Skip to main content

Table 4 System sensitivity (cps/MBq) for each collimator and energy window for 177Lu

From: Characterisation of a hand-held CZT-based gamma camera for 177Lu imaging

 

Circular ROI

Full FOV

 

55 keV

113 keV

208 keV

55 keV

113 keV

208 keV

MEGP

12.8

17.6

13.0

14.0

19.1

14.7

LEHR

20.0

27.5

26.0

24.2

32.2

43.5

LEHS

61.7

84.4

83.3

72.6

96.7

127.2