PET system | Contrast (%) | CV (%) | SD (clinical vs. 30 min) | Contrast gain (%) |
---|
PMT-PET | 58.5 | 11.1 | 0.021/0.018 | − 1.7 |
SiPM-PET | 72.4 | 11.5 | 0.015/0.009 | 3.5 |
- JSNM criteria: contrast ratio > 55%, CV < 13%, SD < 0.0249
- CV Coefficient of variation, JSNM Japanese Society of Nuclear Medicine, PET Positron emission tomography, PMT Photomultiplier tube, SD Standard deviation, SiPM Silicon photomultiplier