIC is compatible with all SiPM types tested in this study (see Table 1). We observe that high discriminator thresholds (vth_t1=40−50), thicker reflector layers, and lower data rates (larger distances to the sources and diffusive inter-crystal reflexion) result in an improved performance regarding both CRT and energy resolution. Among the SiPM types used, so far, Broadcom AFBR-S4N44P643S and Hamamatsu S14161-3050-HS-08 SiPMs show the most promising performance results in combination with the TOFEPT2 ASIC.

The appearance of satellite peaks is confirmed for multi-channel time difference spectra. A performance dependency related to the configuration of the delay element is observed. A first indication of a γ-rate-dependent ASIC performance was found in experiments with different distances to a Na22 source. Adjustments of the integrator gain can be used to linearize the energy value spectra acquired in qdc-mode or to improve the energy resolution.

Even without optimizing the SiPM-to-ASIC coupling, the CRTs and energy resolutions reported for a coincidence setup of two detector blocks lay in the order of the performance of clinical and pre-clinical ToF-PET systems, while providing a uniform and stable readout of multiple channels at the same time. Therefore, it will be considered for integration in (whole-body) ToF-PET/MRI applications.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ASIC:

application-specific integrated circuit

BGO:

bismuth germanate

CRT:

coincidence resolution time

CT:

computed tomography DOI: depth of interaction

FBK:

Fondazione Bruno Kessler

FEB:

front end board

FWHM:

full width at half maximum

FWTM:

full width at tenth maximum

GbE:

Gigabit Ethernet

HV-DAC:

high-voltage digital-to-analog converter

LOR:

line of response

LYSO:

lutetium-yttrium oxyorthosilicate

MRI:

magnetic resonance imaging

PDE:

photo detection efficiency PET: positron emission tomography

QDC:

charge-to-digital converter

SiPM:

silicon-photomultiplier

TDC:

time-to-digital converter

ToF:

time-of-flight

References

  1. Bailey DL, Townsend DW, Valk PE, Maisey MN. Positron Emission Tomography: Basic Sciences, 1st ed: Springer Science and Business Media; 2005. https://doi.org/10.1007/b136169.

  2. Phelps ME. PET: Physics, Instrumentation, and Scanners, 1st ed: Springer; 2006. https://doi.org/10.1007/0-387-34946-4.

  3. Vaquero JJ, Kinahan P. Positron Emission Tomography: Current Challenges and Opportunities for Technological Advances in Clinical and Preclinical Imaging Systems. Annu Rev Biomed Eng. 2015; 17(1):385–414. https://doi.org/10.1146/annurev-bioeng-071114-040723.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Surti S, Karp JS. Advances in time-of-flight PET. Phys Med. 2016; 32(1):12–22. https://doi.org/10.1016/j.ejmp.2015.12.007.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Vandenberghe S, Mikhaylova E, D’hoe E, Mollet P, Karp J. Recent developments in time-of-flight PET. EJNMMI Phys. 2016; 3(1). https://doi.org/10.1186/s40658-016-0138-3.

  6. Conti M. Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging. 2011; 38(6):1147–57. https://doi.org/10.1007/s00259-010-1711-y.

    Article  PubMed  Google Scholar 

  7. Gundacker S, Knapitsch A, Auffray E, Jarron P, Meyer T, Lecoq P. Time resolution deterioration with increasing crystal length in a TOF-PET system. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2014; 737:92–100. https://doi.org/10.1016/j.nima.2013.11.025.

    Article  CAS  Google Scholar 

  8. Surti S. Update on time-of-flight PET imaging. J Nucl Med. 2015; 56(1):98–105. https://doi.org/10.2967/jnumed.114.145029.

    Article  PubMed  Google Scholar 

  9. Spencer BA, Berg E, Schmall JP, Omidvari N, Leung EK, Abdelhafez YG, Tang S, Deng Z, Dong Y, Lv Y, Bao J, Liu W, Li H, Jones T, Badawi RD, Cherry SR. Performance evaluation of the uexplorer total-body pet/ct scanner based on nema nu 2-2018 with additional tests to characterize long axial field-of-view pet scanners. J Nucl Med. 2020. https://doi.org/10.2967/jnumed.120.250597.

  10. UC Davis. Explorer Product Website; 2019. https://explorer.ucdavis.edu/about-explorer/. Accessed 16 Jan 2019.

  11. Schug D, Wehner J, Dueppenbecker PM, Weissler B, Gebhardt P, Goldschmidt B, Salomon A, Kiessling F, Schulz V. PET performance and MRI compatibility evaluation of a digital, ToF-capable PET/MRI insert equipped with clinical scintillators. Phys Med Biol. 2015; 60(18):7045. https://doi.org/10.1088/0031-9155/60/18/7045.

    Article  PubMed  Google Scholar 

  12. Bugalho R, Di Francesco A, Ferramacho L, Leong C, Niknejad T, Oliveira L, Pacher L, Rolo M, Rivetti A, Silveira M, et al. Experimental results with TOFPET2 ASIC for time-of-flight applications. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2017. https://doi.org/10.1016/j.nima.2017.11.034.

  13. Chen H, Briggl K, Fischer P, Gil A, Harion T, Munwes Y, Ritzert M, Schimansky D, Schultz-Coulon H-C, Shen W, et al.A dedicated readout ASIC for time-of-flight positron emission tomography using silicon photomultiplier (SiPM). In: Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE. IEEE: 2014. p. 1–5. https://doi.org/10.1109/NSSMIC.2014.7431045.

  14. Rolo MD, Bugalho R, Goncalves F, Rivetti A, Mazza G, Silva JC, Silva R, Varela J. A 64-channel ASIC for TOFPET applications. In: 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC). IEEE: 2012. p. 1460–4. https://doi.org/10.1109/NSSMIC.2012.6551353.

  15. Nemallapudi MV, Gundacker S, Lecoq P, Auffray E, Ferri A, Gola A, Piemonte C. Sub-100 ps coincidence time resolution for positron emission tomography with LSO:Ce codoped with Ca. Phys Med Biol. 2015; 60(12):4635. https://doi.org/10.1088/0031-9155/60/12/4635.

    Article  CAS  PubMed  Google Scholar 

  16. Sarasola I, Nemallapudi MV, Gundacker S, Sánchez D, Gascón D, Rato P, Marín J, Auffray E. A comparative study of the time performance between NINO and FlexToT ASICs. J Instrum. 2017; 12:04016. https://doi.org/10.1088/1748-0221/12/04/P04016.

    Article  Google Scholar 

  17. Gundacker S, Turtos RM, Auffray E, Paganoni M, Lecoq P. High-frequency SiPM readout advances measured coincidence time resolution limits in TOF-PET. Phys Med Biol. 2019; 64(5):055012. https://doi.org/10.1088/1361-6560/aafd52.

    Article  CAS  PubMed  Google Scholar 

  18. Lecoq P. Development of new scintillators for medical applications. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2016; 809:130–9. https://doi.org/10.1016/j.nima.2015.08.041.

    Article  CAS  Google Scholar 

  19. Borghi G, Tabacchini V, Schaart DR. Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation. Phys Med Biol. 2016; 61(13):4904–28. https://doi.org/10.1088/0031-9155/61/13/4904.

    Article  CAS  PubMed  Google Scholar 

  20. Müller F, Schug D, Hallen P, Grahe J, Schulz V. A novel DOI Positioning Algorithm for Monolithic Scintillator Crystals in PET based on Gradient Tree Boosting. IEEE Trans Radiat Plasma Med Sci. 2019; 3(4):465–74. https://doi.org/10.1109/TRPMS.2018.2884320.

    Article  Google Scholar 

  21. Müller F, Schug D, Hallen P, Grahe J, Schulz V. Gradient Tree Boosting-Based Positioning Method for Monolithic Scintillator Crystals in Positron Emission Tomography. IEEE Trans Radiat Plasma Med Sci. 2018; 2(5):411–21. https://doi.org/10.1109/TRPMS.2018.2837738.

    Article  Google Scholar 

  22. Peng P, Judenhofer MS, Cherry SR. Compton PET: a layered structure PET detector with high performance. Phys Med Biol. 2019. https://doi.org/10.1088/1361-6560/ab1ba0. http://iopscience.iop.org/10.1088/1361-6560/ab1ba0.

  23. Gross-Weege N, Schug D, Hallen P, Schulz V. Maximum likelihood positioning algorithm for high-resolution pet scanners. Med Phys. 2016; 43:3049–61. https://doi.org/10.1118/1.4950719.

    Article  PubMed  Google Scholar 

  24. Schug D, Weissler B, Gebhardt P, Schulz V. Crystal Delay and Time Walk Correction Methods for Coincidence Resolving Time Improvements of a Digital-Silicon-Photomultiplier-Based PET/MRI Insert. IEEE Trans Radiat Plasma Med Sci. 2017; 1(2):178–90. https://doi.org/10.1109/TNS.2017.2654920.

    Article  Google Scholar 

  25. Bisogni MG, Morrocchi M. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2016; 809:140–8. https://doi.org/10.1016/j.nima.2015.09.114.

  26. Bisogni MG, Guerra AD, Belcari N. Medical applications of silicon photomultipliers. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2019; 926:118–28. https://doi.org/10.1016/j.nima.2018.10.175.

    Article  CAS  Google Scholar 

  27. Weissler B, Gebhardt P, Lerche CW, Wehner J, Solf T, Goldschmidt B, Mackewn JE, Marsden PK, Kiessling F, Perkuhn M, Heberling D, Schulz V. MR compatibility aspects of a silicon photomultiplier-based PET/RF insert with integrated digitisation. Phys Med Biol. 2014; 59(17):5119–39. https://doi.org/10.1088/0031-9155/59/17/5119.

    Article  PubMed  Google Scholar 

  28. Weissler B, Gebhardt P, Dueppenbecker PM, Wehner J, Schug D, Lerche CW, Goldschmidt B, Salomon A, Verel I, Heijman E, Perkuhn M, Heberling D, Botnar RM, Kiessling F, Schulz V. A digital preclinical pet/mri insert and initial results. IEEE Trans Med Imaging. 2015; 34(11):2258–70. https://doi.org/10.1109/TMI.2015.2427993.

    Article  PubMed  Google Scholar 

  29. Frach T, Prescher G, Degenhardt C, de Gruyter R, Schmitz A, Ballizany R. The digital silicon photomultiplier — principle of operation and intrinsic detector performance. In: 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC): 2009. p. 1959–65. https://doi.org/10.1109/NSSMIC.2009.5402143.

  30. Haemisch Y, Frach T, Degenhardt C, Thon A. Fully Digital Arrays of Silicon Photomultipliers (dSiPM) – a Scalable Alternative to Vacuum Photomultiplier Tubes (PMT). Phys Procedia. 2012; 37:1546–60. https://doi.org/10.1016/j.phpro.2012.03.749.

    Article  CAS  Google Scholar 

  31. Schaart DR, Charbon E, Frach T, Schulz V. Advances in digital sipms and their application in biomedical imaging. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2015; 809:31–52. https://doi.org/10.1016/j.nima.2015.10.078.

    Article  CAS  Google Scholar 

  32. Hallen P, Schug D, Weißler B, Gebhardt P, Salomon A, Kiessling F, Schulz V. PET performance evaluation of the small-animal Hyperion II PET/MRI insert based on the NEMA NU-4 standard. Biomed Phys Eng Express. 2018; 4. https://doi.org/10.1088/2057-1976/aae6c2.

  33. Schug D, Lerche C, Weissler B, Gebhardt P, Goldschmidt B, Wehner J, Dueppenbecker PM, Salomon A, Hallen P, Kiessling F, Schulz V. Initial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology. Phys Med Biol. 2016; 61(7):2851. https://doi.org/10.1088/0031-9155/61/7/2851.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  34. Omidvari N, Topping G, Cabello J, Paul S, Schwaiger M, Ziegler S. MR-compatibility assessment of MADPET4: a study of interferences between an SiPM-based PET insert and a 7 T MRI system. Phys Med Biol. 2018; 63(9). https://doi.org/10.1088/1361-6560/aab9d1.

  35. Omidvari N, Cabello J, Topping G, Schneider FR, Paul S, Schwaiger M, Ziegler SI. PET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2017; 62(22):8671–92. https://doi.org/10.1016/j.nima.2017.11.097.

    CAS  Google Scholar 

  36. Ahmad S, Fleury J, de l Taille C, Seguin-Moreau N, Dulucq F, Martin-Chassard G, Callier S, Thienpont D, Raux L. Triroc: A Multi-Channel SiPM Read-Out ASIC for PET/PET-ToF Application. IEEE Trans Nucl Sci. 2015; 62(3):664–8. https://doi.org/10.1109/TNS.2015.2397973.

    Article  Google Scholar 

  37. Ahmad S, de l Taille C, Fleury J, Seguin-Moreau N, Raux L, Callier S, Martin-Chassard G, Dulucq F, Thienpont D. Triroc, a Versatile 64-Channel SiPM Readout ASIC for Time-of-Flight PET. IEEE Nucl Sci Symp Med Imaging Conf Room-Temperature Semicond Detector Work (NSS/MIC/RTSD). 2016. https://doi.org/10.1109/NSSMIC.2016.8069882.

  38. Sportelli G, Ahmad S, Belcari N, Bisogni MG, Camarlinghi N, Pasquale AD, Dussoni S, Fleury J, Morrochi M, Zaccaro E, Guerra AD. The TRIMAGE PET Data Acquisition System: Initial Results. IEEE Trans Radiat Plasma Med Sci. 2017; 1(2):168–77. https://doi.org/10,1109/TNS.2016.2633237.

    Article  Google Scholar 

  39. Shen W, Briggl K, Chen H, Fischer P, Gil A, Harion T, Ritzert M, Schultz-Coulon HC. STiC – a Mixed Mode Chip for SiPM ToF Applications. 2012 IEEE Nucl Sci Symp Med Imag Conf Rec (NSS/MIC). 2012; N14-37:877–81.

    Article  Google Scholar 

  40. Corsi F, Foresta M, Marzocca C, Matarrese G, Guerra AD. ASIC development for SiPM readout. J Instrum. 2009; 4. https://doi.org/10.1088/1748-0221/4/03/P03004.

  41. Orita T, Koyama A, Yoshino M, Kamada K, Yoshikawa A, Shimazoe K, Sugawara H. The current mode Time-over-Threshold ASIC for a MPPC module in a TOF-PET system. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2017. https://doi.org/10.1016/j.nima.2017.11.097.

  42. Fleury J. PETIROC2A : New measurement results on fast ToF SiPM read-out chip. Beijing: Talk at TIPP 2017; 2017.

    Google Scholar 

  43. Fischer P, Peric I, Ritzert M, Koniczek M. Fast Self Triggered Multi Channel Readout ASIC for Time- and Energy Measurement. IEEE Trans Nucl Sci. 2009; 53(3):1153–8. https://doi.org/10.1109/TNS.2008.2008807.

    Article  Google Scholar 

  44. Fischer P, Peric I, Ritzert M, Solf T. Multi-Channel Readout ASIC for ToF-PET. In: 2006 IEEE Nuclear Science Symposium Conference Record, vol. 4: 2006. p. 2523–7. https://doi.org/10.1109/NSSMIC.2006.354423.

  45. Piemonte C, Gola A, Tarolli A, Fisher P, Ritzert M, Schulz V, Solf T. Performance of FBK SiPMs coupled to PETA3 read-out ASIC for PET application. Elsevier Nucl Inst Methods Phys Res A. 2012; 718:345–6. https://doi.org/10.1016/j.nima.2012.10.012.

    Article  CAS  Google Scholar 

  46. Sacco I, Fischer P, Ritzert M. PETA4: a multi-channel TDC/ADC ASIC for SiPM readout. J Instrum. 2013; 8. https://doi.org/10.1088/1748-0221/8/12/C12013.

  47. Sacco I, Dohle R, Fischer P, Piemonte C, Ritzert M. A compact, high-density gamma-detection module for Time-of-Flight measurements in PET applications. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2015; 824. https://doi.org/10.1016/j.nima.2015.11.004.

  48. Schug D, Gebhardt P, Weissler B, Gross-Weege N, Dey T, Schulz V. Measurements with a PET Coincidence Setup Based on the PETA5 ASIC and FBK RGB-HD SiPMs. In: 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). IEEE: 2017. p. 1–3. https://doi.org/10.1109/NSSMIC.2017.8532761.

  49. PETsys Electronics S.A.TOFPET 2 SiPM readout ASIC (rev.2). PETsys Electronics SA, Taguspark - Lisboa Science and Technology Park, Edificio Tecnologia I, 26, 2740-257 PORTO SALVO, Portugal. 2018. http://www.petsyselectronics.com/.

  50. PETsys Electronics S.A.TOFPET ASIC v1 – Short Data Sheet (Rev. 1.2). PETsys Electronics SA, Taguspark - Lisboa Science and Technology Park, Edificio Tecnologia I, 26, 2740-257 PORTO SALVO, Portugal. 2014. http://www.petsyselectronics.com/.

  51. PETsys Electronics S.A.Website; 2019. http://www.petsyselectronics.com/. Accessed 27 Feb 2019.

  52. Schug D, Nadig V, Weissler B, Gebhardt P, Schulz V. Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC. IEEE Trans Radiat Plasma Med Sci. 2018:1–1. https://doi.org/10.1109/TRPMS.2018.2884564.

  53. Bugalho R, Francesco AD, Ferramacho L, Leong C, Niknejad T, Oliveira L, Rolo M, Silva JC, Silva R, Silveira M, Tavernier S, Varela J. Experimental characterization of the TOFPET2 ASIC. J Instrum. 2019; 14(03):03029–03029. https://doi.org/10.1088/1748-0221/14/03/p03029.

    Article  Google Scholar 

  54. Di Francesco A, Bugalho R, Oliveira L, Pacher L, Rivetti A, Rolo M, Silva J, Silva R, Varela J. TOFPET2: a high-performance ASIC for time and amplitude measurements of SiPM signals in time-of-flight applications. J Instrum. 2016; 11(03):03042. https://doi.org/10.1088/1748-0221/11/03/C03042.

    Article  Google Scholar 

  55. Li M, Abbaszadeh S. Depth-of-interaction study of a dual-readout detector based on TOFPET2 application-specific integrated circuit. Phys Med Biol. 2019. https://doi.org/10.1088/1361-6560/ab3866.

  56. Lamprou E, Gonzalez AJ, Sanchez F, Benlloch JM. Exploring TOF capabilities of PET detector blocks based on large monolithic crystals and analog SiPMs. Phys Med. 2020; 70:10–8. https://doi.org/10.1016/j.ejmp.2019.12.004.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Lamprou E, Gonzalez-Montoro A, Canizares G, Ilisie V, Barrio J, Sanchez F, Gonzalez AJ, Benlloch JM. Characterization of TOF-PET Detectors Based on Monolithic Blocks and ASIC-Readout. arXiv preprint arXiv:1806.08715. 2018.

  58. PETsys Electronic S.A.PETsys TOF ASIC Evaluation Kit (Flyer), v15 ed. 2018. http://www.petsyselectronics.com/. Accessed Oct 2018.

  59. Nadig V, Weissler B, Radermacher H, Schulz V, Schug D. Investigation of the Power Consumption of the PETsys TOFPET2 ASIC. IEEE TRPMS Spec Issue 2019. 2019. https://doi.org/10.1109/TRPMS.2019.2955032.

  60. PETsys Electronics SA.TOFPET2 ASIC Evaluation Kit - Hardware User Guide (v1.2), v1.2 ed. 2018. http://www.petsyselectronics.com/. Accessed May 2018.

  61. PETsys Electronics SA.TOFPET2 ASIC Evaluation Kit - Software User Guide V2018.04, v2018.04 ed. 2018. http://www.petsyselectronics.com/. Accessed Apr 2018.

  62. PETsys Electronics S.A.Personal Communication. Emails, conferences. 2018.

  63. Lecoq P. Pushing the limits in time-of-flight pet imaging. IEEE Trans Radiat Plasma Med Sci. 2017; 1(6):473–85. https://doi.org/10.1109/TRPMS.2017.2756674.

    Article  Google Scholar 

  64. Burgess D, Tervo R. Background estimation for gamma-ray spectrometry. Nucl Inst Methods Phys Res. 1983; 214:431–4. https://doi.org/10.1016/0167-5087(83)90612-9.

    Article  CAS  Google Scholar 

  65. Morháč M, Kliman J, Matoušek V, Veselský M, Turzo I. Background elimination methods for multidimensional coincidence γ-ray spectra. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 1997; 401(1):113–32. https://doi.org/10.1016/S0168-9002(97)01023-1.

    Article  Google Scholar 

  66. Ryan CG, Clayton E, Griffin WL, Sie SH, Cousens DR. SNIP, a statistics-sensitive background treatment for the quantitative analysis of PIXE spectra in geoscience applications. NIM B. 1988; 34(3):396–402. https://doi.org/10.1016/0168-583X(88)90063-8.

    Article  Google Scholar 

  67. Morháč M, Kliman J, Matoušek V, Veselský M, Turzo I. Identification of peaks in multidimensional coincidence γ-ray spectra. Nucl Inst Methods Phys Res A. 2000; 443:108–25. https://doi.org/10.1016/S0168-9002(99)01005-0.

    Article  Google Scholar 

  68. PETsys Electronics SA.TOFPET2 ASIC Evaluation Kit - Software User Guide V2019.01, v2019.01 ed. 2019. http://www.petsyselectronics.com/. Accessed Jan 2019.

  69. Huizenga J, Seifert S, Schreuder F, Dam HT, Dendooven P, Loehner H, Vinke R, Schaart D. A fast preamplifier concept for sipm-based time-of-flight pet detectors. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2012; 695:379–84. https://doi.org/10.1016/j.nima.2011.11.012.

    Article  CAS  Google Scholar 

  70. Acerbi F, Gundacker S. Understanding and simulating sipms. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2019; 926:16–35. https://doi.org/10.1016/j.nima.2018.11.118.

    Article  CAS  Google Scholar 

  71. Goertzen AL, Van Elburg D. Performance characterization of mppc modules for tof-pet applications. IEEE Trans Radiat Plasma Med Sci. 2019; 3(4):475–82. https://doi.org/10.1109/TRPMS.2018.2885439.

    Article  Google Scholar 

  72. Belcari N, Bisogni MG, Camarlinghi N, Carra P, Cerello P, Morrocchi M, Patera A, Sportelli G, Del Guerra A. Design and detector performance of the pet component of the trimage pet/mr/eeg scanner. IEEE Trans Radiat Plasma Med Sci. 2019; 3(3):292–301. https://doi.org/10.1109/TRPMS.2019.2906407.

    Article  Google Scholar 

  73. Siemens Healthcare. Biograph Vision Product Website. Siemens Healthcare GmbH, Henkestr. 127, 91052 Erlangen, Germany. https://www.siemens-healthineers.com/molecular-imaging/pet-ct/biograph-vision. Accessed 04 Sept 2019.

  74. Siemens Healthineers - Biograph Vision. Website. 2019. https://www.siemens-healthineers.com/nl/molecular-imaging/pet-ct/biograph-vision. Accessed 12 Nov 2019.

  75. Vandenberghe S, Marsden PK. PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging. Phys Med Biol. 2015; 60(4). https://doi.org/10.1088/0031-9155/60/4/R115.

  76. Wehner J, Weissler B, Dueppenbecker PM, Gebhardt P, Goldschmidt B, Schug D, Kiessling F, Schulz V. MR-compatibility assessment of the first preclinical PET-MRI insert equipped with digital silicon photomultipliers. Phys Med Biol. 2015; 60(6):2231–55. https://doi.org/10.1088/0031-9155/60/6/2231.

    Article  CAS  PubMed  Google Scholar 

  77. Wehner J, Weissler B, Dueppenbecker P, Gebhardt P, Schug D, Ruetten W, Kiessling F, Schulz V. PET/MRI insert using digital SiPMs: Investigation of MR-compatibility. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2014; 734:116–21. https://doi.org/10.1016/j.nima.2013.08.077.

    Article  CAS  Google Scholar 

  78. KETEK GmbH. Product Data Sheet SiPM – Silicon Photomultiplier Array PA3325-WB-0808. 2017. https://www.ketek.net/. Accessed 10 Aug 2018.

  79. SensL. J-Series - High PDE and Timing Resolution SiPM Sensors in a TSV Package. 2017. https://www.sensl.com/. Accessed 01 Aug 2018.

  80. Hamamatsu. Product Flyer MPPC for scintillation S14160/S14161 series. 2017. https://www.hamamatsu.com/. Accessed 29 Oct 2018.

  81. Broadcom. AFBR-S4N44P643 - 8x8 NUV-HD Silicon Photo Multiplier Array - Target Data Sheet. 2018. https://www.broadcom.com/. Accessed 29 Oct 2018.

Download references

Acknowledgements

We thank Ricardo Bugalho and Luis Ferramacho from PETsys Electronics S.A. for kindly answering our many questions. We thank Torsten Solf from Philips Digital Photon Counting for sharing his expertise and for providing additional supplies.

Funding

This project has received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No 667211. Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

All authors participated in the design of the study. VN contributed the experiments and data evaluation and wrote the manuscript. DS established a framework for data analysis. BW designed additional electronic parts for the setup. All authors contributed to the manuscript by discussing and revising its content. All authors gave their approval for the final version of the manuscript.

Corresponding author

Correspondence to Volkmar Schulz.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visithttp://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nadig, V., Schug, D., Weissler, B. et al. Evaluation of the PETsys TOFPET2 ASIC in multi-channel coincidence experiments. EJNMMI Phys 8, 30 (2021). https://doi.org/10.1186/s40658-021-00370-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-021-00370-x

Keywords