From: Performance evaluation of a novel multi-pinhole collimator on triple-NaI-detector SPECT/CT for dedicated myocardial imaging
Collimator and Detector Configuration
50Â mm diameter VOI Mean [counts]
50Â mm diameter VOI St.Dev. [counts]
Image uniformity (%)
\({100\times \left(\frac{St.Dev.}{Mean}\right)}^{2}\)
LEHR 90°
37.55
2.03
0.292
MPH-CARDIAC 75°
287.16
13.29
0.214