Fig. 3 | EJNMMI Physics

Fig. 3

From: Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators

Fig. 3

Schematic visualization of uEXPLORER (I) (blue) [8, 10, 13]) and sparse geometry (transparent yellow) with 194.8 cm of AFOV and dual layers. TB J-PET (II) with axially arranged plastic scintillators (gray) coupled with SiPMs (black) at each end and arrays of WLS (green) between each layer. The oblique LORs (with large values of \(\theta\)) have a negative contribution in the spatial resolutions due to the parallax error. To avoid it, uEXPLORER uses a ring-based cut that accepts the coincidences within a maximum of 5 rings. As shown in figure (II), TB J-PET uses continuous plastic scintillators (gray), \(\theta _{AA}\) denotes the acceptance angle applied for it to cut oblique LORs. \(\theta _{Max}\) demonstrate the largest angle of detectable oblique coincidences