Table 2 Routine X-ray CT quality control testing; daily, monthly and quarterly/annually

From: Understanding the importance of quality control and quality assurance in preclinical PET/CT imaging

Test

Purpose

Frequency

Comments

CT

Daily procedures

Daily

Follow manufacturer’s recommendations (e.g. tube warm-up)

CT/PET

Determine co-registration vector of PET

and CT field of view

Daily

Check PET and CT image alignment

CT HUs

Determine CT HUs

accuracy

Daily

CT HUs accuracy: water and air and standard deviation (noise)

CT artefacts

Evaluation of artefacts

Daily

Visual inspection of the water image slices from the CT QC phantom

CT performance

Check X-ray performance

and radiation exposure

Based on national legislation

Follow national legislation guidelines

CT grey level performance

Check scanner acquisition display monitors

Monthly

Check quality of monitors grey scale for image display (see also Society of Motion Picture and Television Engineers (SMPTE) test) [48, 49]

  1. Modified, EANM Physics Group on Nuclear Medicine Instrumentation Quality Control 2010 [24, 50]