Table 2 Contrast, noise, and CNR of images with different acquisition parameters

From: Detection of prostate cancer bone metastases with fast whole-body 99mTc-HMDP SPECT/CT using a general-purpose CZT system

Acquisition time (min)

Energy window width (%)

Matrix size

Contrast, median (IQR)

Noise, median (IQR)

CNR, median (IQR)

50

15

128 × 128

2.0 (1.4–3.2)

0.09 (0.08–0.11)

22.2 (15.7–31.1)

50

10

128 × 128

2.0 (1.4–3.4)

0.12 (0.09–0.16)

17.5 (12.0–24.3)

50

8

128 × 128

2.1 (1.4–3.5)

0.13 (0.10–0.18)

17.4 (12.6–25.8)

50

6

128 × 128

2.2 (1.4–3.5)

0.15 (0.12–0.18)

14.4 (10.4–21.3)

50

4

128 × 128

2.3 (1.4–3.6)

0.18 (0.15–0.22)

12.8 (8.2–20.1)

50

15

256 × 256

2.1 (1.4–3.2)

0.12 (0.10–0.15)

17.1 (12.1–24.5)

41

15

128 × 128

2.0 (1.4–3.3)

0.12 (0.10–0.16)

18.0 (12.0–25.8)

32

15

128 × 128

2.1 (1.4–3.3)

0.15 (0.11–0.19)

14.0 (9.4–20.3)

26

15

128 × 128

2.1 (1.4–3.3)

0.17 (0.14–0.23)

12.7 (8.3–19.7)

20

15

128 × 128

2.2 (1.5–3.6)

0.22 (0.18–0.29)

10.8 (7.1–15.3)