Validation of image-derived input function using a long axial field of view PET/CT scanner for two different tracers

Abstract

Background

Accurate image-derived input function (IDIF) from highly sensitive large axial field of view (LAFOV) PET/CT scanners could avoid the need of invasive blood sampling for kinetic modelling. The aim is to validate the use of IDIF for two kinds of tracers, 3 different IDIF locations and 9 different reconstruction settings.

Methods

Eight [18F]FDG and 10 [18F]DPA-714 scans were acquired respectively during 70 and 60 min on the Vision Quadra PET/CT system. PET images were reconstructed using various reconstruction settings. IDIFs were taken from ascending aorta (AA), descending aorta (DA), and left ventricular cavity (LV). The calibration factor (CF) extracted from the comparison between the IDIFs and the manual blood samples as reference was used for IDIFs accuracy and precision assessment. To illustrate the effect of various calibrated-IDIFs on Patlak linearization for [18F]FDG and Logan linearization for [18F]DPA-714, the same target time-activity curves were applied for each calibrated-IDIF.

Results

For [18F]FDG, the accuracy and precision of the IDIFs were high (mean CF ≥ 0.82, SD ≤ 0.06). Compared to the striatum influx (Ki) extracted using calibrated AA IDIF with the updated European Association of Nuclear Medicine Research Ltd. standard reconstruction (EARL2), Ki mean differences were < 2% using the other calibrated IDIFs. For [18F]DPA714, high accuracy of the IDIFs was observed (mean CF ≥ 0.86) except using absolute scatter correction, DA and LV (respectively mean CF = 0.68, 0.47 and 0.44). However, the precision of the AA IDIFs was low (SD ≥ 0.10). Compared to the distribution volume (VT) in a frontal region obtained using calibrated continuous arterial sampler input function as reference, VT mean differences were small using calibrated AA IDIFs (for example VT mean difference = -5.3% using EARL2), but higher using calibrated DA and LV IDIFs (respectively + 12.5% and + 19.1%).

Conclusions

For [18F]FDG, IDIF do not need calibration against manual blood samples. For [18F]DPA-714, AA IDIF can replace continuous arterial sampling for simplified kinetic quantification but only with calibration against arterial blood samples. The accuracy and precision of IDIF from LAFOV PET/CT system depend on tracer, reconstruction settings and IDIF VOI locations, warranting careful optimization.

Introduction

The standardized uptake value (SUV) is a widely used semi-quantitative metric referring to the measurement of tracer uptake based on a single-time-point [1]. However, even for the most widespread tracer 18F-fluorodeoxyglucose ([18F]FDG), there are still concerns regarding its accuracy for quantification of uptake in lesions, mostly due to the variation of the availability of [18F]FDG in the blood, which is poorly accounted in the SUV calculation [2].

Dynamic PET imaging offers information about the tracer concentration over time and enables the extraction of kinetic parameters such as the net influx rate (Ki) for tracers with irreversible binding or the distribution volume (VT) for tracers with reversible binding. Compared to SUV, studies have shown that Ki is a more robust and accurate index of glucose metabolic rate for [18F]FDG [3, 4]. However, an accurate arterial input function is needed to determine this parameter. The reference method for extracting an accurate arterial input function is the use of a continuous arterial blood sampler and calibrated with several manual arterial samples [5] which is invasive and challenging as arterial blood sampling is laborious and could have some adverse effects [6, 7]. For [18F]FDG, studies using short axial field-of-view (FOV) showed previously that image derived input function (IDIF) from aorta or left ventricle can replace continuous arterial blood sampler [8,9,10] and IDIF from aorta or left ventricle can replace manual blood samples for kinetic modelling [11]. However, the short axial FOV ( 20 cm) PET scanner limits the ability to obtain accurate IDIF because aorta or left ventricle can be located outside the FOV. Due to the limited spatial resolution of the PET scanners, measured activity from small arteries for brain dynamic scans (carotids) or pelvic dynamic scans (iliac or femoral arteries) are affected to spillover and partial-volume effects [12]. Some authors developed correction methods to mitigate the impact of these effects on quantification [12,13,14,15,16,17]. However, application of these methods is scarce. Moreover, in order to perform a whole-body simplified kinetic quantification using a short axial FOV PET system, Karakatsanis et al. developed a method which starts with a single dynamic cardiac-bed position acquisition followed by multiple whole-body acquisitions [18]. However, this method allows only low temporal sampling due to the iterative bed motion in the PET system.

Long axial field-of-view (LAFOV) (> 100 cm) PET scanner may overcome these limitations because major vessels and distant lesions are included in the FOV during the total dynamic acquisition time. LAFOV PET scanner offers up to a 3 times increase in peak axial sensitivity [19], leading to lower image noise and higher temporal sampling, opening up the possibility of better characterization of whole-body pathologies. However, extension of axial field-of-view increases the rates of random and scatter coincidences that could lead to quantification bias. To the best of our knowledge, no validation of using IDIF from the LAFOV PET scanner Biograph Vision Quadra with manual blood samples has been done yet, especially for [18F]FDG which is currently the most widely used radiotracer in clinical PET imaging. Regarding the location of choice for defining IDIF, ascending aorta (AA), descending aorta (DA) or left ventricular cavity (LV) are the most widely used regardless of the tracer applied. However, the IDIF location could impact the accuracy of IDIF and affect the quantification of kinetic parameters [8, 11]. Furthermore, the use of different reconstruction settings (matrix size, post-reconstruction filtering, scatter correction method or number of iterations/subsets) could potentially modify the time-activity curve of IDIF [20].

The aim of this study is to validate the use of IDIF for two kinds of tracers ([18F]FDG for irreversible uptake and [18F]DPA-714 for reversible uptake), 3 different IDIF VOI locations and 9 different reconstruction settings with a LAFOV PET/CT system.

Materials and methods

Study population

Between November 2022 and July 2023, 8 consecutive patients with [18F]FDG injection from a prospective study including patients with various diseases (TBPETCT001 study) and 10 consecutive patients with post COVID syndrome with [18F]DPA-714 injection from the VeCosCO study [21] scanned on the Siemens Biograph Vision Quadra at the Amsterdam UMC, location VUmc (the Netherlands) were included. Informed consent was obtained for all patients. These studies were approved by the Medical Ethics Review Committee of the Amsterdam UMC. The patients with [18F]DPA-714 injection were all high affinity binders based on TSPO genotype analysis.

PET/CT data

Patients received 241 ± 61 MBq of [18F]FDG and 222 ± 27 MBq of [18F]DPA-714. All PET scans were performed after a low dose CT scan acquired without contrast agents. Acquisition of list-mode PET data started immediately after injection of the tracer. For [18F]FDG PET, the 70-min dynamic imaging data were binned into 20 frames (1 × 15 s, 3 × 5 s, 3 × 10 s, 4 × 60 s, 2 × 150 s, 2 × 300 s, 5 × 600 s). For [18F]DPA-714 PET, the 60-min dynamic imaging data were binned into 19 frames (1 × 15 s, 3 × 5 s, 3 × 10 s, 4 × 60 s, 2 × 150 s, 2 × 300 s, 4 × 600 s). For each scan, 9 different PET reconstructions were done using a dedicated offline reconstruction workstation (E7, Siemens Healthineers, Knoxville, TN, USA). All of the reconstructions shared the following settings: PSF + TOF OSEM algorithm, 5 subsets, decay correction, scatter correction and CT-based attenuation correction. The voxel size in the z direction was CT-matched (2 mm) or not CT-matched (1.65 mm). For each scan, this value was identical between reconstructions. For the scatter correction, relative scaling of the scatter correction is the default recommended option, but the use of absolute scatter scaling was explored as well. The other settings for the 9 reconstructions are detailed in Table 1.

Table 1 Reconstruction settings

IDIF validation using calibration assessment with manual blood samples

For all patients, IDIFs were taken from a 2 × 2 (with matrix 220 × 220) or 4 × 4 (with matrix 440 × 440) voxels region placed centrally on five adjacent slices in the DA and LV on the updated European Association of Nuclear Medicine Research Ltd. standard reconstruction (EARL2) [22] used as reference and in the AA on all of the reconstructions (11 IDIFs in total: 9 derived from AA with different reconstruction settings, 1 IDIF from DA with EARL2 reconstruction and 1 IDIF from LV with EARL 2 reconstruction) using the ACCURATE software tool (developed in IDLversion 8.4 (Harris Geospatial Solutions, Bloomfield, USA)) [23]. For patients with [18F]FDG injection, 3 manual venous samples were used for calibration of the 11 different IDIFs (35, 45 and 60 min post-injection because the arterio-venous [18F]FDG concentration equilibrium is reached around 40 min post-injection [24]). For patients with [18F]DPA-714 injection, 4 manual arterial samples (at 20, 40, 60 and 75 min post-injection) were used for calibration of the 11 different IDIFs and for the input function obtained from an arterial continuous sampler (Comecer, Joure, the Netherlands). The calibration factor (CF) was the mean of the ratio between the whole blood activity from the manual samples and the whole blood activity from the IDIF (or the input function from the arterial continuous sampler for [18F]DPA-714) at the corresponding time points [12]. The CF was used for IDIFs accuracy and precision assessment.

Uptake ratio between organs and aorta

In order to quantify the uptake for each tracer in the main organs close to the aorta which may affect the accuracy and the precision of IDIF, a VOI was manually drawn in the liver, lung, myocardium and spleen on the EARL2 reconstruction. The standardized uptake value normalized to body weight (SUVbw) for each organs VOI and AA was based on 50–60 min post-injection time interval.

Illustration of the impact of different calibrated-IDIFs using kinetic modelling

Blood processing

For [18F]FDG, calibrated IDIFs were corrected for plasma to whole-blood ratios to obtain plasma input functions. The calibrated whole-blood curve was multiplied with the function obtained from interpolation between plasma-to-whole blood ratio values from the manual samples to generate the corresponding whole-plasma curve.

For [18F]DPA-714, calibrated IDIFs and input function from continuous sampler were corrected for plasma to whole-blood ratios and metabolites to obtain metabolite corrected plasma input functions. The method for plasma/whole blood correction was the same method as described above for [18F]FDG. Additionally, for metabolite corrections, parent fractions measured in the manual plasma samples were fitted to a Hill function [25]. Each individual whole-plasma curve was then multiplied by the corresponding fitted Hill function.

Kinetic modelling

For [18F]FDG, Patlak linearization [26] (using t* = 30 min) was performed to estimate the net tracer flux Ki (mL·cm− 3·min− 1) using the 11 different calibrated IDIFs and using a same target tissue time-activity-curve (TAC) from a VOI manually drawn using the EARL2 reconstruction on the striatum. Four patients did not have any high tumor uptake so the striatum was chosen as target region because it is a region with high level uptake similar to that seen in cancers. The EARL2_AA IDIF was used as reference because arterial continuous sampler was not available for [18F]FDG. Moreover, EARL2 is the standard reconstruction for whole-body and because a previous study concluded that AA is the structure of choice for defining IDIF for [18F]FDG [8]. Therefore, the Ki extracted with the 10 others calibrated IDIFs were compared to the Ki extracted with the calibrated EARL2_AA IDIF.

For [18F]DPA-714, Logan linearization [27] (using t* = 30 min) was performed to estimate the distribution volume (VT) (mL·cm− 3) using the 11 different calibrated IDIFs and using a same target tissue time-activity-curve from a VOI manually drawn using the EARL2 reconstruction on the cerebral frontal region. The brain was chosen because post COVID syndrome might be associated with cerebral neuroinflammation on [18F]DPA-714 PET [28]. The VT extracted with the 11 calibrated IDIFs were compared to the VT extracted with the calibrated input function obtained from an arterial continuous sampler used as reference. The kinetic modelling was performed using in-house developed software.

Results

Study population

For [18F]FDG, 2 female patients and 6 male patients were included in this study, with a mean age of 62 ± 10 years. For [18F]DPA-714, 3 female patients and 7 male patients were included in this study, with a mean age of 49 ± 5 years. Table 2 summarizes the patient characteristics for each tracer. Images of a patient for each tracer are given in Fig. 1.

Table 2 Patient demographics (N = 18)
Fig. 1
figure 1

Example of [18F]FDG and [18F]DPA-714 PET Maximum Intensity Projection images on the 25–30 s post-injection frame with ascending aorta (AA), descending aorta (DA) and left ventricular cavity (LV) volume of interests and the 50–60 min post-injection frame on the EARL2 reconstruction

IDIF validation using calibration assessment with manual blood samples

For each tracer, examples of IDIF time-activity curves derived at different locations and using different reconstruction settings are displayed in Fig. 2. For [18F]FDG, the accuracies and the precisions of estimated IDIF CFs were high (mean calibration factor (CF) ≥ 0.82 & SD ≤ 0.06). Compared to the mean CF using EARL2_AA IDIF (0.87), the mean CF value of other IDIFs was similar, e.g. the maximum difference was observed for the mean CF using the it10_nogauss_abs IDIF (mean CF = 0.82). For [18F]DPA714, the results showed also high accuracy in the estimation of IDIF CFs (mean CF ≥ 0.86) except when using it10_nogauss_abs, EARL2_DA and EARL2_LV (respectively mean CF = 0.68, 0.47 and 0.44). Furthermore, the precision of the different AA IDIFs was low (SD ≥ 0.10). The CF improved using more iterations at the cost of increased variability (mean CF = 0.99 (SD = 0.21) for it10_nogauss_AA). The mean CF value using the maximum ring difference (MRD) 322 compared to MRD85 and the mean CF value using the matrix 440 × 440 (high-resolution) compared to the matrix 220 × 220 were similar (respectively difference = -1.1% and − 4.1%). The results of CF are summarized in Table 3; Fig. 3.

Fig. 2
figure 2

Example of the tail of the time-activity curves from imaging-derived input functions at different locations and different reconstruction settings for [18F]FDG (A) and for [18F]DPA-714 (B) (AA: Ascending Aorta, DA: Descending Aorta, LV: Left Ventricular Cavity, it: iterations, gauss: Gaussian post-filtering, mat440: matrix 440 × 440, abs: absolute scatter correction method). All IDIFs were extracted using AA location except for EARL2_DA and EARL2_LV

Table 3 Mean value and standard deviation of the calibration factors using blood samples for the input functions using different locations or reconstruction settings
Fig. 3
figure 3

Boxplots illustrating the spread of calibration factors for [18F]FDG (A) and for [18F]DPA-714 (B) using the different input functions (gauss: Gaussian post-filtering, AA: Ascending Aorta, DA: Descending Aorta, LV: Left Ventricular Cavity, MRD: maximum ring difference, it: iterations, mat440: matrix 440 × 440, abs: absolute scatter correction method)

Uptake ratio between organs and aorta

The mean SUVbw on the 50–60 min post-injection frame of the AA_EARL2 IDIF was 5.5 times lower for [18F]DPA-714 than for [18F]FDG. The average liver, lung, myocardium, and spleen uptake ratio between organs and AA were respectively 9.8, 19.5, 8.9, 16.4 times higher for [18F]DPA-714 than for [18F]FDG. The results are summarized in Table 4.

Table 4 Average (SD) SUVbw of organs and blood pool on the 50–60 min post-injection frame on the EARL2 reconstruction and average (SD) SUVbw ratio of organs with blood pool

Illustration of the impact of different calibrated-IDIFs using kinetic modelling

For [18F]FDG, one patient was excluded for this analysis due to movements of the head during the acquisition. Compared to striatum Ki extracted using calibrated EARL 2_AA IDIF as reference, striatum Ki mean differences were < 2% using other calibrated IDIFs. The maximum mean difference was for IDIF using calibrated EARL 2_LV (+ 1.6%). For [18F]DPA-714, two patients were excluded for this analysis because the arterial continuous sampler acquisition failed. Compared to frontal VT extracted using calibrated continuous sampler input function as reference, frontal VT mean differences was low using calibrated AA IDIFs (for examples, VT mean difference = -5.3% using calibrated EARL2_AA IDIF or VT mean difference = -8.2% using calibrated IDIF with the it8_nogauss_mat440_AA reconstruction, which is the high-resolution reconstruction usually applied for brain tracers) but higher using calibrated EARL2_DA and EARL2_LV IDIFs (respectively + 12.5% and + 19.1%). The results of the calibrated-IDIF effect on simplified kinetic parameters are summarized in Table 5; Fig. 4. For both tracers, typical examples of parametric images using Patlak or Logan linearization with different calibrated-IDIF derived at different locations and using different reconstruction settings are displayed in Fig. 5.

Table 5 Mean value and standard deviation of the percentage of difference of the simplified kinetic parameters values (striatum Ki using Patlak linearization for [18F]FDG and frontal VT using Logan linearization for [18F]DPA-714) extracted from the different calibrated IDIF compared to the simplified kinetic parameters values as reference obtained using calibrated IDIF from EARL2 reconstruction for [18F]FDG and using calibrated input function from continuous blood sampler for [18F]DPA-714, with the same tissue time-activity curve from EARL 2
Fig. 4
figure 4

Boxplots illustrating the spread of simplified kinetic parameters values for [18F]FDG (A) and [18F]DPA-714 (B) obtained using the different calibrated image-derived input functions compared to the simplified kinetic parameters values obtained using the reference input function (EARL2_AA for [18F]FDG and continuous sampler for [18F]DPA-714) (gauss: Gaussian post-filtering, AA: Ascending Aorta, DA: Descending Aorta, LV: Left Ventricular Cavity, it: iterations, mat440: matrix 440 × 440, MRD: maximum ring differences, abs: absolute scatter correction method)

Fig. 5
figure 5

Coronal views of [18F]FDG Ki parametric images from Patlak linearization (A) and [18F]DPA-714 VT parametric images from Logan linearization (B) using different calibrated input functions and the same whole body reconstruction (EARL 2). All IDIFs were extracted using AA location except for EARL2_DA and EARL2_LV. For [18F]FDG, no significant visual differences are seen between Ki parametric images using the different IDIFs. For [18F]DPA-714, compared to the VT parametric images obtained with input function using continuous arterial sampler (reference), visual differences are seen using EARL2_DA and EARL2_LV IDIFs, significantly in lungs (orange arrows) and liver (blue arrows) (AA: Ascending Aorta, DA: Descending Aorta, LV: Left Ventricular Cavity, it: iterations, mat440: matrix 440 × 440, MRD: maximum ring differences, abs: absolute scatter correction methods)

Discussion

This study demonstrates that the accuracy and precision of IDIF obtained with the LAFOV PET/CT system are tracer, reconstruction settings and IDIF VOI locations dependent, which could affect quantitative assessment of kinetic parameters.

Regarding tracer dependency, even if our study showed that the accuracy of AA IDIFs was high and similar for both tracers, the precision of IDIF varies according to the tracer. For [18F]FDG, the precision of IDIFs was high so IDIF does not need calibration using manual blood samples, confirming the results already published for [18F]FDG using short AFOV PET scanners [11]. For [18F]DPA-714, the precision of IDIF was lower suggesting that IDIF needs to be calibrated using a few manual blood samples. Additionally, in case of [18F]DPA-714, manual blood samples are also needed for plasma to whole-blood and metabolite corrections. The lower precision could have several explanations. Firstly, we noticed that [18F]DPA-714 blood clearance is higher than for [18F]FDG, leading to lower count statistics in the blood pool. PET regions with lower counts are more sensitive to the noise than PET regions with higher counts. This also explains why the precision of IDIF was even lower using more iterations for [18F]DPA-714 in our study, as more iterations improved convergence but at the cost of increased image noise. Secondly, we showed that contrasts in tracer uptake among organs neighboring the aorta are much higher for [18F]DPA-714 than for [18F]FDG, due to lower count statistics in the blood pool for [18F]DPA-714 but also due to the clearly higher liver, lung, myocardium and spleen uptakes for [18F]DPA-714. These higher contrasts lead to bias from the scanners scatter correction algorithm [29]. As reported previously, the accuracy is lower as the tail regions have comparatively low counts using the relative scatter correction method [30]. Thirdly, bias on manual arterial samples used for calibration could also explain the CF variability of IDIF. However, the precision of the whole blood IF from the continuous sampler was high in our study using the same manual arterial samples for calibration. Using calibrated AA IDIFs, our results exhibited similar frontal VT values compared to frontal VT values extracted using calibrated continuous sampler input function, so AA IDIF can replace continuous sampler for quantification with Logan graphical method after calibration against manual arterial samples. However, for [18F]DPA-714 quantification using non-linear method, further validation against continuous blood sampling is needed because it is more sensitive to the initial part of the input function than linear method. Linear method for macroparameter estimation is based on the whole area under the curve (AUC) of the input function in which the peak contributes very little [31] whereas microparameters estimation using non-linear method needs stringent requirements of the initial part of the input function. LAFOV PET/CT system may improve microparameters estimation because of increased sensitivity leading to better capturing rapid kinetics of the tracer in the blood compartment and the multiple organs.

Regarding IDIF location for [18F]FDG, our results showed similar CF for AA, DA or LV IDIFs. Using the LAFOV Siemens Biograph Vision Quadra, Sari et al. showed also good agreement between the IDIFs derived from AA, DA, LV, and left atrium. However, authors exhibited that AUC derived from LV had higher relative standard deviation than AUC derived from the other IDIF regions suggesting worse precision of the estimated blood pool [3]. Using the LAFOV uEXPLORER, Zhang et al. exhibited in a figure a similar peak amplitude between IDIF derived from DA and IDIF derived from LV but a smoother and lower tail of the IDIF derived from DA compared to the IDIF derived from LV [32]. A previous study which evaluated the accuracy of IDIF from AA, LV, and abdominal aorta IDIFs in 136 [18F]FDG PET/CT scans including patients with various cancers showed that the metabolic rate of glucose (MRGlu) of lesions using AA IDIF had the strongest correlation with the MRGlu using arterial sampling [11]. However, this strong relationship was similar for all IDIFs. Van der Weerdt et al. showed a better accuracy for AA and DA than LV and left atrium IDIFs and a better precision for AA than DA IDIF in 18 cardiac [18F]FDG PET scans with arterial samplings [8]. The higher accuracy differences between AA and LV compared to our report could be due to spillover effect linked to higher [18F]FDG uptake in myocardium in the cardiac study. Indeed, in the cardiac study, patients had a light breakfast before [18F]FDG injection which induces endogenous insulin secretion and then high myocardium uptake. In our study, patients with oncological diseases were fasting at least 6 h so the myocardium uptake was probably lower than in the cardiac study. This effect comes from high uptake regions close to IDIF regions because of the limited spatial resolution of PET scanners and movements. Since those reports, spatial resolution of PET scanners has improved substantially and could be also a reason why we found lower differences between AA and LV IDIFs for [18F]FDG. For [18F]DPA-714, our study exhibited that the accuracy of DA and LV IDIFs were lower than the accuracy of AA IDIFs. Due to the small size of the VOI, IDIFs potentially could suffer from partial-volume effect. However, the size of the 3 different IDIFs VOIs was the same. The main factor of this lower accuracy for DA IDIF and LV IDIF is probably the spillover effect. The reason why LV and DA IDIFs are more sensitive to this effect for [18F]DPA-714 than for [18F]FDG is due to the respectively markedly higher [18F]DPA-714 myocardium and posterior lung uptakes than [18F]FDG.

Regarding reconstruction settings, for [18F]DPA-714, our results exhibited that CF improved using more iterations because more iterations improves the contrast in the images [33]. We observed that the variability also increased using more iterations. This lower precision is due to higher image noise using more iterations [34]. Our study showed a significant lower IDIF accuracy for the absolute scatter correction method compared to the relative scatter correction set as default. Our results are consistent with a previous study showing an underestimation of scatter using the absolute scatter correction method [29]. However, the absolute scatter correction method improved the precision of IDIF in our study. There are two different methods currently available to perform scatter scaling of the most widely used scatter correction method algorithm, based on the single scatter simulation (SSS) [35]: the absolute method which utilizes the SSS scatter distribution directly and the relative method. The relative method is the default technique implemented in clinical PET imaging because it compensates for multiple scatter and for scatter contributions, scaling the activity located outside the axial FOV. The higher contrast between normophysiological uptakes for [18F]DPA-714 than for [18F]FDG probably explains why differences were greater for the CF IDIF between the use of relative and absolute scatter correction for [18F]DPA-714 compared to [18F]FDG data, which is consistent with two previous studies [29, 36].

We hypothesize that the accuracy of IDIF for tracers such as [18F]DPA-714 with high uptake in the main organs close to the aorta could sometimes also be low due to high variability of IDIF and need manual blood samples for calibration contrary to tracers such as [18F]FDG with low contrast. Moreover, for tracers such as [18F]DPA-714 with high myocardium and/or posterior lung uptake, DA and/or LV IDIFs are not accurate due to the spillover effect. Therefore, we conclude that the use of IDIF has to be validated for each tracer using LAFOV PET/CT system.

Conclusion

For [18F]FDG, IDIF do not need calibration against manual blood samples. For [18F]DPA-714, AA IDIF can replace the continuous arterial sampling for simplified kinetic quantification but only with calibration against manual arterial blood samples. Findings from this study suggest that the accuracy and precision of IDIF obtained using a LAFOV PET/CT system depend on tracer, reconstruction settings and IDIF location. Therefore, the use of IDIF needs to be validated for each tracer and optimization is warranted.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Huang S. Anatomy of SUV. Standardized uptake value. Nucl Med Biol. 2000;27:643–6.

    Article  CAS  PubMed  Google Scholar 

  2. Ronald Boellaard. Need for standardization of 18F-FDG PET/CT for treatment response assessments. J Nucl Med. 2011;52(Supplement 2):93S. https://doi.org/10.2967/jnumed.110.085662

    Article  PubMed  Google Scholar 

  3. Sari H, Mingels C, Alberts I, et al. First results on kinetic modelling and parametric imaging of dynamic 18F-FDG datasets from a long axial FOV PET scanner in oncological patients. Eur J Nucl Med Mol Imaging. 2022;49(6):1997–2009. https://doi.org/10.1007/s00259-021-05623-6

    Article  CAS  PubMed  Google Scholar 

  4. Freedman N, Sundaram S, Kurdziel K, et al. Comparison of SUV and Patlak slope for monitoring of cancer therapy using serial PET scans. Eur J Nucl Med Mol Imaging. 2003;30:46–53. https://doi.org/10.1007/s00259-002-0981-4

    Article  PubMed  Google Scholar 

  5. Greuter H, Boellaard R, Lingen A, Franssen E, Lammertsma A. Measurement of 18F-FDG concentrations in blood samples: comparison of direct calibration and standard solution methods. J Nucl Med Technol. 2004;31:206–9.

    Google Scholar 

  6. Everett B, Oquendo M, Abi-Dargham A, et al. Safety of radial arterial catheterization in PET research subjects. J Nucl Med. 2009;50:1742. https://doi.org/10.2967/jnumed.109.063206

    Article  PubMed  Google Scholar 

  7. Chim H, Bakri K, Moran SL. Complications related to radial artery occlusion, radial artery harvest, and arterial lines. Hand Clin. 2015;31(1):93–100. https://doi.org/10.1016/j.hcl.2014.09.010

    Article  PubMed  Google Scholar 

  8. van der Weerdt AP, Klein LJ, Boellaard R, Visser CA, Visser FC, Lammertsma AA. Image-derived input functions for determination of MRGlu in cardiac (18)F-FDG PET scans. J Nucl Med. 2001;42(11):1622–9.

    PubMed  Google Scholar 

  9. Ohtake T, Kosaka N, Watanabe T, et al. Noninvasive method to obtain input function for measuring tissue glucose utilization of thoracic and abdominal organs. J Nucl Med. 1991;32(7):1432–8.

    CAS  PubMed  Google Scholar 

  10. Huisman MC, van Golen LW, Hoetjes NJ, et al. Cerebral blood flow and glucose metabolism in healthy volunteers measured using a high-resolution PET scanner. EJNMMI Res. 2012;2(1):63. https://doi.org/10.1186/2191-219X-2-63

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. de Geus-Oei LF, Visser EP, Krabbe PFM, et al. Comparison of image-derived and arterial input functions for estimating the rate of glucose metabolism in therapy-monitoring 18F-FDG PET studies. J Nucl Med. 2006;47(6):945–9.

    PubMed  Google Scholar 

  12. Mourik JEM, Lubberink M, Klumpers UMH, Comans EF, Lammertsma AA, Boellaard R. Partial volume corrected image derived input functions for dynamic PET brain studies: methodology and validation for [11 C]flumazenil. NeuroImage. 2008;39(3):1041–50. https://doi.org/10.1016/j.neuroimage.2007.10.022

    Article  PubMed  Google Scholar 

  13. Teo BK, Seo Y, Bacharach SL, et al. Partial-volume correction in PET: validation of an iterative postreconstruction method with phantom and patient data. J Nucl Med. 2007;48(5):802–10. https://doi.org/10.2967/jnumed.106.035576

    Article  PubMed  Google Scholar 

  14. Kim E, Shidahara M, Tsoumpas C, et al. Partial volume correction using structural-functional synergistic resolution recovery: comparison with geometric transfer matrix method. J Cereb Blood Flow Metab. 2013;33(6):914–20. https://doi.org/10.1038/jcbfm.2013.29

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Sari H, Erlandsson K, Law I, et al. Estimation of an image derived input function with MR-defined carotid arteries in FDG-PET human studies using a novel partial volume correction method. J Cereb Blood Flow Metab. 2017;37(4):1398–409. https://doi.org/10.1177/0271678X16656197

    Article  PubMed  Google Scholar 

  16. Alavi A, Werner TJ, Høilund-Carlsen PF, Zaidi H. Correction for partial volume effect is a must, not a luxury, to fully exploit the potential of quantitative PET imaging in clinical oncology. Mol Imaging Biol. 2018;20(1):1–3. https://doi.org/10.1007/s11307-017-1146-y

    Article  PubMed  Google Scholar 

  17. Cysouw MCF, Golla SVS, Frings V, et al. Partial-volume correction in dynamic PET-CT: effect on tumor kinetic parameter estimation and validation of simplified metrics. EJNMMI Res. 2019;9:12. https://doi.org/10.1186/s13550-019-0483-z

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Karakatsanis NA, Zhou Y, Lodge MA, et al. Generalized whole-body patlak parametric imaging for enhanced quantification in clinical PET. Phys Med Biol. 2015;60(22):8643. https://doi.org/10.1088/0031-9155/60/22/8643

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Surti S, Pantel AR, Karp JS, Total Body PET. Why, how, what for? IEEE Trans Radiat Plasma Med Sci. 2020;4(3):283–92. https://doi.org/10.1109/trpms.2020.2985403

    Article  PubMed  PubMed Central  Google Scholar 

  20. Vriens D, Visser EP, de Geus-Oei LF, Oyen WJG. Methodological considerations in quantification of oncological FDG PET studies. Eur J Nucl Med Mol Imaging. 2010;37(7):1408–25. https://doi.org/10.1007/s00259-009-1306-7

    Article  PubMed  Google Scholar 

  21. Verveen A, Verfaillie SCJ, Visser D, et al. Neurobiological basis and risk factors of persistent fatigue and concentration problems after COVID-19: study protocol for a prospective case-control study (VeCosCO). BMJ Open. 2023;13(6):e072611. https://doi.org/10.1136/bmjopen-2023-072611

    Article  PubMed  PubMed Central  Google Scholar 

  22. Kaalep A, Sera T, Rijnsdorp S, et al. Feasibility of state of the art PET/CT systems performance harmonisation. Eur J Nucl Med Mol Imaging. 2018;45(8):1344–61. https://doi.org/10.1007/s00259-018-3977-4

    Article  PubMed  PubMed Central  Google Scholar 

  23. Ronald Boellaard. Quantitative oncology molecular analysis suite: ACCURATE. J Nucl Med. 2018;59(supplement 1):1753.

    Google Scholar 

  24. Wakita K, Imahori Y, Ido T, et al. Simplification for measuring input function of FDG PET: investigation of 1-point blood sampling method. J Nucl Med. 2000;41(9):1484–90.

    CAS  PubMed  Google Scholar 

  25. Gunn RN, Sargent PA, Bench CJ, et al. Tracer kinetic modeling of the 5-HT1AReceptor ligand [carbonyl-11 C]WAY-100635 for PET. NeuroImage. 1998;8(4):426–40. https://doi.org/10.1006/nimg.1998.0379

    Article  CAS  PubMed  Google Scholar 

  26. Patlak CS, Blasberg RG. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. Generalizations. J Cereb Blood Flow Metab. 1985;5(4):584–90. https://doi.org/10.1038/jcbfm.1985.87

    Article  CAS  PubMed  Google Scholar 

  27. Logan J. Graphical analysis of PET data applied to reversible and irreversible tracers. Nucl Med Biol. 2000;27(7):661–70. https://doi.org/10.1016/s0969-8051(00)00137-2

    Article  MathSciNet  CAS  PubMed  Google Scholar 

  28. Visser D, Golla SSV, Verfaillie SCJ et al. Long COVID is associated with extensive in-vivo neuroinflammation on [18F]DPA-714 PET. Published online June 4, 2022:2022.06.02.22275916. https://doi.org/10.1101/2022.06.02.22275916

  29. Bal H, Kiser JW, Conti M, Bowen SL. Comparison of maximum likelihood and conventional PET scatter scaling methods for 18 F-FDG and 68 Ga-DOTATATE PET/CT. Med Phys. 2021;48(8):4218–28. https://doi.org/10.1002/mp.14988

    Article  PubMed  Google Scholar 

  30. Lindemann ME, Guberina N, Wetter A, Fendler WP, Jakoby B, Quick HH. Improving 68Ga-PSMA PET/MRI of the prostate with unrenormalized absolute scatter correction. J Nucl Med. 2019;60(11):1642–8. https://doi.org/10.2967/jnumed.118.224139

    Article  CAS  PubMed  Google Scholar 

  31. Zanotti-Fregonara P, Chen K, Liow JS, Fujita M, Innis RB. Image-derived input function for brain PET studies: many challenges and few opportunities. J Cereb Blood Flow Metab. 2011;31(10):1986–98. https://doi.org/10.1038/jcbfm.2011.107

    Article  PubMed  PubMed Central  Google Scholar 

  32. Zhang X, Xie Z, Berg E, et al. Total-body dynamic reconstruction and parametric imaging on the uEXPLORER. J Nucl Med. 2020;61(2):285–91. https://doi.org/10.2967/jnumed.119.230565

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Karlberg AM, Sæther O, Eikenes L, Goa PE. Quantitative comparison of PET performance-Siemens biograph mCT and mMR. EJNMMI Phys. 2016;3(1):5. https://doi.org/10.1186/s40658-016-0142-7

    Article  PubMed  PubMed Central  Google Scholar 

  34. Boellaard R, van Lingen A, Lammertsma AA. Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med. 2001;42(5):808–17.

    CAS  PubMed  Google Scholar 

  35. Watson CC. New, faster, image-based scatter correction for 3D PET. IEEE Trans Nucl Sci. 2000;47(4):1587–94. https://doi.org/10.1109/23.873020

    Article  ADS  Google Scholar 

  36. Heußer T, Mann P, Rank CM, et al. Investigation of the halo-artifact in 68Ga-PSMA-11-PET/MRI. PLoS ONE. 2017;12(8):e0183329. https://doi.org/10.1371/journal.pone.0183329

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The Biograph Vision Quadra PET/CT system at Amsterdam UMC was made possible because of contributions from the European Regional Development Fund (ERDF, “Kansen voor West 2”), the City of Amsterdam, the Department of Economic Affairs of the Netherlands, and the Province of Noord-Holland. It was actively supported by Amsterdam Economic Board and Amsterdam Marketing. In addition, the Biograph Vision Quadra PET/CT system is financially supported by the Amsterdam Oncology and Neuroscience Research (ADORE) institute. The VeCosCO study was funded by a ZonMw grant (number: 10430302110003). This study was also supported by the French League against cancer.

Author information

Authors and Affiliations

Authors

Contributions

RB and MY contributed to the conception and design. DV, NT, CS, GZ, EvdG, MdH, FB, BW and BNMvB contributed to the data acquisition. XPN, MY and RB contributed to the data analysis. XPN drafted the manuscript. Other authors revised the manuscript.

Corresponding author

Correspondence to Xavier Palard-Novello.

Ethics declarations

Ethics approval and consent to participate

Patients were included in studies approved by the Medical Ethics Review Committee of the Amsterdam UMC (NL80924.029.22 for the TBPETCT001 study and NL77033.029.21 for the VECOSCO study).

Consent for publication

Informed consent was obtained from all individual participants included in the study.

Competing interests

Siemens Healthineers provided e7 tools and contributed to the acquisition of the total body PET-CT system in the Amsterdam UMC Imaging Center. Competing interests: Frederik Barkhof is a steering committee or data safety monitoring board member for Biogen, Merck, ATRI/ACTC and Prothena, is a consultant for Roche, Celltrion, Rewind Therapeutics, Merck, IXICO, Jansen, Combinostics, has research agreements with Merck, Biogen, GE Healthcare, Roche and is the co-founder and shareholder of Queen Square Analytics LTD. No other potential conflict of interest relevant to this article was reported.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Palard-Novello, X., Visser, D., Tolboom, N. et al. Validation of image-derived input function using a long axial field of view PET/CT scanner for two different tracers. EJNMMI Phys 11, 25 (2024). https://doi.org/10.1186/s40658-024-00628-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-024-00628-0

Keywords