Skip to main content
Fig. 3 | EJNMMI Physics

Fig. 3

From: A proper protocol for routine 18F-FDG uEXPLORER total-body PET/CT scans

Fig. 3

Total-body 18F-FDG PET MIP images of different scan times reconstructed by OSEM and HYPER Iterative in a patient with lung cancer injected with half dose18F-FDG. a 18F-FDG PET MIP images of 10 s, 30 s, 1 min, 2 min, 3 min, 4 min and 5 min, respectively, reconstructed by OSEM. The quality of OSEM-10 s, OSEM-30 s, OSEM-1 min and OSEM-2 min images was low due to low counting rate and large signal noise, especially the OSEM-10 s and OSEM-30 s. The OSEM-3 min image reached the high quality, either that of OSEM-4 min and OSEM-5 min. b 18F-FDG PET MIP images of 10 s, 30 s, 1 min, 2 min, 3 min, 4 min and 5 min, respectively, reconstructed by HYPER Iterative. The quality of HYPER Iterative-10 s, HYPER Iterative-30 s and HYPER Iterative-1 min had lower signal noise and higher quality than the counterparts by OSEM although they were still not excellent. The HYPER Iterative-2 min image reached the high quality, either that of HYPER Iterative-3 min to HYPER Iterative-5 min

Back to article page