Table 2 Mean-, max-, and min-percentages of contrast recovery, background variability, and lung residual for the 4:1 sphere-to-background ratio for three sequential measurements of IEC image quality phantom

From: Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard

Sphere diameter(mm)

10

13

17

22

28

37

Lung residual

Contrast recovery

Mean

46.5%

76.2%

79.3%

82.8%

82.1%

83.9%

 

Max

48.5%

79.5%

83.0%

85.0%

84.9%

84.5%

Min

43.4%

72.7%

76.4%

80.8%

80.4%

82.8%

Background variability

Mean

6.4%

5.5%

4.7%

4.0%

3.3%

2.5%

3.1%

Max

6.9%

6.0%

5.1%

4.5%

3.7%

2.6%

3.5%

Min

5.7%

5.0%

4.3%

3.6%

2.9%

2.3%

2.7%