Skip to main content

Table 3 Visual scores of PET images reconstructed by HYPER Iterative and OSEM at different scan times in the half-dose group

From: A proper protocol for routine 18F-FDG uEXPLORER total-body PET/CT scans

Time

Reconstruction algorithm

Reader 1

Reader 2

Average score

P

Score 5

Score 4

Score 3

Score 2

Score 1

Score 5

Score 4

Score 3

Score 2

Score 1

10 s

OSEM

0

0

0

0

10

0

0

0

0

10

1.00 ± 0.00

0.000

 

HYPER Iterative

0

0

0

10

0

0

0

0

10

0

2.00 ± 0.00

30 s

OSEM

0

0

0

8

2

0

0

0

7

3

1.75 ± 0.44

0.000

 

HYPER Iterative

0

0

10

0

0

0

0

8

2

0

2.90 ± 0.31

1 min

OSEM

0

0

7

3

0

0

0

7

3

0

2.70 ± 0.47

0.000

 

HYPER Iterative

0

8

2

0

0

0

7

3

0

0

3.75 ± 0.44

2 min

OSEM

0

8

2

0

0

0

7

7

0

0

3.75 ± 0.44

0.000

 

HYPER Iterative

8

2

0

0

0

8

2

0

0

0

4.80 ± 0.41

3 min

OSEM

9

1

0

0

0

8

2

0

0

0

4.85 ± 0.37

0.250

 

HYPER Iterative

10

0

0

0

0

10

0

0

0

0

5.00 ± 0.00

4 min

OSEM

10

0

0

0

0

10

0

0

0

0

5.00 ± 0.00

1.000

 

HYPER Iterative

10

0

0

0

0

10

0

0

0

0

5.00 ± 0.00

5 min

OSEM

10

0

0

0

0

10

0

0

0

0

5.00 ± 0.00

1.000

HYPER Iterative

10

0

0

0

0

10

0

0

0

0

5.00 ± 0.00