Skip to main content

Table 1 Standard clinical reconstruction parameters for the different PET scanners

From: Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer

Iterative reconstruction

Scanner type

Iterations × subsets

Gauss filter (mm)

Voxel size (mm3)

OSEM

Vision

10 × 5

4

1.7 × 1.7 × 2.0

mCT

3 × 24

3

2.0 × 2.0 × 2.0

mMR

3 × 21

4

2.1 × 2.1 × 2.0

OSEM-TOF

Vision

4 × 5

4

1.7 × 1.7 × 2.0

mCT

2 × 21

3

2.0 × 2.0 × 2.0

mMR

–

–

–

OSEM-TOF+PSF

Vision

4 × 5

4

1.7 × 1.7 × 2.0

mCT

2 × 21

3

2.0 × 2.0 × 2.0

mMR

–

–

–