Table 2 Background CV (%) in the central position and useful axial FOV with background CV < 15% for the reconstructions after applying different Gaussian filter kernels

From: Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system for 18F-FDG

Gaussian filter FWHM (mm)

Equivalent acquisition time (s)

Background CV (%)

aFOV with CV < 15% (cm)

0

600

7.9

99.8

0

120

17.3

NA

2

120

14.6

83.0

3

120

12.5

91.6

4

120

9.3

97.0

5

120

7.5

100.4

6

120

6.1

102.8

0

60

24.6

NA

2

60

20.7

NA

3

60

16.6

NA

4

60

13.3

88.5

6

60

8.7

95.1

0

30

35.2

NA

2

30

28.8

NA

3

30

22.8

NA

4

30

18.1

NA

5

30

14.0

81.8

6

30

11.2

90.5

0

15

49.3

NA

2

15

39.9

NA

3

15

31.3

NA

4

15

24.7

NA

5

15

19.5

NA

6

15

15.6

NA

  1. NA denotes settings where a CV < 15% could not be reached at any position throughout the aFOV