Skip to main content

Table 5 The full-width half maximum (FWHM) of the Gaussian fits to each profile of the spatial resolution measurements

From: Quantitative validation of Monte Carlo SPECT simulation: application to a Mediso AnyScan GATE simulation

 

FWHM (mm)

 

Experiment

Simulation

Position name

  

Centre

10.14 ± 0.22

11.15 ± 0.15

CentreLeft

9.64 ± 0.20

10.64 ± 0.13

TopRight

10.50 ± 0.21

11.39 ± 0.12

Detector radius (mm)

  

106.5

10.14 ± 0.22

11.15 ± 0.15

150.0

11.24 ± 0.23

12.13 ± 0.13

250.0

14.50 ± 0.29

15.05 ± 0.12

350.0

18.81 ± 0.51

18.38 ± 0.14