Skip to main content

Table 6 SPECT volume sensitivity, uniformity, and variability

From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

Reporting parameter

Measurement

Simulation

Collimator

MPH

SPH

SPH

System volume sensitivity SVS (cps/(MBq/\(\hbox {cm}^3\)))

2200

329

354

Volume sensitivity per axial centimeter VSAC (cps/(MBq/\(\hbox {cm}^2\)))

1570

157

169

Total system response (cps/(MBq/\(\hbox {cm}^3\)))

2200

901

970

Uniformity (%)

96.5

48.2

38.2

Coefficient of variation CV (%)

38.7

11.6

10.2