Skip to main content

Table 1 Cross-calibration factors for the planar, SPECT/CT and SPECT acquisitions

From: Quantitative analysis of 99mTc-pertechnetate thyroid uptake with a large-field CZT gamma camera: feasibility and comparison between SPECT/CT and planar acquisitions

Acquisition

Cross-calibration factors (counts.MBq−1 s−1)

Type

Mean ± SD

Planar

66.2 ± 4.9

SPECT/CT (Tomo-AC)

60.7 ± 0.7

SPECT (Tomo-NoAC)

26.5 ± 0.3