Skip to main content

Table 2 Percent contrast and percent background variability evaluated with the brain-size image-quality phantom

From: Evaluation of the performance of a high-resolution time-of-flight PET system dedicated to the head and breast according to NEMA NU 2-2012 standard

Parameter

Sphere diameter (mm)

10

13

17

22

28

37

Percent contrast for hot spheres (%)

49.9 ± 23.3

63.2 ± 22.6

72.4 ± 22.0

79.5 ± 21.5

Percent contrast for cold spheres (%)

81.8 ± 6.7

84.8 ± 7.1

Percent background variability (%)

4.2

  1. Data are presented as mean ± standard deviation