Table 1 Target activity concentrations for compartments in the Probe-IQ phantom

From: Quantitative evaluation of PSMA PET imaging using a realistic anthropomorphic phantom and shell-less radioactive epoxy lesions

Compartment

Volume (mL)

Activity concentration (kBq/mL)

Relative concentration

Bladder

440 (377 mL)

176.3

98:1

Liver

1025

9.9

5.5:1

Lungs (L)

875

0.67

0.37:1

Lungs (R)

1152

0.67

0.37:1

Pelvis

9000

1.8

1.0:1

Thorax

17,300

1.8

1.0:1

Ureter (L)

10

92.8

51.6:1

Ureter (R)

10

66.5

37:1