Skip to main content

Table 3 Summary of the supply and net administration doses, and the leakage from the face mask system

From: System evaluation of automated production and inhalation of 15O-labeled gaseous radiopharmaceuticals for the rapid 15O-oxygen PET examinations

 

C15O scan

15O2 scan

C15O2 scan

Supplied radioactivity per scan

2574 ± 255 MBq

2220 ± 766 MBq

1763 ± 174 MBq

Net administration per scan

541 ± 149 MBq

320 ± 103 MBq

523 ± 137 MBq

%Leakage

0.166 ± 0.493%

0.181 ± 0.525%

0.148 ± 0.545%

# of scans with no detected leakage

417

395

529

# of scans

749

749

749