Skip to main content

Table 1 Peak noise equivalent count rates (NECRs), system type, and references of scanners shown in Figs. 1 and 2

From: Technical opportunities and challenges in developing total-body PET scanners for mice and rats

Name of scanner

Peak NECR (kcps)

System type

References

Mouse phantom

Rat phantom

NanoPET

430

130

PET/CT

[45]

A-PET

~ 500

~ 230

Standalone

[40]

Inveon

1670

590

PET/CT

[43]

Metis

1344

640

PET/CT

[47]

IRIS

185

40

PET/CT

[46]

ClairvivoPET

415 kcps (50 mm 18F line source)

Standalone

[41]

HiPET

179

63

Standalone

[30]

LabPET-12

362

156

PET/CT

[42]

ClearPET

Unknown

Unknown

PET/CT

[44]

Albira Si PET

486

239

PET/SPECT/CT

[29]

SIAT aPET

324

144

MR-Compatible

[31]

β-CUBE

300

160

PET/CT

[28]

SimPET-X

348

Unknown

MR-Compatible

[13]

quadHIDAC32

67

52

Standalone

[38]

Yamaya’s PET

Unknown

Unknown

Standalone

[21]

Mouse J-PET

Unknown

Unknown

Unknown

[48]

Rat J-PET

Unknown

Unknown

Unknown

[48]

HS-BGO PET

Unknown

Unknown

MR-Compatible

[49]

H2RS110

Unknown

Unknown

MR-Compatible

[12]

H2RS160

Unknown

Unknown

Unknown

[12]

UHS-PET

Unknown

Unknown

Unknown

[50]