Fig. 6 | EJNMMI Physics

Fig. 6

From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

Fig. 6

System spatial resolution without scatter presented in terms of FWHM and FWTM for the SPH collimator. Measured results are shown with solid red carets and solid lines of best fit, and simulated results are shown with empty black carets and dashed lines of best fit. Equations for lines of best fit are distinguished in the legend with abbreviated subscripts. The measured (simulated) system resolution without scatter at the center of rotation was 1.87 mm (1.80 mm)