Skip to main content

Table 1 Scores and failure rates for lesion visibility and image quality with different acquisition parameters

From: Detection of prostate cancer bone metastases with fast whole-body 99mTc-HMDP SPECT/CT using a general-purpose CZT system

Acquisition time (min)

Energy window width (%)

Matrix size

N

Lesion visibility score, mean (SD)

Overall image quality score, mean (SD)

Lesion visibility failure rate

Overall image quality failure rate

50

15

128 × 128

45

3.8 (0.8)

3.6 (0.9)

4%

11%

50

10

128 × 128

45

3.8 (0.9)

3.5 (1.0)

4%

16%

50

8

128 × 128

45

3.7 (1.0)

3.4 (1.1)

11%

27%

50

6

128 × 128

45

3.6 (0.9)

3.1 (1.0)a

9%

31% a

50

4

128 × 128

45

3.5 (0.9)a

2.7 (1.1)a

7%

49% a

50

15

256 × 256

45

3.7 (0.9)

3.3 (1.0)a

7%

22%

50

15

128 × 128

90

4.2 (0.8)

3.8 (1.0)

0%

13%

38

15

128 × 128

90

3.8 (0.8)b

3.4 (1.0)b

2%

19%

32

15

128 × 128

90

3.4 (0.9)b

2.9 (1.0)b

11%b

37%b

26

15

128 × 128

90

3.0 (1.0)b

2.2 (1.0)b

32%b

67%b

20

15

128 × 128

90

2.3 (1.0)b

1.4 (0.6)b

58%b

92%b

  1. N  Total number of scores, i.e., the number of readers multiplied by the number of images. Statistically significant difference (p < 0.05) compared with original images (N = 45). Statistically significant difference (p < 0.05) compared with original images (N = 90)