Skip to main content

Table 3 Reference values measured in a healthy liver for OSEM and BSREM

From: Noise reduction using a Bayesian penalized-likelihood reconstruction algorithm on a time-of-flight PET-CT scanner

MeasurebOSEMa BSREM 
β300Β400β500Β600β750β1000β1500β3000
Volume (ml)14.137        
SUVmean2.2122.1872.1842.1922.1932.1972.1932.2042.212
SUVSD0.4800.4410.3620.3010.2720.2410.2060.1520.092
  1. aTOF PSF OSEM (3 iterations and 16 subsets and 5 mm Gaussian filter)
  2. bMeasured in a sphere placed in liver for 10 min (total) acquisition time