Skip to main content

Table 8 Comparison of measured planar System Sensitivity with Monte Carlo results. All parameters have been measured at distance of 10 cm from collimator

From: Validation of \(^{99m}\)Tc and \(^{177}\)Lu quantification parameters for a Monte Carlo modelled gamma camera

Radioisotope

Main peak [keV]

Sensitivity [cps/MBq]

Experimental

Simulated

Tc-99m

140.5

88.0 ± 4.4

89.4 ± 0.5

Lu-177

113.0

9.9 ± 0.5

9.7 ± 0.1

Lu-177

208.0

9.6 ± 0.5

10.9 ± 0.2