Skip to main content

Table 2 PET/CT scan and reconstruction algorithms

From: Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients

PET Scanner

Biograph mCT

Gemini TF 16

CT

Voltage/Intensity

120 kV/ ref mAs

120 kV/ 100 mAs

Collimation/pitch

16 × 1.2 mm/pitch 0.8

16 × 1.5 mm/pitch 0.813

PET scan parameter

Bed motion speed

FlowMotion mode 2 mm/s (~ 100 s/bed)

40 s/bed

PET reconstruction algorithm

Reconstruction algorithm

OSEM

OSEM + PSF + TOF (Clinical)

BLOB-OS-TOF (Clinical)

BLOB-OS-TOF

Iterations/Subsets

2/24

2/21

3/33

3/33

Smoothing

No

5 mm

Normal*

Smooth A*

Matrix

200 × 200

200 × 200

144 × 144

144 × 144

EQ.PET filter (mm)

8

7.8

4.9

4.4

  1. *The ‘normal’ and ‘smooth A’ smoothing options correspond to a relaxation parameter of 1.0 and 0.6 in the BLOB-OS-TOF algorithm, respectively