Skip to main content

Table 3 Photopeak Window (PW), Low Scatter Window (LSW), Upper Scatter Window (USW) for \(^{99m}\)Tc and \(^{177}\)Lu main peaks are listed

From: Validation of \(^{99m}\)Tc and \(^{177}\)Lu quantification parameters for a Monte Carlo modelled gamma camera

Radionuclide

Main peak [keV]

PW range [keV]

LSW range [keV]

USW range [keV]

Tc-99m

140.5

123.9–151.4

103.2–123.9

not used

Lu-177

113

102.7–119.4

86.1–102.7

119.4–128.2

Lu-177

208

189.3–220.0

168.8–189.3

Not used