Skip to main content

Table 3 Results of image evaluation

From: Development of a bespoke phantom to optimize molecular PET imaging of pituitary tumors

Reconstruction

Confusion matrix

Confidence

Combined

Sensitivity

Specificity

Accuracy

True positive confidence

True negative confidence

False positive confidence

True negative confidence

Summed confidence

OSEM 128 TOF 3i24s

62

95

83

1.86

2.86

− 0.25

− 1.59

30.8

OSEM 128 TOF PSF 3i24s

78

90

89

2.78

2.32

− 0.88

− 0.44

37.1

OSEM 256 TOF PSF 3i24s

78

90

88

3.02

2.51

− 1.38

− 1.22

39.1

OSEM 256 TOF PSF 5i24s

81

79

84

3.09

2.30

− 1.47

− 2.67

34.4

OSEM 256 TOF PSF 7i24s

84

60

74

3.01

1.98

− 0.84

− 1.00

30.1

OSEM 256 TOF PSF 9i25s

86

56

73

2.94

2.04

− 0.94

− 1.00

29.1

BPL 256 TOF PSF β50

86

32

60

3.03

1.31

− 1.18

− 4.00

19.2

BPL 256 TOF PSF β100

89

47

70

3.14

1.79

− 1.12

0.00

27.3

BPL 256 TOF PSF β200

86

74

84

3.16

2.24

− 0.95

0.00

37.3

BPL 256 TOF PSF β300

83

84

88

3.07

2.52

− 1.54

0.00

39.7

BPL 256 TOF PSF β400

77

93

90

2.98

2.51

− 1.33

− 0.90

39.6

BPL 256 TOF PSF β500

68

95

86

3.02

2.86

− 0.25

− 1.65

39.7

BPL 256 TOF PSF β600

68

95

86

2.93

2.77

− 1.25

− 1.76

37.7

BPL 256 TOF PSF β800

67

95

86

2.83

2.82

− 1.00

− 1.72

37.3

BPL 256 TOF PSF β1000

64

95

86

2.71

2.86

− 1.00

− 1.75

37.1