Clinical feasibility study of early 30-minute dynamic FDG-PET scanning protocol for patients with lung lesions

Abstract

Purpose

This study aimed to evaluate the clinical feasibility of early 30-minute dynamic 2-deoxy-2-[18F]fluoro-D-glucose (18F-FDG) positron emission tomography (PET) scanning protocol for patients with lung lesions in comparison to the standard 65-minute dynamic FDG-PET scanning as a reference.

Methods

Dynamic 18F-FDG PET images of 146 patients with 181 lung lesions (including 146 lesions confirmed by histology) were analyzed in this prospective study. Dynamic images were reconstructed into 28 frames with a specific temporal division protocol for the scan data acquired 65 min post-injection. Ki images and quantitative parameters Ki based on two different acquisition durations [the first 30 min (Ki-30 min) and 65 min (Ki-65 min)] were obtained by applying the irreversible two-tissue compartment model using in-house Matlab software. The two acquisition durations were compared for Ki image quality (including visual score analysis and number of lesions detected) and Ki value (including accuracy of Ki, the value of differential diagnosis of lung lesions and prediction of PD-L1 status) by Wilcoxon’s rank sum test, Spearman’s rank correlation analysis, receiver operating characteristic (ROC) curve, and the DeLong test. The significant testing level (alpha) was set to 0.05.

Results

The quality of the Ki-30 min images was not significantly different from the Ki-65 min images based on visual score analysis (P > 0.05). In terms of Ki value, among 181 lesions, Ki-65 min was statistically higher than Ki-30 min (0.027 ± 0.017 ml/g/min vs. 0.026 ± 0.018 ml/g/min, P < 0.05), while a very high correlation was obtained between Ki-65 min and Ki-30 min (r = 0.977, P < 0.05). In the differential diagnosis of lung lesions, ROC analysis was performed on 146 histologically confirmed lesions, the area under the curve (AUC) of Ki-65 min, Ki-30 min, and SUVmax was 0.816, 0.816, and 0.709, respectively. According to the Delong test, no significant differences in the diagnostic accuracies were found between Ki-65 min and Ki-30 min (P > 0.05), while the diagnostic accuracies of Ki-65 min and Ki-30 min were both significantly higher than that of SUVmax (P < 0.05). In 73 (NSCLC) lesions with definite PD-L1 expression results, the Ki-65 min, Ki-30 min, and SUVmax in PD-L1 positivity were significantly higher than that in PD-L1 negativity (P < 0.05). And no significant differences in predicting PD-L1 positivity were found among Ki-65 min, Ki-30 min, and SUVmax (AUC = 0.704, 0.695, and 0.737, respectively, P > 0.05), according to the results of ROC analysis and Delong test.

Conclusions

This study indicates that an early 30-minute dynamic FDG-PET acquisition appears to be sufficient to provide quantitative images with good-quality and accurate Ki values for the assessment of lung lesions and prediction of PD-L1 expression. Protocols with a shortened early 30-minute acquisition time may be considered for patients who have difficulty with prolonged acquisitions to improve the efficiency of clinical acquisitions.

Background

2-deoxy-2-[18F]fluoro-D-glucose(18F-FDG) positron emission tomography (PET) /CT imaging has a great impact on the diagnostics and management of oncological patients and has gained tremendous use worldwide [1, 2]. Two 18F-FDG PET image acquisition frameworks are commonly used, namely static and dynamic acquisitions [3]. At present, static imaging is more widely used, and partly for convenience, the patients are scanned 60 min after administration of the tracer [4]. The standard uptake values (SUVs), especially maximum standard uptake value (SUVmax), and mean standard uptake value (SUVmean), represent semiquantitative measures of glucose uptake as the main quantitative indicators in clinical practice [5]. Although useful, SUVs are highly dependent on the interval between 18F-FDG injection and image acquisition. It can also be affected by blood glucose levels, and non-perfect injections [5, 6]. And SUVs have some limitations in a range of clinical tasks, including distinguishing between malignant vs. benign (e.g., inflammatory) uptake and assessment of treatment response [5, 7]. The need for an additional quantification to compensate for the shortcomings of semiquantitative assessment of PET is constantly increasing.

Dynamic 18F-FDG PET scanning has also been in use for a long time, which continuously acquires imaging data over usually 60 min. Compared with the single semi-quantitative SUVs provided by the commonly used static PET/CT, dynamic 18F-FDG PET can provide quantitative evaluation with kinetic rate constants, e.g., net influx rate Ki, tumor blood influx rate K1, tumor blood efflux rate k2, phosphorylation rate k3 [7,8,9]. Kinetic analysis of dynamic PET imaging with a more accurate assessment of changes in tumor metabolism has been shown to be effective in improving diagnostic accuracy and achieving appropriate therapy monitoring in many different types of cancer [10]. While comparing the rate constant values of K1, k2, and k3 estimated by 18F-FDG, Ki was more commonly used and had been shown to be useful for characterizing tumor metabolism and assessing therapy response by reflecting more accurate changes after therapy [11,12,13,14,15]. The Ki was superior to SUV in differential diagnosing solitary lung nodules and lymph nodes (LNs), as well as in delineating tumor volume [6,7,8, 16]. However, the long acquisition time, single bed-position dynamic acquisition (axial extent of view of 15-25 cm), invasive arterial blood sampling, large volume of post-processed data, and the relatively complex process limit the routine use of dynamic PET in clinical practice [5, 17,18,19]. With the update of new acquisition equipment (e.g., total-body PET) and the development of computational models, the problems of short acquisition fields, invasiveness, and data processing have basically been solved [9, 19], but shortening the acquisition time still requires further research.

A range of time windows involved in shortened dynamic 18F-FDG PET scanning protocols have been considered by several groups. Three scanning protocols are used in most studies: (i) only late-time dynamic scanning (such as 30-60 min post-injection), which lacks early-time data. It often needs another plasma input function (IF), such as population-based IF (PBIF). However, it may introduce errors into the parametric analysis as it does not take into account the specificity of the individual IF [20]; (ii) two-short-dynamic-scanning, which requires two separate scans after 18F-FDG injection, such as (0-6 min + 60-75 min) scans, (0-10 min + 55-60 min) scans, (0-10 min + 40-60 min) scan and two 5-min scans [21,22,23,24,25]. These studies showed that the generated parameter metabolic rate of glucose (MRglu) [21] or Ki [22, 24, 25] is highly correlated with the reference parameter calculated from the full scan. However, such a protocol requires a break in the middle, and the problem of image registration can limit the feasibility; and (iii) only early-time scanning, with dynamic analysis carried out based on the acquired data from the early frames, such as 0-30 min scans. The use of early dynamic PET data to calculate Ki is based on the hypothesis that the Patlak plot, which represents the metabolic rate of MRglu in lesions, enters a linear phase at an early stage for which the slope can theoretically be extracted. During this period, the dynamic trend can be measured and is consistent with that provided by a standard scan [26]. These studies published by Torizuka et al. [7] and Visser et al. [27] showed that the MRglu values or Ki derived from the standardized scans and the 30 min scans had strong correlation, suggesting that a shorter imaging duration of 0-30 min may represent a clinically viable alternative to an imaging sequence of 0-60 min for kinetic modeling of FDG-PET. The early-time scanning protocol also had the advantage of providing accurate individual PBIF estimates, avoiding the problem of inaccurate image registration, and more accessible for patients. Although some studies have shown a strong correlation between the dynamic parameter Ki obtained with a shortened acquisition time of 0-30 min and that obtained with 60 min, these are small sample size studies and the consistency of the diagnostic results for benign and malignant lesions has not been investigated. Simplified acquisition protocols always involve a trade-off between clinical convenience and quantitative accuracy. In balancing these competing considerations, how the protocol is adapted depends on the clinical study or diagnostic application [17].

The development and clinical use of immune checkpoint inhibitors (ICIs) in recent years has opened new frontiers in the treatment of non-small cell lung cancer (NSCLC) [28]. National Comprehensive Cancer Network (NCCN) guidelines recommend treatment based on expression of programmed death-ligand 1 (PD-L1) in tumor as determined by immunohistochemistry (IHC) [29]. PD-L1 positivity was associated with significantly higher objective response rate, longer progression-free survival (PFS), and longer overall survival (OS) [30]. However, for patients who are unable to provide histological samples or who fail IHC testing, an alternative non-invasive method of measuring PD-L1 status would have important implications for clinical decision support. Some studies have demonstrated that the predictive value of SUVmax on FDG PET/CT in PD-L1 expression from the lung cancer patients at the initial diagnosis [31,32,33,34]. But data on the application of dynamic PET/CT in immunotherapy are limited, e.g., the predictive value of Ki in PD-L1 expression.

Accordingly, our study intends to further investigate the clinical feasibility of early 30-minute dynamic PET in terms of image quality, consistency of quantitative parameters, diagnostic efficacy in benign and malignant lung lesions, and predictive value in PD-L1 expression.

Materials and methods

Patient demographics

This prospective study was approved by the ethics committee of Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences (Clinical Trial Number: KYLH2022-1), which followed the 1964 Helsinki Declaration ethical standards and its subsequent amendments. All patients were provided written informed consent. Dynamic FDG-PET scans of the chest region were performed on 229 patients with clinical suspicion of lung cancer without treatment from May 2021 to September 2023. The following criteria were used to determine inclusion: (i) successful completion of dynamic FDG-PET, breath-holding chest CT, and whole-body static PET/CT scan; (ii) lung lesions with FDG avid confidentially identified by the readers. The exclusion criteria were as follows: patients failed to obtain histological confirmation of at least one lung lesion within two weeks (by biopsy or surgery). Finally, 146 patients were enrolled, including 87 males and 59 females, with age of 59.71 ± 10.75 (35–81) years. In these 146 patients, a total of 181 FDG avid lung lesions that were completely located in the field-of-view of the scanner were identified, of which 36 have maximum diameter (dmax) < 1.5 cm, 77 have dmax≥1.5 cm and < 3.0 cm, 68 have dmax≥3.0 cm. Out of 181 lung lesions, 146 lesions were histologically confirmed by biopsy or surgery (each patient had one lesion), including 128 malignant and 18 benign lesions. Among 128 malignant lesions, 73 NSCLC lesions had definite PD-L1 expression results. PD-L1 expression was positive in 39 lesions and negative in 34 lesions. The characteristics of the patients and lung lesions are shown in Table 1.

Table 1 Characteristics of the patients and lung lesions

Dynamic FDG-PET acquisition and reconstruction

All patients avoided strenuous exercise for 24 h and fasted for at least 6 h prior to the PET/CT scan (Discovery MI PET/CT, GE Healthcare, Milwaukee, USA). At the time of 18F-FDG injection, blood glucose was below 8.0mmol/L. Firstly, the breath-holding chest CT and whole-body CT scans (from the head to the mid-femur in the supine position with the arms raised) were performed using the following parameters: tube voltage, 120 kV; tube current, 10-220 mA; pitch, 1.375:1; noise index, 20. Then, the dynamic PET scans of the chest region (encompassing a 20-cm axial field of view) were obtained immediately after the injection of 18F-FDG (mean ± SD, 288.78 ± 49.69MBq, range, 203.15-423.65MBq) from an intravenous indwelling needle. The total dynamic scan lasted for 65 min, and its data were partitioned into 28 frames as follows: 6 × 10s, 4 × 30s, 4 × 60s, 4 × 120s, 10 × 300s. The 21st frame represents the last frame of the first 30 min of data. Lastly, a whole-body static PET scan was performed with the speed of 1.5 min/bed at the end of the dynamic acquisition. The attenuation correction was performed using CT data, and PET image reconstruction was performed using the block sequential regularized expectation maximization (BSREM) reconstruction algorithm with 25 iterations and 2 subsets, matrix size, 256 × 256.

PET/CT data analysis

Dynamic parameter Ki analysis

The dynamic parameters Ki of the first 30 min post-injection (Ki-30 min) and the first 65 min post-injection (Ki-65 min) were obtained based on the two-tissue irreversible compartment model by using the imaging frames for the first 30 min and the first 65 min post-injection, respectively. The image-derived input function (IDIF) was extracted from the ascending aorta by drawing a 10-mm-diameter region-of-interest on six consecutive slices in an image obtained by combining early time frames (0-60s), where the effects of motion and partial volume were less prominent than in the left ventricle. The uptake difference in blood and plasma was not accounted for. In this model, we assumed unidirectional uptake of 18F-FDG (i.e., k4 = 0), with irreversible trapping in tissue as 18F-FDG-6-PO [35]. Ki-65 min and Ki-30 min parametric images of each dynamic scan were generated using voxel-based analysis. Given the large number of voxels in a PET image, the Lawson-Hanson non-negative least squares algorithm was applied to solve a linearized problem instead of the conventional nonlinear one [36]. The 3D volume of interest (VOI) of each lesion in the SUV images of 65 min and 30 min were delineated, respectively, using semi-automatic methods with a threshold of 40% SUVmax in the ITK-snap software (version 4.9). For lesions with physiological uptake in the periphery tissue, two experienced nuclear medicine physicians manually delineated 3D VOIs on a slice-by-slice basis. The segmented VOI was then applied to the Ki images of 65 min and 30 min, respectively, to extract quantitative measurements from each scan.

Visual quality assessment and lesion detectability of Ki images

The Ki images were independently evaluated by two clinical nuclear medicine experts with more than 3 years of experience. The Ki images of 65-min and 30-min scans of each subject were anonymized and presented in random order to them. According to the Likert quintile [37], the subjective scores ranging from 1 to 5 of 5 categories were as follows: artifact reduction (ranging from 1 [enormous image artifact] to 5 [no image artifact]), noise suppression (ranging from 1 [enormous image noise] to 5 [no perceivable image noise]), contrast retention (ranging from 1 [hard to distinguish lesion edge] to 5 [very sharp lesion edge]), lesion discrimination (ranging from 1 [difficulty in lesion detection] to 5 [high confidence for small, low uptake lesion]), overall image quality (ranging from 1 [poor overall image quality] to 5 [excellent overall image quality]). Image quality scores of 3 or higher were qualified, indicating that the needs of clinical diagnosis could be met, whereas image quality scores of 1–2 did not meet the needs of clinical diagnosis [38, 39]. The final score is the average score from two clinical nuclear medicine experts.

All FDG avid lesions confidentially identified in Ki images by the readers were counted and the maximum diameter (dmax) of the lesion was measured in breath-holding chest CT. The SUVmax of all FDG-avid lesions in standard static PET/CT were recorded. In the case of peripheral lung lesions combined with obstructive pneumonia or atelectasis, the dmax of the lesion was measured by delineating the lesion on the static PET image with a threshold of 40% SUVmax. The partial-volume effect (PVE) could introduce large quantitative bias, especially in lesions with diameters less than 3 times the resolution of the imaging system [40]. Therefore, the lung lesions were classified into the group of dmax<1.5 cm, 1.5 ≤ dmax<3.0 cm, dmax≥3 cm based on the spatial resolution of the PET/CT system used in this study of 0.5 cm. The result of 65 min’s scan served as the reference to test the lesion detectability. The lesion detectability was determined by assessing the lesion detection rate of the lesion in this study.

Pathological evaluation

Biopsy or surgical specimens were reviewed by two independent pathologists with more than 10 years of experience in lung cancer pathology. All specimens were sectioned and examined conventionally using hematoxylin-eosin staining. IHC staining was also performed at the pathologist’s discretion. The platform of Ventana BenchMark ULTRA and the antibody of Dako 22C3 were used for PD-L1 staining to quantify the presence of PD-L1. Tumor proportion score (TPS) was recorded as the percentage of PD-L1 positive tumor cells over all tumor cells, and TPS ≥ 1% were considered PD-L1 positive expression [34].

Statistical analysis

MedCalc 20.010 software (MedCalc Software Ltd) was used for statistical analysis. Continuous group data were all non-normally distributed by Kolmogorov-Smirnov test and presented as mean ± SD as appropriate. The percentage difference (D%) in Ki-65 min and Ki-30 min were calculated (D%=absolute value of difference of Ki-65 min and Ki-30 min divided by the Ki-65 min) [7]. The differences in subjective scores of Ki image quality, Ki value, and D% with different groups were compared by the Wilcoxon rank-sum test. Spearman’s rank correlation coefficient was used to assess the correlation between Ki-30 min and Ki-65 min. Receiver operating characteristic (ROC) curve analyses were performed to evaluate the diagnostic accuracy of Ki-30 min, Ki-65 min and SUVmax in differentiating benign and malignant lung lesions and predicting PD-L1 expression. The differences in area under the curve (AUC) were determined by Delong’s test. The significant testing level (alpha) was set to 0.05.

Results

Comparison of visual quality assessment and lesion detectability of Ki images

All the image quality scores were above 3 in terms of five categories (artifact reduction, noise suppression, contrast retention, lesion discrimination, and overall quality) that were determined to meet clinical needs. Table 2 presents the subjective scores in five categories of the Ki-30 min images and the Ki-65 min images. No significant differences were found for different aspects of image quality between the Ki-30 min images and the Ki-65 min images (P > 0.05) (Fig. 1).

Fig. 1
figure 1

Representative Ki-65 min images and Ki-30 min images at different planes (the transverse, coronal and sagittal plane). Ki-65 min images and Ki-30 min images were found with good quality and showed no visual distinction between the two

In terms of lesion detectability, 181 lung lesions were identified in both Ki-65 min and Ki-30 min images.

Table 2 Quantitative subjective scores in terms of five categories of the Ki-30 min images and the Ki-65 min images

Assessment of quantitative dynamic parameter Ki

Among 181 lesions, there were very high correlations in Ki-65 min and Ki-30 min (r = 0.977, P < 0.05) (Fig. 2A). Figure 2B-D also show that there was a very high correlation of Ki-65 min and Ki-30 min within the group of dmax<1.5 cm (r = 0.947, P < 0.05), 1.5 cm ≤ dmax<3.0 cm (r = 0.959, P < 0.05), and dmax≥3.0 cm (r = 0.978, P < 0.05).

Fig. 2
figure 2

Correlations in Ki-65 min and Ki-30 min in all lesions(A), the groups of dmax<1.5 cm (B), 1.5 cm ≤ dmax<3.0 cm(C), and dmax≥3.0 cm (D)

The mean ± SD Ki-65 min and Ki-30 min of 181 lesions were 0.027 ± 0.017 ml/g/min and 0.026 ± 0.018 ml/g/min, respectively, and the difference was statistically significant (P < 0.05). The differences between Ki-65 min and Ki-30 min in the group of dmax<1.5 cm (0.014 ± 0.009 ml/g/min vs. 0.013 ± 0.009 ml/g/min), 1.5 cm ≤ dmax<3.0 cm (0.024 ± 0.014 ml/g/min vs. 0.023 ± 0.014 ml/g/min), and dmax≥3.0 cm (0.037 ± 0.019 ml/g/min vs. 0.035 ± 0.020 ml/g/min) were statistically significant (P < 0.05).

The mean ± SD D% of the 181 lesions was 13.03%±10.38% (0.16-44.87%), and that in the groups of dmax<1.5 cm, 1.5 cm ≤ dmax<3.0 cm, and dmax≥3.0 cm were 16.95%±10.29% (1.69-44.87%), 13.93%±11.70% (0.16-44.20%), and 9.94%±7.75% (0.21-31.70%), respectively. D% in the group of dmax≥3.0 cm was significantly lower than that of dmax<1.5 cm and 1.5 cm ≤ dmax<3.0 cm (P < 0.05), while no significant difference was found between the groups of dmax<1.5 cm and 1.5 cm ≤ dmax<3.0 cm (P > 0.05).

The diagnostic performance of quantitative dynamic parameter Ki

The ROC curves were plotted in 146 histologically confirmed lung lesions to determine the diagnostic accuracy of Ki-65 min, Ki-30 min and SUVmax in differentiating between benign and malignant lesions (Fig. 3). The optimal cut-off value of Ki-65 min was 0.022 ml/g/min, with an AUC of 0.816 (95%CI:0.744–0.876), a sensitivity of 66.40%, and a specificity of 83.30%. The optimal cut-off value of Ki-30 min was 0.018 ml/g/min, with an AUC of 0.816 (95%CI:0.743–0.875), a sensitivity of 69.50%, and a specificity of 83.30%. The optimal cut-off value of SUVmax was 9.65, with an AUC of 0.709 (95%CI:0.628–0.781), a sensitivity of 64.10%, and a specificity of 72.20%. And when the cut-off value of SUVmax was 2.50, the sensitivity and specificity were 98.40% and 5.6%, respectively. According to the results of the Delong test, no significant difference in the diagnostic accuracy was found between Ki-65 min and Ki-30 min (P > 0.05), while the diagnostic accuracies of Ki-65 min and Ki-30 min were both significantly higher than that of SUVmax (P < 0.05).

Fig. 3
figure 3

The ROC curve of Ki-65 min, Ki-30 min and SUVmax in the differential diagnosis of malignant and benign lung lesions

The relationship between quantitative dynamic parameter Ki and PD-L1 expression

In 73 NSCLC lesions with definite PD-L1 expression results, the Ki-65 min (0.035 ± 0.018 ml/g/min vs. 0.024 ± 0.018 ml/g/min, P < 0.05), Ki-30 min (0.034 ± 0.019 ml/g/min vs. 0.023 ± 0.019 ml/g/min, P < 0.05), and SUVmax (14.80 ± 7.32 vs. 9.05 ± 6.31, P < 0.05) in PD-L1 positive lesions were significantly higher than that in PD-L1 negative lesions. The ROC curves were plotted to determine the value of Ki-65 min, Ki-30 min, and SUVmax in predicting PD-L1 positive NSCLC lesions (Fig. 4). The optimal cut-off value of Ki-65 min, Ki-30 min, and SUVmax was 0.020 ml/g/min (AUC of 0.704, sensitivity of 84.60%, and specificity of 58.80%), 0.018 ml/g/min (AUC of 0.695, sensitivity of 84.60%, and specificity of 58.80%), and 9.55 (AUC of 0.737, sensitivity of 79.50%, and specificity of 64.70%), respectively. No significant differences in predicting PD-L1 positive NSCLC lesions were found among Ki-65 min, Ki-30 min, and SUVmax (P > 0.05), according to the results of the Delong test.

Fig. 4
figure 4

The ROC curve of Ki-65 min, Ki-30 min and SUVmax in predicting PD-L1 positive NSCLC lesions

Discussion

The main challenge of dynamic FDG-PET imaging to achieve clinical translation is partly due to the long scanning time [5, 41]. Optimal acquisition times for various applications remain to be determined. In this study of patients with lung lesions, we evaluated the clinical feasibility of shortening the dynamic acquisition time from 65 min to 30 min by comparing the image quality and quantitative dynamic parameters Ki of 6-5 min and 30-minute dynamic acquisitions. Our results showed that Ki-65 min images and Ki-30 min images both have good visual quality in terms of artifact reduction, noise suppression, contrast retention, lesion discrimination, and overall quality. Quantitative analyses showed that Ki-65 min and Ki-30 min were very highly correlated, and had similar values in differentiating benign and malignant lung lesions and predicting PD-L1 positive NSCLC lesions.

In terms of image quality, our study found that the quality of Ki-30 min images is as good as that of Ki-65 min images by visual quality assessment. In a different setting by Wang et al., they compared the MRglu images of two-short-dynamic-scanning protocols (0-6 min + 60-75 min) and standard scanning protocols concluding that both protocols produced good-quality MRglu images with no visual distinction [21]. In addition, our study found that the number of lesions detected in the two image types is consistent. This indicated that the Ki-30 min images will be able to meet the clinical requirements and early 30-minute dynamic FDG-PET acquisition might have the potential to improve clinical workflow.

In addition, regarding the assessment of the quantitative parameter Ki, our study showed a significant correlation with moderate variability between Ki-30 min and Ki-65 min (r = 0.977, D%=13.03%±10.38%), which was consistent with the study of 20 patients with untreated primary lung cancer published by Torizuka et al. (r = 0.966, D%=11.42%±11.31%) [7]. The same results were also found in the study that concluded 35 lesions of 15 different tumor patients by Chen et al. [26], which showed that Ki-30 min and Ki-60 min have an excellent agreement with r = 0.987. These results suggested that a 30-min shortened dynamic acquisition is sufficient to calculate Ki values comparable to 60-min dynamic acquisitions with clinical feasibility. On the other hand, our results also showed that the value of Ki-65 min is significantly higher than the Ki-30 min (0.027 ± 0.017 ml/g/min vs. 0.026 ± 0.018 ml/g/min, P < 0.05). The average uptake values within tumor isocontours contain statistical noise, which is reflected in uncertainties in the slope, and accordingly in the Ki values [27]. This might partly explain the difference between Ki-65 min and Ki-30 min.

In terms of diagnostic efficacy, the previous study performed by Skawran et al. in 60 cancer lesions and 17 inflammatory/infectious lesions has shown that using a cut-off value for Kimax-60 min of 0.026 ml/g/min delivers a sensitivity of 63.3% and a specificity of 82% for the detection of cancer lesions [42]. And previous research by our team also found that a cut-off value of Ki-65 min of 0.022 ml/g/min was identified as the optimal compromise point between sensitivity and specificity, with values of 39.50% and 91.80%, respectively, in discriminating between metastatic (n = 86) and non-metastatic LNs (n = 49) of lung cancer, and concluded that Ki with high specificity provided a complementary value to SUVmax [8]. The study performed by Ye Q et al. strongly indicated that Ki from dynamic PET can provide superior discrimination between benign and malignant lung nodules than SUV [43]. The current study showed that the Ki-65 min, Ki-30 min, and SUVmax cut-off values for distinguishing malignant from benign lung lesions were 0.022 ml/g/min (AUC of 0.816, sensitivity of 66.40%, and specificity of 83.30%), 0.018 ml/g/min (AUC of 0.816, sensitivity of 69.50% and specificity of 83.30%), and 9.65(AUC of 0.709, sensitivity of 64.10% and specificity of 72.20%), respectively. In concordance with previous work, our study showed that both Ki-65 min and Ki-30 min have higher diagnostic efficacy than SUVmax, especially in specificity (Fig. 5). Furthermore, our study also revealed that Ki-65 min and Ki-30 min have similar diagnostic efficacy in lung lesions. In clinical practice, we preferred to improve the specificity of 18F-FDG PET/CT in the diagnosis of lung lesions by measuring Ki values, given that most lung cancer often shows high SUVmax values in static 18F-FDG PET/CT images which gives considerable diagnostic sensitivity of 87-100%, and slightly lower specificity of 50-88% with a cut-off value of 2.5 [44]. So, this potential improvement in specificity may support the use of Ki-65 min and Ki-30 min in the differential diagnosis between benign and malignant lung lesions and is a valuable addition to static 18F-FDG PET/CT. Such observation will require careful validation in the future. These results also further illustrated the feasibility of 30-min dynamic scanning in terms of quantitative dynamic parameters.

Fig. 5
figure 5

A 60-year-old male patient with clinical suspicion of lung cancer. PET/CT scan showed FDG-avid pulmonary nodules in the upper lobe of the left lung(red arrow), with a size of 2.2 × 1.3 cm, SUVmax of 12.2, Ki-65 min of 0.016 ml/g/min, and Ki-30 min of 0.014 ml/g/min. Surgical pathology confirmed a granulomatous lesion. In addition, PET/CT scan showed many FDG-avid LNs in the mediastinum. (A, static PET maximum intensity projection; B, static PET SUV image; C, static PET/CT fusion image; D, dynamic PET Ki-65 min image; E, dynamic PET Ki-30 min image)

In the previous study of the correlation between PET/CT metabolic parameters and PD-L1 expression, it was indicated that SUVmax, total lesion glycolysis, standard uptake value ratio of PD-L1 positivity were higher than PD-L1 negativity [31,32,33,34], and the multivariate analysis revealed that only SUVmax as an independent predictor of PD-L1 positivity, and the best SUVmax cut-off value was determined to be 12.5 with sensitivity and specificity of 65.4% and 86.7%, respectively [34]. Our study also showed that PD-L1 status of NSCLC could be predicted by SUVmax at the cut-off value of 9.55 with sensitivity and specificity of 79.50% and 64.70%, respectively. In addition, the results of the present study showed that both Ki-65 min and Ki-30 min could predict PD-L1 status of NSCLC and both had the same predictive performance compared with SUVmax. Chang et al. showed PD-L1 positivity could increase the expression of glycolysis enzymes and promote the utilization of glucose in tumor [45]. Thus, for NSCLC, it may explain why SUVmax, Ki-65 min, and Ki-30 min all can predict PD-L1 expression. More importantly, the results of the present study revealed that Ki-30 min has the similar predictive power as Ki-65 min in PD-L1 expression. Therefore, the dynamic quantitative parameters obtained from a 30-min dynamic scan are sufficient to meet the clinical needs.

Interestingly, we also observed the differences in Ki-30 min and Ki-65 min between the lesion’s dmax in different groups. Compared with the group of dmax<1.5 cm and 1.5 cm ≤ dmax<3.0 cm, the group of dmax≥3 cm had a stronger correlation between Ki-30 min and Ki-65 min, with a lower D%. While the difference between the group of dmax<1.5 cm and 1.5 cm ≤ dmax<3.0 cm in D% was not found. PVE could introduce large quantitative bias, especially in lesions with diameters less than 3 times the resolution of the imaging system. And PVE is affected by noise and strongly depends on the size of the lesion. The smaller the lesion, the greater the underestimation of the uptake value. Previous studies have revealed that lesions larger than 2.8 cm, are weakly or not at all affected by PVE [40]. The kinetic analyses also are subject to the PVE, which varied over time due to blood pool activity and changing tumor contrast [40, 46]. In the Ki-30 min image, the lesions of dmax<3.0 cm might be more severely affected by PVE because of high image noise, high blood pool background uptake, and low tumor uptake. This might lead to the difference in Ki-30 min and Ki-65 min between the different groups of the lesion’s dmax but needs further research.

The main limitations of our study are: (i) Only chest dynamic scanning was performed, and the dynamic quantitative parameter Ki value of lung lesions was analyzed. (ii) Motion correction was not performed. (iii) Input function was not derived from blood samples. (iv) A considerable but imbalanced number of benign and malignant lung lesions were included in the study, which may lead to a statistical bias. Further study with more lesions is needed.

Conclusion

This study indicates that an early 30-minute dynamic FDG-PET acquisition appears to be sufficient to provide good quality quantitative images and accurate dynamic parameter-Ki for quantitative assessment of lung lesions and prediction PD-L1 expression of NSCLC. Protocols with a shortened 30-minute acquisition time may be considered for patients who have difficulty with prolonged acquisitions because of it being more time-saving, patient-friendly, and without sacrificing accurate quantitative parameters.

Data availability

The datasets used or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

18F-FDG:

2-deoxy-2-[18F]fluoro-D-glucose

AC:

Adenocarcinoma

AUC:

Area under the curve

BSREM:

Block sequential regularized expectation maximization

D%:

Percentage difference

dmax :

Maximum diameter

ICIs:

Immune checkpoint inhibitors

IDIF:

Image-derived input function

IF:

Input function

IHC:

Immunohistochemistry

Ki-30 min:

Ki of the first 30 min post-injection

Ki-65 min:

Ki of the first 65 min post-injection

LNs:

Lymph nodes

MRglu :

Metabolic rate of glucose

NCCN:

National comprehensive cancer network

NSCLC:

Non-small cell lung cancer

OS:

Overall survival

PBIF:

Population-based input function

PD-L1:

Programmed death-ligand 1

PET:

Positron emission tomography

PFS:

Progression-free survival

PVE:

Partial-volume effect

ROC:

Receiver operating characteristic

SCC:

Small cell lung cancer

SCC:

Squamous cell carcinoma

SUVmax:

Maximum standard uptake value

SUVmean:

Mean standard uptake value

SUVs:

Standard uptake values

TPS:

Tumor proportion score

VOI:

Volume of interest

References

  1. Dimitrakopoulou-Strauss A. PET-based molecular imaging in personalized oncology: potential of the assessment of therapeutic outcome. Future Oncol. 2015;5:127–42.

    Google Scholar 

  2. Dimitrakopoulou-Strauss A, Pan L, Strauss LG. Quantitative approaches of dynamic FDG-PET and PET/CT studies (dPET/CT) for the evaluation of oncological patients. Cancer Imaging. 2012;12:283–9.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Bertoldo A, Rizzo G, Veronese M. Deriving physiological information from PET images: from SUV to compartmental modeling. Clin T Ransl Imaging. 2014;2(3):239–51.

    Article  Google Scholar 

  4. Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumor imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54.

    Article  CAS  PubMed  Google Scholar 

  5. Rahmim A, Lodge MA, Karakatsanis NA, Panin VY, Zhou Y, McMillan A, et al. Dynamic whole-body PET imaging: principles, potentials and applications. Eur J Nucl Med Mol Imaging. 2019;46(2):501–18. https://doi.org/10.1007/s00259-018-4153-6

    Article  PubMed  Google Scholar 

  6. Dias AH, Pedersen MF, Danielsen H, Munk OL, Gormsen LC. Clinical feasibility and impact of fully automated multiparametric PET imaging using direct patlak reconstruction: evaluation of 103 dynamic whole-body 18F-FDG PET/CT scans. Eur J Nucl Med Mol Imaging. 2021;48(3):837–50. https://doi.org/10.1007/s00259-020-05007-2

    Article  PubMed  Google Scholar 

  7. Torizuka T, Nobezawa S, Momiki S, Kasamatsu N, Kanno T, Yoshikawa E, et al. Short dynamic FDG-PET imaging protocol for patients with lung cancer. Eur J Nucl Med. 2000;27(10):1538–42. https://doi.org/10.1007/s002590000312

    Article  CAS  PubMed  Google Scholar 

  8. Wumener X, Zhang Y, Wang Z, Zhang M, Zang Z, Huang B, et al. Dynamic FDG-PET imaging for differentiating metastatic from non-metastatic lymph nodes of lung cancer. Front Oncol. 2022;12:1005924. https://doi.org/10.3389/fonc.2022.1005924

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Liu G, Yu H, Shi D, Hu P, Hu Y, Tan H, et al. Short-time total-body dynamic PET imaging performance in quantifying the kinetic metrics of 18F-FDG in healthy volunteers. Eur J Nucl Med Mol Imaging. 2022;49(8):2493–503. https://doi.org/10.1007/s00259-021-05500-2

    Article  CAS  PubMed  Google Scholar 

  10. Takesh M. The potential benefit by application of kinetic analysis of PET in the clinical oncology. ISRN Oncol. 2012;2012:349351. https://doi.org/10.5402/2012/349351

  11. Gjedde A, Diemer NH. Autoradiographic determination of regional brain glucose content. J Cereb Blood Flow Metab. 1983;3:303–10.

    Article  CAS  PubMed  Google Scholar 

  12. Lammertsma AA, Hoekstra CJ, Giaccone G, Hoekstra OS. How should we analyse FDG PET studies for monitoring tumour response? Eur J Nucl Med Mol Imaging. 2006;33(Suppl S1):16–21.

    Article  PubMed  Google Scholar 

  13. Freedman M, Sundaram SK, Kurdziel K, Carrasquillo JA, Whatley M, Carson JM, et al. Comparison of SUV and Patlak slope for monitoring of cancer therapy using serial PET scans. Eur J Nucl Med Mol Imaging. 2003;30:46–53.

    Article  PubMed  Google Scholar 

  14. Nakajo M, Ojima S, Kawakami H, Tani A, Hirayama A, Jinguji M, et al. Value of Patlak Ki images from (18)F-FDG-PET/CT for evaluation of the relationships between disease activity and clinical events in cardiac sarcoidosis. Sci Rep. 2021;11:2729.

    Article  ADS  CAS  PubMed  PubMed Central  Google Scholar 

  15. Mankoff DA, Muzi M, Krohn KA. Quantitative positron emission tomography imaging to measure tumor response to therapy: what is the best method? Mol. Imaging Biol. 2003;5:281–5.

    Article  Google Scholar 

  16. Visser EP, Philippens ME, Kienhorst L, Kaanders JH, Corstens FH, de Geus-Oei LF, et al. Comparison of tumor volumes derived from glucose metabolic rate maps and SUV maps in dynamic 18F-FDG PET. J Nucl Med. 2008;49(6):892–8. https://doi.org/10.2967/jnumed.107.049585

    Article  PubMed  Google Scholar 

  17. Muzi M, O’Sullivan F, Mankoff DA, Doot RK, Pierce LA, Kurland BF, et al. Quantitative assessment of dynamic PET imaging data in cancer imaging. Magn Reson Imaging. 2012;30(9):1203–15. https://doi.org/10.1016/j.mri.2012.05.008

    Article  PubMed  PubMed Central  Google Scholar 

  18. Liu G, Xu H, Hu P, Tan H, Zhang Y, Yu H, et al. Kinetic metrics of 18F-FDG in normal human organs identified by systematic dynamic total-body positron emission tomography. Eur J Nucl Med Mol Imaging. 2021;48(8):2363–72. https://doi.org/10.1007/s00259-020-05124-y

    Article  CAS  PubMed  Google Scholar 

  19. Dimitrakopoulou-Strauss A, Pan L, Sachpekidis C. Parametric imaging with dynamic PET for oncological applications: protocols, interpretation, current applications and limitations for clinical use. Semin Nucl Med. 2022;52(3):312–29. https://doi.org/10.1053/j.semnuclmed.2021.10.002

    Article  PubMed  Google Scholar 

  20. Van Sluis J, Yaqub M, Brouwers AH, Dierckx RAJO, Noordzij W, Boellaard R. Use of population input functions for reduced scan duration whole-body patlak 18F-FDG PET imaging. EJNMMI Phys. 2021;8(1):11. https://doi.org/10.1186/s40658-021-00357-8

    Article  PubMed  PubMed Central  Google Scholar 

  21. Wang H, Miao Y, Yu W, Zhu G, Wu T, Zhao X, et al. Improved clinical workflow for whole-body patlak parametric imaging using two short dynamic acquisitions. Front Oncol. 2022;12:822708. https://doi.org/10.3389/fonc.2022.822708

    Article  PubMed  PubMed Central  Google Scholar 

  22. Wang Z, Wu Y, Li X, Bai Y, Chen H, Ding J, et al. Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body 18F-FDG PET imaging. EJNMMI Phys. 2022;9(1):63. https://doi.org/10.1186/s40658-022-00492-w

    Article  PubMed  PubMed Central  Google Scholar 

  23. Wang T, Li B, Shi H, Li P, Deng Y, Wang S, et al. Short-term PET-derived kinetic estimation for the diagnosis of hepatocellular carcinoma: a combination of the maximum-slope method and dual-input three-compartment model. Insights Imaging. 2023;14(1):98. https://doi.org/10.1186/s13244-023-01442-5

    Article  PubMed  PubMed Central  Google Scholar 

  24. Indovina L, Scolozzi V, Capotosti A, Sestini S, Taralli S, Cusumano D, et al. Short 2-[18F]Fluoro-2-Deoxy-D-Glucose PET Dynamic acquisition protocol to evaluate the influx rate constant by regional patlak graphical analysis in patients with non-small-cell lung cancer. Front Med (Lausanne). 2021;8:725387. https://doi.org/10.3389/fmed.2021.725387

    Article  PubMed  Google Scholar 

  25. Wu J, Liu H, Ye Q, Gallezot JD, Naganawa M, Miao T, et al. Generation of parametric Ki images for FDG PET using two 5-min scans. Med Phys. 2021;48(9):5219–31. https://doi.org/10.1002/mp.15113

    Article  CAS  PubMed  Google Scholar 

  26. Chen Z, Cheng Z, Duan Y, Zhang Q, Zhang N, Gu F, et al. Accurate total-body Ki parametric imaging with shortened dynamic 18 F-FDG PET scan durations via effective data processing. Med Phys. 2023;50(4):2121–34. https://doi.org/10.1002/mp.15893

    Article  CAS  PubMed  Google Scholar 

  27. Visser EP, Kienhorst L, de Geus-Oei LF, Oyen WJ. Shortened dynamic FDG-PET protocol to determine the glucose metabolic rate in non-small cell lung carcinoma. 2008 IEEE Nuclear Sci Symp Conf Record Dresden Ger. 2008:4455–8. https://doi.org/10.1109/NSSMIC.2008.4774271

  28. Rittmeyer A, Barlesi F, Waterkamp D, Park K, Ciardiello F, von Pawel J, et al. Atezolizumab versus docetaxel in patients with previously treated non-small-cell lung cancer (OAK): a phase 3, open-label, multicentre randomised controlled trial. Lancet. 2017;389(10066):255–65. https://doi.org/10.1016/S0140-6736(16)32517-X

    Article  PubMed  Google Scholar 

  29. National Comprehensive Cancer Network (NCCN). NCCN clinical practice guidelines in oncology. Non-small Cell Lung Cancer version 4.2021. 2021:4.

  30. palian SL, Hodi FS, Brahmer JR, Gettinger SN, Smith DC, McDermott DF, et al. Safety, activity, and immune correlates of anti-PD-1 antibody in cancer. N Engl J Med. 2012;366(26):2443–54. https://doi.org/10.1056/NEJMoa1200690

    Article  CAS  Google Scholar 

  31. Hu B, Jin H, Li X, Wu X, Xu J, Gao Y. The predictive value of total-body PET/CT in non-small cell lung cancer for the PD-L1 high expression. Front Oncol. 2022;12:943933. https://doi.org/10.3389/fonc.2022.943933

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Takada K, Toyokawa G, Okamoto T, Baba S, Kozuma Y, Matsubara T, et al. Metabolic characteristics of programmed cell death-ligand 1-expressing lung cancer on 18F-fluorodeoxyglucose positron emission tomography/computed tomography. Cancer Med. 2017;6(11):2552–61. https://doi.org/10.1002/cam4.1215

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Kaira K, Shimizu K, Kitahara S, Yajima T, Atsumi J, Kosaka T, et al. 2-Deoxy-2-[fluorine-18] fluoro-d-glucose uptake on positron emission tomography is associated with programmed death ligand-1 expression in patients with pulmonary adenocarcinoma. Eur J Cancer. 2018;101:181–90. https://doi.org/10.1016/j.ejca.2018.06.022

    Article  CAS  PubMed  Google Scholar 

  34. Wu X, Huang Y, Zhao Q, Wang L, Song X, Li Y, et al. PD-L1 expression correlation with metabolic parameters of FDG PET/CT and clinicopathological characteristics in non-small cell lung cancer. EJNMMI Res. 2020;10(1):51. https://doi.org/10.1186/s13550-020-00639-9. PMID: 32430866; PMCID: PMC7237589.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood FlowMetab: off J Int Soc Cereb Blood Flow Metab. 1983;3:1–7. https://doi.org/10.1038/jcbfm.1983.1

    Article  CAS  Google Scholar 

  36. Lawson CL, Hanson RJ. Solving least squares problems. Linear Least Squares Linear Inequal Constraints. 1995;23:158–73. https://doi.org/10.1137/1.9781611971217

    Article  Google Scholar 

  37. Van Sluis J, Boellaard R, Somasundaram A, van Snick P, Borra R, Dierckx R, et al. Image quality and semiquantitative measurements on the biograph vision PET/CT system: initial experiences and comparison with the biograph mCT. J Nucl Med. 2020;61(1):129–35.

    Article  PubMed  Google Scholar 

  38. Wang H, Wu Y, Huang Z, Li Z, Zhang N, Fu F, et al. Deep learning-based dynamic PET parametric Ki image generation from lung static PET. Eur Radiol. 2023;33(4):2676–85. https://doi.org/10.1007/s00330-022-09237-w

    Article  PubMed  Google Scholar 

  39. Huang Z, Wu Y, Fu F, Meng N, Gu F, Wu Q, et al. Parametric image generation with the uEXPLORER total-body PET/CT system through deep learning. Eur J Nucl Med Mol Imaging. 2022;49(8):2482–92. https://doi.org/10.1007/s00259-022-05731-x

    Article  CAS  PubMed  Google Scholar 

  40. Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med. 2007;48(6):932–45. https://doi.org/10.2967/jnumed.106.035774

    Article  PubMed  Google Scholar 

  41. Sun T, Wu Y, Wei W, Fu F, Meng N, Chen H, et al. Motion correction and its impact on quantification in dynamic total-body 18F-fluorodeoxyglucose PET. EJNMMI Phys. 2022;9(1):62. https://doi.org/10.1186/s40658-022-00493-9

    Article  PubMed  PubMed Central  Google Scholar 

  42. Skawran S, Messerli M, Kotasidis F, Trinckauf J, Weyermann C, Kudura K, et al. Can dynamic whole-body FDG PET imaging differentiate between malignant and inflammatory lesions? Life (Basel). 2022;12(9):1350. https://doi.org/10.3390/life12091350

    Article  ADS  CAS  PubMed  Google Scholar 

  43. Ye Q, Wu J, Lu Y, Naganawa M, Gallezot JD, Ma T, et al. Improved discrimination between benign and malignant LDCT screening-detected lung nodules with dynamic over static 18F-FDG PET as a function of injected dose. Phys Med Biol. 2018;63(17):175015. https://doi.org/10.1088/1361-6560/aad97f

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  44. Li W, Pang H, Liu Q, Zhou J. The role of 18F-FDG PET or 18F-FDG-PET/CT in the evaluation of solitary pulmonary nodules. Eur J Radiol. 2015;84(10):2032–7. https://doi.org/10.1016/j.ejrad.2015.06.008

    Article  PubMed  Google Scholar 

  45. Chang CH, Qiu J, O’Sullivan D, Buck MD, Noguchi T, Curtis JD, et al. Metabolic competition in the tumor microenvironment is a driver of cancer progression. Cell. 2015;162(6):1229–41. https://doi.org/10.1016/j.cell.2015.08.016

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  46. Cysouw MCF, Golla SVS, Frings V, Smit EF, Hoekstra OS, Kramer GM, et al. Partial-volume correction in dynamic PET-CT: effect on tumor kinetic parameter estimation and validation of simplified metrics. EJNMMI Res. 2019;9(1):12. https://doi.org/10.1186/s13550-019-0483-z

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by the Shenzhen Science and Technology Program of China (JCYJ20220818101804009), Shenzhen High-level Hospital Construction Fund, National Cancer Center, National Clinical Research Center for Cancer, Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen (E010322003, SZ2020MS008), the Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province (2023B1212060052), Shenzhen Science and Technology Innovation Committee (20220531100209020), Department of Science and Technology of Guangdong Province (2022A1515110716).

Author information

Authors and Affiliations

Authors

Contributions

Ying Liang and Tao Sun contributed to the conception and design of the study, and revision and review of the manuscript. Material preparation and data collection were performed by Fen Du, Xieraili Wumener and Yarong Zhang, Rongliang Wu. Image processing was performed by Maoqun Zhang, Jiuhui Zhao, Jinpeng Zhou, Yiluo Li, and Bin Huang. Data analysis was performed by Fen Du, Xieraili Wumener, Zeheng Xia and Zhiheng Yao. The first draft of the manuscript was written by Fen Du and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Tao Sun or Ying Liang.

Ethics declarations

Ethics approval and consent to participate

This prospective study was approved by the ethics committee of Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences (Clinical Trial Number: KYLH2022-1), which followed the 1964 Helsinki Declaration ethical standards and its subsequent amendments. All patients were provided written informed consent.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Du, F., Wumener, X., Zhang, Y. et al. Clinical feasibility study of early 30-minute dynamic FDG-PET scanning protocol for patients with lung lesions. EJNMMI Phys 11, 23 (2024). https://doi.org/10.1186/s40658-024-00625-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-024-00625-3

Keywords