Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body 18F-FDG PET imaging

Abstract

Purpose

Efforts have been made both to avoid invasive blood sampling and to shorten the scan duration for dynamic positron emission tomography (PET) imaging. A total-body scanner, such as the uEXPLORER PET/CT, can relieve these challenges through the following features: First, the whole-body coverage allows for noninvasive input function from the aortic arteries; second, with a dramatic increase in sensitivity, image quality can still be maintained at a high level even with a shorter scan duration than usual. We implemented a dual-time-window (DTW) protocol for a dynamic total-body 18F-FDG PET scan to obtain multiple kinetic parameters. The DTW protocol was then compared to several other simplified quantification methods for total-body FDG imaging that were proposed for conventional setup.

Methods

The research included 28 patient scans performed on an uEXPLORER PET/CT. By discarding the corresponding data in the middle of the existing full 60-min dynamic scan, the DTW protocol was simulated. Nonlinear fitting was used to estimate the missing data in the interval. The full input function was obtained from 15 subjects using a hybrid approach with a population-based image-derived input function. Quantification was carried out in three areas: the cerebral cortex, muscle, and tumor lesion. Micro- and macro-kinetic parameters for different scan durations were estimated by assuming an irreversible two-tissue compartment model. The visual performance of parametric images and region of interest-based quantification in several parameters were evaluated. Furthermore, simplified quantification methods (DTW, Patlak, fractional uptake ratio [FUR], and standardized uptake value [SUV]) were compared for similarity to the reference net influx rate Ki.

Results

Ki and K1 derived from the DTW protocol showed overall good consistency (P < 0.01) with the reference from the 60-min dynamic scan with 10-min early scan and 5-min late scan (Ki correlation: 0.971, 0.990, and 0.990; K1 correlation: 0.820, 0.940, and 0.975 in the cerebral cortex, muscle, and tumor lesion, respectively). Similar correlationss were found for other micro-parameters. The DTW protocol had the lowest bias relative to standard Ki than any of the quantification methods, followed by FUR and Patlak. SUV had the weakest correlation with Ki. The whole-body Ki and K1 images generated by the DTW protocol were consistent with the reference parametric images.

Conclusions

Using the DTW protocol, the dynamic total-body FDG scan time can be reduced to 15 min while obtaining accurate Ki and K1 quantification and acceptable visual performance in parametric images. However, the trade-off between quantification accuracy and protocol implementation feasibility must be considered in practice. We recommend that the DTW protocol be used when the clinical task requires reliable visual assessment or quantifying multiple micro-parameters; FUR with a hybrid input function may be a more feasible approach to quantifying regional metabolic rate with a known lesion position or organs of interest.

Introduction

18F-FDG positron emission tomography (PET) imaging is widely used for tumor characterization, staging, restaging, and therapy monitoring [1, 2]. Currently, FDG PET quantification is mostly limited to standard uptake value (SUV), a semiquantitative measure derived from static acquisition. Several alternative measures exist to better quantify the differences in subjects, including normalizing the activity concentration by using the lean body mass (SUL) [3] and the body surface area [4], rather than the total body mass. SUV-based measures have a number of disadvantages [5,6,7]. An optimal scan window, for example, may differ between subjects; inter-study variability in blood supply may also influence quantification; and an uptake image may contain both specific and nonspecific components. To reduce the effect of nonspecific uptake, SUV was proposed to be normalized to the uptake in the background region (SUR) [8] or to the integral of the arterial input function (fractional uptake ratio [FUR]) [9]. To improve the diagnosis, it has also been proposed to assess the relative SUV change between early and late scans [10].

On the other hand, net influx rate Ki is a full-quantitative parameter that outperforms SUV in differentiating malignant from benign lesions and delineating tumor volume [11,12,13,14,15,16]. Accurate Ki estimation necessitates full dynamic PET acquisition. A standard dynamic acquisition protocol warrants more than 60 min of list-mode acquisition beginning with tracer injection, together with sequential arterial blood sampling. The two most common methods for calculating Ki are irreversible two-tissue compartment model (2T3k) fitting [17] and Patlak graphical analysis [18, 19]. If the arterial input function of the full scan can be approximated, for example, with a population-based one sampled and averaged from previous full dynamic scans, the scan time for Patlak graphical analysis can be significantly reduced [20].

Many efforts have been made to avoid invasive blood sampling and shorten the duration of scans. Several noninvasive methods for determining the input function have been proposed, including image-derived input function (IDIF) [21, 22], population-based input function (PBIF) [23, 24], model-based input function [25], and hybrid input function [26] (combination of IDIF and PBIF). The difficulty in obtaining the input function for a bed position with no large artery in the field of view (FOV) is a concern when using these methods. IDIF derived from carotid artery scanning, for example, is known to be underestimated in brain PET. The FlowMotion technique can obtain the input function for the entire body from the aortic arteries, allowing for whole-body Ki imaging [27]. While for reducing scan time, one option is to replace the long dynamic acquisitions with two static acquisitions, i.e., the dual-time-window protocol. This protocol has been shown to be reliable for obtaining reliable Ki quantification and images [8, 28,29,30]. The majority of the preceding studies, however, are based on Patlak graphical analysis for FDG, from which only the macro-parameter Ki (net FDG influx rate) can be obtained. On the other hand, nonlinear estimation based on 2T3k model can also yield micro-kinetic parameters like K1, k2, and k3. Previous research has shown that K1 is a useful marker for identifying tumor subgroups [31] and assessing chemotherapy response [32]. k3 was proven to be effective in subtyping of pheochromocytoma and paraganglioma [33]. The combination of K1 and Ki is also useful in gaining access to metabolic tumors [34]. However, a whole-body assessment of these micro-parameters would require a total-body dynamic scan. Nonlinear parametric estimation is known sensitive to the quality of dynamic images; thus, region of interest (ROI)-based kinetic analysis is frequently used instead of voxelized parametric images. One can imagine that the quality of kinetic modeling would be even less reliable with a simplified imaging protocol because the scan data would be less reliable.

Total-body scanners, such as the uEXPLORER or Biograph Vision Quadra, may be able to address these issues in dynamic imaging. First, the whole-body coverage allows for the noninvasive input function from the aortic arteries to be obtained, which has already been shown to be close to the arterial sampling, at least for FDG [35]. Zhang et al. [36] and Sari et al. [37] investigated the feasibility of extracting an input function from dynamic images of the left atrium, left ventricle, aortic artery, and carotid artery, for example. This capability can improve the dependability of IF estimation methods such as PBIF approximation and the hybrid method. Second, with a significant increase in sensitivity, image quality can be maintained at a high level even when scanning for a shorter period of time than usual [38,39,40]. This will enable the use of more flexible simplified protocols with significantly reduced scan time, allowing for more confident linear or nonlinear parametric estimation. uEXPLORER was used to investigate simplified dynamic FDG total-body imaging protocols. Feng et al. [41] investigated early FDG kinetics using only 2-min post-injection data to obtain whole-body K1 images with IDIF. Wu et al. [26] proposed two simplified acquisition protocols for producing whole-body Ki images with hybrid input functions, reducing the scan time to 10 min. Liu et al. [39] obtained 45-min scan data and reported that the quantification of Ki was comparable to the 60-min scan.

In this study, we used a dual-time-window (DTW) total-body FDG scan protocol to obtain multiple kinetic parameters in key areas. The protocol included two dynamic acquisitions: An early acquisition performed 10 min after injection and a late acquisition performed with a fixed end time of 60 min. Nonlinear fitting with rational functions was used to fill the missing data in the interval. The full input function was obtained using a hybrid approach with a population-based image-derived function (PB-IDIF) from 15 subjects. The processed data at each ROI were then used to estimate micro- and macro-kinetic parameters K1, k2, k3, Ki, and Vb in 2T3k model. Fast nonlinear least squares fitting at each voxel was also used to generate whole-body parametric images. The proposed DTW protocol quantification was evaluated and compared with existing simplified quantification methods such as SUV, FUR, and Patlak analysis. Finally, we discuss which simplified quantification method is appropriate for specific clinical applications. The novelties of the present work are twofold: First, with 10-min early acquisition and a 5-min late acquisition on a total-body PET, relatively accurate quantifications of Ki, K1, k3, and Vb at whole-body level were obtained. Second, quantification of DTW was first time compared with several existing simplified quantification methods, including Patlak analysis, FUR, and SUV, as surrogates for the net influx rate Ki.

Materials and methods

Study subjects and image generation

The present study was approved by the Institutional Review Board of Henan Provincial People’s Hospital, China. After obtaining informed consent, 28 human subjects were enrolled in the study. Table 1 shows the demographic information for all participants. Among the 28 participants, 18 were used for normal organ/tissue quantification, and 10 were used for tumor lesion quantification (one lesion from each). A total-body PET/CT scan was performed on each subject at Henan Provincial People's Hospital using an uEXPLORER PET/CT scanner (United Imaging Healthcare, Shanghai, China). The following are the scan procedures. Following the injection of 18F-FDG in the lower extremity vein, a CT scan was performed for attenuation correction, followed by a 60-min list-mode acquisition. The list-mode data were partitioned and reconstructed to produce 66 images (24 × 5 s, 6 × 10 s, 6 × 30 s, 6 × 60 s, 24 × 120 s). Each frame was reconstructed with manufacturer-supplied reconstruction software (TOF-OSEM with 3 iterations and 24 subsets) with the point-spread-function modeled. The reconstruction process compensated for random, scatter, attenuation, normalization, and dead time effects. The reconstructed image had a total of 192 × 192 × 672 voxels. In the transaxial plane, the reconstructed image had a slice thickness of 2.89 mm and a voxel size of 3.125 mm. Visual examinations were performed by an experienced operator to guarantee the enrolled subjects free of visible body motion artifacts. Using frame-by-frame movies, the status of patient movement was inspected in axial, coronal, and sagittal view planes. The exclusion criteria were the positional displacement cannot exceed 3 voxels for several consecutive frames. 3D ROIs were drawn in both normal tissues (cerebral cortex, muscle) and tumor lesions for ROI-based quantifications. The time-activity curve (TAC) at each ROI was created by averaging the activity uptake of all voxels over time and was then used for the quantifications described below. Pixel-by-pixel, whole-body parametric images were also created.

Table 1 Demographical information (mean ± s.d.)

Proposed DTW protocol and data processing

The DTW protocol, as shown in Fig. 1, consists of two short dynamic acquisitions: an early acquisition performed after injection for 10 min and a late acquisition performed after a break with a fixed end time of 60 min. To determine the shortest possible scan time, various late scan durations (5, 10, and 20 min) were tested. DTW (10 + 5 min) refers to the DTW protocol with a 10-min early scan and a 5-min late scan for convenience; similar notations were used for the other two late scan durations. The protocol was tested by discarding the corresponding data in the middle of a 60-min dynamic scan. As a result, the first step in data processing was to fill in the gaps. To estimate the data in the interval, nonlinear fittings to the 3rd degree rational function were used, and the function was as follows:

$$C_{E} \left( t \right) = \frac{{p_{1} + p_{3} \times t + p_{5} \times t^{2} + p_{7} \times t^{3} }}{{p_{2} + p_{4} \times t + p_{6} \times t^{2} + p_{8} \times t^{3} }}$$
(1)

where CE(t) is the fitted tissue activity concentration over time and p1p8 are the parameters that affect the shape of the TAC. (p1 was fixed to 0, p2 was fixed to 1, and others were restricted to be positive.) The fitted and the original measured data were then combined to complete the 60-min TACs at each ROI.

Fig. 1
figure 1

Scan protocols of different quantification methods used in this study

Due to the lack of imaging data in the scan interval, the input function was derived using an approach as follows that is similar to the hybrid approach described in the Introduction. From the reconstructed images, IDIF was extracted from the ascending aorta for each subject, where cardiac and respiratory motion was less prominent than at the left ventricle; hence, the partial volume effect (PVE) was less. A 10-mm-diameter ROI was drawn on 6 consecutive slices during early time frames (0–30 s). The hybrid IF was combined with the IDIF from the scan period and scaled with PB-IDIF for the missing part. The PB-IDIF was obtained by averaging the normalized IDIFs of 15 subjects (10 males, 5 females, \(53.5 \pm 12.7\) years, \(65.7 \pm 9\).2 kg, \(249.5 \pm 42.4\) MBq) from the 60-min full dynamic scans. The equation to generate a hybrid IF was as follows:

$$C_{P} \left( t \right) = \left\{ {\begin{array}{*{20}c} {C_{{{\text{image}}}} \left( t \right),} & {\left( {0 \le t < t_{1} } \right)} \\ {\mu e^{{ - \gamma \left( {t - t_{1} } \right)}} C_{p0} \left( t \right), } & { \left( {t_{1} \le t < t_{2} } \right)} \\ {C_{{{\text{image}}}} \left( t \right),} & {\left( {t_{2} \le t \le 60 \min } \right)} \\ \end{array} } \right.$$
(2)

where Cimage(t) is from the measured data, Cp0(t) is the PB-IDIF, t1 is the end time of the early scan, t2 is the start time of the late scan, and \(\mathrm{and }\mu\) and \(\gamma\) are the scaling factors. Blood-to-plasma correction was not performed in this study, i.e., Cp(t) used the whole blood TAC derived from the images. Delay correction was performed for each ROI as in [42]: The input function was shifted from − 60 to + 60 s, and a 2T3k model was fitted to each regional TAC with the shifted input function. For each region, the delay time that provides the best fitting was thus selected. The dispersion was not considered in the present study. Kinetic modeling was then performed for the complete TAC and input function at each ROI with the 2T3k model. In this model, the dynamic activity change was described by a set of linear ordinary differential equations:

$$\frac{{{\text{d}}C_{1} \left( t \right)}}{{{\text{d}}t}} = K_{1} C_{p} \left( t \right) - \left( {k_{2} + k_{3} } \right)C_{1} \left( t \right) + k_{4} C_{2} \left( t \right)$$
$$\frac{{{\text{d}}C_{2} \left( t \right)}}{{{\text{d}}t}} = k_{3} C_{1} \left( t \right) - k_{4} C_{2} \left( t \right)$$
(3)

where Cp(t), C1(t), and C2(t) correspond to activity concentration in plasma, free FDG, and phosphorylated FDG, respectively, and K1, k2, k3, and k4 denote the transform rates between the compartments. Because the phosphorylation of FDG is considered an irreversible reaction in most tissues, the above-mentioned model was simplified by assuming k4 = 0. The intensity value of each voxel in a PET image represented the sum of activity concentrations in all compartments:

$$C_{{{\text{PET}}}} \left( t \right) = V_{b} C_{b} \left( t \right) + \left( {1 - V_{b} } \right)\left[ {C_{1} \left( t \right) + C_{2} \left( t \right)} \right]$$
(4)

where CPET(t) is the measured tissue activity concentration, \(C_{b} \left( t \right)\) is the activity concentration in blood, and Vb is the vascular volume fraction. By taking Eq. 3 to Eq. 4, kinetic parameters were estimated by minimizing the objective function:

$$\chi^{2} = \mathop \sum \limits_{i = 1}^{N} \omega_{i} \left[ {C_{{{\text{PET}}}} \left( {t_{i} } \right) - C_{t} \left( {t_{i} ,\overline{p}} \right)} \right]^{2}$$
(5)

where Ct is the simulated activity concentration by applying the estimated kinetic parameters into Eq. (4), \(\overline{p}\) represents K1, k2, k3, and Vb; i is the frame index; and \({\omega }_{i}\) is the weight at each frame.

Apart from the ROI-based analysis, we also performed a voxel-based analysis aiming to generate parametric images for each simplified dynamic scan. Given the large number of voxels in a whole-body image, the conventional nonlinear problem in Eq. 3 was reformed into a linearized problem [43, 44], as shown in the following equation:

$$C_{T} (t) = P_{1} C_{P} (t) + P_{2} \int_{0}^{t} {C_{P} (\tau )} {\text{d}}\tau + P_{3} \int_{0}^{t} {\int_{0}^{\tau } {C_{P} (s){\text{d}}s{\text{d}}\tau } } + P_{4} \int_{0}^{t} {C_{T} (\tau )} {\text{d}}\tau + P_{5} \int_{0}^{t} {\int_{0}^{\tau } {C_{T} (s){\text{d}}s{\text{d}}\tau } }$$
(6)

where P1, P2, P3, P4, and P5 are the functions of kinetic parameters to be estimated. Lawson–Hanson NNLS (nonnegative least squares) algorithm was then applied to solve the above equation. This will accelerate the estimation dramatically, thereby allowing the generation of whole-body multiparametric images within 1 min in our experience. After obtaining K1, k2, and k3, one can compute the macro-kinetic parameter (image) Ki:

$$K_{i} = \frac{{K_{1} \times k_{3} }}{{k_{2} + k_{3} }}$$
(7)

which represents the net FDG influx rate and is a surrogate to the glucose influx rate. The unit for Ki and K1 is ml/g/min, while for k2 and k3 is min−1. In this study, Lawson–Hanson NNLS was used to generate multiparametric (K1, k2, k3, Vb, Ki) images using full dynamic data, and Patlak Ki images using 30-min data were generated for comparison.

Other simplified quantification methods

Standardized uptake value (SUV): SUV was calculated using the following equation:

$${\text{SUV}} = \frac{{C_{{{\text{PET}}}} \left( T \right)}}{A/W}$$
(8)

where CPET(T) is the mean activity concentration in 50–60 min, and A and W are the total injected activity (Bq/cc) and body weight (kg) of the subject, respectively.

Fractional uptake ratio (FUR): FUR is defined as in [9, 45]:

$${\text{FUR}} = \frac{{C_{{{\text{PET}}}} \left( T \right)}}{{\mathop \smallint \nolimits_{0}^{T} C_{p} \left( t \right){\text{d}}t}}$$
(9)

where CPET(T) is normalized by the integral of plasma activity concentration from the start to time T. Like SUV, FUR can be calculated from a single static PET scan (50–60 min in this study). However, the input function is still required to calculate the integral in the denominator. Similar to the proposed protocol, a hybrid input function was used here with a different method of scaling:

$$C_{p} \left( t \right) = \left\{ {\begin{array}{*{20}c} {\alpha \cdot C_{p0} \left( t \right), \left( {0 \le t < 50 \min } \right)} \\ {C_{{{\text{PET}}}} \left( t \right), \left( {50 \le t \le 60 \min } \right)} \\ \end{array} } \right.$$
(10)

where the area under the curve (AUC) in 50–60 min was used to scale PB-IDIF Cp0(t) with a factor α to obtain the full input function.

Patlak graphical analysis: Patlak analysis [18, 19] can be expressed as a linear regression process:

$$\frac{{C_{{{\text{PET}}}} \left( T \right)}}{{C_{p} \left( T \right)}} = K_{i} \times \frac{{\mathop \smallint \nolimits_{0}^{T} C_{p} \left( t \right){\text{d}}t}}{{C_{p} \left( T \right)}} + {\text{int}} \left( {t \ge t^{*} } \right)$$
(11)

where CPET(T) and Cp(T) are the activity concentration in tissue and plasma at time T(0–60 min), respectively; Ki denotes the net FDG influx rate; int is the y-axis intercept of the regression plot; and \({t}^{*}\) is the time point (30 min in this study) when an equilibrium between blood and tissue is reached. The input function for Patlak analysis was obtained in the same manner as that for FUR.

Statistical analysis

All statistical analyses were performed using Statistical and Machine Learning Toolbox in MATLAB R2018b. Twelve subjects (5 males, 7 females, 57.8 ± 9.1 years, 62.5 ± 12.3 kg, injection dose \(230.3 \pm 47.6\) MBq) were randomly chosen from the subjects listed in Table 1 to evaluate the comparability of the hybrid IFs to IDIFs with correlation analysis. To verify the accuracy of the complete TAC, the mean absolute percentage error (MAPE) of all time points in the interval was calculated for each subject:

$${\text{MAPE}}\_{\text{TAC}}\left( \% \right) = \frac{{\mathop \sum \nolimits_{i} \left| {C_{E} \left( {t_{i} } \right) - C_{M} \left( {t_{i} } \right)} \right|/C_{M} \left( t \right)}}{n} \times 100\%$$
(12)

where i is the index of the frame in the interval; ti is the sample time point; n is the total number of sample time points; and CE and CM are the estimation and true measured activity concentrations, respectively. The Bland–Altman distribution of the MAPE in TAC estimation for each ROI was plotted. The parameters computed by 2T3k from the entire 60-min scan were used as the reference for the ROI-based quantification. For each ROI, the kinetic parameters (K1, k2, k3, Vb, and Ki) were calculated and compared to the reference. Correlation determination and bias analysis were carried out for various scan duration configurations. The percentage bias of Ki and K1 from the DTW protocol was calculated as:

$${\text{Bias}}\_K\left( \% \right) = \frac{{K_{{{\text{ref}}}} - K_{{{\text{DTW}}}} }}{{K_{{{\text{ref}}}} }} \times 100\%$$
(13)

where \(K_{{{\text{ref}}}}\) and \(K_{{{\text{DTW}}}}\) denote a specific parameter from the standard full scan protocol and the DTW protocol, respectively. Correlation analysis was performed for each simplified quantification method to the true quantification from the full scan, where P < 0.05 was considered as a significant correlation.

Results

Hybrid IFs and TAC completion

The IFs for both DTW and FUR/Patlak analysis were generated according to the steps described in Methods section. Correlation of AUCs between the hybrid IFs and the reference IDIFs from the full scan was determined. As shown in Fig. 2, relatively high correlations were observed for both DTW (R2 = 0.995, y = 1.04*x-0.0271) and FUR/Patlak (R2=0.931, y = 1.01*x-0.0193).

Fig. 2
figure 2

Example of hybrid IFs used for DTW (a) and Patlak/FUR quantifications (c). Their corresponding correlations to the true IDIFs are shown in (b) and (d), respectively

The estimated TACs in ROIs were compared to the reference measured ones from a full scan. Figure 3d shows the distribution of the MAPE for each ROI. Even for the protocol with the shortest scan duration (a 10-min early scan and a 5-min late scan), the TACs after completion were generally close to the reference ones. It is true, however, that estimation accuracy decreases as scan time decreases.

Fig. 3
figure 3

For an example subject, nonlinear fitting of the missing data was performed using the 3rd degree rational function. ac correspond to the TACs in the cerebral cortex, muscle, and tumor, respectively; d Bland–Altman plots of the MAPE (mean absolute percentage error, Eq. 12) in TAC estimation for each ROI with different scan durations

ROI-based quantification for the DTW protocol

The correlation results of Ki are shown in Fig. 4. In general, as the late scan duration decreased, the correlation also decreased. Nonetheless, even for the 5-min late scan, a relatively high correlation was observed, with correlation coefficients and linear regression equation being (0.971, y = 1.04*x − 0.0013), (0.990, y = 0.990*x + 0.0001), and (0.990, y = x − 0.0002) for the cerebral cortex, muscle, and tumor lesion, respectively. Figure 5a depicts the corresponding percentage bias distributions in Ki. The bias varied across the ROIs, with the cerebral cortex, muscle, and tumor lesion being − 0.1 ± 3.2%, − 3.3 ± 6.7%, 1.2 ± 2.5%, respectively. As shown in Additional file 1: Fig. S1, the correlation coefficients and linear regression equations of K1 derived from DTW (10 + 5 min) comparing with the reference were (0.820, y = 1.03*x + 0.0006), (0.940, y = 0.890*x + 0.0028), and (0.975, y = 0.98*x + 0.0018) for the cerebral cortex, muscle, and tumor lesion, respectively. Figure 5b depicts the bias analysis for K1. The bias of K1 in estimation varied across ROIs, with the cerebral cortex, muscle, and tumor being 5.2 ± 8.6%, 2.6 ± 9.1%, 1.7 ± 7.9%. In all regions, the bias in K1 was greater than that in Ki. In Additional file 1: Figs. S2 and S3 show the correlation analysis for k3 and Vb.

Fig. 4
figure 4

Correlation analysis for Ki derived from the DTW protocol with different scan durations. The ROIs were sampled in the cerebral cortex, muscle, and tumor lesion. The correlation plot for K1 is shown in Additional file 1: Fig. S1. The associated Bland–Altman plots are shown in Additional file 1: Fig. S5

Fig. 5
figure 5

Distribution of the percentage bias in Ki (a) and K1 (b) derived from the DTW protocol

Comparison among different simplified quantification methods

The proposed protocol's quantification was compared to three existing FDG quantification methods, namely SUV, FUR, and Patlak analysis. The Ki estimated from a full dynamic scan was used as a reference. Table 2 displays Ki correlations. The scattering plots for these correlations are shown in Additional file 1: Fig. S4. The DTW protocol (10 + 5 min) had the highest quantification accuracy relative to the reference Ki of all the simplified quantification methods, followed by FUR and Patlak analysis. In all three sampled regions, SUV had the lowest correlation.

Table 2 Correlation analysis for Ki and its surrogate parameters from different quantification methods; the squared correlation coefficient (R2), p value(p), and linear regression equations with each quantification method were listed in the table and the results of strongest correlation was marked in bold

The results for the visual appearance of the parametric images were overall consistent with the ROI-based quantifications. Figure 6a shows the example Ki images generated by the reference 60-min full scan, the DTW protocols, and Patlak analysis; Fig. 6b shows the difference images between the Ki derived from DTW protocols or Patlak analysis and the reference Ki images. It can be seen that the bias of Ki increases as the duration of the DTW protocol decreases; however, even with a 15-min total scan time, the Ki image is still more consistent with the reference Ki image than the one generated by Patlak analysis. The coefficient of variation in sampled muscle region (red box in Fig. 6a) was used to quantify the noise level of Ki images. For the reference Ki images, the mean and standard deviation of the coefficient of variation are 0.21 ± 0.003; for Ki images derived from DTW protocols, the values are 0.24 ± 0.005 (10 + 20 min), 0.28 ± 0.005 (10 + 10 min), 0.31 ± 0.007 (10 + 5 min), respectively; and for Patlak Ki images, the value is 0.83 ± 0.18, which is higher than others and indicated an overall higher level of noise. K1 images derived from the DTW protocols were almost visually identical to the reference, as shown in Fig. 6c.

Fig. 6
figure 6

Whole-body parametric images from a dynamic scan: (a) Ki images generated from the 60-min full scan, the DTW protocols, and Patlak analysis; the muscle region inside the red box was sampled to quantify the image noise; (b) difference images between the ones and the reference Ki image from 60-min full scan; (c) K1 images generated from the 60-min full scan and the DTW protocols;

Discussion

The DTW FDG scan protocol was implemented on a state-of-the-art uEXPLORER scanner in this study, and multiple kinetic parameters were quantified. Whole-body imaging allows for a simultaneous scan of the entire human body with unprecedented sensitivity, as well as the extraction of IDIF directly from large arteries such as the aorta. Multiple kinetic parameters can be obtained by using the processed TACs and the input function at the designated organs. The results showed that with a 10-min early acquisition and a 5-min late acquisition, relatively accurate quantifications of Ki, k1, k3, and Vb could be obtained. The estimation bias could be reduced with a longer scan duration. Furthermore, we compared the quantification of DTW to several existing simplified quantification methods, including Patlak analysis, FUR, and SUV, as a surrogate for the net influx rate Ki. The proposed DTW protocol produced the most accurate quantification, followed by FUR and Patlak analysis, while SUV had the lowest correlation, which is consistent with previous study findings [20, 26, 46]. In terms of visual performance, the DTW protocol generated parametric Ki and K1 images that were more consistent with the reference and had less noise than the Patlak analysis had.

To reduce the total scan time while maintaining an acceptable quantification accuracy, nonlinear estimation was applied to the shortened dynamic scan data. DTW protocol allows for the efficient use of hospital facilities while maintaining quantification accuracy. For example, interleaved scans become possible, during which the first scan of the second patient can be acquired, while the first patient rests outside the scanning room. The potential inaccuracy of the estimated IFs and TACs is a disadvantage of this method. However, it was found that combining the scaled PB-IDIF with the existing part IDIF from the dual-phase scan can provide accurate IF. TACs estimated using nonlinear fitting with a rational function were also close to the true ones. Inferring parameters from reduced scan data did introduce bias in Ki and K1 estimation. Using the tumor lesion as an example, Ki estimation was quite accurate with a relative bias of less than 5%, while K1 estimation had a relative bias of less than 20%. One possible explanation is that the size of some tumors is small and hence TACs are quite unstable compared to uniform regions such as a muscle. This is especially true for the early stage of the scan, where estimation of K1 depends mostly on. Another possibility is that some tumors had not reached true equilibrium even after 60 min of scanning. Estimation based on 2T3k may not be optimal in such cases. Regarding the nonlinear fitting to complete the missing data in the interval, it took 1 min to run over a single dynamic scan.

There are some limitations to the current study: (1) A direct comparison between the hybrid input function and the arterial sampled input function is missing. (2) Although the primary goal of this work is to compare against DTW to existing simplified measures, it may be an unfair comparison since they have different amount of data available for kinetic modeling. For instance, Patlak, FUR, SUV all use single scan. DTW protocol with more scan information in theory can achieve more accurate estimation in kinetic parameters. In the future, it may be necessary to compare the current DTW protocol with other published dual-time-window methods, such as dual-time-point Patlak analysis [29] and retention index [46]. (3) A 60-min scan may not be long enough to capture the kinetics in some organs, such as the brain and some malignant tumors, to reach the equilibrium state. To avoid the potential bias in Ki and K1 quantification caused by this factor, a longer scan time may be required. (4) The dispersion effect for the input function was not modeled by our method. This could have resulted in quantitative inaccuracies in head and neck organ measurements [47]. Furthermore, scatter corrections may be insufficiently accurate due to the very concentrated and rapidly changing tracer distribution in the first few seconds after injection. (5) There could be a misalignment in patient position between the early and late phases of the DTW protocol’s actual implementation that affects the accuracy of the kinetic parameter estimation. It may be necessary to align and perform new reconstructions with the re-aligned attenuation image. To avoid two attenuation CT scans, one option is to use the TOF information (e.g., MLAA [48]) or deep learning techniques [49]) to generate the pseudo-attenuation map for the second scan. Then, one can register the first CT image to this attenuation map and hence apply the transformation to the corresponding PET data. Low-dose CT could be another solution to reduce the total amount of radiation exposure, which will require dedicated denoising techniques such as deep learning. (6) The current DTW protocol was implemented on a total-body scanner, which is available in few institutions for now. In theory, it can also be applied on a conventional scanner with short FOV. In such case, full kinetic analysis can only be performed near the chest region since conventional scanner need to capture the input function from the early dynamics of the descending aorta. Therefore, at least the micro-parameters cannot be obtained for non-chest regions. Another potential issue when applying DTW protocol on a conventional scanner would be the relatively lower sensitivity compared with the uEXPLORER which could hamper the estimation from the partial scan data, especially when assessing voxelized parametric images. (7) The number of patients included in the study is limited, given the difficult in acquiring a dynamic scan without patient motion affected. A dynamic total-body scan may be vulnerable to patient movement, especially when considering the increased sensitivity. Motion could potentially affect not only the visual quality but also the accurate quantification across the entire body. In this work, we only performed the simple visual inspection to exclude the datasets with obvious motion artifacts. Therefore, an even more thorough quantitative quality control on the effect of motion may be required other than simply checking visually the images frame by frame. In case when motion is severe, motion compensation [50,51,52] will be essential before applying the data analysis to a given dynamic scan.

The study’s primary goal was to investigate and improve the feasibility of using the total-body dynamic imaging protocol in clinics. Although a total-body scanner greatly aids in this goal, we believe that not all applications are suitable for dynamic imaging, even with proper scan time reduction. When weighing the benefits of increased absolute quantification accuracy versus the additional effort required, some existing alternatives to SUV may be preferable. As demonstrated by the findings, when the clinical task is to quantify the regional metabolic rate for a known lesion position or organs of interest, FUR with the hybrid IF is the most feasible protocol because it requires a regular scan time with an acceptable bias in Ki estimation. However, when detecting an unknown lesion that necessitates a reliable visual assessment or quantifying micro-kinetic parameters such as K1, the DTW protocol may be preferable. It would be also interesting to investigate which simplified quantification should be used for applications such as assessing the treatment response. More research for specific applications should be done to minimize the trade-off between accuracy achieved and the extra effort required for dynamic imaging.

Conclusion

In this study, we evaluated a DTW protocol for FDG quantification and compared its accuracy to that of existing simplified quantification methods, e.g., Patlak analysis, FUR, and SUV. The results showed that by using the DTW protocol, the dynamic total-body FDG scan time could be reduced to 15 min. It is possible to achieve accurate Ki and K1 quantification as well as acceptable visual quality in parametric images. We anticipate seeing benefits from applying a similar concept to dynamic imaging with other radiotracers, such as 68 Ga-PSMA. Although scanners such as uEXPLORER can help to shorten the scan time, more time and effort are still required when compared to traditional SUV static imaging. As a result, other existing simplified quantifications, such as FUR or SUR, may be more appropriate at least in certain applications.

Availability of data and materials

The data and material will be available upon reasonable request.

Abbreviations

PET:

Positron emission tomography

CT:

Computerized tomography

FDG:

Fluorodeoxyglucose

SUV:

Standardized uptake value

FUR:

Fractional uptake ratio

TOF:

Time of flight

OSEM:

Ordered subset expectation–maximization

2T3k:

Irreversible 2-tissue compartment model

DTW:

Dual time window

IDIF:

Image-derived input function

PBIF:

Population-based input function

PB-IDIF:

Population-based image-derived input function

TAC:

Time activity curve

ROI:

Region of interest

PVE:

Partial volume effect

References

  1. Krause BJ, Schwarzenböck S, Souvatzoglou M. FDG PET and PET/CT. In: Schober O, Riemann B, editors. Molecular imaging in oncology. Berlin: Springer Berlin Heidelberg; 2013. p. 351–69. https://doi.org/10.1007/978-3-642-10853-2_12.

    Chapter  Google Scholar 

  2. Townsend DW, et al. PET/CT today and tomorrow. J Nucl Med. 2004;45(1 suppl):4S-14S.

    PubMed  Google Scholar 

  3. Tahari AK, et al. Optimum lean body formulation for correction of standardized uptake value in PET imaging. J Nucl Med. 2014;55(9):1481–4.

    Article  PubMed  Google Scholar 

  4. Kim CK, et al. Standardized uptake values of FDG: body surface area correction is preferable to body weight correction. J Nucl Med. 1994;35(1):164–7.

    CAS  PubMed  Google Scholar 

  5. Hamberg LM, et al. The dose uptake ratio as an index of glucose metabolism: useful parameter or oversimplification? J Nucl Med Off Publ Soc Nucl Med. 1994;35(8):1308–12.

    CAS  Google Scholar 

  6. Huang S-C. Anatomy of SUV. Nucl Med Biol. 2000;27(7):643–6.

    Article  CAS  PubMed  Google Scholar 

  7. Keyes JW. SUV: standard uptake or silly useless value? J Nucl Med. 1995;36(10):1836–9.

    PubMed  Google Scholar 

  8. van den Hoff J, et al. The PET-derived tumor-to-blood standard uptake ratio (SUR) is superior to tumor SUV as a surrogate parameter of the metabolic rate of FDG. EJNMMI Res. 2013;3(1):1–8.

    Article  CAS  Google Scholar 

  9. Rutland M, Que L, Hassan IM. “FUR”–one size suits all. Eur J Nucl Med. 2000;27(11):1708–13.

    Article  CAS  PubMed  Google Scholar 

  10. Parghane RV, Basu S. Dual–time point 18F-FDG-PET and PET/CT for differentiating benign from malignant musculoskeletal lesions: opportunities and limitations. In: Seminars in nuclear medicine. Elsevier; 2013

  11. Dunnwald LK, et al. PET tumor metabolism in locally advanced breast cancer patients undergoing neoadjuvant chemotherapy: value of static versus kinetic measures of fluorodeoxyglucose uptake. Clin Cancer Res. 2011;17(8):2400–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Fahrni G, et al. Does whole-body Patlak 18 F-FDG PET imaging improve lesion detectability in clinical oncology? Eur Radiol. 2019;29(9):4812–21.

    Article  PubMed  Google Scholar 

  13. Huang Y-E, et al. Solitary pulmonary nodules differentiated by dynamic F-18 FDG PET in a region with high prevalence of granulomatous disease. J Radiat Res. 2012;53(2):306–12.

    Article  CAS  PubMed  Google Scholar 

  14. Wangerin KA, et al. A virtual clinical trial comparing static versus dynamic PET imaging in measuring response to breast cancer therapy. Phys Med Biol. 2017;62(9):3639.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Ye Q, et al. Improved discrimination between benign and malignant LDCT screening-detected lung nodules with dynamic over static 18F-FDG PET as a function of injected dose. Phys Med Biol. 2018;63(17): 175015.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  16. Zaidi H, Karakatsanis N. Towards enhanced PET quantification in clinical oncology. Br J Radiol. 2017;91(1081):20170508.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Carson RE. Tracer kinetic modeling in PET. In: Bailey DL, Townsend DW, Valk PE, Maisey MN, editors. Positron emission tomography. London: Springer-Verlag; 2005. p. 127–59. https://doi.org/10.1007/1-84628-007-9_6.

    Chapter  Google Scholar 

  18. Patlak CS, Blasberg RG. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. Generalizations. J Cereb Blood Flow Metab. 1985;5(4):584–90.

    Article  CAS  PubMed  Google Scholar 

  19. Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood Flow Metab. 1983;3(1):1–7.

    Article  CAS  PubMed  Google Scholar 

  20. van Sluis J, et al. Use of population input functions for reduced scan duration whole-body Patlak 18 F-FDG PET imaging. EJNMMI Phys. 2021;8(1):1–8.

    Article  Google Scholar 

  21. Zanotti-Fregonara P, et al. Image-derived input function for brain PET studies: many challenges and few opportunities. J Cereb Blood Flow Metab. 2011;31(10):1986–98.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Zanotti-Fregonara P, et al. Comparison of eight methods for the estimation of the image-derived input function in dynamic [18F]-FDG PET human brain studies. J Cereb Blood Flow Metab. 2009;29(11):1825–35.

    Article  PubMed  Google Scholar 

  23. Eberl S, et al. Evaluation of two population-based input functions for quantitative neurological FDG PET studies. Eur J Nucl Med. 1997;24(3):299–304.

    CAS  PubMed  Google Scholar 

  24. Takikawa S, et al. Noninvasive quantitative fluorodeoxyglucose PET studies with an estimated input function derived from a population-based arterial blood curve. Radiology. 1993;188(1):131–6.

    Article  CAS  PubMed  Google Scholar 

  25. Vriens D, et al. A curve-fitting approach to estimate the arterial plasma input function for the assessment of glucose metabolic rate and response to treatment. J Nucl Med. 2009;50(12):1933–9.

    Article  PubMed  Google Scholar 

  26. Yaping Wu, et al. Whole-body parametric imaging of FDG PET using uEXPLORER with reduced scan time. J Nucl Med. 2022;63(4):622–8.

    Article  Google Scholar 

  27. Hu J, et al. Design and implementation of automated clinical whole body parametric PET with continuous bed motion. IEEE Trans Radiat Plasma Med Sci. 2020;4(6):696–707.

    Article  Google Scholar 

  28. Hofheinz F, et al. Comparative evaluation of SUV, tumor-to-blood standard uptake ratio (SUR), and dual time point measurements for assessment of the metabolic uptake rate in FDG PET. EJNMMI Res. 2016;6(1):1–9.

    Article  CAS  Google Scholar 

  29. Wu J, et al. Generation of parametric Ki images for FDG PET using two 5-min scans. Med Phys. 2021;48(9):5219–31.

    Article  CAS  PubMed  Google Scholar 

  30. Zhu W, et al. Patlak image estimation from dual time-point list-mode PET data. IEEE Trans Med Imaging. 2014;33(4):913–24.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Sugawara Y, et al. Germ cell tumor: differentiation of viable tumor, mature teratoma, and necrotic tissue with FDG PET and kinetic modeling. Radiology. 1999;211(1):249–56.

    Article  CAS  PubMed  Google Scholar 

  32. Song S-L, et al. 18F-FDG PET/CT-related metabolic parameters and their value in early prediction of chemotherapy response in a VX2 tumor model. Nucl Med Biol. 2010;37(3):327–33.

    Article  CAS  PubMed  Google Scholar 

  33. van Berkel A, et al. Metabolic subtyping of pheochromocytoma and paraganglioma by 18F-FDG pharmacokinetics using dynamic PET/CT scanning. J Nucl Med. 2019;60(6):745–51.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  34. Wang G, et al. Total-body dynamic PET of metastatic cancer: first patient results. Soc Nucl Med. 2020

  35. Lodge MA, et al. Measurement of PET quantitative bias in vivo. J Nucl Med. 2021;62(5):732–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  36. Zhang X, et al. Total-body dynamic reconstruction and parametric imaging on the uEXPLORER. J Nucl Med. 2020;61(2):285–91.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  37. Sari H, et al. First results on kinetic modelling and parametric imaging of dynamic 18F-FDG datasets from a long axial FOV PET scanner in oncological patients. Eur J Nucl Med Mol Imaging. 2022;49(6):1997–2009.

    Article  CAS  PubMed  Google Scholar 

  38. Badawi RD, et al. First human imaging studies with the EXPLORER total-body PET scanner. J Nucl Med. 2019;60(3):299–303.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. Liu G, et al. Short-time total-body dynamic PET imaging performance in quantifying the kinetic metrics of 18F-FDG in healthy volunteers. Eur J Nucl Med Mol Imaging. 2022;49(8):2493–503.

  40. Zhang Y-Q, et al. The image quality, lesion detectability, and acquisition time of 18F-FDG total-body PET/CT in oncological patients. Eur J Nucl Med Mol Imaging. 2020;47(11):2507–15.

    Article  PubMed  CAS  Google Scholar 

  41. Feng T, et al. Total-body quantitative parametric imaging of early kinetics of (18)F-FDG. J Nucl Med. 2021;62(5):738–44.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. Meyer E. Simultaneous correction for tracer arrival delay and dispersion in CBF measurements by the H215O autoradiographic method and dynamic PET. J Nucl Med. 1989;30(6):1069–78.

    CAS  PubMed  Google Scholar 

  43. Blomqvist G. On the construction of functional maps in positron emission tomography. J Cereb Blood Flow Metab. 1984;4(4):629–32.

    Article  CAS  PubMed  Google Scholar 

  44. Cai W, et al. Generalized linear least squares algorithms for modeling glucose metabolism in the human brain with corrections for vascular effects. Comput Methods Progr Biomed. 2002;68(1):1–14.

    Article  Google Scholar 

  45. Ishizu K, et al. Effects of hyperglycemia on FDG uptake in human brain and glioma. J Nucl Med. 1994;35(7):1104–9.

    CAS  PubMed  Google Scholar 

  46. Prando S, et al. Comparison of different quantification methods for 18F-fluorodeoxyglucose-positron emission tomography studies in rat brains. Clinics (Sao Paulo). 2019;74: e1273.

    Article  Google Scholar 

  47. Chen C-J, et al. Dual-phase 18F-FDG PET in the diagnosis of pulmonary nodules with an initial standard uptake value less than 2.5. Am J Roentgenol. 2008;191(2):475–9.

    Article  Google Scholar 

  48. Rezaei A, et al. Simultaneous reconstruction of activity and attenuation in time-of-flight PET. IEEE Trans Med Imaging. 2012;31(12):2224–33.

    Article  PubMed  Google Scholar 

  49. Wang T, et al. A review on medical imaging synthesis using deep learning and its clinical applications. J Appl Clin Med Phys. 2021;22(1):11–36.

    Article  PubMed  Google Scholar 

  50. Sun T, et al. Body motion detection and correction in cardiac PET: phantom and human studies. Med Phys. 2019;46(11):4898–906.

    Article  PubMed  Google Scholar 

  51. Kyme AZ, Fulton RR. Motion estimation and correction in SPECT, PET and CT. Phys Med Biol. 2021;66(18):18TR02.

    Article  Google Scholar 

  52. Sun T, et al. Motion correction and its impact on quantification in dynamic total-body 18F-Fluorodeoxyglucose PET. EJNMMI Physics. 2022; accepted.

Download references

Acknowledgements

Not applicable.

Funding

This work is supported by the Scientific Instrument Innovation Team of the Chinese Academy of Sciences (GJJSTD20180002) and the Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province (2020B1212060051).

Author information

Authors and Affiliations

Authors

Contributions

TS, YW, YY, and MW contributed to conception and design. TS, ZH, YY, DL, XL, and HZ provided administrative support. TS, YW, XL, and MW were involved in provision of study materials or patients. ZW, TS, YW, XL, YB, JD, and CS contributed to collection and assembly of data. TS, HC, YW, FG, and YZ were involved in data analysis and interpretation. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Meiyun Wang or Tao Sun.

Ethics declarations

Ethics approval and consent to participate

This study was performed in line with the principles of the Declaration of Helsinki. All scans were approved by the Ethics Committee of Henan Provincial People’s Hospital and the People’s Hospital of Zhengzhou (IRB2020123). Informed consent was obtained from all individual participants in the study.

Consent for publication

All authors agree to publish the paper.

Competing interests

The authors have no conflict of interests to declare.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

 Fig. S1-S3: Correlation analysis between micro-parameters (K1k3Vb) derived from the proposed DTW protocol and  reference values. Fig. S4: Correlation analysis between all simplified quantification methods - DTW (10+5 min), Patlak Ki, FUR and SUV to the reference Ki in all ROIs. Fig. S5: Bland-Altman plots of the percentage bias in Ki and K1 derived from the DTW protocol. Fig. S6: Correlation analysis between Ki estimated by non-linear fitting and Lawson-Hanson NNLS fitting.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, Z., Wu, Y., Li, X. et al. Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body 18F-FDG PET imaging. EJNMMI Phys 9, 63 (2022). https://doi.org/10.1186/s40658-022-00492-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-022-00492-w

Keywords