Skip to main content

Table 5 SPECT reconstructed spatial resolution with scatter

From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

Reporting parameter

Measurement

Simulation

Collimator

MPH

SPH

SPH

SPH

SPH

Reconstruction algorithm

MLEM

MLEM

FBP

MLEM

FBP

Central FWHM (mm)

\(1.29\pm 0.05\)

\(1.52\pm 0.04\)

\(2.23\pm 0.07\)

\(1.54 \pm 0.05\)

\(2.21 \pm 0.08\)

Radial FWHM (mm)

\(1.27\pm 0.06\)

\(1.39\pm 0.06\)

\(2.34\pm 0.11\)

\(1.40\pm 0.04\)

\(2.30\pm 0.06\)

Tangential FWHM (mm)

\(0.99\pm 0.07\)

\(1.41\pm 0.03\)

\(2.03\pm 0.11\)

\(1.43 \pm 0.03\)

\(2.06 \pm 0.10\)

Average in-plane FWHM (mm)

\(1.18\pm 0.15\)

\(1.44\pm 0.07\)

\(2.20\pm 0.16\)

\(1.46 \pm 0.07\)

\(2.19 \pm 0.12\)