Skip to main content

In-line high resolution PET and 3T MRI hybrid device for preclinical multimodal imaging

In this work we evaluate a combination of a high-performance PET with a superconductive 3T MRI system. This PET/MRI system is implemented in a tandem configuration similar to some scanners used in clinical imaging but with both subsystems placed in proximity, what allows sharing a common animal holder that travels between both FOVs.

The PET detector consists in two rings, each one having 18 DOI detectors, which provide a high sensitivity with a 1.0mm3 resolution at the centre and minimal degradation toward the edges of the 68mm diameter FOV.

The MRI system is a compact, cryogen-free, auto-shielded and with a very tight fringe-field and no special room requirements. These characteristics enable the close location of the PET detector without deteriorating its performance.

We have evaluated the system obtaining images of three mice in cardiology, oncology and neurology applications. All PET scans were performed 30m after intraperitoneal FDG injection of 25.2±5.8MBq and were reconstructed using a 3D-OSEM algorithm. For the cardiac MRI a gated FLASH sequence and a body coil were used. Tumor MRI was using an FSE T1w sequence and a body coil. Finally, brain MRI was done with an FSE T2w sequence.

In all the cases registration of PET and MRI images was obtained following an equivalent protocol to the standard PET/CT preclinical system, and since the animals didn’t move from the bed the results were satisfactory. Attenuation correction was not applied to any image. The preliminary results from this prototype suggest that the tandem solution for preclinical PET/MRI is feasible and has the advantage of preserving the best image quality in both modalities without requiring any trade-off of specifications. The limitation of not being able to perform simultaneous acquisitions is compensated by the high quality results of studies that do not actually require such simultaneous data collection.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Vaquero, J.J., Montesinos, P., Cussó, L. et al. In-line high resolution PET and 3T MRI hybrid device for preclinical multimodal imaging. EJNMMI Phys 1 (Suppl 1), A7 (2014). https://doi.org/10.1186/2197-7364-1-S1-A7

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A7

Keywords