Skip to main content

Table 6 The sensitivity for the emission window (EM) and scatter window (SC) images of the \(^{99\text {m}}\)Tc phantoms with standard uncertainties

From: Quantitative validation of Monte Carlo SPECT simulation: application to a Mediso AnyScan GATE simulation

Isotope

Phantom

Window

Sensitivity (counts/(s\(\times\)MBq))

Sensitivity ratio

   

Simulation

Experiment

(Sim/Exp)

\(^{99\text {m}}\)Tc

Cylinder

EM

36.33 ± 0.73

37.39 ± 0.33

0.972 ± 0.021

  

SC

8.74 ± 0.19

7.865 ± 0.071

1.111 ± 0.026

\(^{99\text {m}}\)Tc

NEMA

EM

24.77 ± 0.50

27.06 ± 0.24

0.915 ± 0.020

  

SC

4.464 ± 0.095

4.376 ± 0.039

1.020 ± 0.023

  1. The last column gives the sensitivity ratio (simulation/experiment) which was used to normalise the simulation to validate projection and profile data