From: Evaluation of the PETsys TOFPET2 ASIC in multi-channel coincidence experiments
 | KETEK | SensL | Hamamatsu |
---|---|---|---|
 | PA3325-WB-0808 | Array-J-30020-64P-PCB | S14161-3050-HS-08 |
Threshold vth_t1 | Â | Â | Â |
10 | 273.8±0.7 ps | 265.2±0.6 ps | 224.1±0.7 ps |
20 | 271.4±0.7 ps | 265.4±0.6 ps | 221.2±0.7 ps |
30 | 268.6±0.7 ps | 263.7±0.6 ps | 221.1±0.7 ps |
40 | 266.1±0.6 ps | 262.8±0.6 ps | 219.9±0.7 ps |
50 | 264.6±0.6 ps | 262.4±0.6 ps | 220.0±0.7 ps |
Segmentation layer | Â | Â | Â |
BaSO4 (360 µm) | 264.6±0.6 ps | 262.4±0.6 ps | 220.0±0.7 ps |
BaSO4 (110 µm) | 282.8±0.6 ps | 306.5±0.6 ps | 242.9±0.5 ps |
ESR (67 µm, glued) | 371.4±0.8 ps | 463.4±0.8 ps | 328.1±0.7 ps |
Distance | Â | Â | Â |
18 mm | 277.3±0.4 ps | - | - |
38 mm | 264.6±0.6 ps | 262.4±0.6 ps | 220.0±0.7 ps |
58 mm | 245.3±0.8 ps | 253.3±0.8 ps | 224.4±0.7 ps |
Clinical (19 mm height) | Â | Â | Â |
ESR (155 µm, air-coupled) | 317.3±1.9 ps | 325.1±1.7 ps | 247.5±2.4 ps |