Skip to main content
  • Short communication
  • Open access
  • Published:

Multicentre quantitative 68Ga PET/CT performance harmonisation

Abstract

Purpose

Performance standards for quantitative 18F-FDG PET/CT studies are provided by the EANM Research Ltd. (EARL) to enable comparability of quantitative PET in multicentre studies. Yet, such specifications are not available for 68Ga. Therefore, our aim was to evaluate 68Ga-PET/CT quantification variability in a multicentre setting.

Methods

A survey across Dutch hospitals was performed to evaluate differences in clinical 68Ga PET/CT study protocols. 68Ga and 18F phantom acquisitions were performed by 8 centres with 13 different PET/CT systems according to EARL protocol. The cylindrical phantom and NEMA image quality (IQ) phantom were used to assess image noise and to identify recovery coefficients (RCs) for quantitative analysis. Both phantoms were used to evaluate cross-calibration between the PET/CT system and local dose calibrator.

Results

The survey across Dutch hospitals showed a large variation in clinical 68Ga PET/CT acquisition and reconstruction protocols. 68Ga PET/CT image noise was below 10%. Cross-calibration was within 10% deviation, except for one system to overestimate 18F and two systems to underestimate the 68Ga activity concentration. RC-curves for 18F and 68Ga were within and on the lower limit of current EARL standards, respectively. After correction for local 68Ga/18F cross-calibration, mean 68Ga performance was 5% below mean EARL performance specifications.

Conclusions

68Ga PET/CT quantification performs on the lower limits of the current EARL RC standards for 18F. Correction for local 68Ga/18F cross-calibration mismatch is advised, while maintaining the EARL reconstruction protocol thereby avoiding multiple EARL protocols.

Introduction

The use of 68Gallium (68Ga)-labelled peptides for PET imaging has increased in the past years with the market authorisation for 68Ga/68Ge-generators. The main applications include imaging of neuroendocrine tumours using somatostatin analogues and prostate cancer imaging using the prostate-specific membrane antigen [1, 2]. Though the interpretation of 68Ga-PET/CT is mainly based on visual assessment, quantitative measures should be used to evaluate or predict therapy response.

Previous experience with 18Fluorine (18F) expressed the need for standardisation of acquisition and reconstruction protocols in order to retrieve comparable quantitative imaging data. The EANM Research Ltd. (EARL) provides an accreditation programme to ensure PET/CT system harmonisation in multicentre 18F-FDG PET/CT studies [3]. This approach is based on standardizing the recovery coefficient (RC) for six phantom spheres with different sizes, thereby minimising inter- and intra-institute variability. For other isotopes, quantification should be evaluated separately as isotope characteristics can result in different image quality and quantification accuracy. For example, Makris et al. studied 89Zirconium (89Zr) PET and showed the need for a specific harmonisation step including post-reconstruction smoothing to enable comparable quantitative measures among PET/CT systems [4]. In contrast, a recent 18F performance study showed that post-reconstruction filtering is not required for state-of-the-art PET/CT systems in relation to this isotope [5]. However, for 68Ga, such studies are not yet available.

In general, PET quantification accuracy depends on reconstructions, noise, and spatial resolution [6]. For 68Ga, the lower positron yield (89%), long positron range due to high initial positron energy (max 1.90 MeV, mean 0.84 MeV), short physical half-life (68 min) and small prompt gamma branching (3.2%, 1.077 MeV) may result in an inferior image quality compared to 18F [7]. Therefore, the aim of this study was to assess 68Ga-PET/CT quantification accuracy and reproducibility in a multicentre setting based on EARL standards.

Materials and methods

Clinical protocol evaluation

A survey among eight Dutch hospitals was performed to evaluate factors that affect quantification and to assess variability in clinical 68Ga-PET/CT acquisition protocols. Questions focussed on administered activity, PET/CT system, and acquisition- and reconstruction settings.

18F and 68Ga PET/CT phantom acquisitions

Eight European hospitals with 13 PET/CT systems performed phantom acquisitions, of which 11 systems were EARL accredited, but all had recoveries within the published EARL specifications. Six Biograph mCT systems (Siemens Healthineers, Erlangen, Germany), three Discovery systems (GE Healthcare, Milwaukee, WI, USA) and four Philips systems (Philips Healthcare, Eindhoven, The Netherlands) were included.

18F and 68Ga acquisitions were performed at the end of 2017 and beginning of 2018 with two phantoms which were prepared using a standardised procedure by experienced staff from each centre. First, the NEMA PET cylindrical phantom was filled with 6ā€“13 kBq/ml of 18F and 68Ga. Second, the NEMA NU-2 Image Quality (IQ) phantom was imaged using a 1:10 ratio with 2.0 and 20.0 kBq/ml of 18F and 68Ga in background compartment and spheres (37, 28, 21, 17, 13, and 10 mm diameter), respectively. Acquisitions of both phantoms were performed with minimal two bed positions and at least 5 min per bed position. Images were reconstructed according to local settings, including corrections for decay, randoms, dead time, CT-based attenuation, and scatter.

Data analysis

Image noise was characterized for 68Ga only using the coefficient of variation (CoV) along a 30 Ɨ 30 Ɨ 160 mm bar in the centre of the cylindrical phantom. Image quality was based on the RC of all six spheres, analysed by the EARL semi-automatic tool [5, 8]. The RCmax, RCpeak and RCmean were determined as a function of sphere size based on the maximum voxel value (RCmax), the 1.0 cm3 volume with the maximised average value (RCpeak) and the mean value of 50% isocontour of the maximum voxel value (RCmean) with contrast correction, respectively. A spherical volume-of-interest (VOI) of ~ā€‰300 ml in the centre of the cylindrical phantom and ten VOIs in the background of the IQ phantom were used for local PET and dose calibrator cross-calibration. IQ phantom background volume was 9400 ml, unless specified otherwise by the institute.

Results

Eight Dutch hospitals provided their clinical acquisition- and reconstruction protocols (Table 1), which showed to be different.

Table 1 Acquisition and reconstruction settings of clinical 68Ga PET/CT imaging for prostate cancer and neuroendocrine tumours. One hospital per row is presented

An overview of all PET/CT systems and reconstruction settings is provided in Table 2. For local cross-calibration, most systems performed within 10% deviation of the dose calibrator (Fig. 1). The median [IQR] ratio was 0.93 [0.91ā€“0.98] and 0.99 [0.97ā€“1.01] for 68Ga and 18F, respectively. Two systems showed identical calibration accuracy for both isotopes (system 2 and 11), all other show a consistent underestimation for 68Ga. The 68Ga CoV in the centre of the cylindrical phantom was below 10% (Fig. 2).

Table 2 PET/CT reconstruction settings for phantom measurements
Fig. 1
figure 1

Accuracy of the measured activity by the PET/CT system and local dose calibrator, based on the average between the cylindrical and IQ phantom. Numbers correspond to Table 2

Fig. 2
figure 2

Noise across the cylindrical phantom filled with 68Ga, visualized as coefficient of variation (CoV)

The 18F RC-curves of all PET/CT systems satisfied the current EARL specifications (Fig. 3aā€“c). However, for 68Ga the RC-curves were located around the lower limit of the EARL specifications (Figure 3d-f). In addition, 68Ga showed a reduced mean recovery and larger variation between PET/CT systems compared to the 18F. The variation for all spheres of the RCmean, RCmax and RCpeak for 18F was 6%, 6% and 8%, respectively. For 68Ga, the mean range was 11%, 11% and 15% (largest variation was 19%). Furthermore, the mean RCmax and RCmean were both 11% lower compared to the mean EARL specifications for 18F. The mean 68Ga/18F calibration difference within one scanner was 7% (range 1ā€“13%).

Fig. 3
figure 3

RC for 18F with the current EARL standards and RC of 68Ga. Solid lines: maximum and minimum values according to EARL limits as applicable before 2019

After correction for the local difference between 68Ga/18F cross-calibration (Fig. 1), the 68Ga RC curve was within EARL limits for all but two scanners (Figure 4). The mean 68Ga RCmax and RCmean were accordingly 5% lower compared to mean EARL standards.

Fig. 4
figure 4

68Ga RC-curves corrected for the 18F/68Ga calibration mismatch according to local cross-calibration. Solid lines: maximum and minimum values according to EARL limits as applicable before 2019

Discussion

In this study, quantitative 68Ga PET/CT performance was evaluated in a multicentre setting. In a survey across Dutch hospitals, differences in clinical acquisition and reconstruction protocols were observed, underlining the need for clinical harmonisation. Although 11 out of the 13 PET/CT systems were EARL accredited, all systems showed 18F recovery performance within EARL standards. For this reason, all systems were included for 68Ga evaluation.

The absence of local and central dose calibrator cross-calibration for 68Ga is a limitation in this study. This would increase local calibrator harmonisation and improves PET/CT comparability across sites. Most institutes use a long-lived (137Ceasium) source to assess constancy and accuracy of the dose calibrator on a daily basis, and perform actual cross-calibration with the PET/CT system at least once a year using 18F. Still, in all but three PET/CT systems the measured 18F and 68Ga activity concentrations were within 10% deviation from the local dose calibrator. High energy prompt gammas emitted by 68Ga are likely detected by the dose calibrator causing a disconcordance, yet in fewer extent by the PET system. Because of this, the dose calibrator overestimates 68Ga-activity, and a persistent underestimation for 68Ga compared to 18F is seen in Fig. 1. A recent study by Bailey et al. also showed an underestimation of Ā±ā€‰15% for 68Ga, which was primarily related to an inaccurate scaling factor for the dose calibrator of a specific vendor [9]. To avoid these issues, they calibrated the dose calibrator towards the PET, after verifying that the scanner has a good response for 18F. These results are also supported by the fact that on specific Siemens scanners (scanners 1 and 2), a traceable 68Germanium (68Ge) source was used to verify absolute PET response independent of a dose calibrator. When imaging the 68Ge-source, the PET/CT system did not show the same offset as was observed when imaging the 68Ga cross-calibration phantom (roughly a deviation of <ā€‰1% vs. 6% and 7%, respectively). For the sake of simplicity, we would suggest to correct the RC curve for the local 68Ga/18F discrepancy, as after correction for this 68Ga/18F difference (Fig. 4) all but two scanners were within EARL specifications. This correction has to be performed offline in multicentre quantitative studies. The 68Ga used for this study was produced either locally or by a pharmaceutical institution and was therefore not traceable to a central dose calibrator. We expect that the response between the dose calibrator and the PET-system could be uniform in future clinical 68Ga-PET/CT studies if a traceable (NIST) source is used to harmonise protocols between centres.

68Ga image noise was below 10% for all PET/CT systems which is in concordance with the EANM/EARL guidelines [3, 8]. The RC variation is larger for 68Ga compared to 18F (Fig. 3). However, 68Ga performance nearly reached EARL performance specifications after correction for the local 68Ga/18F ratio. Surprisingly, the RCpeak variation (8% and 15%) is larger in contrast to RCmax and RCmean (both 6% and 11%) for both 18F and 68Ga, respectively. The study of Kaalep et al. showed the opposite result in RCpeak variation [5]. The RCpeak is expected to be less prone to noise compared to RCmax; therefore, it was expected to be more comparable over all PET-systems. The difference could be explained by the fact that the standard deviation of RCmax and RCpeak are similar: 8.4% and 8.6% for 68Ga and 4.8% and 5.0% for 18F, respectively. Yet, the mean RCpeak value is lower; therefore, resulting in a higher CoV. Next to that, the larger 68Ga variation in the RC-curves compared to 18F is likely related to the higher positron energy of 68Ga and thereby revealing a lower signal-to-noise ratio. This effect is enhanced by post-reconstruction filtering. Finally, previous single-centre studies show 68Ga RC-curves similar [10] or somewhat better due to point spread function reconstruction [11] as observed in the current study. The EARL limits as applicable before 2019 (EARL1) are shown in Figs. 3 and 4, as all acquisitions were acquired before 2019 and therefore site-specific acquisition and reconstruction protocols are designed to meet the EARL1 limits. RCpeak specifications are not available for EARL1 and are therefore not shown in Figs. 3 and 4. EARL2 limits (applicable from 2019) for RCmax and RCmean increased with ~ā€‰25% in comparison to EARL1. We expect that the gap between 18F and 68Ga recoveries will further increase with these new limits, as already for EARL1 not all scanners agreed to EARL1 limits after 68Ga/18F correction (Fig. 4).

Based on the results, we propose to correct 68Ga recovery towards the 18F recovery to correct for the current dose calibrator deviation. We suggest, therefore, to apply the EARL acquisition and reconstruction protocol and to correct for 68Ga/18F cross-calibration mismatch. One can assume that 68Ga recovery is steady if 18F specifications of a PET-system are stable during regular yearly assessment. Unless the acquisition and reconstruction protocol is changed or major maintenance is performed to the PET/CT-system, we recommend to perform additional 68Ga IQ acquisitions only when regular 18F evaluations are deviating. An EARL accreditation programme for 68Ga can thus be based on the 18F accreditation but extended with a cross-calibration verification between 68Ga measured by the dose calibrator and PET/CT system only, similarly as proposed by Kaalep et al. for 89Zr [12]. In addition, frequent 18F cross-calibration acquisitions using the cylindrical phantom are advised, especially after PET/CT system maintenance.

Conclusion

This evaluation of multicentre 68Ga PET/CT performance showed that 68Ga RCs perform at the lower limits of current 18F EARL standards. For practical reasons, we recommend to use the 18F EARL approved reconstruction settings and to correct for 68Ga/18F calibration mismatch based on local cross-calibration. Finally, we suggest to evaluate 68Ga PET/CT recovery performance once and repeat only when 18F specifications are changed.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

18F:

18Fluorine

68Ga:

68Gallium

89Zr:

89Zirconium

CoV:

Coefficient of variation

EARL:

EANM Research Ltd

IQ:

Image quality

RC:

Recovery coefficient

VOI:

Volume-of-interest

References

  1. Singh S, Poon R, Wong R, Metser U. 68Ga PET imaging in patients with neuroendocrine tumors: a systematic review and meta-analysis. Clin Nucl Med. 2018;43:802ā€“10.

    ArticleĀ  Google ScholarĀ 

  2. LĆ¼tje S, Heskamp S, Cornelissen AS, Poeppel TD, van den Broek SAMW, Rosenbaum-Krumme S, et al. PSMA ligands for radionuclide imaging and therapy of prostate cancer: clinical status. Theranostics. 2015;5:1388ā€“401.

    ArticleĀ  Google ScholarĀ 

  3. Boellaard R, Willemsen A, Arends B, Visser EP. EARL procedure for assessing PET/CT system specific patient FDG activity preparations for quantitative FDG PET/CT studies. 2013. p. 1ā€“3. Available from: http://earl.eanm.org/html/img/pool/EARL-procedure-for-optimizing-FDG-activity-for-quantitative-FDG-PET-studies_version_1_1.pdf.

  4. Makris NE, Boellaard R, Visser EP, de Jong JR, Vanderlinden B, Wierts R, et al. Multicenter Harmonization of 89Zr PET/CT Performance. J Nucl Med. 2014;55:264ā€“7.

    ArticleĀ  CASĀ  Google ScholarĀ 

  5. Kaalep A, Sera T, Rijnsdorp S, Yaqub M, Talsma A, Lodge MA, et al. Feasibility of state of the art PET/CT systems performance harmonisation. Eur J Nucl Med Mol Imaging. 2018;45:1344ā€“61.

    ArticleĀ  Google ScholarĀ 

  6. Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519ā€“27.

    PubMedĀ  Google ScholarĀ 

  7. Sanchez-Crespo A. Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography. Appl Radiat Isot. 2013;76:55ā€“62.

    ArticleĀ  CASĀ  Google ScholarĀ 

  8. Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328ā€“54.

    ArticleĀ  CASĀ  Google ScholarĀ 

  9. Bailey D, Hofman M, Forwood N, Oā€™Keefe G, Scott A, van Wyngaardt W, Howe B, Kovacev O, Francis R. Accuracy of dose calibrators for 68Ga PET imaging: unexpected findings in a multicenter clinical pretrial assessment. J Nucl Med. 2018;59:636ā€“8.

    ArticleĀ  CASĀ  Google ScholarĀ 

  10. Preylowski V, Schlƶgl S, Schoenahl F, Jƶrg G, Samnick S, Buck AK, et al. Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas? PLoS One. 2013;8:1ā€“8.

    ArticleĀ  Google ScholarĀ 

  11. Jƶnsson L, Stenvall A, Mattsson E, Larsson E, Sundlƶv A, Ohlsson T, et al. Quantitative analysis of phantom studies of 111In and 68Ga imaging of neuroendocrine tumours.

  12. Kaalep A, Huisman M, Sera T, Vugts D, Boellaard R. Feasibility of PET/CT system performance harmonisation for quantitative multicentre 89Zr studies. EJNMMI Phys. 2018;5:26.

    ArticleĀ  Google ScholarĀ 

Download references

Acknowledgements

The authors thank A. Eek for coordinating the BetaCure study. Furthermore, we thank the hospitals of the Dutch PSMA consortium and the other centres who have sent their clinical 68Ga PET/CT acquisition protocols.

Funding

The research leading to these results have received funding from the European Communityā€™s Seventh Framework Programme (FP7/2007-2013) under grant agreement no. 602812 (BetaCure study).

Author information

Authors and Affiliations

Authors

Contributions

DH performed data collection, analysis and drafted the manuscript. DH, DK, LWV, MS and JvD discussed the methodology. RB provided the analysis tools and discussed methodology. All authors critically reviewed the manuscript and approved the final version of the manuscript.

Corresponding author

Correspondence to Berlinda J. de Wit-van der Veen.

Ethics declarations

Ethics approval and consent to participate

This article does not contain any studies with human participants or animals performed by any of the authors.

Consent for publication

Not applicable.

Competing interests

RB is a scientific advisor and chair of the EARL accreditation programme. TS is an associate of the EARL accreditation programme. All other authors declare that they have no conflict of interest.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Huizing, D.M.V., Koopman, D., van Dalen, J.A. et al. Multicentre quantitative 68Ga PET/CT performance harmonisation. EJNMMI Phys 6, 19 (2019). https://doi.org/10.1186/s40658-019-0253-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-019-0253-z

Keywords