Skip to main content

Table 2 Comparison of spatial resolution and image quality parameters obtained between dedicated SA-PET devices and Vision clinical SiPM-based PET system

From: PET imaging and quantification of small animals using a clinical SiPM-based camera

 

MicroPET focus 120 [22]

MosaicHP [23]

Inveon [24]

Flex triumph [25]

Clear PET[26]

rPET [26]

NanoScan [27]

Iris [28]

PET Insert Si 198 [29]

Biograph Vision 450

PSF

PSF + G2 mm

Radial FWHM centre FOV (mm)

1.92

2.32

1.81

 ~ 1.8

1.9

1.4

1.5

1.05

0.75

1.80

-

Radial FWTM centre FOV (mm)

3.66

5.3

3.57

 ~ 3.6

3.8

2.5

3.29

2.35

1.38

3.35

-

SORwater

0.018

0.063

0.046

0.09

0.27

0.15

0.097

0.11

0.22

0.10

0.10

SORair

0.003

0.027

0.047

0.1

0.37

0.24

0.086

0.11

0.12

0.05

0.05

Uniformity (%SD)

6

5.1

7.447

6.01

-

-

6.53

7.0

6.5

5.54

3.47

RC 5 mm

0.93

0.84

1.059

0.89

0.89

0.81

0.92

0.89

0.94

1.67

1.1

RC 4 mm

0.86

0.7

0.911

0.68

0.73

0.76

0.8

0.82

0.95

1.58

0.91

RC 3 mm

0.75

0.56

0.816

0.56

0.42

0.66

0.65

0.73

0.91

0.9

0.56

RC 2 mm

0.48

0.36

0.604

0.43

0.21

0.46

0.49

0.58

0.64

0.5

0.28

RC 1 mm

0.15

0.16

0.367

0.15

0.14

0.11

0.24

0.14

0.14

0.08

0.06