From: Development of a bespoke phantom to optimize molecular PET imaging of pituitary tumors
Reconstruction | Confusion matrix | Confidence | Combined | |||||
---|---|---|---|---|---|---|---|---|
Sensitivity | Specificity | Accuracy | True positive confidence | True negative confidence | False positive confidence | True negative confidence | Summed confidence | |
OSEM 128 TOF 3i24s | 62 | 95 | 83 | 1.86 | 2.86 | − 0.25 | − 1.59 | 30.8 |
OSEM 128 TOF PSF 3i24s | 78 | 90 | 89 | 2.78 | 2.32 | − 0.88 | − 0.44 | 37.1 |
OSEM 256 TOF PSF 3i24s | 78 | 90 | 88 | 3.02 | 2.51 | − 1.38 | − 1.22 | 39.1 |
OSEM 256 TOF PSF 5i24s | 81 | 79 | 84 | 3.09 | 2.30 | − 1.47 | − 2.67 | 34.4 |
OSEM 256 TOF PSF 7i24s | 84 | 60 | 74 | 3.01 | 1.98 | − 0.84 | − 1.00 | 30.1 |
OSEM 256 TOF PSF 9i25s | 86 | 56 | 73 | 2.94 | 2.04 | − 0.94 | − 1.00 | 29.1 |
BPL 256 TOF PSF β50 | 86 | 32 | 60 | 3.03 | 1.31 | − 1.18 | − 4.00 | 19.2 |
BPL 256 TOF PSF β100 | 89 | 47 | 70 | 3.14 | 1.79 | − 1.12 | 0.00 | 27.3 |
BPL 256 TOF PSF β200 | 86 | 74 | 84 | 3.16 | 2.24 | − 0.95 | 0.00 | 37.3 |
BPL 256 TOF PSF β300 | 83 | 84 | 88 | 3.07 | 2.52 | − 1.54 | 0.00 | 39.7 |
BPL 256 TOF PSF β400 | 77 | 93 | 90 | 2.98 | 2.51 | − 1.33 | − 0.90 | 39.6 |
BPL 256 TOF PSF β500 | 68 | 95 | 86 | 3.02 | 2.86 | − 0.25 | − 1.65 | 39.7 |
BPL 256 TOF PSF β600 | 68 | 95 | 86 | 2.93 | 2.77 | − 1.25 | − 1.76 | 37.7 |
BPL 256 TOF PSF β800 | 67 | 95 | 86 | 2.83 | 2.82 | − 1.00 | − 1.72 | 37.3 |
BPL 256 TOF PSF β1000 | 64 | 95 | 86 | 2.71 | 2.86 | − 1.00 | − 1.75 | 37.1 |