Fig. 4From: NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scannerIntrinsic count rate performance in air. Measured results are shown with red solid carets and simulated results with black empty carets. Additional lines illustrate the maximum OCR and 20% loss OCR for measurement (solid lines) and simulation (dashed lines). The count rates are in agreement below the maximum OCR, while above the maximum, the measured OCR falls off the trend line as the photopeak shifted to lower energiesBack to article page