Table 5 Comparison of the calibration factors obtained with standard method and image-based FAF method, computed on patient data

From: Image-based SPECT calibration based on the evaluation of the Fraction of Activity in the Field of View

Patient SFAF (cps/MBq) Rel. diff. with Sstd(%) Arms in same position
1 1335 -1.29 Yes
2 1411 4.19 No
3 1374 1.58 No
4 1481 8.72 No
5 1347 -0.39 Yes
6 1486 9.00 No