Skip to main content

Table 3 Spatial resolution FWHM for design B

From: Monte Carlo simulation of the system performance of a long axial field-of-view PET based on monolithic LYSO detectors

 

Radius (cm)

Radial (mm)

Tangential (mm)

Axial (mm)

aFOV centre

1

1.45

1.48

1.36

10

1.46

1.44

1.32

20

1.5

1.34

1.24

3/8 aFOV

1

1.37

1.37

1.31

10

1.39

1.35

1.20

20

1.44

1.31

1.17