Optimal clinical protocols for total-body 18F-FDG PET/CT examination under different activity administration plans

Abstract

Background

Highly sensitive digital total-body PET/CT scanners (uEXPLORER) have great potential for clinical applications and fundamental research. Given their increasing sensitivity, low-dose scanning or snapshot imaging is now possible in clinics. However, a standardized total-body 18F-FDG PET/CT protocol is still lacking. Establishing a standard clinical protocol for total-body 18F-FDG PET/CT examination under different activity administration plans can help provide a theoretical reference for nuclear radiologists.

Methods

The NEMA image quality (IQ) phantom was used to evaluate the biases of various total-body 18F-FDG PET/CT protocols related to the administered activity, scan duration, and iterations. Several objective metrics, including contrast recovery (CR), background variability (BV), and contrast-to-noise ratio (CNR), were measured from different protocols. In line with the European Association of Nuclear Medicine Research Ltd. (EARL) guidelines, optimized protocols were suggested and evaluated for total-body 18F-FDG PET/CT imaging for three different injected activities.

Results

Our NEMA IQ phantom evaluation resulted in total-body PET/CT images with excellent contrast and low noise, suggesting great potential for reducing administered activity or shortening the scan duration. Different to the iteration number, prolonging the scan duration was the first choice for achieving higher image quality regardless of the activity administered. In light of image quality, tolerance of oncological patients, and the risk of ionizing radiation damage, the 3-min acquisition and 2-iteration (CNR = 7.54), 10-min acquisition and 3-iteration (CNR = 7.01), and 10-min acquisition and 2-iteration (CNR = 5.49) protocols were recommended for full-dose (3.70 MBq/kg), half-dose (1.95 MBq/kg), and quarter-dose (0.98 MBq/kg) activity injection schemes, respectively. Those protocols were applied in clinical practices, and no significant differences were observed for the SUVmax of large/small lesions or the SUVmean of different healthy organs/tissues.

Conclusion

These findings support that digital total-body PET/CT scanners can generate PET images with a high CNR and low-noise background, even with a short acquisition time and low administered activity. The proposed protocols for different administered activities were determined to be valid for clinical examination and can maximize the value of this imaging type.

Background

Positron emission tomography combined with computed tomography (PET/CT) is a powerful modality to noninvasively collect information about the biodistribution of radiotracers in organs/tissues and anatomic structures in vivo [1,2,3,4]. This imaging modality shows enormous potential for various clinical applications including oncology [5, 6], cardiology [7, 8], and neurology [9]. 18F-fluoro-deoxyglucose (18F-FDG, an analog of glucose) is a common radiotracer widely used in routine oncological PET/CT examinations. Aggressive tumors commonly show a glucose-avid high metabolic feature termed the Warburg effect [10]. 18F-FDG PET/CT examination thus plays a crucial role in tumor localization [11], tumor staging [12, 13], and monitoring treatment responses [14,15,16], with diagnostic accuracy intimately depending on PET images.

PET image quality is greatly restricted by the performance of the PET/CT scanner and reconstruction algorithm, which directly determines whether it is necessary to increase or decrease the injection activity and acquisition time [17]. Over the past two decades, commercial PET/CT scanner detectors have transitioned from using a combination of bismuth germanate oxide scintillators, and photomultiplier tubes (BGO-PMTs) to digital mode—lutetium-based scintillators read out by silicon photomultipliers (SiPMs) [18,19,20]. This evolution has led to huge improvement in detector detection efficiency and made the application of time-of-flight (TOF) technology possible [21, 22]. uEXPLORER (United Imaging Healthcare, Shanghai, China), a model of total-body digital PET/CT scanner, utilizes SiPMs and tiles them along the axial field-of-view (AFOV) to 194 cm [23]. This design enables entire human body PET coverage in one-bed scan and increases the effective detection angle, thereby improving the sensitivity of the total PET system [24,25,26]. With this improved sensitivity (40-fold gain) [27], both the acquisition time and administered activity can be dramatically reduced for PET scans [28] while maintaining image quality. These are favorable in terms of both the economic aspect and patient radiation dose management.

The uEXPLORER has used in clinical practice since 2018 and is now used worldwide. Several prior studies have reported advantages of digital PET/CT scans with a long axial field-of-view in clinical practice, including the potential of low doses [29, 30] and snapshot PET/CT imaging in adults [31], fast scanning for pediatric patients [32], small lesion localization [33, 34], etc. Despite that, how to effectively use total-body PET/CT scanners to meet the ultra-high throughput of patients at PET centers in China is still unclear. Therefore, it is of great importance to unify the scanning protocol to achieve repeatability and reproducibility in total-body 18F-FDG PET scans.

At present, the protocol of total-body PET/CT scanners is adjusted to meet the demands of a highly diagnostic performance and to maximize patient throughput in clinical practice. To our knowledge, a protocol for total-body PET/CT (with improved sensitivity) to meet different imaging conditions has not yet been proposed. To present a standard procedure for static 18F-FDG PET/CT imaging examination, EARL’s guideline [11] detailed patient preparation and precautions, recommendations for 18F-FDG administered activity, and data acquisition protocols. In the present study, we determine the optimal administered activity and reconstruction setting (including scanning duration and iteration numbers) by evaluating the contrast recovery (CR) and noise properties of NEMA IQ phantom images [35, 36] and then verify the optimized protocol in total-body PET/CT clinical examinations.

Materials and methods

PET/CT system

The PET component of the uEXPLORER PET/CT device comprises 8 PET units (24.02 cm axial length) along the axial direction, with the units arranged cylindrically in a ring with a 78.6 cm diameter (detector face-to-face). Each unit consists of 24 detector modules, each of which contains an array of 70 block detectors. Each detector block is composed of a 7 × 6 array of lutetium-yttrium oxyorthosilicate (LYSO) crystals coupled to four SiPMs (SensL J-series). In total, the PET part of the uEXPLORER comprises 13,440 detector blocks with a total of 564,480 LYSO crystals and 53,760 SiPM channels. The energy window of the scanner is 430–645 keV, while the time-of-flight (TOF) resolution of the system is ~ 430 ps. The coincidence time window ranges from 4.5–6.9 ns varies with different PET units.

NEMA IQ phantom preparation

All measurements were performed and evaluated with respect to the NEMA NU2-2007 manual [37]. The NEMA IQ phantom has a background compartment with a measured volume of 9.8 L, a low-density lung insert in the center, and 6 fillable spheres (with diameters of 10, 13, 17, 22, 28, and 37 mm) arranged circumferentially around the lung insert. The background compartment of the phantom was filled with 18F-FDG solution with an activity concentration of ~ 5.2 MBq/L. The 37 and 28 mm cold spheres were filled with water, while the four remaining smaller hot spheres were filled with 18F-FDG solution with an activity concentration four-fold denser than the background concentration. To obtain PET images derived using different administered activities, PET examination of NEMA IQ phantom can be performed at the same bed position on several timepoints to cross few half-lives of 18F radionuclide. Based on the decay rate equation, the injected activity can be estimated by the instantaneous activity concentration multiplied by a decay correction factor of 1.460. Thus, a series of injected activities was estimated, e.g., 7.61 MBq/L (0.206 mCi/kg) for the original measurement, 3.81 MBq/L (0.103 mCi/kg, group full-dose) after the first half-life, 1.90 MBq/L (0.051 mCi/kg, group half-dose) after the second half-life, 0.95 MBq/L (0.026 mCi/kg, group quarter-dose) after the third half-life, 0.48 MBq/L (0.013 mCi/kg, group 1/8-dose) after the fourth half-life, 0.24 MBq/L (0.006 mCi/kg, group 1/16-dose) after fifth half-life, and 0.12 MBq/L (0.003 mCi/kg, group 1/32-dose) after the sixth half-life.

Phantom measurements

NEMA IQ phantom measurement was performed at the center of the PET system, as the axial sensitivity profiles of the uEXPLORER showed a peak sensitivity (up to 147 kcps/MBq) plateau ± 48.5 cm from the center [38]. All necessary corrections were performed, e.g., attenuation and scatter correction. Before PET measurement, a low-dose CT scan (100 kV tube voltage, 10 mAs exposure, 1.0125 pitch) was performed for CT-based attenuation correction (CTAC). To mimic the conditions of clinical practice, all images were reconstructed using the list-mode ordered-subsets expectation maximization algorithm incorporating TOF and point-spread function modeling (OSEM-TOF-PSF) in which several reconstruction parameter settings are fixed, e.g., FOV = 600 mm with a Gaussian post-filter (full width at half maximum = 3 mm), matrix 192 × 192 (3.125 × 3.125 mm2), and 2.89 mm slice thickness. For the different injected doses, the corresponding PET data of NEMA IQ were reconstructed into a series of images (e.g., 40 s, 1 min, 2 min, 3 min, 5 min, 10 min and 30 min) by truncating the acquisition duration. Additionally, we tuned the iteration number from 2 to 15 to explore the effect of iteration convergence on image quality.

Image analysis and interpretation

Using software developed in-house that runs on MATLAB (MathWorks, MA, USA), regions of interest (ROI) were automatically drawn to include the spheres and background of the phantom image according to NEMA NU2-2007 criteria. To optimize the total-body 18F-FDG PET/CT scanning protocol, the background variability (BV) and CR were considered the measures of image noise and quantitation accuracy, respectively. BV was defined by the standard deviation (SD) of 12 ROIs located at five slices (a total of 60 background ROIs of each size)—one central slice crossing the sphere center and four close slices at ± 1 cm and ± 2 cm—divided by the mean activity of these 60 background ROIs. CR was defined as follows [39]:

$${\text{CR}} = \frac{{{\raise0.7ex\hbox{${M_{{\text{H}}} }$} \!\mathord{\left/ {\vphantom {{M_{{\text{H}}} } {M_{{\text{B}}} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${M_{{\text{B}}} }$}} - 1}}{{{\raise0.7ex\hbox{${C_{{\text{H}}} }$} \!\mathord{\left/ {\vphantom {{C_{{\text{H}}} } {C_{{\text{B}}} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${C_{{\text{B}}} }$}} - 1}} \times 100\% ,$$
(1)

where M is the count from the PET image and C is the activity in hot spheres and background ROIs (subscripts H and B, respectively). Notably, the contrast-to-noise ratio (CNR), which is considered a key measure of PET imaging performance, is defined as CR divided by BV. First, we regrouped the subsets based on the BV value, where the upper threshold of BV was 15% [40]. We then calculated the CR in each filtered condition and produced the CNR that provides a quantitative measure of balance for both BV and RC with the same weight ratio [41, 42]. A subjective method of assessment—namely the 5-point Likert scale—was also used to evaluate overall impression of image quality and image noise. All of the center slice images were independently rated by two nuclear radiologists (a senior radiologist with > 5-year experience and a junior radiologist with 1 year of experience reading PET/CT scans) in blinded. The 5-point Likert scale of overall image quality comprises five categories: (1) poor, (2) barely diagnostic, (3) clinically acceptable, (4) superior to the regular quality of daily practice, and (5) excellent. The relationship plot between the CNR value and 5-point Likert scale combines objective and subjective assessment for optimal protocol recommendation. Specifically, the protocol that generates images with high quality (e.g., a score greater than 4- on the 5-point Likert scale) and simpler acquisition settings (e.g., a shorter scan duration and lesser iterations) was preferred in this study.

To facilitate the localization of small lesions, our results—such as BV, CR, and CNR—are mainly demonstrated for hot-spheres with a diameter of 10 mm.

Clinical PET/CT imaging

A total of 15 oncological patients were enrolled in this prospective study. Patient characteristics are listed in Table 1. The exclusion criteria included exercise in the prior 24 h; history of diabetes; and inability to lie supine and still inside the scanner for the acquisition duration. The patients were divided into three groups of three dose levels (full-dose, half-dose, and quarter-dose); each group included five patients. To avoid imaging failure, the lower injected dose levels were not evaluated. Patients in full-dose group, half-dose group, and quarter-dose group were injected with activities of 3.70 MBq/kg (0.1 mCi/kg), 1.85 MBq/kg (0.05 mCi/kg), and 0.93 MBq/kg (0.025 mCi/kg), respectively. All patients fasted for more than 6 h before the 18F-FDG PET examination. Total-body PET/CT examinations were performed with 10–15-min acquisition 60 min after intravenous injection of 18F-FDG. On the PET images, a spherical volume of interest (VOI) with a diameter of 2 cm was manually placed on the aortic arch, normal right liver lobe, muscle of left inner thigh, and normal lung tissue to calculate the organ/tissue SUVmean value. Similarly, a size-adaptive VOI was manually placed on the major suspect lesion and one small metastatic lesion to calculate the lesion SUVmax value, respectively. Statistical comparisons were made using the paired t-test performed using the R statistical package. p values < 0.05 were considered statistically significant.

Table 1 Clinical characteristics of the enrolled oncological patients

Results

NEMA IQ phantom

Figure 1 depicts a duration-iteration-dependent PET image of the NEMA IQ phantom with different injected activities from the transverse view. Visually, the higher image quality requires a longer acquisition time, while the background noise increases with the increasing iteration number.

Fig. 1
figure 1

The schematic figure of transverse image slices of the NEMA IQ phantom achieved at the different acquisition time, iteration number, and injected activity. The PET raw-data were measured from the NEMA IQ phantom with a background activity of ~ 5.2 MBq/L (hot sphere-to-background ratio = 4:1) scanned for 30 min. By truncating the list-mode PET raw-data, a series of PET images were reconstructed into 40 s, 1 min, 2 min, 3 min, 5 min, 10 min and 30 min (ordered from left to right column), where the iteration number applied in OSEM-TOF-PSF reconstruction process was 2, 3, 4, 5, and 15 from top to bottom line, respectively. The activity per body weight was estimated from the instantaneous background activity concentration taking into account of the nature decay of 18F nuclide for a waiting time of 60 min after injection. The activity per body weight ranged from 0.24 to 7.61 MBq/L

Figure 2 illustrates the BV values at different injected activities from 0.47–3.81 MBq/L. As the number of iterations increases so does image noise, ultimately leading to a high BV value; this relationship is consistent with previously published reports [40]. For example, with a BV at injection activity of 3.81 MBq/L and 10-min acquisition duration, image noise increased at a factor of 2.66 compared with 2 and 15 iterations (3.35% at 2 iterations vs. 8.77% at 15 iterations). In contrast, a longer acquisition time results in minimum image noise. Specifically, extending the acquisition time from 10 to 30 min enables a reduction in BV from 8.77% to 4.57% for 15 iterations at 3.81 MBq/L. As the injected activity decreases, the BV value in the same condition increases, indicating worse image quality. However, the trend for the CR value (plotted in Additional file 1: Fig. 1) is quite different from the BV values presented. In general, an increase in iterations results in larger CR values. Such differences hinder the use of BV or CR values as independent indicators to evaluate image quality.

Fig. 2
figure 2

Background variability (BV) map, which is derived from NEMA IQ phantom measurement (hot sphere: background, 4:1) at four estimated doses, as a function of both factors—acquisition time and iterations. The dark area represents that BV value is more than 15%

Figure 3 demonstrates an iteration-and-duration-dependent CNR for different injected activities while Fig. 4 presents the relationship between the objective image CNR value and subjective 5-point Likert scale score. In addition to the total body PET/CT scanner data, similar experimental results using a conventional short AFOV PET/CT scanner (Siemens Biograph mCT) are presented in Fig. 4 (the experimental setting and relative results are detailed in the Additional file 1). No substantial difference was found in the mean CNR value for each 5-point Likert scale grade between the long and short AFOV scanner. Hence, we intuitively equally exchanged these two indicators and defined a CNR of 5, 7, and larger than 9 as a grade of 3, 4, and 5 on the 5-point Likert scale for later analysis. Notably, grade 3 indicates clinical routine image quality using a conventional short AFOV digital PET/CT scanner, as described by Zhao et al. [32].

Fig. 3
figure 3

Contrast-to-noise ratio (CNR) map as a function of both factors—acquisition time and iterations at different estimated injected dose. The dark area represents the corresponding BV value that is more than 15%. The CNR values that excessed 5.0 were considered as acceptable for diagnosis

Fig. 4
figure 4

Relationship between the objective indicator (CNR) and the subjective assessment (5-point Likert scale). A The reference example images of grades 1–5 in the 5-point Likert scale (L1: poor, L5: excellent). B Scatter plot for the CNR value vs. 5-point Likert scale upon the total body PET/CT scanner and the conventional short AFOV PET/CT scanner (derived from Siemens Biograph mCT scanner in this study), respectively. The 5-point Likert scale was independently rated by two nuclear radiologists in blind

As shown in Fig. 3, the 30-min duration condition generally achieves higher CNR values than the other shorter durations; however, it is not recommended for clinical application considering patient tolerance and departmental throughput. For the injected activity of 3.81 MBq/L (full-dose), several protocols are promising (CNR > 7, i.e. 5-point Likert scale larger than grade 4) in different acquisition durations, e.g., 3 iterations for 10 min, 3 iterations for 5 min, and 2 iterations for 3 min. Accounting for the shortened scan duration and reconstruction time consumption, the 2 iterations for 3-min acquisition are more suitable for a full-dose scheme in routine clinical examination since it slightly degrades the image quality but accelerates data acquisition speed nearly one- to two-fold that for the 3 iterations for 5 min and 3 iterations for 10-min settings, respectively. As the injected activity decreased to a half-dose (1.90 MBq/L), the PET images of the phantom were similar quality (CNR = 7.01) to those for full-dose under the conditions of 3-iteration and 10-min scanning, and is the optimal protocol for the half-dose scheme. When the injected activity is 0.95 MBq/L (a quarter of a full-dose) the image degraded; in this case, the 10-min and 2-iteration acquisition were suggested, in which a maximal CNR of 5.49 was found. Notably, when the injected activity is 0.48 MBq/L (1/8 of a full-dose), the CNR could reach 5 in the protocol with 10 min and 2 iterations, implying the feasibility of a low-dose PET scan with total-body PET scanner.

Examples of clinical images

The aforementioned protocols were verified and evaluated in each dose group and then compared with the standard uptake values (SUVs) for several organs/tissues.

For the FDG injection with full dose, there was no significant difference for the mean SUV (SUVmean) in the blood pool (p > 0.1507), lung (p > 0.4206), muscle (p > 0.4206), and liver (p > 0.4206), showing robust stability that background did not change dramatically with the acquisition time and iterative number (Fig. 5). Although the median value increased as the iterative number increased in the intra-group comparison with the same acquisition time, the maximum SUV (SUVmax) at suspected lesions and small lesions were not significantly different (all p > 0.2222). The similar behavior was observed in the groups with injected activity of half-dose and quarter-dose level (for further details, see Additional file 1: Fig. 2 and Fig. 3).

Fig. 5
figure 5

Iteration-dependent SUVmean at muscle, lung, liver, and blood pool, and SUVmax at suspected major lesions and metastatic small lesions upon full-dose group for an acquisition time from 40 to 600 s, where the iterative numbers set at 2, 3, 4, 5, and 15

To clarify the clinical feasibility of low-dose protocols, we compared the SUVs of the blood pool, lung, muscle, liver, suspected major lesions, and small lesions derived from the aforementioned protocols (as shown in Fig. 6). Although the median SUVmean values of lung, muscle, blood pool, and liver derived from the quarter-dose protocol with 600-s acquisition and two iterations were relatively lower than others, a significant difference was only observed in lung and muscle between the quarter-dose and full-dose (p < 0.0317). For other organs/tissues, the differences were not significant (all p > 0.0555). For more intuitive presentation, three clinical PET examination results with different injected activities (full-dose, half-dose, and quarter-dose) are shown in Additional file 1: Fig. 4, Fig. 5, and Fig. 6, respectively.

Fig. 6
figure 6

Comparison of SUVmean at lung, muscle, blood pool and liver, and SUVmax at suspected major lesions and small metastatic lesions among six different suggested protocols that involves full-dose, half-dose, and quarter-dose

Discussion

Total-body 18F-FDG PET/CT has the potential to provide fast acquisition with a low injected dose while assuring the PET images are of clinically acceptable quality. Based on NEMA IQ phantom studies, CNR can serve as a measure to quantitatively evaluate different PET/CT examination protocols. To achieve the best image quality using a total-body PET/CT scanner (uEXPLORER), several plausible protocols were proposed with three injected-activities schemes (0.925–3.7 MBq/kg, 0.025–0.1 mCi/kg) and successfully verified in the clinical practice.

The acquisition duration should be prolonged for images with a less noisy background. However, PET images with a long duration may be blurry because oncological patients are often in intense pain and tend to move during acquisition. The increased injected activity enables faster acquisition, thus avoiding motion artefacts that compromise image quality. However, it also results in a higher risk of ionizing radiation for patients. Therefore, achieving a balance between image quality, acquisition time, and injected activity is of great clinical significance. Although the full-dose (3.7 MBq/L, 0.1 mCi/L), 30-min acquisition, and 5-iteration scheme provided excellent image quality (CNR = 18.7), it is still not clinically feasible. Compared to a 20–30-min whole-body scan using a conventional short AFOV scanner, the full-dose scheme with 3-min acquisition using a total body scanner has tremendous clinical potential since it greatly increases patient throughput and achieves images with relative high quality (CNR > 7). Given the rapidly increasing sensitivity of uEXPLORER, it is able to further shorten the acquisition duration of the PET signal. A PET image, which is similar to an image derived from the conventional PET-CT scanner, has been found with reconstruction settings of 1 min and 2 iterations (CNR = 5.29, with a 5-point Likert scale larger than grade 3) using total-body PET/CT scanner. This fast scanning might improve the accuracy in pediatric diagnosis—eliminating the false-positive lesions caused by the motion, by only reconstructing with the data before the motion. Example of fast scanning eliminating the motion-induced false-positive lesions is shown in Additional file 1: Fig. 7. The results revealed that fast 18F-FDG PET/CT scans are feasible and useful for examining both adult and pediatric patients. Our findings are consistent with those of Li et al. and Tan et al. that fast-scan and low-dose schemes are helpful for adults with cancer pain [28, 43] and pediatric PET/CT examination under clinical settings [32].

To minimize radiation-induced DNA damage, the injected activity of 18F-FDG can be greatly reduced with total-body PET/CT scanners. This has been discussed by Shi et al. in previous studies [29, 30, 44], but the optimal protocols have not been systematically investigated. Based on the calculation of CNR, we designed several low-dose protocols to yield a PET image with clinically acceptable quality. The CNR calculated from 18F-FDG PET imaging with half-dose injection, 5-min acquisition, and 2 iterations was approximating similar to 5 (clinically acceptable image quality). Although the image is more blurred than that reconstructed using 10-min acquisition and 3 iterations (BV: 6.37% vs. 5.66%; CNR: 7.01 vs. 5.11), it met the diagnostic requirements while greatly enhancing patient throughput in our department. The injected activity can be further decreased to a quarter of the full-dose, whereas two iterations and more than 10-min acquisition time are suggested to generate a diagnosable image. Notably, the associated CNR is 5.49, which is slightly higher than that of the half-dose PET imaging with 5 min and 2 iterations. This relationship respects the rule of time-activity product. In clinical practice, oncological patients often have to undergo several PET examinations (interim PET) to monitor their treatment response. Such interim PET is especially important for immunotherapy, which may introduce side effects such as hyperprogression or pseudoprogression [45, 46]. With low-dose interim PET, the treatment response can be carefully evaluated between cycles so that adjustments to the treatment plan can be made accordingly and the patients can be protected from overexposure to radiation. In this context, the protocol of 1/8-dose (administered activity of 0.48 MBq/kg, 0.0125 mCi/kg, 10-min acquisition, 2-iteration) could be used in clinical practice with CNR > 5. However, the accuracy of SUV in the low-dose scheme should be further analyzed and evaluated using a large number of patients.

Reproducibility and repeatability are of great importance for PET multi-center studies [2, 47]. Given the injected activity of 3.7 MBq/kg (0.1 mCi/kg), our results indicate that with a reconstruction setting of 2 iterations and 1-min acquisition (CNR = 5.29, i.e., grade 3 on the 5-point Likert scale) for the OSEM algorithm, a total-body scanner generates images with similar quality and performance as conventional PET/CT scanner. This result agrees with Zhao et al., who determined the equivalent state between uEXPLORER and other modern digital PET/CT scanners in terms of signal-to-noise ratio and tumor-to-background ratio [31]. For the half-dose, the PET images were reconstructed at two iterations with 5-min data. The half-dose images had a CNR of 5.11, which was consistent with images derived from the full-dose scheme using a total-body PET/CT scanner (CNR = 5.29). This half-dose protocol for the total-body scanner may generate equivalent images as a short AFOV (20–30 cm) PET/CT since they are at the same CNR level (~ 5). Furthermore, the images acquired with the total-body PET/CT scanner under the quarter-dose scheme had a similar CNR (CNR = 5.49) as the short AFOV scanner.

Based on the NEMA IQ phantom study, six protocols involving full-dose, half-dose, and quarter-dose methods were applied in clinical PET/CT examinations. Our results demonstrated that the SUVmean at background and SUVmax at lesions were nearly robust. Nevertheless, SUVmean at the lung and muscle exhibited a significant difference between the quarter-dose and full-dose protocols. However, this did not impede clinical diagnosis in our study.

There are some limitations of this study that should be noted. First, we should mention that the scatter and attenuation patterns, as well as the activity distribution of the NEMA phantom, are significantly different from patient scans due to differences in the body size. Such a translated result may induce biases in practical application; however, even for the quarter-dose clinical patient examination, there were no difficult diagnoses in our study. Besides, to avoid examination failure, ultra-low-dose conditions—such as 1/8 and 1/16 activity administered cases—were not performed in clinical patients. A larger standard human body-like phantom would be needed for further evaluation before guaranteeing clinical safety.

We should also mention that the aforementioned optimized protocols involved in this study merely adapt to the OSEM-TOF-PSF reconstruction algorithm for 18F-FDG PET/CT examinations. The application of different reconstruction algorithms, pixel sizes, and smoothing kernel are not considered. The optimal protocols for other radiotracers on total-body PET/CT scanners—such as 68 Ga-DOTA-TATE, 68 Ga-FAPI-04, and 68 Ga-PSMA—are not discussed in this study. Moreover, the involved cancer types were limited and the patient cohort was small. As a result, the results may be influenced by patient characteristics such as BMI, cancer type, age, etc.

Conclusion

Utilizing a digital PET/CT scanner with a 194-cm-long axial field-of-view, total-body 18F-FDG PET images were generated using the standard protocol to yield high image quality. The optimal acquisition duration and iteration number for different injected activity conditions were recommended, which do help maximize the potential clinical applications of total body PET/CT scanners.

Availability of data and materials

Not applicable.

References

  1. Weber WA, Grosu AL, Czernin J. Technology Insight: advances in molecular imaging and an appraisal of PET/CT scanning (in eng). Nat Clin Pract Oncol. 2008;5(3):160–70.

    Article  CAS  PubMed  Google Scholar 

  2. Boellaard R, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42(2):328–54.

    Article  CAS  PubMed  Google Scholar 

  3. Stauss J, et al. Guidelines for 18 F-FDG PET and PET-CT imaging in paediatric oncology. Eur J Nucl Med Mol Imaging. 2008;35(8):1581–8.

    Article  CAS  PubMed  Google Scholar 

  4. Vaquero JJ, Kinahan P. Positron emission tomography: current challenges and opportunities for technological advances in clinical and preclinical imaging systems. Annu Rev Biomed Eng. 2015;17:385–414.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Rohren EM, Turkington TG, Coleman RE. Clinical applications of PET in oncology (in eng). Radiology. 2004;231(2):305–32.

    Article  PubMed  Google Scholar 

  6. Weissleder R. Molecular imaging in cancer. Science. 2006;312(5777):1168–71.

    Article  CAS  PubMed  Google Scholar 

  7. Murthy VL, et al. Clinical quantification of myocardial blood flow using PET: joint position paper of the SNMMI cardiovascular council and the ASNC (in eng). J Nucl Med. 2018;59(2):273–93.

    Article  CAS  PubMed  Google Scholar 

  8. Machac J. Cardiac positron emission tomography imaging. Semin Nucl Med 2005;35(1):17–36.

  9. Tai YF, Piccini P. Applications of positron emission tomography (PET) in neurology. J Neurol Neurosurg Psychiatry. 2004;75(5):669–76.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Hanahan D, Weinberg RA. Hallmarks of cancer: the next generation. Cell. 2011;144(5):646–74.

    Article  CAS  PubMed  Google Scholar 

  11. Bar-Shalom R, et al. Clinical performance of PET/CT in evaluation of cancer: additional value for diagnostic imaging and patient management. J Nucl Med. 2003;44(8):1200–9.

    PubMed  Google Scholar 

  12. Paydary K, et al. The evolving role of FDG-PET/CT in the diagnosis, staging, and treatment of breast cancer (in eng). Mol Imaging Biol. 2019;21(1):1–10.

    Article  CAS  PubMed  Google Scholar 

  13. Kandathil A, Kay FU, Butt YM, Wachsmann JW, Subramaniam RM. Role of FDG PET/CT in the eighth edition of TNM staging of non-small cell lung cancer (in eng). Radiographics 2018;38(7):2134–2149.

  14. Bastiaannet E, et al. The value of FDG-PET in the detection, grading and response to therapy of soft tissue and bone sarcomas; a systematic review and meta-analysis (in eng). Cancer Treat Rev. 2004;30(1):83–101.

    Article  CAS  PubMed  Google Scholar 

  15. Aide N, Hicks RJ, Le Tourneau C, Lheureux S, Fanti S, Lopci E. FDG PET/CT for assessing tumour response to immunotherapy: report on the EANM symposium on immune modulation and recent review of the literature (in eng). Eur J Nucl Med Mol Imaging. 2019;46(1):238–50.

    Article  PubMed  Google Scholar 

  16. Kim J, Lee JO, Paik JH, Lee WW, Kim SE, Song YS. Different predictive values of interim (18)F-FDG PET/CT in germinal center like and non-germinal center like diffuse large B-cell lymphoma (in eng). Ann Nucl Med. 2017;31(1):1–11.

    Article  CAS  PubMed  Google Scholar 

  17. Eriksson L, et al. An investigation of sensitivity limits in PET scanners. Nucl Instrum Methods Phys Res Sect A. 2007;580(2):836–42.

    Article  CAS  Google Scholar 

  18. Jones T, Townsend DW. History and future technical innovation in positron emission tomography. J Med Imaging. 2017;4(1): 011013.

    Article  Google Scholar 

  19. S. Vandenberghe, P. Moskal, and J. S. Karp, "State of the art in total body PET," EJNMMI Phys, vol. 7, no. 1, p. 35, May 25 2020.

  20. Van Eijk CW. Radiation detector developments in medical applications: inorganic scintillators in positron emission tomography. Radiat Prot Dosimetry. 2008;129(1–3):13–21.

    Article  PubMed  Google Scholar 

  21. Cherry SR, Jones T, Karp JS, Qi J, Moses WW, Badawi RD. Total-body PET: maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med. 2018;59(1):3–12.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Alberts I, et al. Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the Biograph Vision Quadra with the Biograph Vision PET/CT (in eng). Eur J Nucl Med Mol Imaging 2021.

  23. Badawi RD, et al. First human imaging studies with the EXPLORER total-body PET scanner. J Nucl Med. 2019;60(3):299–303.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Borasi G, Fioroni F, Del Guerra A, Lucignani G. PET systems: the value of added length. Eur J Nucl Med Mol Imaging. 2010;37(9):1629–32.

    Article  PubMed  Google Scholar 

  25. Cherry SR, Badawi RD, Karp JS, Moses WW, Price P, Jones T. Total-body imaging: transforming the role of positron emission tomography. Sci Transl Med 2017;9(381).

  26. Zhang X, Cherry SR, Xie Z, Shi H, Badawi RD, Qi J. Subsecond total-body imaging using ultrasensitive positron emission tomography. Proc Natl Acad Sci. 2020;117(5):2265–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Zhang X, Zhou J, Cherry SR, Badawi RD, Qi J. Quantitative image reconstruction for total-body PET imaging using the 2-meter long EXPLORER scanner. Phys Med Biol. 2017;62(6):2465–85.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  28. Tan H, et al. Total-body PET/CT: current applications and future perspectives. AJR Am J Roentgenol. 2020;215(2):325–37.

    Article  PubMed  Google Scholar 

  29. Liu et al. Ultra-low-activity total-body dynamic PET imaging allows equal performance to full-activity PET imaging for investigating kinetic metrics of 18F-FDG in healthy volunteers. Eur J Nucl Med Mol Imaging 01/22 2021.

  30. Tan H et al. Total-body PET/CT using half-dose FDG and compared with conventional PET/CT using full-dose FDG in lung cancer. Eur J Nucl Med Mol Imaging Nov 27 2020.

  31. Hu P et al. Total-body 18F-FDG PET/CT scan in oncology patients: How fast could it be? Eur J Nucl Med Mol Imaging 2021/04/18 2021.

  32. Zhao YM, et al. Image quality and lesion detectability in low-dose pediatric (18)F-FDG scans using total-body PET/CT (in eng). Eur J Nucl Med Mol Imaging 2021.

  33. Fu F, et al. Total-body dynamic PET/CT of micro-metastatic lymph node in a patient with lung cancer (in eng). Eur J Nucl Med Mol Imaging. 2021;48(5):1678–9.

    Article  PubMed  Google Scholar 

  34. Zhang YQ, et al. The image quality, lesion detectability, and acquisition time of (18)F-FDG total-body PET/CT in oncological patients (in eng). Eur J Nucl Med Mol Imaging. 2020;47(11):2507–15.

    Article  PubMed  Google Scholar 

  35. Performance Measurements of Positron Emission Tomographys (PET): NEMA, Jun.20.2018. [Online]. Available.

  36. Ziegler S, Jakoby BW, Braun H, Paulus DH, Quick HH. NEMA image quality phantom measurements and attenuation correction in integrated PET/MR hybrid imaging (in eng). EJNMMI Phys. 2015;2(1):18.

    Article  PubMed  PubMed Central  Google Scholar 

  37. N. E. M. A. J. N. S. P. N. 2-, "Performance measurement of positron emission tomographs," 2007.

  38. Spencer BA, et al. Performance evaluation of the uEXPLORER total-body PET/CT scanner based on NEMA NU 2-2018 with additional tests to characterize PET scanners with a long axial field of view 2021;62(6):861–870

  39. Vallot D, et al. Evaluation of PET quantitation accuracy among multiple discovery IQ PET/CT systems via NEMA image quality test (in eng). EJNMMI Phys 2020;7(1):30.

  40. Gnesin S, et al. Phantom-based image quality assessment of clinical 18F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CT. EJNMMI Phys 2020;7(1):1.

  41. Koopman D, et al. Technical note: how to determine the FDG activity for tumour PET imaging that satisfies European guidelines (in eng). EJNMMI Phys. 2016;3(1):22.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Fukukita H, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 2.0 (in eng). Ann Nucl Med 2014;28(7)L693–70.

  43. Li Y, et al. Ultra-short time imaging of total-body PET/CT for cancer pain-induced untenable body position (in eng). Eur J Nucl Med Mol Imaging 2021.

  44. Sui X, et al. Total-body PET/computed tomography highlights in clinical practice: experiences from Zhongshan Hospital, Fudan University. PET Clin. 2021;16(1):9–14.

    Article  PubMed  Google Scholar 

  45. Burggraaff CN, et al. Predictive value of interim positron emission tomography in diffuse large B-cell lymphoma: a systematic review and meta-analysis (in eng). Eur J Nucl Med Mol Imaging. 2019;46(1):65–79.

    Article  PubMed  Google Scholar 

  46. Sachpekidis C, Hassel JC, Kopp-Schneider A, Haberkorn U, Dimitrakopoulou-Strauss A. Quantitative dynamic (18)F-FDG PET/CT in survival prediction of metastatic melanoma under PD-1 inhibitors (in eng). Cancers (Basel) 13(5).

  47. Busemann Sokole E, Płachcínska A, Britten A. Acceptance testing for nuclear medicine instrumentation (in eng). Eur J Nucl Med Mol Imaging 2010;37(3):672–81.

Download references

Acknowledgements

Not applicable.

Funding

This work is supported by the National Natural Science Foundation of China (No. 11205081), Science and Technology Planning Project of Guangdong Province of China (No.2014A020212175) and President Foundation Project of Nanfang Hospital, Southern Medical University (No.2020L002 and No. 2021C017).

Author information

Authors and Affiliations

Authors

Contributions

YH contributed to conceptualization; YH, MW, LJ, and LW were involved in methodology; LC and QW were involved in drug injection and data collection. YH, LW, HW, JF and JW contributed to formal analysis and investigation. WX and YH were involved in the radiopharmaceutical production and quality control. YH, JF and JW were involved in writing—original draft preparation; HW and YH contributed to writing—review and editing. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yanjiang Han.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This prospective study was approved by the clinical research ethics committee of Nanfang Hospital Southern Medical University. Written informed consent was obtained from all enrolled patients.

Consent for publication

Not applicable.

Competing interests

Jiatai Feng and Jingyi Wang are employees of United Imaging Healthcare. Wanbang Xu works at the Department of Traditional Chinese Medicine in Guangdong Institute for Drug Control. The other authors working at Nanfang PET Center in Nanfang Hospital Southern Medical University have full control of the data and Nanfang hospital do not have any revenue sharing with other two Institutes. No other potential conflicts of interest relevant to this article exist including employment, royalties, stock options, or patients.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. Supplementary file.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Huang, Y., Wang, M., Jiang, L. et al. Optimal clinical protocols for total-body 18F-FDG PET/CT examination under different activity administration plans. EJNMMI Phys 10, 14 (2023). https://doi.org/10.1186/s40658-023-00533-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-023-00533-y

Keywords