Skip to main content

Table 2 Intrinsic spatial resolution and linearity

From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

Reporting parameter

Region of interest

Measurement

Simulation

Resolution PSF FWHM (mm)

Middle of FOV

\(0.851\pm 0.010\)

\(0.850\pm 0.003\)

Resolution PSF FWTM (mm)

Middle of FOV

\(1.559\pm 0.014\)

\(1.591\pm 0.007\)

Resolution LSF FWHM (mm)

UFOV

\(0.912\pm 0.098\)

\(0.916\pm 0.026\)

CFOV

\(0.953\pm 0.091\)

\(0.924\pm 0.029\)

Resolution LSF FWTM (mm)

UFOV

\(1.73\pm 0.15\)

\(1.66\pm 0.03\)

CFOV

\(1.80\pm 0.14\)

\(1.68\pm 0.03\)

Differential linearity (mm)

UFOV

0.023

0.001

CFOV

0.024

0.002

Absolute linearity (mm)

UFOV

0.102

0.003

CFOV

0.055

0.003