From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
Reporting parameter
Region of interest
Measurement
Simulation
Integral uniformity (%)
UFOV
2.96
1.72
CFOV
2.79
1.96
Row differential uniformity (%)
2.75
1.69
2.52
1.77
Column differential uniformity (%)
1.67
2.11