Table 3 Image contrasts at the central position (POS-0) for the different sphere sizes, SBRs and acquisition times. Of note, the slightly higher contrast for the 22 mm sphere size than found for the 28 mm sphere size is expected to be related to Gibbs artefacts well known in PSF reconstructions, in particular evident in unfiltered reconstructions [32]

From: Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system for 18F-FDG

 

SBR = 8:1

SBR = 4:1

SBR = 2:1

Acquisition time

600 s

120 s

120 s filtered

60 s

30 s

15 s

600 s

120 s

120 s filtered

60 s

30 s

15 s

600 s

120 s

120 s filtered

60 s

30 s

15 s

Sphere size (mm)

37

85.3

85.4

83.4

85.3

85.2

85.9

84.8

85.1

83.6

85.3

85.0

83

86.7

86.2

85.2

87.2

88.3

87.9

28

79.9

79.4

76.9

78.6

78.1

77.6

79.4

79.1

77.3

79.2

79.8

76.0

79.5

79.8

78.7

81.0

78.3

80.6

22

81.5

81.8

78.2

82.1

83.0

82.4

80.8

80.3

77.8

80.9

80.7

78.5

81.4

80.7

79.3

78.8

76.3

77.2

17

77.8

77.8

73.3

79.3

80.1

78.2

75.8

77.7

74.3

77.3

78.6

75.9

76.8

76.3

74.4

73.8

71.5

76.1

13

67.9

69.1

63.9

68.5

66.0

65.2

65.6

65.9

62.3

62.4

64.8

61.5

68.3

69.3

68.1

69.2

61.2

61.9

10

61.1

60.4

54.2

59.3

62.6

53.4

59.6

59.3

54.7

55.7

56.4

55.4

67.1

68.1

65.2

70.2

70.7

72.5