Skip to main content

Table 1 ESVTrue, EDVTrue and EFTrue measured from gated CT images

From: Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR

Phantom settings bpm EDVTrue
[ml]
ESVTrue
[ml]
EDV-ESV
[ml]
EFTrue
[%]
EF 45% 60 56.9 31.2 25.8 45.3
EF 55% 60 73.1 33.1 40.0 54.8
EF 65% 40 94.3 33.2 61.2 64.8
EF 65% 50 95.2 33.3 61.9 65.0
EF 65% I 60 93.6 32.6 61.0 65.2
EF 65% II 60 94.9 32.9 62.0 65.1
EF 65% 70 95.2 33.2 61.9 65.1
EF 70% 60 110.4 32.5 77.9 70.6